Today I measured the dark noise of the new focus 1811 RF PD used in the unbalanced MZI. I connected the RF signal to the mixer together with the RF LO from the generator, then the output is filtered with the 1.9MHz filer.
I first measured on the oscilloscope the signal with the laser light on, to determine the calibration factor from Volt to radians: ASD[V] = A * ASD[rad], where A is the peak amplitude of the signal, which is A = 0.255 V/rad. Then I measured the spectrum of the output of the mixer with the SR785 without laser light on the PD, in Vrms/rtHz. The spectrum calbrated in rad/rtHz is shown in Fig. 1.
Wen made this diagram of the LASTI squeezer a few years ago. Uploading to log for posterity
[Eleonora, Wen]
Yesterday, we acquired the OLTF of the unbalanced MZI with the SR785 to characterize the loop.
We inject a sweep sine from 100 Hz to 10 kHz and we measured the TF in different configurations:
Before lunch
After lunch
Something that can be noticed is that the noise amplitude, needed to perform this measurement, incresed between before and after lunch. It can be due to a change of the demodulation phase, since it is not controlled. This is not the case unlocking and re-locking the loop. The TF shape did not change.
[Eleonora, Wen]
We detail here the instructions to perform a TF measurement with the SR785, how to extract / save data on a USB stick or floppy and how to convert the file.78D to file.asc.
Display Options -› Format -> Dual
Freq -> Type -> Log
Display Setup -> Measurement Group -> Swept Sine
Display Setup -> Meausrement -> Freq Resp
Active display to change from the first to the second subscreen
First Display: Display Setup -> View -> Log Mag
Second Display: Display Setup -> View -> Phase or Unwrp. Phase
Select the span in the frequency menu (Start and Stop)
Select the resolution of the swept sine in the average menu
Select the sweped sine amplitude in Source -> Amplitude
Connect input in Channel one A
Connect output in Channel two A
Save:
Disk -> File name -> Alt -> type file name
Disk -> Display to Disk
Convert data with the software on the NW22 control room computer (from floppy or USB stick):
Open Command Prompt
C:\SRTRANS> srtrans.exe ./FolderName/*.78D
It is possible to save the data in .TXT, with the following commands:
Output -> Hard Copy Button -> ASCII Dump
Output -> File start Number -> select output number
It saves the file as SCRN0001.TXT.
(Masaya) Turned on Diabolo for 2nd Squeezer. According to a data sheet, I set the values on the drivers as the below. Injection current : 2.15 [A] Diode A temperature : 22.47 [℃] Diode B temperature : 26.52 [℃] Laser Crystal Temperature : 24.16[℃] Doubling Crystal Temperature : 103.49[℃] Offset level read on 10 turns potentiometers : 5.18 Gain level read on 10 turns potentiometers : 2.88
I've noticed that most of the fiber components around the lab are from Thorlabs. In my experience, I've found Thorlabs fiber components to be higher loss and more expensive than other sources. I recently bought a batch of components from this company, https://www.afwtechnologies.com.au/, which I used extensively during my postdoc and industry job for 1064 nm and 1550 nm work. Their selection is most extensive at 1550 and 1310 nm, but most of their components have 1064 nm versions as well. Their components are lower loss than any other vendor I've tried and cost 3-5 times less than the equivalent part from thorlabs. The only downside is that you have to plan ahead a bit: Delivery usually takes about a month. I recently bought a batch of components for GRAVITES for ~$8000 that would have cost over $40000 from thorlabs.
After realizing with Nick and Slawek that we do not have a temperature acquisition system in the NW22 lab, I ordered a thermometer that allows us to record and save temperature regardless of distance. This can be done through the Govee Home smartphone app. The thermometer is shown in Fig.1, which also shows that it is way too hot in the lab at the moment (even I confirm)! Fig.2 shows the interface of the application.
If there are issues again with the lab heat/humidity (I'd heard it got above 30 C at one point), I would be conservative about running the Diabolo. Unfortunately, that laser is no longer manufactured and I'm unsure if anyone will do repairs given that Innolight was purchased by Coherent after they discontinued the Diabolo. In general lasers don't like high temp/humidity and running them under such conditions can shorten their lifespan. That laser, in particular, is irreplaceable, so we should be extra careful.
Yesterday, I connected a LabJack T7 to the cymac computer in NW22 and I configured it. After soldering two wired to the BNC adaptor (black for ground and red for V+), I connected the red to AIN1 and the black to AIN0 as shown in Fig. 1. I downloaded the python libraries and I wrote a python script to plot the voltage and the temperature as a function of time. It starts recording data from the time you run the script from a jupyter notebook called LabJack_vdiff.ipynb in controls/Documents/eleonora.polini/labjack. It is not a continuous monitor, but one can run the script remotely (by ssh to the cymac) and acquire the data any moment. In Fig.2 and Fig. 3, 5 minutes of data are shown for the differential voltage vs time and its correponding temperature.
(Hudson, Masaya, and Wen) It looks like we get about 0.3dB of squeezed light from the compact squeezer. We measured up to 7dB of anti-squeezing and observe a large amount of phase noise at 8kHz in the LO loop's error signal from the OPO dither lock. This phase noise looks to be at the level of about a radian RMS, so we will need to change the OPO locking scheme to have any hope of seeing high levels of squeezing. We're planning to switch the lock scheme to a Miller-Evans lock and have some new ideas about how to clip the beam properly to get a good Miller-Evans error signal. The LO error signal also has a component at 80kHz about 1/10th the size of the 8kHz component. This 80kHz component comes from the SHG dither lock, so this lock may also need to be altered. Since light is delivered to the SHG from a fiber, we have more options for modifying this lock, including potentially placing a fiber-coupled EOM before the SHG and locking it with a true PDH lock with sidebands outside of the cavity bandwidth. We could also consider switching the SHG locking scheme to a Miller-Evans lock.
With the 1/2" retroreflector, 751-VSMY2890GX01CT-ND LED, 6 cm tube, 1/2" f = 25 mm lens, and K857PE QPD.
Total photocurrent: 4.2 mA
Since the temperature shown by the termistor was not correct, I investigated the calibration of the termistor checking the read-out circuit with the help of Alvaro's thesis. The termistor is a TDK G1540/10k/F Negative Temperature Coefficient (NTC) and from its datasheet we can derive the relationship between the resistance R and the temperature T, as shown in Fig. 1. The circuit used to read out the temperature as a voltage is shown in Fig. 2. I verified that the resistors in this drawing correspond to the resistors in the integrated circuit and they do. The Vout is given by: Vout = (R2 + R3)/R2 * R1/(Rntc + R1) * Vdd - Vref * R3/R2, where R1 = 7.5 kOhm, R2 = 1 kOhm, R3 = 6.8 kOhm, Vdd = 9 V, Vref = 5 V and Rntc = 0.006 * T**2 - 0.67*T + 23.408 from the fit in Fig.1. Vout as a function of the temperature is shown in Fig. 3 and I considered a restricted range (15°C - 40°C) wrt Alvaro's thesis (15°C - 55°C), to have a better linear fit: Vout = 0.6*T - 19.5. The calibration is consistent to what was used before and according to Eric a possible problem of this circuit could be that the resistor R1 is heated by the current passing through it, increasing the overall temperature. This would not explain the discrepancy we were reading yesterday, but it could be a reason why the Vout is higher than what we expect.
I put back the termistor on the table and I taped it. I recorded the data with the cymac and in Fig. 4 the last hour of temperature data are shown on ndscope (checkign the X5:ISC-STATE_OBS4_INMON, sampled at 16Hz). The absolute value in counts can be converted in Volts and then in temperature with the following formulas:
Volts = counts * 40V / 2^16
T = (Vout + 19.512) / 0.6 ~ 25°C.
The temperature in the lab is around 20°C, meaning that the absolute value cannot be trusted. I will install some easier monitor system to check longterm temperature fluctuations. In the last hour the temperature changes are of the order of 0.12V which corresponds to a change in temperature of 0.2°C, which makes more sense wrt last time. Maybe the problem before was in the acquisition of data with the matlab script or some wrong connection in the circuit during the last data acquisition time.
I threw together a couple of general purpose broadband InGaAs fiber coupled PDs for GRAVITES. PD should have ~250 MHz BW (still need to take a TF). Measured the dark noise and rough gain. Looks like it hits a minimum around 80 MHz. Corresponds to ~3 pA/rtHz, consistant with the spec for the transimpedance IC at 100 MHz.
Currently we only have one HP4395A in NW17, so I've relocated the instrument cart from the purple lab to the green lab temporarily while I test some circuits.
I set up a new reflow oven in the EE Shop. It is venting into the fume hood for now (we cleared it out completely first). I've purchased a HEPA unit for the exhaust, per Brian Surette's recommendation, and plan to set it up on one of the tables once we clear a bit of space. I slightly overcooked the first couple of photodiode circuits (still seem to work fine), but soon got the sequence parameters dialed in. The demod board in the attached photo actually looks pretty good.
Maya used this to great effect for the new RetroSense circuits. The solder paste is ChipQuik TS391LT Sn42/Bi57.6/Ag0.4. I bought a 50 g jar.
Readout setup shifted from mirror B (power at readout 0.02mW) to mirror C (power at readout 4.5mW)
Readout PD (new focus 1811) spectrum at low power:
Shifting to higher power can some what hide the features, but it still pushes noise above shot noise
Spectrum of PDH reflection PD (PDA10CS) when cavity is unlocked, no signal to PZT:
The main differences from v1:
The chosen quad op-amp (OPA4182IDR) has 6 nV/Hz1/2 of input voltage noise and 0.17 pA/Hz1/2 of current noise.
With the choice of 2.4 kΩ transimpedance resistors, and with 1 mA of total photocurrent, the total output noise of the transimpedance stage (i.e., if all four outputs were summed together) comprises 43 nV/Hz1/2 from shot noise, 13 nV/Hz1/2 from Johnson noise, 12 nV/Hz1/2 from op-amp voltage noise, and less than 1 nV/Hz1/2 from op-amp current noise. Since the sum/difference amplifier and the differential amplifier are both unity-gain right now, this compares unfavorably to the >100 nV/Hz1/2 from the AD630.
If we stick with 1 mA of photocurrent, then it is probably advantageous to increase the transimpedance gains to something like 5 or 6 kΩ.
A comment on modulated spectra — even in the idealized case, the spectrum of the square-wave-modulated photocurrent has a noise floor that is clearly above the shot noise floor. The shot-noise-limited performance is recovered after demodulation.
Suggestions for a v3:
I wrote a Matlab script analyze_gracedb_superevents.m (tested on r2023b) that fetches new super events listed in the Gravitational-Wave Candidate Event Database (GraceDB) and creates a cumulative event plot. This code is based on the cumulative_event_plot_O3.m code (see LIGO-G2001862). The code provided here, however, can be run periodically to update the cumulative event plot (O4 events are not hardcoded). There is also the added ability to apply flags related to category (Production vs. MDC), Event Group (Burst vs. CBC), and Detection Pipeline (pycbc, cWb, MBTA, MBTAOnline, gstlal, oLIB, spiir). Superevent definition is based on the alerts-threshold-trials-factor, which sets a false alarm rate less than 6.43x1e-8 Hz (1 in 6 months). This corresponds to the expected effective rate of significant false alarms of 3.9*1e-7 Hz (one per month) for CBC sources. Currently, I include the 2023 engineering run events within the O4 period.
It also fetches the bayestar fits file for these events and generates a distance distribution plot. Interestingly, we had a detection at ~8.7 Gpc!
code: https://git.ligo.org/nikhil-mukund/gw-superevents