Measured, essential for calibration calculations.
[Masaya] We have 2 kinds of beam splitters that might be able to be used for our Heisenberg IFO experiment. I tested the transmitted and reflected lights of these two. The insident light is s-pol(vertical) assuming that these are used in Mach-Zehnder Interferometer. The result is as the below. 1. Designed 99% reflection, 5-20 deg incident angle, superpolished, 2-inch beam splitter Reflected ratio: 99.1 % Transmitted ratio: 0.1 % Ghost: 0.6 % Other loss: 0.2 % 2. Designed 99% reflection, 45 deg incident angle, not superpolished, 1-inch beam splitter Reflected ratio: 98.8 % Transmitted ratio: 0.5 % Loss: 0.7 %
Masaya successfully aligned a pair of AOMs to generate the 3MHz frequency shift for the CLF for the second OPO. This shift was measured by mixing the CLF beam on a beam splitter with light from the same laser that did not pass through the AOMs. As can be seen in the figure, there are sidebands of the 3MHz light offset from the 3MHz signal by about +- 575kHz. These have amplitudes about 40dB below the CLF 3MHz frequency, so they shouldn't be a problem, but we aren't sure where they're coming from. Note: Masaya did the optics alignment and made the measurements. I'm just posing this here for the sake of documentation and in case anyone who sees this has an idea about the origin of the 575kHz sidebands.
Thank you for posting!
[Eleonora, Eric]
We adapted the model on Simulink to the new configuration of the system.
PID SR560
Zeros: []
Poles: 2*pi*[3e3]
Gain: 1500000 (*)
VCO Marconi
Zeros: []
Poles: -2*pi*[0]
Gain: 2*pi*[1.5e3]
Previously, with the minicircuits VCO and the LB1005 we set these parameters to:
PID LB1005
Zeros: -2*pi*[3e3]
Poles: -2*pi*[0]
Gain: 0.0174
VCO minicircuits
Zeros: []
Poles: -2*pi*[0]
Gain: 2*pi*[4.028e6]
(*) We needed to set this gain very high, instead of the measured and expected gain = 100, in order to match the simulated OLTF with the measured OLTF.
We notice that the GAWBS was higher than the correction and error signals measured by a factor 5.
Two observations on this point:
1. Chris is also not confident in the model included in the Optica paper
2. the slope of this noise at high frequency matches perfectly our measurements (scaled by a factor 5).
In Fig. 1 and Fig.2 the error and correction signal plots without adjusting the GAWBS are shown respectively.
In Fig. 3 and Fig. 4 the error and correction signal plots dividing the GAWBS by factor of 5 are shown respectively.
Since the Marconi is very low noise, we already installed for the phase locked measurement and we already measured its noise, we decided to replace the SG384 as LO for the mixer with the Marconi 2026. The SG384 is still used to syncronize both the Marconis. The amplitude of the modulatio is set to 7dBm, as required by the mixer and I lowered the power going to the R port of the mixer by adjusting the power with the power controller in the optical system in order not to deform the error signal shape.
Today I measured the noise of SR560 as: Noise input = Spectrum / TF.
The measurement was performed by setting the SR560 to 3kHz LP, Gain = 100, 6dB/oct LP.
I disconnected the SR560 from the loop, put the 50Ohm terminations to the A and -B inputs and measured the spectrum from the output (after verifying that the input offset was set to zero). The spectrum is shown in V/rtHz in Fig.1 (on the model we input the noise in V/rtHz, but to convert in rad/rtHz one just needs to divide by 0.18 rad/V).
Then, I measured the TF by sending the swept sine signal to the SR560 channel A and the output sent to channel B on the SR785. The souce out was connected to channel A on the SR785. I sent a swept sine of 100mV amplitude from 1Hz to 100kHz. The TF is shown in Fig. 2.
Finally I divided the Spectrum by the TF and I get the noise in V/rtHz, shown in Fig.3. The noise of the SR560 is way lower than the LB1005.
[Eleonora, Slawek, Eric]
On Friday afternoon, we adjusted the settings of the two Marconis and synchronized them to be able to measure their noise phase-locking them.
First of all we synchronized them to the internal clock (10MHz) of the SG384 RF generator since it was the best signal: the first Marconi (2030) intenal clock was saturating at a lower frequency, maybe due to an aliasing problem; the second Marconi (2026) needs an external clock. We divided the SG384 with a power splitter and sent the clock signal to both instruments.
Then we set the Marconis to be symmetric in order to be able to extract the noise of each of them just by diving the spectrum by sqrt(2). We used a phase detector instead of a mixer since it requires the same input powers: we found a minicircuits ZRPD-1+ phase detector that wants 7dBm inputs. We set the Marconis to take the external modulation and the same FM deviation. We performed the noise measurement for both 2kHz and 100kHz FM deviations to check how the noise behave by increasing the actuation range.
1. FM deviation = 2kHz
SR560 settings: LP 100kHz, LP 6dB/oct, Gain = 10
The calibration factor, measured in the previous entry, is 1.48kHz / V.
2. FM deviation = 100kHz
SR560 settings: LP 100kHz, LP 6dB/oct, Gain = 1
The calibration factor in this case is 75kHz / V.
The comparison between the two configurations is shown in Fig. 1. The noise spectrum is the phase-lock loop correction signal. We can conclude that the noise increases with the range, in this case by a factor 50-100 between 2kHz and 100kHz. The noise is in agreement with what we expected from the datasheet. NB 100kHz is the max range for the Marconi.
Today I measured the Marconi's transfer function, which corresponds to the calibration factor from Volts to Hz. I sent a voltage provided by a DC power supply to the external modulation port. The input volage taken by the Marconi accordinag to the datasheet is: 1V RMS (1.414 V peak) sine wave for set deviation and maximum safe input is ±15 V. Fig.1 shows the frequency output of the Marconi measured with the frequency counter as a function of the input voltage. The FM deviation during the measurement was set to 2kHz.
The fit gives a calibration factor of 1.48kHz/V.
Similar measurements on Vienna Marconi 2026B gain vs. modulation deviation span:
| Span | Gain [Hz/v] |
| 500Hz | 710 |
| 1kHz | 1420 |
| 5kHz | 7100 |
| 50kHz | 71100 |
Haocun
[Eleonora, Masaya]
We moved the Marconi from the electronic shop to the NW22 lab, next to MZI setup with its trolley. We cleaned the trolley beforehand. The Marconi in NW22 is shown in the attached picture.
NB. it needs an external clock. I connected it to the other Marconi, but since it had a problematic clock, I will check more carefully tomorrow if that would be an issue.
[Eleonora, Slawek]
Today I changed the setup of the setup in NW22 since we were limited by the servo controller noise and the VCO noise. We replaced the VCO with a Marconi and the LB1005 with an SR560.
We synchronized the Marconi and the SG384 Standford generator (LO) with the internal clock of the Stanford since it was less noisy: the Marconi internal clock had a low frequency oscillation that was saturating at 10V. Then zooming in we saw the 10MHz oscillation. We decided to use the Stanford internal reference since it was just a 10MHz signal.
To lock the loop we set the SR560 to: Gain = 100, Low pass at 3kHz, DC coupled. The error signal is now way less noisy than before and it seems also more responsive on the oscilloscope. The actual electronics setup is shown in Fig. 1. We set the resolution of the Marconi to 2kHz/V.
The correction and error signal in loop are shown in Fig. 2 for this configuration. We see that we have more than a factor 1000 noise suppression now.
Next steps:
Today I measured the OLTF of the system under these conditions.
I added an SR560 with gain = 1 and with a DC configuration, in order to be able to measure the OLTF, before the servo controller in loop. I was taking the output before the new SR560 and the input afterwards. In this way ones bypasses the SR560, measuring the response only of the elements in loop.
I sent a swept sine of 10mV amplitude, 100-100kHz frequency, to the channel -B of the new SR560. I connected the error signal of the loop (output of the mixer) to the channel A of the SR560 and I sent the same channel to SR785 channel B. The SR560 output signal was sent to the SR7585 channel A.
The OLTF is shown in Fig.1 and it gives us an UGF of 7.58kHz.
The Marconi 2030 amplitude power is set to 5dBm, since the amplifier has a 24dB gain and the AOM wants max 29.5dBm RF power.
Attached is a plot showing the p-s splitting of the cavity moving with (very) small changes in one steering and one cavity mirror horizontal angles. Need to do more careful calibration of range for which this tuning scheme could be used.
[Eleonora, Eric]
I summarize here the results of the simulink model interfaced with the measured noises of the system.
Fig. 1: Simulink model of the system / control loop. I underline the noise added at each port (NB. we did not add yet seismic and acoustic noises).
Fig. 2: Error signal closed loop. In black we have the error signal closed loop measured in our system and we see that we are limitedby the PID servo noise.
Fig. 3: Correction signal. In black we have the correction signal we measured in our system. The VCO noise is limiting our system and this is why we cannot suppress more noise. Above 10kHz we might be limited by the PID servo noise.
Since the signal is limited by the servo controller and the actuator (VCO) we can replace these elements with lower noise ones and perform the same measurement in order to understand what would limit us instead. According to the simulation, we should be limited by laser frequency noise.
We are planning to substitute the VCO with a Marconi and the LB1005 with a SR560 and eventually a low pass filter.
Today I did some measurements on the PID servo controller LB1005, I re-measured the OLTF, correction and error signal of the loop with the actual configuration of the system.
I put 50Ohm to A and -B inputs on the LB1005 and I set to zero the input offset. I set to zero the PI corner and the LF gain limit and I set to the maximum the gain. I acquired the spectrum and compared it with the one of last week where I had a low gain and PI corner 3kHz and LF gain limit 60dB. Fig. 1 show the two spectra. I measured the TF of the PID in this configuration as well and it is shown in Fig. 2. The magnitude is flat and it is around 37dB and the LB1005 datasheet says that the gain goes between +40dB and -40dB, i.e. this makes sense. For the measurement of last week the TF was from -20dB and -50dB. I computed the PID noise dividing the PID spectrum by the PID TF. Fig. 3 show the comparison between the PID noise with max gain (blue) and low gain (orange): we notice that the curve in blue is 3 orders of magnitude lower than the curve in orange which corresponds to the actual status of the system. This means that the fact that we operate the system with very low gain makes the PID noise a dominant source of noise.
The DC power at the PD has been set to 160mV. The demodulated signal has a Vpkpk = 360mV (A= 0.18 V/rad). I closed the loop setting the PID servo with PI corner = 3kHz, LF gain limit = 90dB and low gain (checking that it was optimized, because increasing it I was unlocking the loop). Fig. 4 shows the OLTF is shown: we have an UGF of 26.4kHz. The error (in blue) and correction (in orange) signals measured closed loop are shown in Fig. 5. The suppression factor is 100 at 10Hz.
I measured the noise of the LB1005 again in order to have a good noise measurement to include in the Simulink model. The PID TF measured down to 1Hz is shown in Fig. 6 and it is compared with the PID TF derived from the OLTF as PID(s) = OLTF(s) / [ P(s) * S(s) * A(s) ] as shown in the orange curve in Fig. 7. This time the two curves are in agreement. I measured the spectrum of the PID which is shown in V/rtHz in Fig. 8 (since I need to include this noise in V/rtHz in the Simulink model considered where it is added). In Fig. 9 the PID noise spectrum in V/rtHz is shown.
I derived the PID TF, both magnitude and phase from the OLTF of the whole system as:
freq_223 = np.array(freq_223)
mag_223 = np.array(mag_223)
phase_224 = np.array(phase_224)
OLTF_complex = 10**(mag_223/20)*np.exp(1j*np.pi*phase_224/180)
c = 3e8
L = 199.8
n = 1.44
tau = n * L / c
P = np.exp(-1j*2*np.pi*freq_223*tau)
A = 4.028e6 / (-1j*freq_223)
S = 0.18
F = OLTF_complex / (P*A*S)
F = np.array(F)
magnitudes = np.abs(F)
phases = [cmath.phase(element) for element in F]
phases_degrees = [phase * 180 / cmath.pi for phase in phases]
where freq_223, mag_223 and phase_224 are the OLTF data relative to the magnitude and phase. As Eric suggested, I considered the Plant TF = exp(-1j*2*np.pi*freq_223*tau) and I took into account a difference in phase also in the VCO TF = 4.028e6 / (-1j*freq_223).
The comparison of the Mag and Phase TF derived in this way (orange) what I measured yesterday (blue) is shown in Fig. 1.
The main differences from v1:
The chosen quad op-amp (OPA4182IDR) has 6 nV/Hz1/2 of input voltage noise and 0.17 pA/Hz1/2 of current noise.
With the choice of 2.4 kΩ transimpedance resistors, and with 1 mA of total photocurrent, the total output noise of the transimpedance stage (i.e., if all four outputs were summed together) comprises 43 nV/Hz1/2 from shot noise, 13 nV/Hz1/2 from Johnson noise, 12 nV/Hz1/2 from op-amp voltage noise, and less than 1 nV/Hz1/2 from op-amp current noise. Since the sum/difference amplifier and the differential amplifier are both unity-gain right now, this compares unfavorably to the >100 nV/Hz1/2 from the AD630.
If we stick with 1 mA of photocurrent, then it is probably advantageous to increase the transimpedance gains to something like 5 or 6 kΩ.
A comment on modulated spectra — even in the idealized case, the spectrum of the square-wave-modulated photocurrent has a noise floor that is clearly above the shot noise floor. The shot-noise-limited performance is recovered after demodulation.
Suggestions for a v3: