IntegratedPhotonics Chips
melissa - posted 12:27, Wednesday 30 September 2026 (12267)
A note on polishing InGaP squeezers

Today Daniil polished a second chip from the summer InGaP squeezer fab run [ Chip ID: SQA OPA PhM 3 ]
Writing this post to consolidate the polishing specifications for future reference

The attached image shows the polish amount on the narrow tapers which allows 20-30% coupling (780 nm laser) to the bus waveguide from free space. 

Left-side polish goal: center of the last short alignment bar. Historically 15um polish works here but look at the chip edge before making the call.

Right-side polish goal: Anywhere between the last two long vertical bars. Bars have periodicity 5um, GDS screenshot attached. Same note as above, 15um can work, but check the chip.

Images attached to this report
IntegratedPhotonics Chips
melissa - posted 18:36, Tuesday 29 September 2026 - last comment - 11:21, Monday 05 October 2026(12266)
InGaP squeezer microring chips: characterizing the summer 2026 fabrication run

[chaejin, melissa]

Chip ID: SQA PhM 3

PhD student Lilli Thiel from UCSB sent us InGaP microrings designed to produce degenerate squeezed vacuum at 1560 nm when pumped at 780 nm. She fabricated the chips in the UCSB cleanroom. Her intuition is that the fab is botched, no definitive evidence yet. We are measuring the quality factors of the microrings at 1560nm to get the answer.

A plot of measured Q pairs {Qi, Qc} is attached, each the per device median. Unfortunately the gap (300 nm, point coupled) is not swept, so we are unable to definitively identify Qi (intrinsic Q) vs Qc (coupling Q). All devices should have similar coupling regions, so Qc should be roughly uniform across the chip, though both Qs appear to vary. The relation is 1/Q = 1/Qc + 1/Qi where Q is the loaded Q. 

Device ID Ring Width [um] Ring Radius [um]
13 1.02 32.1
23 1.06 31.5
54 1.1 32
55 1.1 32
67 1.14 31.5
76 1.14 32
79 1.14 32.1
86 1.14 32.5

One hint that the lower Q group may be Qc is wavelength dependence: the Q decreases at longer wavelength (negative mode numbers), typical for point coupling (see Sapphire microring example plot, where Qc was confidently identified from the gap sweep). Example summary plot for an InGaP device is attached, where dark blue would be suspected Qc and light blue would be Qi (flip the labels). 

It is mysterious, though, that the Qc would vary so much across the chip (60-130k) for such similar geometry. This would point to large spatial discrepancies in the fabrication, leading to geometric scaling: was the resist reflow non-uniform (temperature non-uniformities)? was the deep UV stepper conditions non-uniform? 

Also mysterious: Devices 54 and 55 are directly next to each other and have fully identical geometric properties (ring width and ring radius) but Qs are different: {80k, 275k} vs {100k, 260k}. This would suggest that the lower Qs are intrinsic, because random local scatterers would substantially change Qi and not Qc.

One more note: Our 3 dB squeezer chip had Qc = 80k and Qi = 400k (escape efficiency 83%), for gap = 350 nm, pulley angle =  5 degrees, ring width = 1.125 um, ring radius = 32 um. 

Daniil will polish another chip for us from this batch, with a larger coupling gap (400 nm); we will compare the Qs and may perhaps put this question to rest. We are sending Lilli the chip we measured today, for her to take 780nm Q measurements using their mode-hop-free laser. 

Images attached to this report
Comments related to this report
melissa - 11:21, Monday 05 October 2026 (12268)

Chip ID: SQA OPA PhM 3

We measured the chip with larger coupling gap of 400 nm and found Qi,Qc both around 300k, an indication that the fabrication was successful and that the low Qs on PhM3 are in fact the overcoupled Qc -- i.e., Qi are sufficiently high for squeezer operation. A plot of the {Qi,Qc} pairs vs ring width is attached. It is not clear what is the origin of the Qc variation across the chip, which we saw on the other chip as well. Does ring width substantially modify the rate despite identical gap? Or are the fabrication conditions non-uniform spatially?

Images attached to this comment
IntegratedPhotonics Chips
melissa - posted 13:08, Friday 25 September 2026 (12265)
Sapphire microrings: OPO, simulated dispersion, soliton crystal generation

[chaejin, melissa, daniil]

Numeric solver
We ran Tidy3D's FDFD eigenmode solver for a bent waveguide with our material and geometric properties (roughly assumed 10 degree sidewall angle from SEMs, example attached). We take the sapphire material dispersion to be Malitson's anisotropic dataset and the SiO2 to be Palik's (lossless) [E. D. Palik. Handbook of Optical Constants of Solids, Academic Press (1998)] -- the latter being the PIC community "default" for SiO2. We will perform ellipsometry on our wafers to have more accurate material dispersion curves. 
The simulated D1 and D2 values for TE0 and TE1 agree quite well (example plot attached):

  TE0 D1 [GHz] TE0 D2 [MHz] TE1 D1 [GHz] TE1 D2 [MHz]
Simulated 110.9 0.49 108.7 1.15
Measured 110.8 0.48 108.6 1.15

OPO threshold power
In the previous logbook post we had predicted an OPO threshold power of 2-3 mW in the waveguide. On Device 19 we observe OPO at measured ~2.2 mW: we directly measure 7 mW in free space before the chip-coupling lens and we estimate 32% coupling efficiency (calibrated using lensed fiber with coupling efficiency 66%). 


Soliton crystal generation Using our "new" Agilent 86142A OSA, we identify frequency comb states by their spectral features. On Device 18, we increase the power and generate a soliton crystal microcomb [photo attached]. Resonator FSR is ~0.9 nm, the teeth are spaced by ~9nm, so this would be something like an N=10 crystal: in the time domain, 10 equally spaced soliton pulses in the ring. We increase the power and observe chaotic states which would likely give way to bright solitons with proper thermal stabilization. The next round of devices will have heaters: we will use them to capture the soliton state, if the fab is successful.

Images attached to this report
IntegratedPhotonics Chips
melissa - posted 14:15, Thursday 24 September 2026 (12264)
Sapphire microring characterization for soliton generation

[chaejin, melissa, daniil]

We're characterizing new microring resonator chips fabricated by Daniil. The guiding layer is sapphire with SiO2 cladding, ring diameter 480 or 240 um (depending on device); ring waveguide width and gap (distance between the ring and the bus waveguide) are swept. Thickness of sapphire is 1um.

We want to know if these microrings could be used for soliton microcomb generation. Solitons are born from non-degenerate cascaded optical parametric oscillation (OPO) processes, so we need to be able to trigger OPO with our available laser power. The laser power requirement scales as V / (n2 * Q^2 ) where V is the mode volume, n2 is the Kerr refractive index (propto chi-3 nonlinearity), and Q is the loaded optical quality factor. To ensure phase matching for the OPO thresholds as the pump is tuned blue to red, we need anomalous dispersion of the pumped azimuthal mode family. Anomalous dispersion means that the FSR between modes increases with optical frequency. Exotic solitons may be generated for other dispersion profiles, though canonical solitons (bright pulses with intensity profile of sech^2 vs freq) are born from anomalous dispersion.

We scan our Toptica CTL from 1515 - 1625 nm and measure the microring transmission. We send a pickoff to a calibrated MZI (period ~20 MHz) to create a mode-hop-free frequency ruler for the scan. We've automated Q factor fitting, mode familiy sorting, dispersion fitting, etc -- example plots from the GUI attached. We can sort the resonance frequencies by mode number where mu=0 represents the center of the scan: f[mu] = f0 + D1 * mu + D2/2 * mu^2 + [higher order terms]. D1 is FSR. Anomalous dispersion: D2 > 0. AMX stands for avoided mode crossing -- a large one is bad for comb generation. Below, Q is taken to be median value where [M] is million. Qi is associated with intrinsic loss (scattering, absorption), whereas Qc is associated with coupling rate to the waveguide. Example transmission scan and dispersion/Q plots attached for device 18. 

For the 480um diameter rings (two mode families, likely TE00 and TE10, but will need to match against Tidy3D numeric solver prediction):

ID Width [um] Gap [um] D1 [GHz] D2 [MHz] Qi [M] Qc [M] D1 [GHz] D2 [MHz] Qi [M] Qc [M] Comment
4 1.5 1.1 109.9 0.07 12 5          
5 1.5 1.3 109.9 -0.05 16 14         AMX
10 2 0.9 110.4 0.42 15 3          
11 2 1.1 110.4 0.43 16 11          
12 2 1.3 110.4 0.46 24 25          
13 2 1.5 110.4 0.47 23 82          
17 2.5 0.9 110.8 0.48 21 6 108.6 1.08 10 19  
18 2.5 1.1 110.8 0.48 21 19 108.5 1.13 11 56  
19 2.5 1.3 110.8 0.5 32 46 108.5 1.16 13 151  
20 2.5 1.5 110.9 0.44 26 148          
24 3.5 0.9 111.6 0.29 26 15 110.1 1.7 14 3 AMX
25 3.5 1.1 111.6 0.23 33 35 110.1 1.63 13 10 AMX
26 3.5 1.3 111.6 0.25 38 140 110.1 1.65 18 30 AMX

Takeaways:

For this coupling geometry, devices with width 2.5 um are the most suitable for comb generation: loaded Q ~ 15 M near critical coupling and anomalous dispersion. We would predict an OPO threshold at 2-3mW of pump power in the bus waveguide. Will measure this directly and compare. 

Images attached to this report
LASTI General
bruno - posted 09:54, Tuesday 22 September 2026 (12263)
Cobri Splitter passes RGA, SENSIT
The 1-4 1550nm splitter cleaning process was more complicated than we thought. 

P/N: PM-PLC-B-1X4
TYPE: PM1550 250µm bare
Connector FC/APC
S/N: 10260717060001

Inside the splitter housing are a few different epoxies which seem difficult to remove. The white epoxy is likely a low outgassing epoxy NTT OS5958.
The black rubber alongside the ends of this splitter remained ambiguous so it was carefully removed with a scalpel and tweezers. 

The RGA results show a HC outgassing rate of 5e-11 Torr-liter/s which comfortably passes under the 4e-10 Torr-liter/s threshold. 

Attached are the RGA results and some photos of the splitter. 
Images attached to this report
CDS General
eoelker - posted 11:03, Friday 18 September 2026 (12262)
CYMAC moved from Yellow lab to Highbay

We're going to use it with the coupled cavity test setup for a couple months.  Will start by taking seismic data on the HAM-Y stack.

 

Not permanent, just until I get our new front end and have time to set everything up.

 

Let me know if you need to borrow it.

CoupledCavity General
mitchell - posted 18:45, Thursday 17 September 2026 (12261)
Test Cavity Locked
Mitchell R, Matthew T

After connecting the PZT driver to the laser PZT, we were able to lock the cavity. All in all we could get it to stay locked for about 5 minutes, which meant we could measure the open loop gain on the SR785. We took a measurement at 20dB of gain on the controller and using the Boost 1 filter. The UGF is at about 46kHz with a phase margin of 150 degrees.
Images attached to this report
CoupledCavity General
mrtodd - posted 09:47, Wednesday 16 September 2026 - last comment - 13:13, Thursday 17 September 2026(12257)
Test Cavity Work

M. Richardson, M. Todd

Mitchell spent some time yesterday aligning the newly input 21.5MHz EOM, and did some cavity alignment to recover the TRANS peaks. Today we wanted to look at the error signal. We are using +7dBm for the EOM and mixer LO drive, and already seeing a massive increase in modulation depth relative to the 532nm EOM (as expected). We did some beam walking to minimize the misalignment peaks, and took a trace of the error signal while scanning. We had to change the LO phase a bit but we were able to get 3FSRs in this trace. Roughly with +7dBm, we are now seeing 15mVpp through the carrier resonance.

 

Analysis of the data will follow in the comments.

---------------  Measurements --------------

RFPD Power: 1.1mW

RFPD DC voltage: 970mV

RFPD DC "Gain" [V/W]: 0.880 V/mW

Comments related to this report
mitchell - 09:53, Wednesday 16 September 2026 (12258)
Here are the saved traces from the oscilloscope.
Non-image files attached to this comment
mrtodd - 13:13, Thursday 17 September 2026 (12260)

I temporarily swapped the 7dBm mixer for the 17dBm mixer and turned up the EOM drive correspondingly. I walked the beam alignment a bit in both pitch and yaw and was able to lower the misalignment peaks quite significantly. Retaking the data from the scope (using the script on the iMac) I was able to get a pretty good fit to our parameters. Plugging the printer output (back) of the oscilloscope into the usb of iMac allows users to grab the data that's on the screen. Just run

cd git
python capture_all_traces.py

As of right now our settings are such:

  • ~1.2 mW of power on the RFPD
  • 17dBm of RF power to EOM and LO port of mixer
  • 16Vpp for PZT scan @ 100Hz
  • 13.3MHz f_mod, even though the RFPD is tuned for 18MHz.

Knowing the modulation frequency well (13.3MHz) you are able to get a very good calibration of the true FSR and corresponding PZT gain (MHz/V). From there you can fit the other parameters, with the uncertainty arising from the multiple FSRs which will not exactly line up with each other.

  1. Cavity Length: 7.4704m +/- 0.9mm
  2. FSR: 20.065MHz +/- 3kHz
  3. TMS: 4.3 +/- 0.1 MHz
  4. Finesse: 15.03 +/- 0.07
  5. Pole: 667kHz +/- 3kHz
  6. Omega (fmod): 13.3 MHz
  7. Mod. Depth. 0.259rad +/- 1mrad
  8. Demod Phase: 9.2 +/- 0.8 deg

 

This morning I also found a cable to act as an adapter of the PZT driver output (SHV) to BNC of the laser PZT input. We have it all hooked up now, so we should be able to feedback to the PZT and try locking.

Non-image files attached to this comment
CoupledCavity General
mitchell - posted 19:15, Tuesday 15 September 2026 (12256)
Power Meter Issue and Replaced EOM
Mitchell R.

Swapped out the EOM with one for the correct wavelength which required fine tuning input alignment to the cavity again. In the process of trying to measure the error signal with the new EOM, I discovered a problem with the power meter we have been using. The "500mw" and "5mW" settings on the sensor were giving different readings. Comparing measurements from a second power meter and sensor showed that the "5mw" setting on ours is wrong for some reason, but the "500mw" agreed. The comparison also involved using our sensor connected to the second meter and vice versa to distinguish between an issue in the sensor vs an issue in the power meter. It appears the sensor has the issue, not the power meter itself.
OEO General
dmdesant - posted 17:52, Monday 14 September 2026 (12255)
LWE-120 NPRO Cavity Response Fit

We were able to fit the cavity response data we took this past week and estimated the output coupler reflectivity and intracavity loss. The rough fit is shown here.

Out-Coupler Reflectivity 97.9%
Intracavity Loss 0.30%
Finesse 42
FWHM 357 MHz
Storage Time 2.8 ns
Injection Efficiency 87.5%

 

Images attached to this report
CoupledCavity General
mrtodd - posted 12:04, Friday 11 September 2026 (12253)
Test Cavity Alignment Work

M. Richardson, M. Todd

Yesterday morning I tested the top phase shifter chassis on the rack with our LO. It seemed to only function when the 2ns toggle was enabled, and then you could use the other switches as desired. In powering the chassis I had to cannibalize the power cable for the LSC PD chassis as there were no others. After moving the cable back to the LSC PD chassis, Mitchell and I were able to generate a pretty good PDH error signal after a little alignment. We are sending 24dBm to the EOM and to the mixer (we may want to attenuate this a bit) and got this error signal.

We then hooked up the error signal to the servo board and tried adjusting the gain and offset from the computer. We were able to see this adjustment well by looking at the scope. We then disabled the feedback and connected the servo output directly the laser PZT. Adjusting the offset a bit and turning up the gain, and then enabling the feedback and boost stage we were able to see some very weak locking. It could hold the TRANS power for a second or so, as seen in the videos.

We will try and touch up the alignment a bit as well as servo settings to see if we can get any better locks before closing the door.

Images attached to this report
Non-image files attached to this report
CoupledCavity General
mrtodd - posted 18:24, Wednesday 09 September 2026 (12252)
Test Cavity Alignment Work

M. Richardson, M. Todd

We did a couple of tasks related to the alignment of the test cavity in efforts towards getting an error signal. We were not seeing very strong dips at REFL even with good peaks at transmission, so we tried to diagnose why that could be. Our first thought was that we needed to align the OC so that the cavity axis was set so that the promptly reflected and cavity REFL beam could overlap well at the same time as being aligned to the cavity axis. We also tried to remedy some of the clipping we were observing on the output path (looking at the TRANS camera), and were more successful with this. We still have some aperture clipping somewhere in the output path it seems but I think it is good enough for our purposes.

This did not really remedy the lack of REFL dips we were seeing, and after awhile of trying to understand what was going on, we think it could be that the expected REFL dips are just so small we weren't looking on the right scale. I think before we were tricking ourselves with larger REFL dips when we are clipping on the PD as the dips to the HG10 are similar magnitude at REFL (while much smaller at TRANS). We will try putting a little bit more power on the RFPD and zooming in to see if we get kind of dip while being well centered and also looking to see if we can get an error signal.

Images attached to this report
Non-image files attached to this report
CoupledCavity Electronics
eoelker - posted 15:26, Wednesday 09 September 2026 - last comment - 16:00, Wednesday 09 September 2026(12250)
Servo for Y-Arm Test cavity

We're using one of my old LabJack servo boards for the test cavity on the Y-arm

 

I set it up yesterday:

1.  LabJack EPICS IOC is running on quadrature.   If it isn't working, try restarting the Daemon:  

2.  There is an MEDM screen here that you can use to control the board:

To acquire lock, sweep the output offset adjust until you go through resonance on the arm cavity, then turn on the "Feedback EN" switch.  Once it grabs lock, turn on the two boost filters.  The sign might need to change.

 

Took TFs of each stage.  I think that the boost filters are just about right for a loop bandwidth of ~30 kHz, but we'll probably need to re-tune the lazer PZT notch filters:

Note: the output of the servo board is connected to the front BNC labeled OPO_SLOW output.  

Needs a bit of troubleshooting:

  1. Variable gain stage is giving the correct gain vs setting behavior except at -10, which it is setting to +30 dB for some reason.  Likely a software bug.
Comments related to this report
eoelker - 16:00, Wednesday 09 September 2026 (12251)

Schematics for the electronics in that box.  The Boost and Filter stages have been stuffed differently than what is shown in the default schematic.  

Non-image files attached to this comment
OEO General
dmdesant - posted 09:48, Wednesday 09 September 2026 (12249)
LWE-120 NPRO Cavity Flashing

Last week, we were able to mode-match a 1064-nm beam from the NP Photonics "Rock" fiber laser into the crystal mounted in its original enclosure (see photo IR attached). In order to accomplish this, the pump diode and pump input optics present on the input path to the NPRO cavity were removed and a 1064-nm beam path is prepared externally.

Mode Matching

The beam parameters for this arrangement assumed an initial approximate eigenmode defined by a cavity with length L = 11mm and an input coupler ROC of 20mm. The computed beam parameters from this geometry are:

Parameter Value
Waist location (inside cystal) 5.5mm
Lasing mode waist @ 1064nm (inside cavity) 41um
Pump waist @ 808nm (inside cavity) 36um
Waist location (outide cavity - image inside cavity)

8.1mm

Lasing mode waist @ 1064nm (outside cavity) 43um
Pump waist @ 808nm (outside cavity) 38um

With minimal adjustment of the mode-matching solution (using thorlabs triplet collimator TC06APC-1064 in alamode), fairly decent visual mode matching can be observed in reflection of the cavity. In this case, the NPRO cavity is very slowly scanned through one FSR using the heater in the laser package. PZT tuning of the "Rock" laser does not provide enough tuning range to observe a full FSR (~13 GHz) unfortunately.

Pump Thermal Lensing

Due to pump absorption, the thermo-refractive effect will slightly lens the lasing mode of the laser. Some back-of-the-envelope calculations (based on a simple thermal model) show that the expected lensing is quite minimal (i.e. the effective focal length of the thermal lens is ~10x larger than the cavity's length) when the laser is putting out 5mW of power at 1064nm while being pumped with 30mW of 808nm light. 

Polarization Matching

Because of the instrinsic polarization rotation that occurs in the NPRO cavity geometry, the eigenmodes of the cavity are not generally linear. In fact, they are typically eliptical (see Kane and Byer attached). In this case, the polarization from the fiber is matched fairly well with a simple half/quarter-waveplate pair. The exact polarization seems to be only slightly eliptical. Without exact details of the output coupler coating and cavity geometry, calculating the eigenpolarizations with Jones calculus has been a guessing game so far.

Next Steps

At this stage, the aligned seed beam through the cavity should be an excellent alignment reference for the pump beam once the parts for it arrive. At that point, if the laser is lasing, we can figure out exactly what the fundamental cavity mode should be by profiling the laser's output beam. In any case, this is an excellent confirmation that we can indeed couple light from an external 1064nm path into the laser cavity, which will be tantamount for injecting squeezed light.

Images attached to this report
Non-image files attached to this report
LASTI General
bruno - posted 12:17, Monday 31 August 2026 - last comment - 17:24, Monday 14 September 2026(12239)
NEW LPS SQS 1-8 SPLITTER MEASUREMENTS
Vermeer took some power measurements of the new 1-8 splitter sent from Birmingham. 
Seems to work well enough. Attached are the measurements. 
Images attached to this report
Comments related to this report
dennis.martynov - 17:24, Monday 14 September 2026 (12254)

Very nice, thank you, Bruno and Vermeer. Would you please set this splitter in the tank for the sensors?

CDS General
Rich - posted 16:08, Wednesday 22 July 2026 - last comment - 14:40, Tuesday 06 October 2026(12211)
random linux/cds bits that i tend to forget

 

from the window:

Simulink files /opt/rtcds/userapps/trunck/SYSTEM/m1/models  where SYSTEM can be isi, sus ............

Overview medm screens  /opt/rtcds/userapps/trunk/SYTEM/m1/medm

Other medm screens /opt/rtcds/mit/m1/medm

CDS Parameters : /home/controls/shared/src/advligorts/src/epics/simLink/lib

 CDS Parts  /home/Controls/lib

 

To run a newly build model:

rtcds build

rtcds install

ssh m1boot

on m1boot: add model name to /diskless/root/etc/rtsystab

back on the other machine: rtcds start

sudo systemctl restart rts-local_dc

Then restart the framebuilder: ssh m1fb0

sudo reboot

 

To get the model into the DAQ/ ndscope/ diaggui:

1. Log into m1boot, change root user to su and navigate to puppert config file (should be /root/puppet/modules/ligo_cds_frontend)
2. grep for the FE name you want to deploy the model onto. For C6D and CRS models they were both on m1hamx, then it would be grep -rn "m1hamx"
3. This will return the location of the .yaml config file that puppet uses to populate the /etc/advligorts/systemd_env file you're looking for. Add the new user model to the models list in the .yaml file
4. Restart the puppet /opt/puppetlabs/bin/puppet apply --modulepath ~/puppet/modules ~/puppet/modules/ligo_cds_frontend/nodes/bootserver.pp

Warning: No server should need restarting to make the changes work. However, both times IGeorge added a new model, the boot server unmounted the controls user directory from all workstations at MIT. George strongly recomends to  that we make sure, either Fred is around, or at least you coordinate with hjim in advance when you restart the puppet, so he can remount all the user folders if needed. Otherwise you risk havin no access to a workstation until this issue is resolved (and we also won't be able to remote access MIT unti workstations have been fixed).

 

To restart DAQ (from m1fb0)

sudo systemctl restart rts-daqd

sudo systemctl restart rts-nds

 

 

For when the workstations lose control sub-directories

Either reboot; do a as root umount -f /home/controls/Desktop,

then a mount -a

Comments related to this report
arnaud - 11:15, Thursday 23 July 2026 (12212)

MIT CDS notes from various users (including this one) have been collected to this dcc page : T2500407

Regina - 14:40, Tuesday 06 October 2026 (12269)

Fred and Erik recently fixed some things on the CDS side to make it more robust. Previously the OS on m1boot was wiped. Fred is still fixing some keys but in the meantime if you ssh into any CDS machines and it gives you the error: "WARNING: REMOTE HOST IDENTIFICATION HAS CHANGED!" use

| ssh-keygen -R m1sus

replace m1sus with the machine you want to log into. Then ssh into the machine again.

Fred has temporarily changed the password to m1boot. Contact him if you need to access it.