I measured the powers circulating in the MZI.
Before 200 meters fiber = 117 uW
After 200 meters fiber = 52 uW
Before 20 cm fiber = 400 mW
After 20 cm fiber = 52 uW (there is a filter in front of the collimator ND 0.2 and I misaligned the collimator to have the same power at the end of both fibers).
I connected the DC channel of the PD 1811 to the oscilloscope and I have around 80 mV per arm, and 160 mV totally in this configuration. I measured with the powermeter the power in the BS output ports: 37 uW in the port going to the PD and 37uW in the other port.
The RF channel pk-pk in this configuration gives 600 mV.
I keep this power level for the next measurements since the error signal is more sinusoidal wrt before. The pk-pk amplitude of the demodulated (after the mixer) signal is 360mV, i.e. the sensor TF is 0.18 V/rad (instead of 0.25 V/rad used before).
We will use 2 Homodyne photodetectors in Heisenberg IFO. One of the two is currently used in cSQZ setup. I tested another detector if it has a peak at 3 MHz by measuring a transfer function. As you can see in the attached photos, there is a peak at 3 MHz as we expected.
The attached image shows the optimized mode matching calculation from the SHG, through an EOM and Faraday, and to the OPO. The calculation includes the measured ray-transfer matrices of the EOM and Faraday (see previous posts by Masaya). The solution has a focus inside the EOM, the lens "Lens 1" is a f = 38.1mm lens to refocus the beam incident on the Faraday, the lenses "Lens 2" and "Lens 3" are then f = -100 mm and f = 88.3 mm lenses to mode match the beam into the OPO. Lenses 2 and 3 will be placed on precision z-axis translation mounts to allow for fine-tuning of the mode matching into the OPO. To align the beam from the SHG to the OPO, we'll want to use mirrors HR2 and HR4, which are separated by 70 degrees of Gouy phase. The maximal beam size is about 1mm radius on Lens 3, which is getting large for a 6.3mm diameter optic, but should be small enough not to clip. If we have problems with clipping on this optic, we could consider using a 1 inch optic for Lens 3 instead of a 1/2 inch optic.
Measured beam parameters from the PDH extension board roughly match the theoretical mode here. I measured a waist location about 150mm from the final PDH mirror and a waist spot size of about 70um. Some of my measurements saturated the camera slightly, so the waist size estimate is likely an over-estimate. I believe the beam shape is good enough that I should see some signal when I try to couple light into the cavity, so for now I'm going to proceed and see if I can see any cavity flashes.
Today I measured the VCO phase noise by phase locking it to a RF generator, as Eric suggested. Since we understood that the VCO is the limiting noise at low frequency in the unbalanced MZI system, it was worth it to measure VCO noise properly. To do so we phase lock the VCO and then we know that the correction signal of this loop corresponds to the VCO noise (at least below few kHz) since it is the noisier component of the system. I connected the LO from the RF generator and the VCO output to a mixer. After the mixer the signal is low passed with a 1.9MHz LP filter. The signal then goes to the LB1005 (we wanted to use the SR560 but we need to send a DC voltage to the VCO in order to set its frequency to 71MHz) and then inputed in the VCO. The LB1005 settings are: PI corne 3KHz, LF gain limit 70dB. I locked the servo and then acquired the spectrum with the SR785 (up to 50kHz because then there was a random problem saving data). I calibrated the spectrum from Vrms/rtHz to rad/rtHz using the V -> Hz calibration measured in the entry 11380 and then Hz -> rad diving by the frequency vector. The VCO noise and the correction signal of the system are shown in Fig. 1. There is still some work to do to match the model and the system but we are getting closer.
Today I measured the VCO TF (calibration), to investigate possible discrepancies between the Smimulink simulation and the system. Before I estimated this parameter from the plot Frequency vs Voltage on the VCO datasheet (minicircuits ZOS 100+). I sent different known voltage values to the VCO using the TENMA DC power supply and I read the output frequencies with a frequency counter (53131A ) borrowed from a shelf in green lab. The plot is shown in Fig. 1 and we get Frequency_out [MHz] / Voltage_in [V] = 4.028 MHz / V (while before I was using 4.2 MHz / V, which is not so far).
[Masaya] I did Beam profile to figure out the optical path length of a new faraday isolator. As a result, the optical path length is extended by 0.87cm for x-axis and 0.97 cm for y-axis compared to the case where there is nothing.
New spectrum at polarimetry readout shows shot noise at frequencies of interest (at and around p-s splitting of 12 MHz) after upgrades. We can now proceed with calibration and measurement.
Note: Some RF amplitude modulation still persists at PDH sideband frequency. We minimized it by tweaking EOM crystal alignment, polarization into the EOM, and sideband frequency. However it still drifts back up after some time (temperature/voltage offsets?). To be monitored in case it becomes a problem, since it is currently dominating the transmission rms.
Problems with old laser (previous alog):
We swapped to new laser:
Modematching done for new laser mode, 60% coupling to cavity achieved.
Today, I measured the noise of the LB1005 servo controller, to include it in the Similink noise budget. I summarize here the steps, discussed with Eric.
I disconnected the PID from the loop and connected 50 Ohm terminations to both the PID inputs (A and -B). I connected the PID Out to the oscilloscope and tuned it to zero with the input signal set knob. Then I connected it to the spectrum analyzer (SR785) and I acquired the spectrum for the following settings: PI corner = 3kHz, LF gain limit = 60dB and Output set = 7.8V (needed to drive the VCO). Fig. 1 shows the spectrum calibrated in rad/rtHz ([V]/0.25).
I measured the TF of the servo controller sending the swept sine to channel A on the LB1005 (100Hz - 100kHz). The SR785 source out was also sent to the channel A of the SR785 as input point of the TF. The OUT of the servo was sent to the channel B of the SR785. During this measurement, the servo controller was set to LFGL, i.e. the P-I filter output is enabled, but the low frequency gain is limited according to the LF Gain Limit switch. (This mode is useful for acquiring lock and situations where limiting the DC gain is necessary), with same settings as during the spectrum measurement. Fig. 2 shows the magnitude and phase of this TF. (NB. the TF needs to be set in dB and not dB/rtHz. To remove the PSD units one needs to go to FFT mode and remove them and then back to swept sine mode).
The nosie of the servo is finally derived by: Noise PID = Spectrum PID / TF.
Since the TF is from 100Hz I needed to interpolate the data and then linearize the TF: TF_lin = 10^(TF_dB/20). Fig.3 shows the noise asd of the PID in rad/rtHz.
I compared the PID TF measured here with the PID TF derived from the open loop transfer function of the system (OLTF).
I know that the system is composed by Plant (MZI), Sensor (PD), Actuator (VCO), Filter (PID Servo). The OLTF = P(s) * S(s) * A(s) * F(s), therefore F(s) = OLTF / ( P(s) * S(s) * A(s) ).
From the output of the PD I can derive P(s) * S(s): V_out = A * sin( x[rad]) ~ A * x[rad], meaning that A is the TF P(s) * S(s) is V_out / x[out] = A, where A is the amplitude of the signal read by the oscilloscope.
A(s) is the TF of the VCO, whose output is in frequency and the input is in Volt : [Hz] / [V] = 4.2e6 Hz / V according to the datasheet. The calibration in radians is obtained diving by the frequency.
The PID TF is obtained as:
TF_PS_VCO_dB = 20*np.log10(TF_PS*TF_VCO)
TF_PID = OLTF_dB - TF_PS_VCO_dB
The resulting PID TF is shown in Fig.1, where the blue curve consider a OLTF with a LF gain limit = 90dB and UGF = 16.8kHz and the orange curve LF gain limit = 60dB and UGF = 18.6kHz (i.e. same configuration as the PID TF measurement of this week). The curve does not match exactly the measurement but maybe there is a small difference in the gain of the PID during the measurements. We wanted to verify that the low gain of the measured PID TF made sense and we can conclude that it does.
Masaya
To design the layout of optics towards PDH lock of SHG of cSQZ,we needed to figure out the optical path length. I measured beam size in the cases with and without EOM. As a result, the ray transfer matrix(ABCD matrix) of the EOM is
{{1,2.3cm},{0,1}}for x-axis, {{1,3.9cm},{0,1}}for y-axis
We're going to move the compact squeezer's electronics rack from the optics table to the floor to free up space. Here are some notes on the setup that should make re-assembly easier. (1) Photodetectors should be powered with the white cable on the three-pronged connector up and the red cable on the two-pronged connector on the left. (See attached picture) (2) PDH loop has 35dB (blue box) high-power RF amp between red pitaya's output and EOM. PDH loop RFPD has 23dB of amplification. (3) CLF path is fiber box -> Polarization controller -> Fiber to free space coupler.
Hudson, Jacques
As a "proof of concept" we managed to PDH lock the SHG.
We placed a RFPD (the one that was previously used for CLF OPO lock) on the reflected path of the 1064 from the SHG and sent a 2W (33dBm), 20MHz modulation signal to the free space EOM.
The code for the SHG dither lock was almost unchanged to lock the PDH, nevertheless the lock was still acquired. Below is an overview of the control scheme, the whole SHG lock being controlled by the SHG RedPitaya.
- The EOM signal is a 20 MHz, 0.18V signal generated by out1 of the RP.
- The error signal is derived from the PD RF port, going into in2 of the RP.
- The correction signal is made from the lock() function and sent to the piezo from out2.
- It was found that the error signal only appears when there is sufficient amplification on the EOM modulation signal. For this we connected the 0.18V out1 signal to a +35dB minicircuits amplifier. Note that the usual +20dBm amplifiers were not enough to provide sufficient SNR on the error signal.
See attached figures for the shapes of the error signals. Nothing has been optimized yet.
Conclusion: we can indeed PDH-lock the SHG in the current setup. For the future we should fine-tune the code for the lock acquision and gain optimization. We should keep in mind that the EOM and RFPD need amplifiers in order to get a non-negligible error signal, and that the small minicircuits amplifiers may not be enough.
Jacques and I rearranged some of the electronics on the CLF path to test a PDH lock for the SHG. Here, we'll document the electronics chain, so it can be reassembled later. The chain was RF Out from BBPD --> ZABP-16+ --> ZFL-1000LN+ --> ZFL-1000LN+ --> 50/50 power splitter --> Red Pitaya input. The ZFL's are 23dB amplifiers and the ZABP is a band-pass filter.
On Thu 10/05 I tested the free space EOM coupling that will be used to generate the sidebands of the SHG PDH lock. This is an 1064 EOM which is borrowed from the Purple lab.
I put the power of the main 1064 laser at ~ 11mW (measured at the exit of the blue fiber, before entering the polarization box), see figure 1 & 2, by mistuning the lambda/2 waveplate inside of the laser box. Note that I measured a max coupling of ~350 mW out of the blue fiber when the halfwave plate was well tuned.
I tuned the fiber couplers to obtain 7.4 mW (so ~70% of power matching) at the output of the fiber after the EOM, see figure 3.
I left the laser with this 11mW output power.
NOTE : This measurement was performed before adding the EOM to the 1064 incident path to the polarization box, although it should not have a big consequence on the green light's profile (but may induce some mode mismatch from the incident 1064 to the SHG cavity).
On Wed 10/04 I measured the beam profile of the GR generated from the SHG. The measurements were taken from the fraction of the GR which is recflected off the first DBS after the SHG.
Figures 1 and 2 show the beam profile with numerical fit for waist and M2, with and without the extra converging lens. The origin of the x scale is taken at the extremity of the csqz table, i.e. when the GR quits the table after being reflected from the DBS.
There is some ellipticity which is expected as the SHG crystal is birefringent.
Today I measured the RIN with the Thorlabs PD free space with the SR785.
The total power read by the PD was 8.7 V.
I acquired 9 spectrum with the SR785 ( 0-400Hz, 0-800Hz, 0-1.6kHz ... 0-102kHz), each of them containing 400 points, in Vrms/rtHz. I cut the second half of them (except the first one) and plot them together.
I normalized the spectrum by the DC value read on the oscilloscope V_DC = 8.7V. The RIN spectrum in [1/rtHz] is shown in Fig. 1 and it is equivalent to a spectrum in [rad/rtHz].
To obtain the RIN spectrum in dB/Hz (as usually shown on the datasheets) I did:
RIN[dB/Hz] = 20 * log10( RIN[1/rtHz] )
The RIN spectrum in [dB/Hz] is shown in Fig. 2.
At some point today the system disaligned and the beam coupling in fiber was completely lost. I recovered the alignment of the system and I set the same optimized power values as before.
Then I took the OLTF of the whole system, shown in Fig. 1. Both curves are taken choosing a PI corner = 3kHz and LF gain limit = 90dB. I changed the gain (which has a broken knob so the value is not known a priori): a lower gain (blue curve) gives an UGF = 1700 Hz, an higher gain (orange curve) gives an UGF = 2500 Hz. The latter shows an oscillation around 20-30kHz that could be due to a too high gain or to a lack of coherence.
In Fig.2 the OLTF acquired some days ago with similar settings PI corner = 1kHz (blue) and 3kHz (orange) and LF gain limit = 90dB. Here the UGF is higher wrt today: UGF(PI = 1kHz) = 8300 Hz, UGF(PI = 3kHz) = 4900 Hz. The gain was the same.
Today I also acquired the error and correction signal spectra in closed loop, but unfortunately after I turned off and on again the laser. I checked that the system was in the same configuration as beforehand.
In Fig. 3 the error signal (blue) and the correction signal (orange) closed loop for a PI corner = 3kHz are shown.
Then I acquired the TF for both PI corner of 3kHz (blue) and PI corner of 10kHz (orange) and they are shown in Fig. 5. Increasing the PI corner I get the weird behavior at high frequency as shown also before. The UGF in this case is 443 Hz for the blue curve and 949 Hz for the orange one.
In Fig.4, the corresponding error signal (blue) and the correction signal (orange) closed loop for a PI corner = 10kHz are shown (the lower resolution at low frequency is due to the fact that for some reason the SR785 did not save the last file of data).
The fact that the TF changes so much for different times of the measurements is weird and I am going to investigate it in the next days.
After speaking with Eric, I realized that one possible issue of this measurement was where the input for the TF was taken.
I was taking the output for the TF from the -B channel of the PID LB1005, same as the swept sine injection. I should indeed take it from the Error Monitor channel.
So in summary I need to connect :
In Fig.1 the OLTF of the system is shown for the following settings: PI corner = 3kHz, LF gain limit = 90dB and reasonable gain. The loop has an UGF = 16.8kHz, which makes more sense according to the correction and error signal closed loop I have been measuring.
The main differences from v1:
The chosen quad op-amp (OPA4182IDR) has 6 nV/Hz1/2 of input voltage noise and 0.17 pA/Hz1/2 of current noise.
With the choice of 2.4 kΩ transimpedance resistors, and with 1 mA of total photocurrent, the total output noise of the transimpedance stage (i.e., if all four outputs were summed together) comprises 43 nV/Hz1/2 from shot noise, 13 nV/Hz1/2 from Johnson noise, 12 nV/Hz1/2 from op-amp voltage noise, and less than 1 nV/Hz1/2 from op-amp current noise. Since the sum/difference amplifier and the differential amplifier are both unity-gain right now, this compares unfavorably to the >100 nV/Hz1/2 from the AD630.
If we stick with 1 mA of photocurrent, then it is probably advantageous to increase the transimpedance gains to something like 5 or 6 kΩ.
A comment on modulated spectra — even in the idealized case, the spectrum of the square-wave-modulated photocurrent has a noise floor that is clearly above the shot noise floor. The shot-noise-limited performance is recovered after demodulation.
Suggestions for a v3: