cSQZ General
masaya - posted 20:20, Thursday 07 December 2023 (11425)
Test of finesse of OPO
[Masaya]

I took alignment of the beam of 532 nm and 1064 nm to the OPO and assessed finesse.

The results were
17.5@532
36.3@1064

In the figures, blue lines show the power of the output of OPO and orange lines show the input signal to PZT.

These results are consistent with the previous results Eric showed in his thesis.
Images attached to this report
cSQZ General
masaya - posted 20:04, Thursday 07 December 2023 (11417)
Cleaning up Newport Broadband EOM
[Slawek, Masaya]

The output beam of an EOM (Newport 4002 for 532nm) was weird. Therefore Slawek and I opened the EOM and cleaned up the internal crystal. The surface of the crysetal was obviously dirty. 
When I saw the beam shape thorough a beam profiler, There were many points of light around the main beam due to defraction but they dissapeared after cleaning.
Images attached to this report
cSQZ General
jacques - posted 20:03, Thursday 07 December 2023 (11424)
PLL error signal

The PLL error signal between the two lasers (beatnote of Diabolo+ Lightwave) been observed on the spectrum analyser, by setting the temperatures of both lasers to a same value (35°C in this test) and adding a small DC voltage to the external laser crystal temperature control of the Diabolo (-0.58V). Below some photos of the time domain and frequency of the beatnote, seen around 111 MHz.

Images attached to this report
GRAVITES General
epolini - posted 20:29, Wednesday 06 December 2023 (11423)
SR560 noise measurement with the AG4395A

[Eleonora, Eric]

We installed the GPIB on the AG4395A in NW22.

We acquired the noise spectrum of the SR560 set in DC with G = 100 after debagging several coding and setting issues. 

Fig. 1 : spectrum of the SR560 acquired putting 50 Ohm terminators to both A and -B channels of the servo. In blue the spectrum acquired with the AG4395A compared with the spectrum acquired with the SR785 as a sanity check. 

Fig. 2: TF acquired with the AG94395A. One cannot go lower than 10Hz with this network analyzer. Power of the sweep sine = -50dBm. The TF was acquired with the servo out of loop. There is an internal LPF in the SR560 at hundreds of Hz.

Fig. 3 : Noise of the SR560 derived as: noise = spectrum / TF. Up to 100 Hz I used the noise measured before (see entry 11393). The difference is at high frequency since now the SR560 is used in a DC configuration and not with a LP.

Images attached to this report
testbedAI General
nikhil - posted 20:17, Tuesday 05 December 2023 (11421)
Finesse3 simulator for a mis-aligned Mach-Zehnder

[Nikhil, Kevin]

Finesse3 has a more intuitive syntax than Finesse2 and includes features like API support, beam propagation, and more. In our case, we wanted our beam to be deliberately off-axis and wanted to match the testbedAI setup as closely as possible. Since finesse uses paraxial approximation, we need to include higher-order modes (n+m, represented as max_tem) to capture what we see in real. The number of modes required depends on the waist position and relative distances of various components. We cannot capture the total beam energy with low max_tem, while larger values slow down the simulation quite a bit. Setting the higher-order modes to a value of 20 provides reasonable results, as can be seen from the attached plot. This simulator will be used to augment the experience buffer of the RL agent that is currently getting trained for testbedAI.

 

# Finesse3 Install Intructions(with jupyter kernel)
conda activate base
conda install nb_conda_kernels
conda create -n finesse python=3.10
conda activate finesse
conda install ipykernel
conda activate base
pip install --upgrade nbconvert
conda activate finesse
conda install -c conda-forge finesse jupyter
conda activate base
jupyter notebook
# then change kernel to finesse inside the notebook (check top right corner)
 
It is also easy to execute (minimal overhead) Finesse3 from MATLAB (ex: r2023b) via pyrun through InProcess code execution. In case someone is interested, the initial bugs can be resolved as given below.
% Resolve GLIBCXX error
REF: https://www.mathworks.com/matlabcentral/answers/1934285-why-do-i-receive-a-glibcxx-not-found-or-unable-to-resolve-the-name-error-when-using-a-python-mod
 
% FIX finesse3 scipy libstdc++ loc
 first get finesse libstdc++ location using ldd
 via
 ldd /home/mitligo/miniconda3/envs/finesse/lib/python3.10/site-packages/scipy/optimize/_highs/_highs_wrapper.cpython-310-x86_64-linux-gnu.so
libstdc++.so.6 => /home/mitligo/miniconda3/envs/finesse/lib/python3.10/site-packages/scipy/optimize/_highs/../../../../../libstdc++.so.6
 then do (or add to ~/.bashrc file & source)
export LD_PRELOAD=/home/mitligo/miniconda3/envs/finesse/lib/python3.10/site-packages/scipy/optimize/_highs/../../../../../libstdc++.so.6:$LD_PRELOAD
Reopen matlab & run the python script via pyrun("finesse3 code")
 
% FIX h5py python-matlab issue (tested on Ubuntu 20)
Ref: https://www.mathworks.com/matlabcentral/answers/327193-calling-python-module-from-matlab-causes-segmentation-fault-in-h5py
open matlab (after the above fix) , then do
 
finally load the right pyenv via inprocess execution.
Images attached to this report
Non-image files attached to this report
General
hudsonl - posted 17:31, Tuesday 05 December 2023 - last comment - 11:51, Friday 08 December 2023(11420)
CSQZ Fiber Box LO Fiber
The LO fiber out of the Fiber box looks to be very lossy. There's 180mW into the fiber box, and we expect the LO fiber to have roughly 90% of this power, or 160mW. Its output is only about 35mW, and the fiber is clearly glowing near where it's spliced when viewed with an IR camera.
Comments related to this report
hudsonl - 14:48, Wednesday 06 December 2023 (11422)
The LO power is also about 25% of the expected power given the 180mW input power. The LO power is 2.5mW vs 9mW expected. This indicates that the problem is not with the new splice, but with a fiber element common to both the LO and the SHG pump paths.
hudsonl - 11:07, Friday 08 December 2023 (11427)
It looks like the 99/1 fiber splitter is damaged. Connecting the PM fiber out of the laser to the fiber splitter, we get about 80mW out of the 99% port with 220mW in, or about 36% coupling. Using a standard single mode patch cable, we get 180mW out, or 82% coupling, more in line with what I would expect. Given this, I believe we should try replacing the 99/1 splitter before we order new AOMs for the fiber box.

The fibers have a damage threshold of about 1 MW/cm^2, and a "practical safe level" of about 250 MW/cm^2 according to thorlabs (see https://www.thorlabs.com/newgrouppage9.cfm?objectgroup_id=1362&pn=P1-1064-FC-1). The fiber we're using has a mode field diameter of about 6 microns, or an effective area of about 1.1 x 10^(-6) cm^2, which gives a damage threshold of about 300 mW and a "practical safe level" of about 75 mW. Based on this, it seems likely we damaged the fiber splitter by sending too much power at it.
hudsonl - 11:17, Friday 08 December 2023 (11428)
While the fiber daage threshold is quoted at 300 mW, the spec sheet for the 99/1 fiber beam splitter says it should be able to handle up to 1 Watt of power. See https://www.thorlabs.com/drawings/830d2b7c309f8c74-E4C1FA78-A61A-FC1E-0E705545D0A363D3/TN1064R1A1B-SpecSheet.pdf
hudsonl - 11:51, Friday 08 December 2023 (11429)
Connecting the fiber-coupled output of the laser directly to the input of the fiber box, we only have about 25% transmission from the input to the pump fiber. We send in about 205 mW and get about 55 mW on the pump path. This indicates that while the 99/1 beam splitter may need to be replaced, there is a sperate problem with the fiber box as well.

Connecting the output of the 99/1 to the input of the fiber box, we have about 200mW in, about 100mW out of the "99%" path, and about 20mW out of the pump path fiber from the fiber box, so the losses all add up, and it looks like there is a problem both with the 99/1 beam splitter and with the fiber box.

The 1% pickoff port of the 99/1 has 1mW of power, so it looks like both paths have about 50% loss. This makes sense, as the damaged fiber looks to be the input to the beam splitter.
DarkMatter General
swadha - posted 16:30, Tuesday 05 December 2023 (11419)
P-s phase splitting of cavity mirrors
Non-image files attached to this report
GRAVITES General
epolini - posted 20:01, Sunday 03 December 2023 - last comment - 18:02, Monday 04 December 2023(11416)
Measurements with servo in DC both with SR785 and 4395A

On Friday, Eric and I realized that the correction signal of FNC loop was flat at higher frequencies (i.e. 1MHz with the Agilent 4395A). We realized that it might be due to the noise of the SR560 which was set to a LP filter @ 3kHz, as shown by the noise spectrum in the entry 11393 . We changed the configuration of the servo to DC, still with a gain = 100. The purpose of this study is to verify if we are limited by GAWBS as we suspected from the latest study we showed in the entry 11396.

I acquired the error signal (closed loop), correction signal and TF of the system, both with the SR785 and 4395A (to be able to verify the calibration).

Fig. 1: Error signal measured with SR785 (blue) from 1Hz - 102 kHz and measured with the Agilent 4395A (orange) from 1Hz - 1.6MHz.

Fig. 2: Correction signal measured with SR785 (blue) from 1Hz - 102 kHz and measured with the Agilent 4395A (orange) from 1Hz - 1.6MHz.

Fig. 3: OLTF of the system measured with SR785 (blue) from 100Hz - 102 kHz and measured with the Agilent 4395A (orange) from 100Hz - 1.6MHz.

Next step:

Images attached to this report
Comments related to this report
epolini - 18:02, Monday 04 December 2023 (11418)

[Eric, Eleonora]

The discrepancy between the TF acquired with the SR785 and Agilent 4395A is due to the fact that the R channel used for the TF (A/R) had a 20dB attenuation. So the discrepancy was 20dB - 6dB that comes from the 50Ohm input impedence. We verified that the two instruments agree.

DarkMatter General
swadha - posted 12:15, Saturday 02 December 2023 (11415)
Measurement process and first measurement for ADBC

Measurement process:

  1. Cavity:
    1. Turn down laser power (input l/2 306deg)
    2. Switch to p-pol (pre-vacuum chamber l/2 45deg, refl post vacuum chamber l/2 65deg)
    3. Sweep laser PZT with triangle wave and check that the 00 mode is being maximally transmitted.
    4. Note down cavity coupling (for coupling factor into cavity)
  2. Calibration:
    1. Turn laser power back up (input l/2 340deg)
    2. Switch to s-pol (pre-vacuum chamber l/2 90deg, refl post vacuum chamber l/2 110deg)
    3. Measure P_in with power meter, maximize and note down
    4. Change pre-vacuum chamber l/2 to ~110deg and note down theta_in
    5. At transmission, note down max transmission to PDA100A and then change transmission l/2 to send ~1mW power to 1811 for LO. Note down new PDA100A value to calculate e_out, P_LO
    6. Use Agilent to measure transfer function while driving the EOM (using backwards RF splitter to add to PDH sidebands). Note down modulation depth of calibration sideband, V_ag.
    7. Perform calibration analysis and verify that it matches the model
  3. Noise:
    1. Remove splitter at EOM driver input (disconnect from Agilent because it causes noise)
    2. Change theta_in back to pure s-pol input
    3. At transmission, note down max transmission to PDA100A and then change transmission l/2 to send ~2mW power to 1811 for LO. Note down new PDA100A value to calculate e_out, P_LO
    4. Take spectrum of 1811 upto 20MHz and check that it is clean and at shot noise
    5. Put 1811 AC output into scope and check that it is less than 300mVp (HF2LI range)
  4. Measurement:
    1. Put 1811 AC output into HF2LI and connect HF2LI to laptop
    2. Perform measurement for ~3.5 hours with T_fft = 0.28s, f_res = 3.5 Hz
       

First measurement data: calibration, noise, measurement

First measurement preliminary result with SNR = 1 approach for calculating axion-photon coupling bounds: plot

Non-image files attached to this report
DarkMatter General
swadha - posted 11:59, Saturday 02 December 2023 (11413)
Configuring the Zurich Instruments Lock-in Amplifier HF2LI
RetroSense General
msredden - posted 18:40, Monday 20 November 2023 (11411)
RetroSense sensitivity measurements

These measurements made with the X and Sum channels which I forgot still have the old op-amps with a slow slew rate on the high-pass filters. Need to replace them with the ADA4610.
Using cat's eye retroreflector.

Sensitivity = 35 V/mm
Sum voltage from QPD = 13 V
Total photocurrent = 2.4 mA
Shotnoise = 150 nV/rt-Hz
Total power from LED = 5 mW
Modulation freq = 10kHz
PDA measurements made with 0dB setting

The signal out of the x-channel of the QPD board (not modulated) around the drive frequency looks like it's shot-noise limited but the noise on the sum-channel and when measured on the PDA36A looks about ~5x above shot noise. Can't figure out why right now. 

Files:
PDA36A.txt - non modulated signal as measured by the PDA36A around 10kHz
diff_12Hz.txt - x-channel signal out of demod board, 0 - 12Hz
diff_400Hz.txt - x-channel signal out of demod board, 0 - 400Hz
LEDoff_12Hz.txt - x-channel dark noise, 0 - 12Hz
LEDoff_400Hz.txt - x-channel dark noise, 0 - 400Hz
diff_nm_10kHz.txt - x-channel signal out of QPD board (not modulated) around 10kHz
sum_nm_10kHz.txt - sum-channel signal out of QPD board (not modulated) around 10kHz

Images attached to this report
Non-image files attached to this report
cSQZ General
hudsonl - posted 17:54, Wednesday 15 November 2023 (11409)
Fit Test of PDH Board for Compact Squeezer
JD manufactured a second compact breadboard to house the optics needed to lock the OPO with a Pound-Drever-Hall lock at 532 nm. I attached all of the optics mounts, radio frequency photodetector enclosure, Faraday isolator, and electro-optic phase modulator to ensure that all of the parts attach to each other properly. We noted a minor issue with some of the holes for the 1/16 inch dowels used to position the mirrors, but other than that all of the parts fit together well.

Once we have the larger baseboard that both the initial compact squeezer board and the Pound-Drever-Hall board will lie on, we'll align the optics, which will e the real test of whether or not this new board is working as expected.

Attached images show the new, compact Pound-Drever-Hall board.
Images attached to this report
cSQZ General
hudsonl - posted 12:15, Tuesday 14 November 2023 (11408)
CSQZ TEC Driver Circuit Fixes
The CSQZ TEC driver circuit V2 needs a few modifications before it works properly:
- The 680 ohm resistor on pad R6 should be replaced with a 680k resistor. This makes the OPO and SHG TEC drives symmetric.
- The 100k resistors on pads R4 and R5 should be removed.
Heisenberg IFO General
hudsonl - posted 11:53, Tuesday 14 November 2023 (11407)
Breadboard OPO Thermistor
The thermistor used to read out temperature on the breadboard OPO is a Siemens 10k B57861 thermistor. It's resistance vs temperature curve is given by 8016 column on page 136 of the attached PDF.
Non-image files attached to this report
Heisenberg IFO General
hudsonl - posted 19:10, Monday 13 November 2023 (11406)
Breadboard OPO PZT Actuation Cables
The pinout for the PZTs on the Breadboard OPO is given in the attached file. Based on this, I made an adapter from the D9 connector to a pair of female BNC connectors.
Images attached to this report
DarkMatter General
swadha - posted 15:54, Thursday 09 November 2023 (11405)
Calibration calculations and measurement
Images attached to this report
Non-image files attached to this report
RetroSense General
evanhall - posted 10:51, Tuesday 26 September 2023 - last comment - 15:02, Thursday 16 November 2023(11342)
RetroSense boards, v2

The main differences from v1:

Non-image files attached to this report
Comments related to this report
evanhall - 19:38, Monday 23 October 2023 (11379)

The chosen quad op-amp (OPA4182IDR) has 6 nV/Hz1/2 of input voltage noise and 0.17 pA/Hz1/2 of current noise.

With the choice of 2.4 kΩ transimpedance resistors, and with 1 mA of total photocurrent, the total output noise of the transimpedance stage (i.e., if all four outputs were summed together) comprises 43 nV/Hz1/2 from shot noise, 13 nV/Hz1/2 from Johnson noise, 12 nV/Hz1/2 from op-amp voltage noise, and less than 1 nV/Hz1/2 from op-amp current noise. Since the sum/difference amplifier and the differential amplifier are both unity-gain right now, this compares unfavorably to the >100 nV/Hz1/2 from the AD630.

If we stick with 1 mA of photocurrent, then it is probably advantageous to increase the transimpedance gains to something like 5 or 6 kΩ.

evanhall - 09:58, Thursday 09 November 2023 (11401)

A comment on modulated spectra — even in the idealized case, the spectrum of the square-wave-modulated photocurrent has a noise floor that is clearly above the shot noise floor. The shot-noise-limited performance is recovered after demodulation.

Non-image files attached to this comment
evanhall - 15:02, Thursday 16 November 2023 (11410)

Suggestions for a v3:

  • Replace OPAx182 with ADA4610 (a JFET) amplifier — from Maya's test it seems like there is some slew rate limiting with the OPAx182 even though it is nominally 10 V/µs. The ADA4610 is typically 25 V/µs. Its current noise amplitude spectral density is given in the attachment — it rises like f and is about 1 pA/Hz1/2 at 1 MHz.
  • Replace the differential amplifier on the demod board with something else — possibly also ADA4610.
  • If rise/fall time is still an issue, try decreasing the 10 pF feedback capacitors on the quad op-amps to 5 pF or smaller.
  • Add test points to the sum/differencing amp on the head board.
  • Add test point to sense voltage drop across LED resistor R9
  • Fix the clocking TTL level issue (demod board switches on a zero crossing, MOSFET switch turns on at 2.5 V or something).
  • If ripple in the LED current is an issue, replace the MOSFET switch with a op-amp BJT current driver.
  • Matt's suggestion: add a phase shifter for clocking the demod.
Non-image files attached to this comment
DarkMatter General
swadha - posted 09:31, Tuesday 01 August 2023 - last comment - 12:02, Saturday 02 December 2023(11273)
Crossover loop to slow input of controller

Problem 1: Error signal drifts over time. Turned up the LB1005 gain and turned on the integrator to stabilize.

Problem 2: Control signal slowly drifts (possibly due to slow laser temperature drifts). For this we put in an integrator to the 'slow' input of the laser controller, with a low frequency crossover between the slow and fast actuators.

Slow actuator: 5GHz/V, pole(s) at ~1 Hz => We want crossover below 0.1 Hz to prevent phase loss

Circuit diagram attached. OpAmp used: LT 1792 (low input current noise)

UGF of integrator = 1/6000 => Crossover frequency = 0.016Hz, well below 1Hz. Put circuit into loop, laser stays locked and control signal no longer drifts.

Images attached to this report
Comments related to this report
swadha - 12:02, Saturday 02 December 2023 (11414)

New NPRO seems to have opposite acutation signs for the slow and fast controllers, so we put an inverter before this circuit.