We're using one of my old LabJack servo boards for the test cavity on the Y-arm
I set it up yesterday:
1. LabJack EPICS IOC is running on quadrature. If it isn't working, try restarting the Daemon:
- ssh quadrature
- sudo systemctl restart CC_LABJACK_SERVO.service
2. There is an MEDM screen here that you can use to control the board:
- /opt/rtcds/userapps/trunk/sqz/m1/medm/ARM.adl
To acquire lock, sweep the output offset adjust until you go through resonance on the arm cavity, then turn on the "Feedback EN" switch. Once it grabs lock, turn on the two boost filters. The sign might need to change.
Took TFs of each stage. I think that the boost filters are just about right for a loop bandwidth of ~30 kHz, but we'll probably need to re-tune the lazer PZT notch filters:
- Stage 1: DIfferential input with 10 dB of upfront gain and sign switch
- Stage 2: Input Offset/Excitation point. 0 dB gain
- Stage 3: Variable gain stage: -10 to +30 dB of Gain.
- Stage 4: Switchable boost filters:
- Boost 1: Pole/Zero at 50 Hz/ 5 kHz, DC Gain 40 dB (ENABLED), Or 0 dB gain (DISABLED)
- Boost 2: Pole/Zero at 100 Hz/ 1 kHz, DC GAIN 20 dB (ENABLED), Or 0 dB gain (DISABLED)
- Stage 5: 2x Bandstop filters. Likely these were notching PZT resonances in the laser above 100 kHz (cant see the bandstop with the SR785)
- Stage 6: Output Offset stage, 0 dB gain
Note: the output of the servo board is connected to the front BNC labeled OPO_SLOW output.
Needs a bit of troubleshooting:
- Variable gain stage is giving the correct gain vs setting behavior except at -10, which it is setting to +30 dB for some reason. Likely a software bug.
Schematics for the electronics in that box. The Boost and Filter stages have been stuffed differently than what is shown in the default schematic.