CoupledCavity Electronics
eoelker - posted 15:26, Wednesday 09 September 2026 - last comment - 16:00, Wednesday 09 September 2026(12250)
Servo for Y-Arm Test cavity

We're using one of my old LabJack servo boards for the test cavity on the Y-arm

 

I set it up yesterday:

1.  LabJack EPICS IOC is running on quadrature.   If it isn't working, try restarting the Daemon:  

2.  There is an MEDM screen here that you can use to control the board:

To acquire lock, sweep the output offset adjust until you go through resonance on the arm cavity, then turn on the "Feedback EN" switch.  Once it grabs lock, turn on the two boost filters.  The sign might need to change.

 

Took TFs of each stage.  I think that the boost filters are just about right for a loop bandwidth of ~30 kHz, but we'll probably need to re-tune the lazer PZT notch filters:

Note: the output of the servo board is connected to the front BNC labeled OPO_SLOW output.  

Needs a bit of troubleshooting:

  1. Variable gain stage is giving the correct gain vs setting behavior except at -10, which it is setting to +30 dB for some reason.  Likely a software bug.
Comments related to this report
eoelker - 16:00, Wednesday 09 September 2026 (12251)

Schematics for the electronics in that box.  The Boost and Filter stages have been stuffed differently than what is shown in the default schematic.  

Non-image files attached to this comment