Last week we implemented our first lock of the length servo for the GRAVITES experiment.
Length feedback is desireable because, unlike frequency feedback, we do not require that the control signal be scaled differently to account for the difference in wavelength between the control (1545 nm) and single photon source (1550 nm) wavelengths:
So far we have implemented piezo feedback only; we will add thermal feedback once the new low-vibration noise fiber spools are complete.
As of now, the loop can only stay locked for a few minutes due to the finite range of the piezo actuator. Our piezo only has about 30 um of range. Adding thermal feedback will boost the range by many orders of magnitude. It may still be desireable to add an intermediate range piezo (with several mm of range) to avoid saturation issues over intermediate timescales. I plan to order a suitable actuator shortly. I'll also add subposts below giving additional detail about the new servo and its performance.
In order to implement length feedback, we needed to add a second PCB to our Mach Zehnder servo to allow for length control. The new servo board splits the feedback into two paths. The first path is much the same as before: It outputs to a VCO (marconi) that shifts the frequency of an AOM in one of the two arms. The second path adds an offset and a 150 V PZT driver and outputs to the piezo fiber stretcher. We will eventually add a third output for the fiber spool temperature. Likely this will incorporate a Red Pitaya to do the thermal stabilization with a voltage-adjustable setpoint.
The servo performance, bandwidth-wise, seems to be in line with what I was expecting based on my servo model. See the dropbox under /Electronics/MZServoModel.ipynb. Attached are measurements of the OLTF (-20 dB corresponds to unity gain here) and crossover between the piezo and VCO paths.