RetroSense General
msredden - posted 17:26, Friday 23 June 2023 - last comment - 17:46, Monday 26 June 2023(11235)
Sensitivity comparisons with varied configurations
Measured the sensitivity and relative intensity of the sensor under different configurations including with/without an f = 25.6mm lens, with/without a 1.5mm iris, and using an OP232 and VSLY5850 LED. 

Found that spot size on the QPD had a significant effect on the sensitivity. Setups with the lens for both LEDs and the iris for the VSLY5850 produced the best results. 
Non-image files attached to this report
Comments related to this report
msredden - 17:46, Monday 26 June 2023 (11237)
6/26/2023 update

Tested the OP232 with the new .5" f = 40.0mm lens and the PLA mount. Found a relative intensity of 1.17 (compare to the OP232 with the 1" f = 25.6mm lens which had a relative intensity of 1.19). 
Non-image files attached to this comment
DarkMatter General
evanhall - posted 18:11, Wednesday 21 June 2023 - last comment - 09:35, Monday 17 July 2023(11233)
ADBC laser swap

Swadha, Evan

Previous laser: NP Photonics Rock (see MIT:11213)

Current laser: Lighwave/JDSU/Lumentum NPRO

The laser worked fine for the first week but starting a few days ago would only run for about an hour before going into a cycle of repeatedly shutting itself off every 30 seconds or so. There is no indication of thermal overload on the diagnostic fault pin, but nonetheless I suspect more heat sinking is needed. We clamped the head to a ~2" thick aluminum block that sits directly on the table. So far it has been running for 2 hours without issue, putting out around 800 mW.

Comments related to this report
evanhall - 14:45, Sunday 09 July 2023 (11254)

Myron fabricated a new mount, but the behavior eventually recurred when the laser was left on for a few hours. We tried another aluminum mount, but the behavior recurred with that too.

To test whether this is a problem with the controller or the head, Swadha swapped the controller to JDSU FQN4113040 (manufactured May 2013).

swadha - 09:35, Monday 17 July 2023 (11260)

Ran for ~2 days with the swapped controller (JDSU FQN4113040) without shutting off. It would seem the problem was with the controller.

GRAVITES General
Anders, Eleonora - posted 18:40, Tuesday 20 June 2023 (11232)
NPRO 126N-1064-500 Laser Beam Profiling

[Anders, Eleonora]

We used the CCD Beam Profiler to measure the average beam diameter at 9-10 different points from the laser tip. The laser was first sent through a HWP and PBS to reduce the intensity of the beam from ~500mW to ~20mW in order to avoid CCD streaking. The measurement was done twice as the first 10 points collected were not at the desired ellipticity (>90%) according to the beam profiling software. To improve that, for the second collection the beam profiler was placed at the maximum distance from the laser source (65cm) and the HWP was tuned to find the best balance between ellipticity and low CCD streaking. This resulted in a better measurement.

The raw data for both trials can be seen here:

Trial 1
Position (cm) Beam Diameter (um) Approx Ellipticity (%)
22.5 757.0 93
27.5 913.6 88
32.5 1075.1 88.3
37.5 1238.9 85.1
42.5 1405.5 84.8
47.5 1545.2 84.4
52.5 1723.0 83
57.5 1832.7 80
62.5 2067.2 83
70 2287.8 81
Trial 2
Position (cm) Beam Diameter (um) Approx Ellipticity (%)
22.5 747.9 84
25.5 883.8 91.4
32.5 1001.9 95
37.5 1107.4 96.3
42.5 1324.6 97
47.5 1481.2 99.3
52.5 1600.0 92.3
57.5 1760.8 94.3
65 1945.0 94.3

The data was then fit to the beam width function for Gaussian beams and the waist (w0) and position of the waist (z0) were extracted. This was done using scipy's curve_fit module in python. The graphs for both trials have been attached below.

The waist was found to be w0 = 218.20 +/- 4.59 um, and the position of the waist was found to be z0 = 3.04 +/- 0.81 cm.

The data sheet for this laser specifies the waist to be 190 um (in x) and 250 um (in y), and to be located at 5 cm from the shutter. The disagreement in beam waist location could be due to approximation error with the positions of measurements from the laser source, as the distances were estimated only with the holes in the optical table and not with a ruler.

Images attached to this report
testbedAI General
epolini - posted 18:37, Thursday 15 June 2023 (11226)
JDSU NPRO laser beam shape
[Eleonora, Nikhil, Anders]

We checked the laser beam profile shape of the 124N-1064-500 laser recently installed in NW22 that will be used for the testbedAI and for the 10-m prototype of GRAVITES. 
In order to avoid beam saturation and streaking on the CCD camera (see attached pictures), we needed to lower the power of the laser.
We set the laser power to 562 mW and then we installed a power controller composed by a half wave plate and a PBS right after the laser. We are dumping the PBS reflection with a beam dump. A transmitted power between 4 mW and 20 mW is good to have a good ellipticity of the beam and avoid the saturation.
Images attached to this report
CDS General
nikhil - posted 17:22, Monday 12 June 2023 - last comment - 14:28, Monday 26 June 2023(11223)
NW22 CyMAC ADC channel check

[Eleonora, Nikhil]

We connected the output of a photodiode to the ADC channels of the cymac to check if they work properly.

We verified all the ADC channels ( X5:IOP-MHZ_MADC0_EPICS_CH{0-15} ) and it seems that only channels 0, 1, 2, 3, 4 are responding to the applied voltage. These 5 channels have been labelled. We suspect that there is a problem in the ADC cards, since we double checked the connectors behaviour. Further verifications will follow.

Comments related to this report
evanhall - 17:32, Friday 16 June 2023 (11228)

Nikhil, Evan

We used a function generator to drive each of the 16 channels with a 1 Vpp sine wave at 1 Hz and got a different answer. Channels 0, 2, 4, 5, and 6 seem to be broken in some way (large offsets, minimal response to signal), while the other channels faithfully reproduced the signal with no offset. We looked at the test points for channel 6 on the AA card directly and verified that the problem appears already in analog, so we suspect that some of the amplifiers on the AA cards are blown.

We also took this opportunity to clean out the blue box a bit (see attachment from before we started). The second 65 kHz ADC, along with the fast ADC, were sitting loose and unused, and they have now been removed. The ribbon cable connecting the AA/ADC adapter to the ADC was not seated securely and was causing vibrational pickup on some of the channels. The timing buffer is still a mess and the clock components are not secured to the chassis.

Some documentation:

  • The serial numbers of the two AA boards are S1101854 and S1400645. Together they supply 16 channels to a single 65 kHz ADC card. We have a second ADC card and room in the PCIe bus, but no extra AA boards and no obvious place in the chassis to put extra AA boards.
  • The DCC filecard for this blue cymac chassis is S1900243.
  • The fast ADC modifications that Chris described from 2021 are attached as a pdf here.
  • The front panel of the blue chassis numbers the channels starting at 1 while the digital system starts at 0.
Images attached to this comment
Non-image files attached to this comment
evanhall - 14:28, Monday 26 June 2023 (11236)

On the assumption that the channel problems are from blown amplifiers, I ordered ten THS4131ID differential amplifiers in SOIC form. We already have an essentially unlimited supply of LT1492 dual op-amps in the EE shop. (Refer to D070081.)

The D-sub to BNC breakout cables that were attached to this chassis were a bad idea (attachment). Aside from the issue of using plugs rather than jacks, and thereby requiring the use of BNC barrels to hook up a normal BNC cable, this cable makes it essentially impossible to isolate the various BNC shields from each other and from the metal rack, which will cause the negative inputs of all the ADC channels to be shorted together.

We have one fully built 8-channel isolating BNC breakout adapter D1900426, which is meant to go into a whitening chassis but can be repurposed to interface with the channel inputs to the blue chassis.

Images attached to this comment
RetroSense General
msredden - posted 14:23, Saturday 03 June 2023 - last comment - 20:50, Monday 19 June 2023(11220)
Noise spectrum analysis of demodulated signal
Noise spectrum analysis of the demodulated signals (prior to any changes made by Evan as noted on the 5/31 report).
Comparisons were made between the normal (led on and signal demodulated) signal noise, noise from the lock-in amplifier and QPD alone (by blocking LED), and noise from the lock-in amplifier alone (by shorting lock-in amplifier input). The noise is also compared to the shot-noise limit. We see the normal signal noise does not reach the shot-noise limit even at high (100 Hz) frequencies. 
Images attached to this report
Comments related to this report
msredden - 20:39, Monday 19 June 2023 (11230)
Updated spectrum:
This new spectrum includes a trace for the noise from the QPD board alone (achieved by removing the previously used QPD which was connected by several long wires and replacing it with a new QPD directly 
attached to the board, this was done to check for any noise possibly caused by the original QPD connection).
There is also an additional shot noise level ("Shot Noise (light off)") which was calculated using the amplitude of the sum signal when the LED was off and the QPD would be detecting only ambient light not blocked by the box placed over the assembly. 
It should also be noted that the "Light Off" trace was achieved by turning off the power to the LED board, instead of simply blocking the light with a card as was used during the previous test.
Images attached to this comment
cSQZ General
Hudson Loughlin - posted 18:59, Friday 02 June 2023 (11219)
3-Axis CLF Polarizaion Control
I added another red pitaya to the MIT network and into our setup to see if actuating on the third axis of the polarization controller might help improve the very long term (scale of hours) polarization control of the CLF. Actuating on this third axis (which is nearly degenerate with the first axis) seemed to significantly degrade the polarization stability for all dither and PID parameters I tried. Even with the gain on the third axis 1/100 of the gain of the first axis, it prevents the loop from actuating to zero and causes a substantial offset. Since this is a fairly low priority, I'm going to set aside further improvements to this loop for now.
cSQZ General
Hudson Loughlin - posted 20:40, Wednesday 31 May 2023 - last comment - 18:34, Thursday 01 June 2023(11216)
Custom Photodetector Noise
The custom RFPD had non-stationary noise that would briefly spike up across about a 1 MHz range by 20-30dB above the usual PD noise level. This was fixed by modifying the RFPD's power cable so that the power voltages and PD bias voltage were picked off of the same cable about 1cm before the PD, instead of about 1 meter before the PD at a patch panel. I believe this was creating a loop between the ground cables and possibly also between the different -15V cables. (The power supply cable has +15V, GND, -15V and the bias cable has -15V and GND.)

The other custom PDs suffer from excessive 60 Hz noise, and re-soldering cables for them will likely cut this noise. This doesn't seem like an essential thing to do in the short term, but should be fixed in a later version of the squeezer and possibly for the PDs used to lock the SHG and OPO cavities.
Comments related to this report
Hudson Loughlin - 15:46, Thursday 01 June 2023 (11217)


		
		
Hudson Loughlin and Evan Hall - 18:34, Thursday 01 June 2023 (11218)
Even with the fix to the power cables, the RFPD has fairly large and non-stationary noise that appears to be from mechanical coupling. The SMA cables going to the RFPD attach to the PCB, and not the case, so any motion of these cables strains the RFPD's PCB. The output signal also oscillates strongly at about 100 MHz, which can be filtered out, but seems indicative of some other problem in the circuit. The best course of action may be to redesign the RFPD and its housing so there is strain relief on the cables, the RFPD's case is electrically isolated from the optical breadboard (probably using a piece of PEEK plastic, and the PCB screws into the PD case more easily.
RetroSense General
evanhall - posted 09:46, Wednesday 31 May 2023 - last comment - 08:46, Wednesday 05 July 2023(11214)
AD630 lock-in circuit

Peter suggested using an AD630 to perform the lock-in demodulation for the RetroSense signals. The datasheet does not give a noise spectrum so I wanted to see how it performed.

I wired an AD630 on a protoboard in the lock-in topology shown in Fig. 35 of the datasheet, omitting the AD711s. I powered the board with ±15 V from a benchtop supply. I used an SR560 with a gain of 20 dB and a 300 Hz 12 dB/octave lowpass in order to do the final filtering step. I drove the comparator with a ±1 V square wave at 30108 Hz, and then drove the signal input (pin 1) with another square wave. I tried various amplitudes and offsets up to 300 mVpp. In all cases the noise at 1 Hz was a few µV/Hz1/2, referred to the AD630's output. This is not good enough for us because we want to be shot noise limited with a few milliamps of photocurrent and a transimpedance of order 1 kΩ, which amounts to a noise of a few tens of nV/Hz1/2. (The gain of the AD630 is of order 0 dB — the internal amplifiers run with a gain of 6 dB and there is some loss due to the demodulation process).

I tried the following, which did not lower the noise:

I didn't add an input buffer to the signal input like shown in the datasheet figure, but I don't see why that should matter.

By using the function generator's internal AM modulation feature, I saw that there was significant harmonic content in the voltage spectrum when the signal was sent to the AD630 with a square wave, but not with a sine wave. I do not understand this behavior since mathematically, square wave demodulation should not have this feature.

Photo shows the AD630 circuit along with the AD711 Sallen-Key filter.

Images attached to this report
Non-image files attached to this report
Comments related to this report
evanhall - 18:40, Wednesday 31 May 2023 (11215)

I retook some of these measurements. First, the SR560 I was using yesterday seems to have excess low-frequency noise, so I suspect that was part of the confusion. Second, the circuit is picking up some kind of environmental noise, so the noise floor is not entirely stationary. At the quietest times it approached 100 nV/Hz1/2 at 1 Hz.

It is probably worth redoing this board as a PCB with on-board voltage regulators and seeing how it performs with actual photocurrent signals.

Non-image files attached to this comment
evanhall - 17:08, Wednesday 07 June 2023 (11221)

I ordered PCBs of this design from JLCPCB, each with three demodulation channels (for X, Y, and sum). This is two layers, with the ground achieved through a polygon pour on the top layer. If we come to regret this then we may want to go for four layers.

Non-image files attached to this comment
evanhall - 22:45, Monday 03 July 2023 (11249)

These PCBs arrived and I stuffed the X channel on one of them. The voltage noise had similar nonstationary behavior when tested as described above outside the NW22 lab, so I moved it into the lab and also added an SR560 at the output with gain.

  • PCB powered with ±18 V
  • Clock drive: 1 Vpp square wave at 33 235.77 Hz
  • Input drive to X: 25 mVpp with 5% modulation depth at 1.112 Hz (AFG3022B screen says 50 mVpp, but it is not... the hi-Z display option seems to disable if you request modulation)... thus 1.25 mVpp modulation envelope. Tried sine and square waves.
  • The PCB contains the AD630 in a gain of 2 demodulator configuration, and an AD711 Sallen-Key filter witha 2-pole rolloff at 290 Hz.
  • Output signal from PCB's X channel goes into SR560, which has gain of 100 V/V and a 12 dB/oct rolloff at 300 Hz.
  • Output of SR560 goes into spectrum analyzer.

I looked on a scope to get the overall gain factors in either the sine or square drive configurations:

  • Output from SR560, sine wave: 140 mVpp at 1.112 Hz, and therefore 1.4 mVpp direct from demodulator.
  • Output from SR560, square wave: 210 mVpp at 1.112 Hz, and therefore 2.1 mVpp direct from demodulator.

Compared to the 1.25 mVpp envelope of the modulated input signal, the gain for sine wave signals is about 1 and the gain for square waves is about 2... we likely want to stick with square waves, since sinusoidal input would probably require bandpassing of the photocurrent signals, which throws away SNR.

In the attached input-referred noise curves there is no indication that square wave input has worse nonlinear content than sine waves. The input-referred noise is about 200 nV/Hz1/2 from 0.1 to 1 Hz. This competes with the expected shot noise of 3 mA of total photocurrent if it is handled with a transimpedance of 4 kΩ, which produces 12 V dc in the total voltage sent to the sum channel of this demodulation board.

I did not build the capability for ac coupling into the input to this circuit, and perhaps I should have, since the dc offset voltage into the demodulator is not needed. Also, the AD711's input voltage noise is somewhat high... though I don't think it's limiting the current performance.

Images attached to this comment
Non-image files attached to this comment
evanhall - 08:46, Wednesday 05 July 2023 (11250)

Tried replacing AD711 with OPA227, did not notice a difference in the 50 Ω input-terminated spectrum. We should consider running the AD630 in a gain 1 configuration by tying pin 14 to pin 13 rather than to ground (refer to the lock-in application section of the AD630 datasheet), which should be doable on the current PCB since the AD630 is a DIP in a socket. It can and should be modified in a reversible way (i.e., do not clip pins) as this part is somewhat expensive.

The AD630 datasheet does not specify a noise performance for this IC.

DarkMatter General
swadha - posted 17:35, Tuesday 30 May 2023 - last comment - 18:14, Wednesday 21 June 2023(11213)
Limited by laser frequency noise
We have a PDH lock crossover loop setup for the ADBC bow-tie cavity. The cavity (FSR 65MHz) has a cavity pole of ~100kHz for p-polarized light and ~1kHz for s-polarized light. The finesse for p and s has been measured using the width of the pdh error signal. We are actuating on the PZT and EOM, attached is the data (oltf.pdf) for the open loop transfer function (UGF 8kHz, PM 50deg, crossover happens around 11kHz).

Also attached is the measured free running noise in the system overplotted with the laser frequency noise (noise_cal.pdf). The plot shows we are currently limited by laser frequency noise. From the error signal spectrum (esigsp.pdf) we see that the rms frequency noise is 4kHz, greater than the s-polarization cavity linewidth, and so we cannot lock in s. We would like to switch to a lower noise laser.
Non-image files attached to this report
Comments related to this report
evanhall - 18:14, Wednesday 21 June 2023 (11234)

This laser is NP Photonics Rock — LIGO-T2000263 for characterization.

cSQZ General
Hudson Loughlin - posted 11:07, Tuesday 30 May 2023 - last comment - 11:25, Tuesday 30 May 2023(11210)
Mog Labs Quad RF Synthesizer Setup
The Mog Labs quad RF synthesizer is now hooked up to the lab desktop. Instructions for connecting the RF synthesizer to a new computer are here: https://www.moglabs.com/support/software/connection the manual for the device is here: https://www.moglabs.com/products/rf-electronics/quad-rf-synth/MOGLabs_QRF_manual.pdf and firmware updates are here: https://www.moglabs.com/support/software/rf
Comments related to this report
Hudson Loughlin - 11:25, Tuesday 30 May 2023 (11211)
I tested the frequency modulation on the Mog Labs RF synthesizer, which will be used as the actuator in the CLF loop, using a function generator and a spectrum analyzer, and it appears to work as designed.
Facilities General
evanhall - posted 17:17, Wednesday 24 May 2023 - last comment - 17:02, Sunday 18 June 2023(11205)
ADBC table floating from N2 tank
Myron plumbed the ADBC table to a nitrogen tank rather than the building air supply. This is a size 300 cylinder.

Valve open at 10 am. The tank pressure started around 2500 psi. Now at 5 pm it is still a little above 2400 psi. The table work today was minimal.

The regulator on the tank reads 45 psi. The gauges on the table legs read more like 15–25 psi.

These are S-2000 series Newport legs.
Comments related to this report
evanhall - 17:26, Thursday 25 May 2023 (11206)
At 5 pm May 25, the tank pressure was just over 2300 psi. The regulator pressure was 44 psi and the table legs were again at 15–20 psi.

The table floats but I suspect not as much as it ultimately should — it seems a bit more sluggish than when we were drawing from the building air at 60 psi. The table is Newport RS-4000-48-12, weight 1110 pounds. The two vacuum tanks, pipe, and inner optical platforms all told are probably north of 700 lbs. The S-2000 manual gives the recommended leg pressure as (Load / (4 × 25.9 sq. in.)) + 10 psi, which for us comes out to 22 psi at least.

It means we may want to repeat this cylinder test with a higher leg pressure. That may require a tank regulator with higher range.
evanhall - 11:26, Tuesday 30 May 2023 (11212)
At 6 am May 30, the tank pressure is about 1990 psi and the regulator pressure is 44 psi. The average tank depletion rate is therefore about 110 psi per day.
evanhall - 17:02, Sunday 18 June 2023 (11229)

On June 18, the regulator pressure is 44 psi and the tank pressure is about 1020 psi.