EOM driver circuit from DCC for broadband EOM (new focus 4004)
PCB ordered from JLCPCB, asked them to assemble parts as given on the DCC. However they have their own stock that usually doesn't include the standard Digikey parts so don't recommend getting IC chips stuffed by them because you probably won’t find the correct replacement part in their stock. Otherwise shipping was fast, most parts were correctly soldered, some had to be changed/separately ordered. Website is kind of slow and keeps losing progress.
Note(!!): U1 is given as Digikey part 296-18037-1-ND on DCC. However, this part is not the correct size for the circuit (it is SOT-353, we need SOT23-5). So we replaced it by 31-AP7375-50SA-7CT-ND, that worked.
Used two inductors (L4) instead of one to get the resonance to <17MHz (within the bandwidth of the PDA10CS for PDH locking). Circuit resonant at 16.8MHz, -35dB
Laser output power: 880mW, after dump: 53mW
Laser beam: beam waist 100um at 4.35 inches (laser_beam.pdf)
Modematched to cavity using wavestate (useful links: [1], [2], some instructions on how to install at [3])
Input setup: diagram, setup on optic table, setup on breadboard
Faraday isolator aperture: 4mm, EOM aperture: 2mm (free space, broadband, new focus model 4004)
Laser head: Lightwave 126-1064-700 S/N 192 (see LIGO-C962734 for test data)
Laser controller: JDSU FQN4113040
See also alog 11233
The default fan (Delta electronics AFB0912VH brushless) on the testbedAI CyMAC makes a lot of noise. So it was replaced a with a new low noise 3-pin NF-A9 FLX 92mm fan.
The Red Pitaya infinite impulse response (IIR) filter has some quirky behavior, which I'll document below. The IIR module wants to be fed zeros and poles, these values correspond to frequencies (not radians/second) and should have negative real parts to make a stable filter. The software will not throw an error if it is fed positive values for the poles and zeros and will just create an unstable filter. The parameter "loops" sets the down-sampling rate of the filter and as far as I can tell, should usually be set to its maximal value of 1023 for audio-band signal processing. With this value, the FPGA's sampling rate will be 125MHz/1023 = 125kHz. The IIR gain sets an overall amplification of the IIR filter. For example, to make an IIR filter sampled at 125kHz with a zero at 100Hz, a pole at 5Hz, and a gain of 10 the commands are: zeros = [-100] poles = [-5] gain = 10 loops = 1023 p.rp.iir.setup(zeros=zeros, poles=poles, gain=gain, loops=loops) where "p" is the name of the red pitaya being used.
[Eleonora, Eric]
After a careful search for another collimator, we found a F220APC-1064 in the Highbay, which has a focal length of 11.17mm and collimates the beam at 2.4mm.
After installing it (see picture attached) and aligning it with the help of the RIO fiber laser, it was easily reached 73% of coupling (6.33mW before the fiber port and 4.66mW after the 2-meters fiber).
This is good enough for our purpose.
The MOG Labs RF synthesizer has a fair bit of cross-talk from its RF outputs to its input with 31dBm outputs, the modulation input channel recieves a 25mV p-p signal at the output frequency of 150 MHz. This made the CLF control signal on the scope look noisier than it really is, but can be filtered out by limiting the scope bandwidth to 20MHz. The MOG labs controller itself has anti-aliasing filters on its modulation inputs with second-order poles at 100kHz, so it should strongly attenuate the 150MHz signal and prevent problems from this cross-talk.
[Eleonora, Anders]
Yesterday and today we worked on the coupling in fiber of the 2 mm collimated NPRO beam (see entry #11239).
First, we installed an ATF 50/50 (super) beam splitter taken from from Slawek bench in the highbay (after asking) to split the previously collimated (logbook entry #11239) beam for testbedAI and the GRAVITES prototype. We placed a beam dump on the reflected beam for testbedAI and left it. We then installed a half-waveplate and PBS on the transmitted beam to allow for further power control into the GRAVITES prototype. We installed two steering mirrors before the fiber port / collimator in order to be able to align the beam properly and freeze the position of the two mirrors in front of the telescope. The new set up is shown in Fig. 1.
We tried to couple the collimated beam with a 2 meters SM fiber, following these steps:
1. install the fiber port / collimator on a tip-tilt mount to be able to align it with the incoming beam.
2. use the RIO fiber laser on the same bench to ease the alignment process: we connected the fiber laser to the back side of the fiber port and we aligned it with the incoming beam by tuning the DOFs of the fiber port (in at least two different reference positions).
3. connect the fiber to a PD and read its power with the oscilloscope and tune the alignment with the steering mirrors in front of the fiber port.
4. once we see some power on the PD, we maximize it with the two steering mirrors in front of the fiber port.
5. we moved the fiber from the PD to the power meter to tune the alignment.
6. we started the mode matching: we moved the lens position on the rails to maximize the alignemnt and then we recovered the alignment if needed.
We did this procedure for two different fiber ports and we could not reach a coupling higher than 30%.
Yesterday, we started with the coupler in Fig.2 (still installed in the same mount) but we could only get around 16% of coupling. As Slawek suggested, this morning, we checked the beam coming out of the coupler when the back port is connected to the RIO fiber laser. The beam diameter was around 1 mm and it was not collimated so we decided to change fiber port.
We tried the fiber port in Fig. 3 and we checked quickly that the beam diameter coming from the coupler was bigger than 1 mm. It was around 2 mm in several positions so we started the fiber port alignment.
The best achieved was 30% but in this configuration the beam diameter in front of the fiber port was no longer 2 mm but even smaller than 1 mm!
We measured the beam parameters coming out of the fiber port in order to try to do an inverse mode matching of the fiber.
The resulting beam profiling is shown in Fig. 4. The mean has a waist radius of 158 um, 30 cm before the coupler telescope which it a very divergent beam. This is extremely far from a collimated beam with 2 mm diameter.
Since we need a collimated beam for the AI experiment, it doesn't worth it to do an inverse mode matching to match the fiber port requirements.
Moreover, we also tested the fiber port in Fig. 5 and it also gives a divegent beam.
If we want to use the NPRO laser beam for the gravites prototype we need to buy a new fiber port / collimator at 2 mm.
[Eleonora, Eric]
We realized that the fiber port / collimator in Fig. 3 is a F240APC-A (Thorlabs), i.e. a collimator for green light. This explains its behaviour at 1064nm.
Moved setup into red lab after doing some noise spectrum analysis at the desk with the setup using the OP232, f = 40.0mm lens, and the 1" retroreflector (pictured below). Noticed the low freq noise appears to be higher by a factor of ~10 with this setup compared to when the LED was positioned immediately beside the QPD (with no lens), this did not change after moving into the lab. Next steps are to mechanically constrain the components more, mostly the LED, and consider other configurations which could miniaturize the apparatus more.
Noise Budget Calculations from last week:
Calculations of noise in QPD circuit components assuming use of an AD8675 op amp and the following values:
| voltage noise density e_n (AD8975) | 2.8 nV/sqrt-Hz |
| current noise density i_n (AD8675) | 0.3 pA/sqrt-Hz |
| temperature T | 298 K |
| feedback resistor R_f | 2000 ohms |
| photocurrent I | 1 mA |
| Noise | Value (nV/sqrt-Hz) |
|---|---|
| Shot Noise | 80 |
| Johnson Noise | 5.7 |
| Amp Voltage Noise | 3.5/f + 2.8 |
| Amp Current Noise | 0.6 |
| Total Noise (at high freq, added in quadrature) | 80.3 |
Attached shows a rough schematic of how we want the QPD readout portion of the RetroSense head circuit to look. Some guiding principles:
The RFPD circuit for the 6MHz CLF readout seems to be working if I use the FFD-100 PD and circuit instead of the FFD-200 PD and circuit. The FFD-200 circuit has semi-regular transient noise spikes that swamp the signal and appear to be coming from some interaction between the FFD-200 photodiode (30pF capacitance) and the RF amplifier. With the RF amplifier disconnected from the PD, the noise problems go away, so they seem to originate with the PD, or some interaction between the PD and the RF amplifier. The FFD-100 PD has 8.5pF capacitance instead of the FFD-200's 30pF, and using this smaller PD instead seems to fix the problem. I am now using the FFD-100 circuit to lock the OPO on co-resonance, and see a clear 6MHz signal, well above the noise on a spectrum analyzer. Now that this PD is working, the next step will be to take a look at the demodulated 6MHz signal from a red pitaya and make sure the error signal is sensible, and then close the CLF feedback loop.
[Anders, Eleonora]
The objective is to output a collimated roughly 2mm diameter beam that could be used for testbedAI and also coupled into a 2mm (diameter) fiber port collimator for the 1064 nm prototype for GRAVITES.
Considering the beam parameters estimated in the logbook #11232, i.e. w0 = 219.99um and z0 = -0.0338m, we simulated a telescope using the software JamMT (a mode-matching software). The desired output beam parameters are w0 = 1000um (waist radius) and z0 between 1 and 8 meters. Other JamMT parameters such as the shifting range were entered based on the table layout, i.e., the last adjustment mirror is roughly 0.975m from z=0 [photo of table dimensions attached below].
The simulation in JamMT suggested a variety of lens combinations considering a discrete list of common lenses found on ThorLabs, the most relevant two options were recorded (screenshots attached below).
Option 1: f=150mm @ z=1.102m + f=125mm @ z=1.402m --> resulting beam w0 = 1007um and z0 = 3.58m
Option 2: f=125mm @ z=1.252m + f=77mm @ z=1.468m --> resulting beam w0 = 1010um and z0 = 2.912m
A ThorLabs order was made for f=125mm and f=150mm lenses, along with optical rails and posts for fine tuning. We had access to a f=77mm lens already.
Both options were assembled and the beams examined (at one location) with the beam profiler. It was determined experimentally that Option 1 produced a more desirable beam for our purposes.
As a result, an inital (rough) measurement of the beam profile after the Option 1 telescope was recorded. The data is presented in the following table:
| Distance from Laser [m] | Beam Diameter [um] |
| 1.45 | 2172 |
| 1.525 | 2209 |
| 1.6 | 2198 |
| 1.675 | 2192 |
| 1.75 | 2240 |
| 1.825 | 2216 |
| 1.9 | 2236 |
| 1.975 | 2288 |
| 2.05 | 2309 |
| 2.125 | 2351 |
| 2.2 | 2335 |
| 2.275 | 2351 |
| 2.35 | 2412 |
Based on this prelimary data, we are happy with the telescope and should be able to fine-tune as needed in the future. The divergence angle was computed to be 0.015 degrees, which is well below the 2 deg ceiling.
Other photos attached include the optical table layout, the lenses as configured, and the beam profile as seen by the beam profiling software for the first measurement at 1.45 meters.
[Nikhil, Slawek, Eleonora]
LASER INFO: M126N-1064-500 JDSU NPRO Laser [Serial Number : 8013]
We checked the intensity noise for the testbedAI laser using SR785 at different power levels. We operated the laser at ~550 mW and dumped most of the light. Tried two methods for the analysis. One uses the in-built SR785 FFT function, and then TimeSeries capture from the buffer. Both results agree (see attached plots). Noise eater is definitely helping. It seems the measurement is noisier (broadband sense) at low incident power (< 10 mW on PD). At higher intensities (above 200 mW), we see some extra noise above 10 kHz. The best noise characteristics were seen at 20 mW. We didn't use a lens to focus the beam onto the REFL2 photodiode. Needs to check if the PD got overfilled and led to extra mechanical vibration coupling. It could also be that there is some gain peaking from the noise eater at higher power.
---------------------------------------------------------
SR785 Settings & Instructions for FFT Measurement
--------------------------------------------------------
FREQ SETTINGS
| Span: | LineWidth | Acquisition Time: | Full Span | FFT Lines: | Base Freq |
|---|---|---|---|---|---|
| 102.4 kHz | 128 Hz | 7.81 ms | 800 | 102.4 kHz |
DISPLAY SETUP
| MeasureGroup | FFT Measurement | View | Units | dB units | Peak Units | PSD Units | dBm Reference Impedance |
|---|---|---|---|---|---|---|---|
| FFT | FFT1 | LogMag | Vrms/rHz | Off | rms | On | 50 Ohm |
DISPLAY OPTIONS: [X-axis Log] [RPM frequency: Off]
INPUT
| Input Source | Analyzer Config: | Input Config | Ch1 Input Mode | Ch1 Grounding | Ch1 Coupling | Ch1 AA Filter |
|---|---|---|---|---|---|---|
| Analog | Independant Channel | A | Ground | AC | On |
:
AVERAGE: [Averaging type: Linear Avg] [Number Average: 100] [Display Average: RMS]
WINDOW: Hanning
Always click AutoRange Ch1 a couple of times when a new connection is made
Also click Start/Reset button a couple of times to refresh/update & redo the measurement
ToSave:
DISK:
Give a filename: a number (like 111) is sufficient. The filetype (.78D) is auto-added
Add Floppy and Click Display To Disk to save the file.
It is a good practice to Export Settings to Disk. It would be saved with the same filename but with .78S extension. This can be recalled later from disk for a similar experiment. Another thing is to first check the SR785 instrument noise floor measurement. It should be around 5-6 nVrms/rtHz
Measured the sensitivity and relative intensity of the sensor under different configurations including with/without an f = 25.6mm lens, with/without a 1.5mm iris, and using an OP232 and VSLY5850 LED. Found that spot size on the QPD had a significant effect on the sensitivity. Setups with the lens for both LEDs and the iris for the VSLY5850 produced the best results.
6/26/2023 update Tested the OP232 with the new .5" f = 40.0mm lens and the PLA mount. Found a relative intensity of 1.17 (compare to the OP232 with the 1" f = 25.6mm lens which had a relative intensity of 1.19).
Swadha, Evan
Previous laser: NP Photonics Rock (see MIT:11213)
Current laser: Lighwave/JDSU/Lumentum NPRO
The laser worked fine for the first week but starting a few days ago would only run for about an hour before going into a cycle of repeatedly shutting itself off every 30 seconds or so. There is no indication of thermal overload on the diagnostic fault pin, but nonetheless I suspect more heat sinking is needed. We clamped the head to a ~2" thick aluminum block that sits directly on the table. So far it has been running for 2 hours without issue, putting out around 800 mW.
Myron fabricated a new mount, but the behavior eventually recurred when the laser was left on for a few hours. We tried another aluminum mount, but the behavior recurred with that too.
To test whether this is a problem with the controller or the head, Swadha swapped the controller to JDSU FQN4113040 (manufactured May 2013).
Ran for ~2 days with the swapped controller (JDSU FQN4113040) without shutting off. It would seem the problem was with the controller.
[Eleonora, Nikhil]
We connected the output of a photodiode to the ADC channels of the cymac to check if they work properly.
We verified all the ADC channels ( X5:IOP-MHZ_MADC0_EPICS_CH{0-15} ) and it seems that only channels 0, 1, 2, 3, 4 are responding to the applied voltage. These 5 channels have been labelled. We suspect that there is a problem in the ADC cards, since we double checked the connectors behaviour. Further verifications will follow.
Nikhil, Evan
We used a function generator to drive each of the 16 channels with a 1 Vpp sine wave at 1 Hz and got a different answer. Channels 0, 2, 4, 5, and 6 seem to be broken in some way (large offsets, minimal response to signal), while the other channels faithfully reproduced the signal with no offset. We looked at the test points for channel 6 on the AA card directly and verified that the problem appears already in analog, so we suspect that some of the amplifiers on the AA cards are blown.
We also took this opportunity to clean out the blue box a bit (see attachment from before we started). The second 65 kHz ADC, along with the fast ADC, were sitting loose and unused, and they have now been removed. The ribbon cable connecting the AA/ADC adapter to the ADC was not seated securely and was causing vibrational pickup on some of the channels. The timing buffer is still a mess and the clock components are not secured to the chassis.
Some documentation:
On the assumption that the channel problems are from blown amplifiers, I ordered ten THS4131ID differential amplifiers in SOIC form. We already have an essentially unlimited supply of LT1492 dual op-amps in the EE shop. (Refer to D070081.)
The D-sub to BNC breakout cables that were attached to this chassis were a bad idea (attachment). Aside from the issue of using plugs rather than jacks, and thereby requiring the use of BNC barrels to hook up a normal BNC cable, this cable makes it essentially impossible to isolate the various BNC shields from each other and from the metal rack, which will cause the negative inputs of all the ADC channels to be shorted together.
We have one fully built 8-channel isolating BNC breakout adapter D1900426, which is meant to go into a whitening chassis but can be repurposed to interface with the channel inputs to the blue chassis.
Peter suggested using an AD630 to perform the lock-in demodulation for the RetroSense signals. The datasheet does not give a noise spectrum so I wanted to see how it performed.
I wired an AD630 on a protoboard in the lock-in topology shown in Fig. 35 of the datasheet, omitting the AD711s. I powered the board with ±15 V from a benchtop supply. I used an SR560 with a gain of 20 dB and a 300 Hz 12 dB/octave lowpass in order to do the final filtering step. I drove the comparator with a ±1 V square wave at 30108 Hz, and then drove the signal input (pin 1) with another square wave. I tried various amplitudes and offsets up to 300 mVpp. In all cases the noise at 1 Hz was a few µV/Hz1/2, referred to the AD630's output. This is not good enough for us because we want to be shot noise limited with a few milliamps of photocurrent and a transimpedance of order 1 kΩ, which amounts to a noise of a few tens of nV/Hz1/2. (The gain of the AD630 is of order 0 dB — the internal amplifiers run with a gain of 6 dB and there is some loss due to the demodulation process).
I tried the following, which did not lower the noise:
I didn't add an input buffer to the signal input like shown in the datasheet figure, but I don't see why that should matter.
By using the function generator's internal AM modulation feature, I saw that there was significant harmonic content in the voltage spectrum when the signal was sent to the AD630 with a square wave, but not with a sine wave. I do not understand this behavior since mathematically, square wave demodulation should not have this feature.
Photo shows the AD630 circuit along with the AD711 Sallen-Key filter.
I retook some of these measurements. First, the SR560 I was using yesterday seems to have excess low-frequency noise, so I suspect that was part of the confusion. Second, the circuit is picking up some kind of environmental noise, so the noise floor is not entirely stationary. At the quietest times it approached 100 nV/Hz1/2 at 1 Hz.
It is probably worth redoing this board as a PCB with on-board voltage regulators and seeing how it performs with actual photocurrent signals.
I ordered PCBs of this design from JLCPCB, each with three demodulation channels (for X, Y, and sum). This is two layers, with the ground achieved through a polygon pour on the top layer. If we come to regret this then we may want to go for four layers.
These PCBs arrived and I stuffed the X channel on one of them. The voltage noise had similar nonstationary behavior when tested as described above outside the NW22 lab, so I moved it into the lab and also added an SR560 at the output with gain.
I looked on a scope to get the overall gain factors in either the sine or square drive configurations:
Compared to the 1.25 mVpp envelope of the modulated input signal, the gain for sine wave signals is about 1 and the gain for square waves is about 2... we likely want to stick with square waves, since sinusoidal input would probably require bandpassing of the photocurrent signals, which throws away SNR.
In the attached input-referred noise curves there is no indication that square wave input has worse nonlinear content than sine waves. The input-referred noise is about 200 nV/Hz1/2 from 0.1 to 1 Hz. This competes with the expected shot noise of 3 mA of total photocurrent if it is handled with a transimpedance of 4 kΩ, which produces 12 V dc in the total voltage sent to the sum channel of this demodulation board.
I did not build the capability for ac coupling into the input to this circuit, and perhaps I should have, since the dc offset voltage into the demodulator is not needed. Also, the AD711's input voltage noise is somewhat high... though I don't think it's limiting the current performance.
Tried replacing AD711 with OPA227, did not notice a difference in the 50 Ω input-terminated spectrum. We should consider running the AD630 in a gain 1 configuration by tying pin 14 to pin 13 rather than to ground (refer to the lock-in application section of the AD630 datasheet), which should be doable on the current PCB since the AD630 is a DIP in a socket. It can and should be modified in a reversible way (i.e., do not clip pins) as this part is somewhat expensive.
The AD630 datasheet does not specify a noise performance for this IC.
Nick, Slawek, Nikhil, Lisa
We have been testing a number of laser heads today in the NW22 lab.
Lightwave M126-1064-700, SN 422 -- only 190 mW with full current, up to 120 mW with nominal setting
⇒ These two laser heads are going to be sent back to be refurbished
Unrelated to the flooding, we also tested two laser heads which have been sitting in the NW22 control room for a while.
For interlocking Lightwave/JDSU/Lumentum laser controllers, suitable connectors and contacts are as follows: