evanhall - posted 17:38, Tuesday 14 January 2025 (11865)
Flow bench

( Sławek, Evan, Swadha )

We vacuumed the soot out of the blower cabinet and installed proper furnace filters. (Refer to MIT:11859)

We noticed the fabric connector connecting the left blower to the ductwork had a gap, which we sealed with tape. There is a more proper fix to be done here by getting new connector fabric and then bolting it in with the provided brackets, but we'll skip this for now.

The left blower also had a bad power connector, which was making it run in reverse. Sławek crimped the wire again and now it works.

Swadha and I moved the flow bench up to the edge of our clean room, so that it can be accessed by rolling up the curtain. We wiped down the interior and exterior with deionized water.

Heisenberg IFO General
eoelker - posted 11:14, Sunday 12 January 2025 (11862)
V2 Fast Homodyne

I put together 2x of my fast homodyne V2.  This version includes a few changes from V1:

  1. The copper planes below the trace connecting the two photodiodes to the balanced transimpedance amplifier has been removed to reduce stray capacitance.  We estimated that this was adding about 15 pF of capacitance on V1, which shares the same layout with the aLIGO homodyne.
  2. U9 has been changed from an OPA818 to a OPA657.  The 818 has a slightly better spec, but its nonstandard package makes it very difficult to work with.

The schematic is attached.  I made a few component value changes. Unfortunately our altium license has lapsed so I cant update the schematic: 

Both boards power up and draw 0.07A from +18V and 0.08A from -18V.  We will test them with photodiodes next week.

Non-image files attached to this report
GRAVITES General
epolini - posted 08:38, Sunday 12 January 2025 - last comment - 11:38, Friday 17 January 2025(11858)
First Acoustic Noise measurements [Vienna]

[Chris, Eleonora]

Yesterday we installed a microphone (Razer Seiren v3 mini) inside the acoustic enclosure in the lab in Vienna (where the 50km interferometer is), between the two spools as shown in Fig. 1.

We acquired the noise floor spectrum turning off the microphone (using the Razer Synapse software) and the acoustic noise spectra with the AC on / off. 

Fig. 2 shows the data from the microphone with AC on and off in Volts.

Fig. 3 shows the data from the microphone with AC on and off in Volts/rtHz.

Fig. 4 compares the control signal from the loop (fast out) with the microphone data, with AC on and off. The microphone data are NOT calibrated, meaning that these data are useful only for a qualitative comparison of the peaks.

Images attached to this report
Comments related to this report
epolini - 13:19, Thursday 16 January 2025 (11867)

[Chris, Eleonora]

Today we also acquired the spectrum of the acoustic noise inside the box after turning on the compressor of the nanowire detectors. 

The attached figure shows the comparison of the past measurements with the one of today with AC on, compressor on and pump on.

 

Images attached to this comment
epolini - 11:38, Friday 17 January 2025 (11868)

I took also the measurement with the compressor on and the pump off. The structure at 1200Hz is gone as expected.

This is the acoustic noise spectrum when we want to measure single photons with the detectors in the current lab. 

I also investigated the shift discepacy between the curve taken at different times. I connected an other microphone and changed the volume level and it makes the level of the curve shift, as shown in the second plot attached. 

Images attached to this comment
GRAVITES General
epolini - posted 07:02, Sunday 12 January 2025 (11861)
Acoustic Frequency Response Inside and Outside the Box [Vienna]

[Chris, Eleonora]

We performed some measurements driving a speaker (WS-17E Driver 8 Ohm) at different frequencies. We sent signals with 20 Vpp amplitude, spanning from 20Hz to 7kHz. We placed the speaker roughly at the height of the microphone on the suspension table inside the box, see Fig. 1. We recorded the data with the microphone inside the box and another one outside the box near the speaker. 

Fig. 2 shows the amplitude of the response of the microphone for the excited frequencies, both inside (orange) and outside (blue) the box. The green curve is the ambient noise inside the box.

Images attached to this report
Facilities General
evanhall - posted 10:29, Saturday 11 January 2025 (11860)
ADBC/TIR cleanroom cleaning

Swadha, Evan

We vacuumed the floor in and around the ADBC/TIR cleanroom. Swadha wiped the curtain interiors with DI water.

GRAVITES Electronics
Haocun - posted 07:57, Thursday 09 January 2025 - last comment - 07:49, Friday 11 April 2025(11857)
GRAVITES Servo Board Updates [Vienna]

[Eric, Haocun]

We were able to lock the whole interferometer (50km fiber on each arm) with the servo board duplicated and shipped by Eric 11773.

Finding that we were limited by VCO noise by using 50kHz span [Fig. 1 VCO noises vs. different modulation deviation span], the following modifications were made to increase the servo gain for using 500Hz VCO span:

Homodyne:

Common Servo:

Split Servo:

       VCO Path:

       PZT Path:

With these cnfigurations, we have OLTF transfer function and crossover measured [Fig 2 & 3]. OLTF UGF is around 15kHz, and the crossover was set to around 200Hz.

TODO:

 

Images attached to this report
Comments related to this report
eoelker - 07:49, Friday 11 April 2025 (11932)

[Dorotea, Eric]

After turning down the input power at 1542 nm to reduce coincidence counts from raman scattering, we found that the overall gain of the servo board is now too low.  This causes the servo to become unstable when the 2nd common integrator is turned on.

To fix this, we are increasing the fixed gain on the common stage of the board by another +20 dB to compensate for the drop in optical power.  This should give us a UGF in the 5-10 kHz range, which should roughly match what we were operating with in January/Febuary.  I think we could push it up further, but we'll start with this to restore the old behavior.  

We increased the gain of the 3rd stage (U6 on the board) by 20 dB while keeping its integrator corner roughly the same.

R18:  10k -> 976 ohm

R31:  0 -> 8.87k

cSQZ General
hudsonl - posted 14:01, Monday 06 January 2025 (11855)
CSQZ State Tomography with Red Pitaya
We've been working to decrease the time needed for state tomography with the compact squeezer, the goals of which are to identify the maximal levels of squeezing and anti-squeezing in the squeezed ellipse and to identify the LO demod phase angles corresponding to each. 

For 100 data points in the phase scan with 2 MB of data each, the scan takes about 9 minutes. Reducing the data per point to 0.5 MB, the scan takes about 3.4 minutes. This indicates that we can likely reduce the time further by reducing the data per point to 1/4 or 1/8 MB, but will then start to be dominated by a constant overhead. This overhead can itself likely be reduced by only setting the amount of data acquired per data point once in the acquisition loop, instead of at every data point.
LASTI General
denis.martynov - posted 07:11, Sunday 05 January 2025 (11851)
ISI stabilisation: slower blends and witness sensors
After Rich's tests on Thu and Fri, I have reengaged ISI stabilisation with C-6D in loop. This time, I lowered the RX/RY blends to 15 mHz and X,Y to 25 mHz (attached figures 'RXY_blend' and 'XY_blend'). These are likely the lowest we can go in X,Y with the current noise level and can achieve isolation of 30 dB at microseism.

This isolation was, indeed, achieved as shown in figure 'X_ISO'. Green shows the ground input as measured by STS-2 on the floor. Black is the C6D (in-loop) signal. Blue is the GS-13 (out-of-loop) signal and red is T240 (out-of-loop) signal. I think the T240 signal is a bit above GS13's and C6D's signals because it couples Z motion with a coefficient of 1:20. C6D and GS13 are less prone to the coupling. There is also a small difference between GS13 and C6D signals around the C6D pendulum frequency of 0.67 Hz. This is due to imperfect calibration of the C6D blended signal and blue should be a better reference here.

I have also looked at X signal without X isolation loop running (but with Z, RX, RY, RZ engaged) to estimate the platform tilt reduction due to the RY loop. The result is shown in figure 'X_RY_ISO'. Similar to above, green is the ground input measured by STS-2 in units of nm/s/sqHz, red is T240's signal. Tilt is reduced above 15 mHz and the X motion is likely limited by the translational motion similar to Z (orange trace). Below 10 mHz, more tilt motion is potentially induced by the RY blend or the gap between green and red may be (at least partially) explained by the difference in responses between T240 and STS-2.

Finally, I noticed that ISI tripped last night. I think the problem may be with one of V coils, likely with V3, because I see lots of induced Y motion when I engage isolation in Z.
Images attached to this report
Non-image files attached to this report
cSQZ General
hudsonl - posted 17:37, Friday 03 January 2025 (11850)
CSQZ Optical Path Stability
I put the compact squeezer in alignment mode after returning from winter break. After about two weeks with no activity in the lab and after building a box around the compact squeezer and likely shaking the device in the process, the visibility was still 93%.

After some re-alignment, I improved the visibility to 97.4%. 

Some observations from tuning up the alignment
(1) At high visibility, there's some degeneracy between the mirror's angle and polarization. It's important to jointly optimize these to achieve very high levels of visibility.
(2) The thorlabs tip-tilt mounts are much more stable when screwed into a post tapped for a 1/4-20 screw using a 1/4-20 screw as opposed to using a #8-32 screw with a washer and a post tapped for #8-32. I was unable to improve the visibility beyond 93% using the mount with a #8-32 screw + washer, and quickly achieved 97.5% visibility after switching to a 1/4-20 screw and mount.
LASTI General
denis.martynov - posted 13:48, Sunday 29 December 2024 (11849)
ISI stabilisation with C-6D
ISI was stabilised relative to C6D in RX, RY, RZ with a blend at 25 mHz and X, Y loops with a blend of 35 mHz. Z was stabilised with GS-13 because C6D's Z mode is stiff. The RX,RY,RZ blending frequency was a bit higher than the design one (8 mHz) because of the actuation noise from the RZ offset and RX/RY degeneracy.

As discussed in log 11846, our stock was welded at a significant angle (~0.15-0.2 rad) relative to the vertical axis. I thought that it would be straightforward to correct the asymmetry digitally. However, it turns out it is a bit complex. Our general strategy is to run 6D controls in its own coordinate system that is aligned to the eigen rotation axes of the fibre. At Lasti, we need to stay aligned with the ISI coordinates and need to rotate C6D's eigen axes. This is tricky because the frequency response of RX and RY are different. First, I tried to digitally reduce the RY eigen mode from 16 mHz down to 7 mHz (to match RX's eigen mode). However, I noticed that the ISI RY -> C6D RY response develops a notch at 16 mHz from the servo. This is probably because of the RY -> X -> RY coupling and requires further investigation.

Then, I decided to do our usual feedforward of X, Y to RY, RX and add another digital filter to make RY's response similar to RX's response to the ISI tilt. This approach worked well above 7 mHz but the transfer functions are not well understood at lower frequencies yet. I think the lesson learnt is that we should disregard any fibre with angled stocks relative to the vertical axis in the future.

The first attached figure 'RXYZ_blend' shows the blending filters for RX,RY, and RZ. I think it can be further improved and my initial goal was to achieve an order of magnitude isolation at 0.1 Hz. From the noise figures discussed in log 11848, we may potentially blend angular degrees of freedom at 15 mHz. However, since I was not quite convinced of my frequency dependent rotation of the C6D coordinates, I blended angular modes at a higher frequency.

The second attached figure 'XY_blend' shows the blending filters for X and Y. Since tilt is suppressed only above 25 mHz, I decided to blend at 35 mHz with the goal to achieve an order of magnitude noise suppression around microseism.

I have also found that C6D's bandwidth is limited by the 18 Hz mechanical mode. It likely comes from the LPS posts. I have, therefore, blended C6D with GS13 signals above 2 Hz to maintain the original bandwidth of the ISI control loops.

The last four figures show the performance of the C6D isolation. Figures 'X_noise' and 'Y_noise' are quite similar. However, the Y-blend is clearly a bit more marginal than X-blend and requires further tuning. Also, I add a red dashed curve that I measured for the ISI isolation with GS-13. This was measured in log 11848 and the blends are as Rich left them before the break: 250 mHz for RX/RY/RZ and 100 mHz for X, and Y. In the future, it would be interesting to look at the out-of-loop T240 sensor.

Figure 'RX_noise' and 'RZ_noise' show  the isolation of the RX and RZ degree of freedom with C6D. Interestingly, the RZ isolation does not show an order of magnitude improvement at 0.1 Hz relative to the uncontrolled state. I think this is likely because ISI X and Y actuation excited RZ by an order or magnitude around microseism and then this extra motion was suppressed by the RZ loop.

Also, it is important to note that LPS run without frequency stabilisation servo because we are missing a temperature controller for our DFB laser. Rich and I borrowed one from the green lab and I guess it was recently claimed back. It does not seem that the LPS frequency noise was a problem in these measurements though because Z degree of freedom was still below all other signals below 5 Hz.

Non-image files attached to this report
LASTI General
denis.martynov - posted 04:35, Thursday 26 December 2024 (11848)
C6D transition into the 1000-sec mode and noise analysis
Once C-6D is transitioned to the low-noise state (or into the medium noise state in our case with the RZ offset), it can be transitioned into the 1000-sec mode. This implies that all angular control bandwidths are reduced from 0.1 Hz down to 1 mHz. We run C-6D in a high-control bandwidth during the installation and pump down. Once in a quiet environment, we need to float the mass above 1 mHz for the ISI control. We steel control the mass below 1 mHz because it may drift over time. Another key difference between the 1000-sec and 10-sec modes is that we utilise only LPS signals in the former regime and shadow sensor signals in the latter. The transition implies that the BOSEM's shadow sensor noise is no longer injected into C-6D. 

The first attached figure "RX_OOL" shows the estimated open loop transfer function of the RX degree of freedom. The servo has two goals: control the mass below 1 mHz and reduce the tilt's Q-factor. In Birmingham, we had a simpler 1/f servo because the Q-factor was only 5. However, since we now use monolithic spacers glued by a thin layer of EpoTec, the Q-factor has increased up to 125. There is another benefit of monolithic spacers regarding classical temperature noise as will be discussed below. Since the rind-down time is 1.5 hours, I have modified the servo to include the damping term for the tilt modes and reduce their Q-factor down to 5-7.

The second attached figure "RX_Y_Z" shows the spectrum measured by C-6D in RX, Y, and Z degrees of freedom. The units are nm/sqHz. There is a high coherence between RX/Y at 10 - 100 mHz that implies that the sensor measures the ground tilt in this frequency band. Above 150 Hz, the signals are limited by the readout noise from LPS. The Z degree of freedom sees the free-running DFB laser noise above 150 Hz (FITEL-FRL15DCWD-A82-19030-C 61408A02). The laser has a ~1 MHz linewidth. The noise is low-passed above 700 Hz to facilite demodulation of the laser modulation harmonics at 1.65 kHz. The noise reaches the level is ~0.7 pm/sqHz at the modulation frequency and then downconverts to lower frequencies. Note that only Z and RZ see the raw laser frequency noise at high frequencies. Other degrees of freedom, including RX and Y are insensitive to the laser phase noise because they subtract signals from different sensing heads and the matrix elements add up to zero.

The third attached figure "C6D_RX_NB" shows an estimation of the RX sensitivity. I utilise the Y-axis signal to estimate the platform tilt and a simple g/omega^2 coupling coefficient. Despite being coherent at 10 - 100 mHz, the tilt estimation does not fully overlap with the RX measurement. I think this is due to the active control of the plantform and the coupling coefficient should take into account the platform actuation with existing GS-13 blends. I think the RX sensitivity below 10 mHz is limited by the H1 actuator noise (that is still in the medium-noise state). I estimate the noise by assuming the coupling factor of 1:6 from H coils to V. I estimated this coefficient by redistributing the RZ control signal among different H coils and measuring the response in V coils.

A positive outcome of the measurement is the observed sensitivity below 4 mHz. In Birmingham, we were limited by the classical temperature noise at the band as shown in Fig 3 of arXiv:2405.13475. The noise level observed in the current version of C-6D is a factor of 100 less than in the previous one at 1 mHz. This is even without the C-6D thermal shields that we did not have time to install. There are two key factor that are now different: vacuum tank and spacers. First, I looked at the vacuum temperature fluctiations. The fourth attached figure "Temp" shows the comparison of the temperature noise in Birmingham and Lasti (the thermisters are on the C-6D frame and on the ISI). The latter noise is better by a factor of 4-7 (red vs blue). I think the residual improvement of the noise comes from the monolithic spacers as discussed above. In Birmingham, we utilised spacers that consisted of several 6.25 mm blocks glued together with Araldite 2014 II that likely tilted the mass during temperature fluctuations. We also had the stock glued to the triangular plate with the same epoxy. We now utilise a monolithic assembly instead.
Images attached to this report
Non-image files attached to this report
LASTI General
denis.martynov - posted 06:37, Wednesday 25 December 2024 (11847)
C-6D low-noise transition
Rich, Den

Last night, Rich and I have partially transitioned C-6D to the low-noise state. We could not go all the way due to the RZ offset. "Low-noise" means actuators in the low range. In air, we operate in the high range and high noise regime with zero impedance of the HAM A coil drivers that control C-6D. This is required to provide enough force (max of 150 uN per coil) to keep the inertial mass controlled in the presence of air currents and during the pump down.

In the low-noise state, we reduce the actuator range and utilise HAM A coil drivers with a bandstop filter that suppresses the DAC noise at 1 - 500 mHz. Before starting the switching procedure, we should, ideally, have all C-6D control signals equal to zero. Otherwise, transients from the switching process may imbalance the mass. We can rebalance RX/RY with magnetobullets as will be discussed below. However, there is no way to rotate RZ remotely in the current design. In Birmingham, we used to vent the tank and manually rotate the top rotation stage. Once done once, the mass comes back to the nominal position after the following pump downs. However, in Lasti, venting is a much more complex and time consuming process. Therefore, we should install a rotation stage with remote control in the future.

Step 1: RX/RY rebalancing. There is one magnetobullet per arm as shown on the attached scheme. In total, there are six Peek coils to move three magnetobullets because each coil can only push. There is only one CDS channel that controls the coils. A particular coil can be selected using the magnetogun switchbox (located inside the 19'' rack). It has 7 positions: 0 means that none of the coils is connected and numbers 1-6 correspond to a particular coil. Each bullet moves perpendicular to its arm. To select the proper coil, we first looked at the control signals V1-3. Positive control signal on a particular vertical coil means that the arm is too high and the coil tries to push it down. Therefore, move mass needs to be moved towards the arm. For example, if V1 = 0 cnts, V2 = 200 cnts, V3 = -200 cnts (all signals add up to zero) then we need to select coil 1 because it will move bullet 1 towards arm 2 and away from arm 3. 

The magnetogun block is embedded into the main C-6D medm screen. Only the bottom row of the two is utilised and the top row is kept as spare. There are 3 fields (cycles, amplitude, frequency) and one command (BANG!). The only parameter that needs to be tuned is "Cycles" that determines the duration of the pulse to a particular magnetogun (Peek coil). The default value is 30 cycles (30 / 16384 = 1.8 msec) that means that a sin wave of max DAC amplitude (set to 33,000 cnts) and frequency of 3800 Hz is transmitted to the coil. The frequency of 3.8 kH is determined by the resonance of the LC-circuit that powers the guns. A pulse of duration of 30 cycles changes the control signal in the vertical coils by ~30 cnts. Note that the movement of the bullet is quadratic with the duraction of the pulse and a 100 cycle pulse would move the bullet ~10 times more than a 30 cycle pulse.

Step 2: once all vertical coil signals are nulled (smaller than 30 cnts), we replaced high-range HAM A drivers with the ones with bandstop filters. In the current design, bandstop filters are not switchable and we just reconnected the cables. In principle, both coil drivers can be reconnected simulteniously and, meanwhile, the mass can float uncontrolled. However, since we have a significant RZ offset, we controlled RZ with coils H2 and H3 (connected to the second HAM A driver) and replaced the first driver (that controls V1-3, H1). We have then engaged filters in the medm coil filter banks to compensate for the bandstop filters and engaged RX/RY control loops. We have then transitioned the RZ control signal to H1 and replaced the second HAM A driver.

Step 3: increased the output impedance of the HAM A coil drivers. The internal output impedance of our HAM A drivers is zero Ohm. There is a separate switchbox that allows to select output impedance of 0 Ohm, 500 Ohm, 1 kOhm, etc. Note that this value means impedance per path (both input and return). For example, if 500 Ohm is selected, this means that the range (and noise) of the driver is reduced by (2 * 500 + 43) / 43 = 24.25 where 43 Ohm is the coil impedance. We have switched all vertical coils to the 500 Ohm state and compensated for the lost gain in MEDM.

Switching the horizontal coils is a bit harder, again, due to the RZ offset. It would not be possible to transition coils H1-3 to the 500 Ohm state. The second HAM A driver that controls H2-3 and magnetoguns is connected to a separate switchbox with two states: 0 Ohm and 150 Ohm (for channels H2-3). This impedance reduced the HAM A noise and range by a factor of (2 * 150 + 43) / 43 = 8 compared to the zero impedance state. We did not have time to modify the impedance of H1 and its lowest value (after 0 Ohm) is still 500 Ohm. We tried to transition H1 to the 500 Ohm state but the control signal would not have enough range to suppress the transition transient. We put the coil back to the 0 Ohm state for now.
Non-image files attached to this report
LASTI General
denis.martynov - posted 03:36, Wednesday 25 December 2024 (11846)
C-6D assembly
Amit, George, Lari, Peter, Rich, Den

On Dec 9 - 17, we have assembled C-6D and installed it on the HAM-X ISI. A few notes related to the installation:

* C-6D consists of a fused silica mass with a diameter of 50 cm suspended by a 55 cm long fused silica fibre. The position of the mass is monitored with 6 BOSEMs and 6 LPS (Laser Position Sensor) relative to the ISI. X,Y modes are at 0.67 Hz, Z mode is at 5.7 Hz and corresponds to the fibre tension of 1.1 GPa. RZ eigen mode is 1.4 mHz.

* RX/RY modes are tunable by the vertical position of the centre of mass relative to the fibre's bending point. Our fibre is a bit elliptical and welded under a noticable angle relative to the anchor. This leads to a noticable nondegeneracy between RX and RY. We tuned the mass such that RX mode is at 7.2 mHz and RY mode is at 16 mHz. Importantly, the eigen modes of the fibre determine the C-6D's local coordiate system. It may not coincide with the ISI's coordinate system (see attached). It is tilted by +30 degrees in RZ relative to the ISI coordiate system. The C-6D controls work in the local coordinate system and then we rotate X,Y and RX, RY signals before sending them to the ISI model.

* The mass hosts 3 "magnetobullets" that can be moved to rebalance the tilt. Each bullet consists of a PFTE sharf with two thin aluminium disks at the ends. The mass of the assembly is 4 g. We typically need to move it by a few um to rebalance the mass. This is achieved by sending short (msec) pulses to one of the two coils installed near the aluminium disks. AC magnetic field create eddy currents on the corresponding aluminium disk and make it work as a magnet for the period of actuation.

* RZ has differente equilibrium positions in air and vacuum. We tried to forecase the effect of buoyancy by adding small masses on the inertial mass and tracking the RZ control signal but still missed the proper position of the mass in vacuum. It seems the best RZ position is its natural one before all sensors and actuators are installed. During the next vent, we should rotate the top rotation stage (that supports the top anchor of the fibre) by 29 degrees anticlockwise (looking from the top).

* During the installation, we suffered with charging noise quite a bit that we did not witness in Birmingham. It turns out that the mass charges positively (up to 2 kV / inch) and our blue suits charged negatively (up to -2 kV/inch). The white suits that we took from the bake-out lab did not charge as much. The cleanroom curtains are charged up to - 1 kV / inch. Peter and Rich found discharging equipment that has successfully reduced the charge down to +/- 20 V / inch that is within error bars of the surface voltage meters.

* Electronics did not survive the transportation very well. Lari fixed most of the damaged cabling. Remaining problems include: high-noise HAM A driver (H1 channel) does not work, actuation from V1-3 is non-linear, one of the sat amps (V1-3, H1) has a high noise level.

* C6D overview medm screen can be accessed from the main MEDM scheen: HAMX-6D window.

Attached are the first C-6D spectra with ISI isolation off and on. The units are um/sqHz as measured by LPS. GS-13s are in loop for the isolation. Next steps should include transitioning C-6D to the low-noise state and stabilise the ISI relative to C-6D.
Images attached to this report
Non-image files attached to this report
cSQZ General
hudsonl - posted 18:31, Friday 20 December 2024 - last comment - 13:52, Monday 06 January 2025(11845)
Compact Squeezer Box
Previous squeezer measurements indicated that the compact squeezer's squeezing levels were drifting on a timescale of several minutes. I built a box around the compact squeezer to try to increase its thermal stability. The bx consists of four 80-20 posts, four walls and a roof made of black hardboard from thorlabs. The hardboard inherently releases dust particles from the material in its center, especially when cut. In an effort to keep the optics in the setup clean, the entire perimeter of the hardboard was covered in kapton tape. The hardboard has windows for the SHG's PDH PD, the CLF 2f PD, and the squeezed beam as well as cut-outs for several clamps and cable assemblies. The roof is held on by four screws + washers and the sides slide in and out of the 80-20 once the roof is removed.
Images attached to this report
Comments related to this report
hudsonl - 13:52, Monday 06 January 2025 (11854)
Unfortunately, the box did not improve the stability of the compact squeezer's squeezing/anti-squeezing levels much. As a result, we will focus on improving the data acquisition time instead of relying on the box or further optimizations of it.
TIR General
evanhall - posted 13:29, Friday 20 December 2024 - last comment - 17:26, Monday 13 January 2025(11844)
Notes for assembling picomotor rig

New Focus 8853 needs ball bearings for kinematic contact, in addition to custom mount.

New Focus 8742 needs new fuse (Littelfuse 229) and a new USB micro-B surface mount connector, which has been inexplicably ripped off the board (candidate replacement on Digikey: UJ2-MIBH-G-SMT-TR). Momentarily bypassing the fuse, the controller lights up but I did not attempt to drive it. The software installs and at least opens on Windows 10. We need a 3-to-1 picomotor concentrator cable (NF8725) unless we can find one around.

Comments related to this report
evanhall - 17:26, Monday 13 January 2025 (11863)

The kinematic mount fits fine into the picomotor portion of the mirror mount with 0.25″ steel balls.

The USB connection on the picomotor driver is not repairable and we'll have to use the ethernet port instead with a switch.

Images attached to this comment
Non-image files attached to this comment
General
evanhall - posted 16:38, Tuesday 17 December 2024 - last comment - 18:05, Monday 13 January 2025(11842)
Interference of Koheras and Teraxion lasers

Sasha, Evan

We are trying to measure the frequency noise of a new Teraxion laser by beating it against the 1550 nm Koheras used by Gravites.

We started with 1.9 mW from the Koheras and 1.8 mW from the Teraxion. We interfered them with a PM 50/50 splitter and found 1.3 mW at each of the two outputs.

We then started scanning the Koheras wavelength while watching for the beat note on an rf spectrum analyzer. At the start, the Koheras was at an offset of 129.6 pm relative to 1550.12 nm. We tuned to 1550.2924 nm (offset of 172.6 pm) and found the beat note at 58 MHz. The nominal frequency of the Teraxion is 1550.27 nm.

The dc out of the NF1811 was 1.5 V, which was enough to totally saturate the rf output with the beat note present. We steered the beam almost completely off the photodiode in order to recover a linear beat note waveform.

We set up an rf frequency synthesizer (IFR 2026) and attempted to make a PLL feeding back to the synthesizer frequency, but it would not lock even at an FM deviation of 100 kHz. We may try to instead lock the lasers to each other.

Comments related to this report
evanhall - 17:01, Wednesday 18 December 2024 (11843)

We are now trying to lock the PLL by feeding back to the Teraxion.

Enabling the modulation input, even when 50 Ω terminated, kicked the Teraxion wavelength by about 7 pm (we tuned the Koheras down from 173 pm of offset relative to 1550.12 nm down to more like 165).

We mixed the beat note (around 80 MHz) with the Aeroflex output using a ZRPD-1+ and a 5 MHz low-pass filter. The LO and RF were both about 7 dBm. This went into an LB1005 with a gain of -20 dB and a PI corner of 1 kHz, then into a series of rf attenuators amounting to -34 dB, then into the Teraxion modulation port. The Teraxion test report gives 5.0 GHz of modulation for a 5 Vpp excitation.

The PLL locked with a bandwidth of 85 kHz and about 60 degrees of phase margin. We measured the spectra of the control and error signals and data analysis will be forthcoming.

evanhall - 18:05, Monday 13 January 2025 (11864)

Fiber collimator: TC06APC-1064 (wrong wavelength)

Focusing lens onto NF1811: LB1901-C 75 mm

PM fiber beamsplitter: AFW PFC-15-2-50-FB

Facilities General
evanhall - posted 14:19, Thursday 12 December 2024 - last comment - 19:28, Friday 10 January 2025(11834)
Particle counts in NW22

Particle counts in three areas in NW22 are as follows, in units of 1/m3. Each is the average of three measurements using a Lighthouse Handheld 3013. This is with one person (me) standing about 2 m from the counter.

  Room CTN ADBC ADBC "night mode"
>0.3 µm 15 600 4300 6300 33 000
>0.5 µm 2900 890 970 6800
>5 µm 0 0 0 0

Can't really explain why ADBC night mode is so horrible. I think from the TUI documentation this is supposed to be a 25% reduction in the amount of blowing, by some measure.

We do not have readings at the >1.0 µm threshold, but the data we do have suggest the two clean rooms (on their "normal" setting, when the LEDs read 02 and ––) are better than class 100, worse than class 10.

Images attached to this report
Comments related to this report
evanhall - 18:06, Wednesday 08 January 2025 (11856)

Sławek, Evan

We measured the flow bench and some other areas with a different particle counter (Airy brand). Counts in 1/m3. The CTN clean room measurement was directly under one of the HEPA filters.

  Room Flow bench Room ADBC (off) CTN
>0.3 µm 9200 0 6400 13400 0
>1.0 µm 6700 0 1060 5300 0
>5 µm 3500 0 710 2500 0

 

evanhall - 19:28, Friday 10 January 2025 (11859)

I vacuumed in and around the flow bench's blower cabinet. The blowers are surrounded by porous black foam, which may be casting off black soot into the blowers. These should be replacted with proper pleated HVAC filters. We need two 20×24×1 filters and four 16×20×1 filters. Some effort is probably also needed to wipe out as much of the soot as possible.

GRAVITES General
epolini - posted 17:43, Monday 30 October 2023 - last comment - 15:13, Sunday 05 January 2025(11391)
Marconi Transfer Function measurement

Today I measured the Marconi's transfer function, which corresponds to the calibration factor from Volts to Hz. I sent a voltage provided by a DC power supply to the external modulation port. The input volage taken by the Marconi accordinag to the datasheet is: 1V RMS (1.414 V peak) sine wave for set deviation and maximum safe input is ±15 V. Fig.1 shows the frequency output of the Marconi measured with the frequency counter as a function of the input voltage. The FM deviation during the measurement was set to 2kHz.

The fit gives a calibration factor of 1.48kHz/V.

Images attached to this report
Comments related to this report
- 15:13, Sunday 05 January 2025 (11853)

Similar measurements on Vienna Marconi 2026B gain vs. modulation deviation span:

Span Gain [Hz/v]
500Hz 710
1kHz 1420
5kHz 7100
50kHz 71100

Haocun