posted 12:16, Tuesday 31 October 2023 (11397)
ADBC In-Cavity Mirror Transmission Measurements
- Cavity input power: 70mW
- Transmitted power on lock at mirror C:
- p-pol: 15.2mW
- s-pol: 0.61mW
- Cavity coupling in p-pol = 1.8/2.8 = 64.2%
- Took in-cavity measurements for mirror transmission. Since some mirrors have a transmission that changes sharply with angle of incidence, this provided more accurate measurements than previously measured out of cavity data.
- Transmission of mirror C in s-pol is hard to measure since it’s very low, and even the input power is low due to low mirror A s-pol transmission. But we can back it out from other mirror transmissions and transmitted power on lock.
- Updated results for T:
- p-pol:
- A: 1.49%
- D: 0.00015
- B: 0.00029
- C: 1.33%
- s-pol:
- A: 0.00053
- C (calculated): 5.7ppm
- Updated p-pol finesse = 200
- s-pol finesse = 9953 (previously measured)
- Cavity math for calculating s-pol T_C