LIGO-MIT Logbook https://emvogil-3.mit.edu/lasti/ General logbook for aLIGO activities en-us Fri, 20 Nov 2026 18:36:11 -0500 IntegratedPhotonics Chips - Squeezing measurement versus NLG prediction; a scan speed-dependent NLG drift https://emvogil-3.mit.edu/lasti/index.php?callRep=12313 <p>Author: melissa</p><p>Report ID: 12313</p><p>[ melissa, chaejin ]<br /> Chip ID: SQA run PhM 1 device 18&nbsp;<br /> <br /> Our BHD visibility drifted down to 80% over the course of our squeezing measurement (limiting squeezing to 2 dB), but we decided to analyze the data and compare to our nonlinear gain (NLG) scans which serve as a classical probe of the parametric gain.<br /> <br /> We start by fitting the 780nm scans to learn the intracavity photon number ( comes into the Hamiltonian like g0 sqrt(n) a^dag a ). These scans are Transmission through the waveguide @ 780nm versus heater voltage (converted to frequency detuning) for different power levels. Then we fit the NLG scans, which are collected simultaneously. This fit parametrizes the nonlinearity g0, nonlinear loss, among other things like waveguide interference, and allows for <a href="https://emvogil-3.mit.edu/lasti/uploads/12313_20261120180104_NLG_to_squeezing.png">a prediction of the measured squeezing vs experiment.</a>&nbsp;Then only the detection efficiency is the final free parameter in the fit against the squeezing data. We plot against NLG contrast (min NLG / max NLG) because we find there is a change in the max NLG when we stabilize on-resonance, likely due to a temperature-dependent interference phase that manifests elsewhere in our model. N<a href="https://emvogil-3.mit.edu/lasti/uploads/12313_20261120181559_device18_1120_maxNLG_vs_dwell_data.png">LG drop vs sweep rate confirmed with measurement.&nbsp;</a></p> <div>g0 = 1.969 MHz/sqrt(photon)</div> <div>gamma1 = 0.0011 kHz/photon&nbsp;</div> <div>gamma2 = 0.0512 &times; 1e-9 GHz/photon^2<br /> <br /> where we have the loss model:<br /> kappa_i(n) = kappa_i_0 + gamma1 * n + gamma_2 * n^2<br /> These two loss rates can act a bit degenerate in our data set so we will need to co-fit data sets versus wavelength and across devices.</div> <p>Detection efficiency fit: 0.55<br /> Loss budget: 0.558</p> <table border="1" cellpadding="1" cellspacing="1" style="height:277px; width:725px"> <tbody> <tr> <td>Loss budget component</td> <td>Value</td> <td>Note</td> </tr> <tr> <td><a href="https://emvogil-3.mit.edu/lasti/index.php?callRep=12311">BHD Visibility,</a> V = 0.80</td> <td>&nbsp;0.640</td> <td>&nbsp;re-measured after squeezing measurement</td> </tr> <tr> <td>Bus waveguide propagation (~1.6 mm)</td> <td>0.94</td> <td>estimated from kappa_i (~0.5 GHz)</td> </tr> <tr> <td>Output facet (Fresnel)</td> <td>0.975</td> <td>measured from transmission Fabry-Perot fit, assuming symmetric facets</td> </tr> <tr> <td>Free-space optics chip -&gt; BHD</td> <td>0.98</td> <td>Nominal free-space optics losses</td> </tr> <tr> <td>Detector Quantum Efficiency</td> <td>0.98</td> <td>nominal spec, within error bars of responsivity measurement</td> </tr> <tr> <td>Dark noise (~20 dB clearance)</td> <td>0.99</td> <td>&nbsp;need dedicated measurement... estimated on the ESA. 9.9 mW LO power</td> </tr> <tr> <td>Loss budget</td> <td>0.558</td> <td>&nbsp;</td> </tr> </tbody> </table> <p><u>Note:</u><br /> -Have since added a PBS before the NLG measurement photodiode to remove background due to orthogonal polarization<br /> <br /> <u>To-do:</u><br /> -Code up a gated swept squeezing measurement, to directly compare against swept NLG data&nbsp;</p> <p>&nbsp;</p> <!--- Output file_1_12313 div --> <div id="file_1_12313" class="commentHdr"> Images attached to this report <!-- Output break div. --> <div class="break"></div> <!--- Close file_1_12313 div --> </div> <!--- Output files_1_12313 div --> <div id="files_1_12313" class="reportDetails"> <a href="https://emvogil-3.mit.edu/lasti/uploads/12313_20261120180104_NLG_to_squeezing.png" target="blank"><img src="https://emvogil-3.mit.edu/lasti/uploads/tn/tn_12313_20261120180104_NLG_to_squeezing.png" class="outputImg" /></a> <a href="https://emvogil-3.mit.edu/lasti/uploads/12313_20261120181559_device18_1120_maxNLG_vs_dwell_data.png" target="blank"><img src="https://emvogil-3.mit.edu/lasti/uploads/tn/tn_12313_20261120181559_device18_1120_maxNLG_vs_dwell_data.png" class="outputImg" /></a> <!-- Output break div. --> <div class="break"></div> <!--- Close files_1_12313 div --> </div> <!-- Output break div. --> <div class="break"></div> https://emvogil-3.mit.edu/lasti/index.php?callRep=12313 melissa Fri, 20 Nov 2026 18:36:11 -0500 IntegratedPhotonics Chips CoupledCavity General - Mid-Y door and test cavity work https://emvogil-3.mit.edu/lasti/index.php?callRep=12312 <p>Author: mrtodd</p><p>Report ID: 12312</p><p><em>M. Todd, M. Richardson, S. Park,&nbsp;R. Mittleman,&nbsp;E. Oelker</em></p> <p>This morning, Eric, Rich, and I put the Mid-Y front door back on after I spent some time at the last minute to get the YAW picomotor working (at this point ITM pit and yaw picos both work). I bagged up a bunch of parts but still need to label them.</p> <p>This afternoon, Mitchell, Sun and I attempted to realign the cavity. I had already mostly aligned the prompt REFL beam using the ITM picos. Mitchell and Sun removed the ETM so that we could realign the input beam to hit the first output steering mirror and keeping the beam on REFL with the ITM pico. Then we got the beam centered on the TRANS camera and PD, before adding the ETM back in. We struggled to find the ETM reflected beam for awhile but Sun recommended we use the IR viewer, and that proved super useful. I could immediately see we were way off the ITM and we used it to move the beam back to the ITM and then continued bringing the sequential bounces on top of one another. Before long, we had the cavity flashing and I touched up the input alignment and ITM pico and we had a great error signal.&nbsp;</p> <p>Then Mitchell and Sun re-wired everything to get the cavity locked. We took an OLG measurement of the lock, and it looks pretty much identical to the previous setup; the only difference is needing a slightly higher overall gain in our servo (20dB), but we still had the same UGF (~30kHz) with 45deg phase margin.</p> https://emvogil-3.mit.edu/lasti/index.php?callRep=12312 mrtodd Fri, 20 Nov 2026 16:42:35 -0500 CoupledCavity General IntegratedPhotonics Chips - BHD specs between SQA squeezer and Helens mode-match chip https://emvogil-3.mit.edu/lasti/index.php?callRep=12311 <p>Author: melissa</p><p>Report ID: 12311</p><p>[ melissa, chaejin ]<br /> Chip ID: SQA run PhM 1 device 18&nbsp;<br /> <br /> Specs of the balanced homodyne detector (BHD) for our Nov 19 squeezing measurements<br /> Power before/after the mode-matching chip (Helens run, straight waveguides): 1.5 mW / 0.94 [ ~63% ] consistent with single-fact coupling efficiency + some fresnel reflection order of a few percent<br /> 1560 AOMs set to -1 dBm each at 80 MHz (up/down), ~216 uW after the squeezer chip (input objective focus optimized for 780nm pump)<br /> Visibility is measured at 91.83 % +/- 0.61 % between the SQA and Helens run chips<br /> Lower than our max of ~97% Helens-Helens, so likely the different chip thicknesses reduce the outcoupled mode overlap.<br /> <br /> Visibility over time (pump blocked):<br /> 13:48&nbsp;&nbsp;| 91.49 % +/- 0.82 %<br /> 14:00 |&nbsp; 91.03 % +/- 0.81 %<br /> 14:15 |&nbsp; 91.14 % +/- 0.78 % [taken before the measurement]<br /> 16:18 | 79.68 % +/- 0.60 % [taken after the measurement]<br /> <br /> At 350 mW setting on the EDFA, the LO power after the chip is 9.9 mW. Results in about 20 dB shot noise clearance @ 5-10 MHz, which is a quiet band for us today. Our RF noise environment picked up by the detector seems to change day to day, we will record the variations in a later log.&nbsp;<br /> <br /> Will perhaps swap in a SQA mode matching chip to try to increase visibility (SQA-SQA) if the squeezing result looks sensible. Post on the squeezing data/modeling to follow.</p> https://emvogil-3.mit.edu/lasti/index.php?callRep=12311 melissa Thu, 19 Nov 2026 20:20:21 -0500 IntegratedPhotonics Chips