SQZ General
dmdesant - posted 17:20, Wednesday 18 February 2026 (12140)
LIGO EtherCAT TEC Controller Functional in LASTI - Loop Parameters for VOPO Under Vacuum

[Daniel, Eric]

Eric and I have been fussing with the Beckhoff and various hardware for the past few days. As of now, were able to bring the LIGO TEC controller and MEDM interface back to life. With the chamber pumped down with just the mechanical pump, we were able to use these loop parameters at a pressur of 4.8*10^-2 torr.

Further testing with the chamber pumped down to ~10^-4 torr (with the turbo pump running) shows very little change in the step response. We are going to run this controller for now and get the squeezer back up and running.

Images attached to this report
GRAVITES General
dorotea - posted 17:35, Friday 13 February 2026 (12139)
Issues of unidentified origin with switched setup

I'm continuing to see weird behavior I don't understand on the IFO. This includes:

I'm pretty lost as to why I'm seeing all of these changes as I don't really see the connection between the new odd behavior and anything specific I've changed. I'm leaving everything turned off over the weekend just in case.

General
richard - posted 03:37, Tuesday 03 February 2026 - last comment - 15:36, Wednesday 11 February 2026(12136)
Some CRS1 data + TFs running overnight

[Bruno, Arnaud]

On Friday, Bruno helped me recover the NKT laser for CRS1 (see log 2560).

[Rich, Michael, Arnaud]

*CRS1 damping, medm and iso script

We commissioned the CRS1 damping using ISI actuators (in lieu of the ESD damping) to reduce the ring down wait time. The damping can be ran via the CRSRX_BLND filter path (FM2) and either injected at the error point of the closed loop (with a gain of 1), or in the open loop state with a gain of -0.01. The filters can be accessed via the 6D blend filter that were updated to include the CRS RX filterbank (see the new rx path on the bottom left in the medm screen attached).
Also, a python script was written to turn on the iso loops (faster than the matlab does). It assumes the ISI damping is on. The script can be found under ~/Desktop/CRS/script/turn_on_iso.py

*CRS calibration cross check

Today we took some data comparing the CRS1 and the other ISI channels in the damped and iso state to verify calibration above 0.1Hz. The attached pdf shows the rx cps, gs13 and crs1 calibrated channels in the off state (page 1), damped state (page 2) and iso state (page 3). The dtt template is under ~/Desktop/CRS/dtt_templates/dtt_hamx.xml

OFF and DAMPED state :

ISO STATE (all loops on, 250mHz CPS-GS13 blend, no sensor correction)

*CRS1 drift

The CRS has been drifting over long periods. We observe a 12h cycle which is likely temperature induced, and a longer trend which might be vacuum induced. Between Jan 31 and Feb 1 it drifted by roughly 7.5mrad (see attached), and has been turning around since then.

*CRS1 TFs

Low resolution TFs will be running overnight to extract the CRS response.

 

Images attached to this report
Non-image files attached to this report
Comments related to this report
richard - 15:36, Wednesday 11 February 2026 (12138)

The CRS1 response was measured, with results and fits posted here.

GRAVITES General
dorotea - posted 18:02, Monday 02 February 2026 (12137)
Servo 1 UGF drift

When I checked the TF on servo 1 this morning, the UGF had jumped from 22kHz to around 100kHz. Eric suggested this might be happening because of some kind of twisting on the piezo-- that as the fiber stretches the polarization might be changing enough to significantly change the contrast. I tried driving the piezo up to 5v (of the total ~8V range) and it didn't seem to change the fringe height that much-- it sits around 150mV and everything acts pretty much as expected. 

Since I'm not sure what could be causing this effect, I'm now working on getting arm 2 locked to see if I can replicate it. Currently both arms are reassembled and recabled but I'm not seeing fringes on the scope anymore, so I've clearly got some connection off to track down.

SQZ General
alossius - posted 21:43, Friday 30 January 2026 (12133)
LASTI SQZ reboot since pump-down, OPO mode-matching and PMC PDH transfer function and noise spectrum

[Daniel, Eric, Anders]

Last week (week of Jan 19, 2026), we resumed work on the LASTI SQZR.

OPO Vacuum Chamber

The VOPO chamber has been under vacuum since Eric and Daniel pumped it down on Nov 20th (see alog). On Thursday (the 22nd), the pressure in the chamber was aroud 5e-7 mbar when we got into the lab. Eric disconnected the pump from the chamber, and we observed the rate at which the pressure increased. By the end of thursday, the chamber pressure had increased to something on the order of 10^-5 mbar. By the end of the day the next day (Friday), the chamber pressure was around 2.5e-4 mbar. This pressure rise indicates the presence of imperfections (somewhere) in the seal, on top of the out-gassing effects. However, with the pressure staying above 10^-3 mbar over the course of 48+ hours, we know now that we can re-pump down the chamber every few days or so to maintain an ideal operating pressure. 

SQZR

It had been a while since the LASTI squeezer has been operational - not since the VOPO chamber was pumped down. We went through the standard procedure of turning everything on (locking SHG, OPO, PMC loops) and were pleased to see that there had been no significant misalignement. We did however, notice that the seed power present on sqz+readout table(s) was very low, which we traced back to a misalignment issue with the seed fiber coupler on the main laser (1064nm) table. This was quicly fixed and everything was back to operation. 

Daniel and I did a quick check of the seed and pump mode matching into the OPO cavity. The results are tabulated below, and the cavity scans can be seen here.

  Power in Fundamental (V) Total Power (V) Mode Alignment (P_Fund/P_Tot)*100
Seed (1064nm) 1.66 2.135 77.7%
Pump (532nm) 1.60 2.026 79%

*** These mode matching values are on-par with what has been measured previously (75.6% for seed/clf (1064nm), 77.4% for pump (532nm)). 

OPO Crystal Temp Control

On a seperate note, we were unable to get the Newport temperature servo for the OPO to lock without oscillating. We will need to set up the LIGO temperature controller again now that we're operating under vacuum.

PMC Loop & Controls

On Friday (23rd), we took spectra of the PMC loop's error signal and an open-loop transfer function (OLTF) using the SR785. The objective was to get a plot of the calibrated, free-running and rms error in the PMC loop. To do this, we needed to convert the y-axis of the noise spectrum to units of meters/sqrtHz and use the OLTF to divide out the loop suppression and get the free noise. To get the proper units for the noise spectra, we measured the PMC's FSR and FWHM using the DC transmission PD to get a conversion from time to meter, and measured the slope of the PMC error signal (in the linear regime near resonance) to get the conversion between volts and time. We fit the OLTF to a model of the TF that Eric had. The summary of this process is outlined below:

* The FSR was measured to be ~449ms, however, scanning the cavity with the PZT does not give us a clean (full) FSR, so this value is most likely imprecise. To get a better reading, we should scan the cavity using the laser's frequency. 

* The linewidth (FWHM) was measured 4 seperate times - each measurement was pretty different: 200, 542, 310 and 788 us - unclear to me at this point why these measuremetns differ so much. 

* Using these two measurements, the PMC cavity (p-pol) finesse can be inferred, however, the significant deviations in linewidths results in equally different values for the finesse: 2245, 828, 1448, 570. Furthermore, none of these finesse values land within 50% of the documented cavity finesse in the PMC DCC document from [year] (see LIGO-T2300300-v5 - link unavailible at this time as I don't have DCC access yet), which is 220. 

* Due to this inconsistency, the cavity finesse was taken to be the 220 value from the DCC document for the remainder of the analysis. 

* To get a conversion between time and meters, we use the cavity finesse and laser wavelength to get the value of the linewidth in meters, and compare this value to the measured linewidths in time (*which I'd like to remeasure soon*). 

* To get a conversion between meters and volts, we use the slope of the error signal with the x-axis converted to units of length using the time-to-meter conversion just mentioned. This slope then has units of volts/meter and gives us this conversion. The data collected and used for this conversion can be seen in the table below:

dY [mV] dX [us] dX [nm] (multiply by time-meter conversion)
768 42 44.1
578.25 98 1.03          *clearly an outlier*
486 48 50.4

* The uncalibrated noise spectrum of the error signal was recorded using the SR785 FFT function. By dividing the y-axis (nominally in units of Vrms/sqrtHz) by this slope (volts/meter), we obtain the calibrated error spectrum in units of meters/sqrtHz. A plot of this calibrated noise spectrum can be seen here.

* The open loop transfer function (measured using the SR785 via the excitation point on Eric's servo) was taken and plotted on top of the transfer function model. In order for the measured TF to match the model, the model's overall gain parameter was bumped up from 25 to 42 dB. A plot of the measured and model TFs can be see here. At the point of this measurement, the UGF was around 1.4kHz, however a few days earlier the loop was stable at a higher gain (~16dB vs this at 4dB) which corresponded to a higher UGF of ~4kHz - we do not fully understand why the loop's lockpoint changed over the few days between locks. 

* Using the model for the TF, we computed the free-running error in the system, by multiplying the calibrated error with |1-OLTF|. We then computed the running rms error from high to low frequencies, and plotted the three curves here.

This data will serve useful for further loop tuning. 


 

Images attached to this report
GRAVITES Electronics
dorotea - posted 18:20, Friday 30 January 2026 (12135)
Servo 1 fixed, locking

Servo 1 is fixed and locking. Eric found some issues with the board and has updated the schematic for future reference. OLTF and crossover measurements are attached-- UGF is around 22kHz with  ~130deg phase margin and crossover is around 200Hz. These measurements were taken with PZT1 in the path; this piezo appears to have a resonance around 4.5kHz so I'll be redoing the notch filter on the split board to deal with that. 

Images attached to this report
LASTI General
ME - posted 13:44, Thursday 29 January 2026 (12134)
HAMX Power

The HAMX DC powewr was turned off for a couple of hours today, there was a problem with the grounding of the Sorenson power supplies. 

The HEPI pier boxes are still not getting power, which is what sparked the power supply investigation.

General
richard - posted 11:31, Friday 16 January 2026 (12132)
CRS Noise Spectra

We took some noise spectra of the CRS with and without sensor correction on, see attached. The CRS is exceeding the ~prad/sqrt(Hz) goal above 20 Hz.

With sensor correction on, the noise roughly matches the T240 noise from 0.1 - 20 Hz and is dominated by residual motion coupling through nonlinearities of the readout. The common signal (i.e. rotation) supression in the noise estimates is typically limits to roughly two orders of magntitude. The signal leaks into the noise due to nonlinearities in the readouts. This is why sensor correction decreases the noise in the CRS. We expect this to decrease even further when we put the sensor in loop. Below 0.1 Hz, the CRS is well below the seismometer noises. The CRS noise in these plots is just the readout noise so it's expected to be less than the goal noise below 0.5 Hz. Below 0.5 Hz, the goal is dominated by the thermal noise of the CRS which would only show up in the angular readout.

Non-image files attached to this report
SQZ General
dmdesant - posted 16:54, Monday 12 January 2026 (12131)
LASTI Homodyne PD Reflection Investigation and LO PMC Realignment

[Eric, Daniel]

Today, we tried to see if we could detect any reflections off the photodiodes in our homodyne and discovered some interesting behavior. 

To do this, we needed to recover our input alignment from the LO fiber into the LO PMC, as it needed to be removed when Eric and I built the box around the readout table next to the chamber. A decent alignment was recovered and the PMC still locks okay with one caveat. The ~1 kHz vibrations from the turbo pump are really pronounced in the transmitted power through the PMC. Before we resume squeezing measurements, it may be a good idea to revisit the PDH loop transfer functions to make sure eveything there is working okay. Nevertheless, a noisy lock was achieved (blue is transmission and yellow is reflection in this scope trace) and we continued to try and detect some reflections from the PDs on the homodyne.

With 2mW of LO leaving the PMC, we measure almost exactly 1mW of LO power incident on each PD. The right-most PD exhibits a very weak reflection at roughly the expected angle from the incoming beam (see detector card in this image). Placing a power meter on this very dim spot revealed that there was only 2uW of reflected power in this beam. That constitutes a 0.2% reflection loss, which is not too bad. However, the left-most detector is more interesting. There appears to be a smeared beam nearly retroreflected from the incident beam (see detector card image here). Since this beam was so close to the incident beam, it was impossible to get a power detector in there to measure the reflected power. I would estimate that the brightness on the card was about 5-10X brighter and over a much larger area, so clearly there is a significant loss here.

Adjusting the turning mirror for the left-most PD affected the position and pattern of the reflection significantly. It seems like this is a result of the input beam on the photodiode somewhere, but it was unclear to me exactly how. It is possible that the beam on the left-most PD and right-most PD are mismatched as they take slightly different path lengths before being focused down, but this should be minimal with our current mode-matching setup. 

In any case, it wouldn't be too hard for me to believe that this accounts for the 1-2% excess loss we discuss here. Some close realignment of the homodyne may be all that is necessary to mitigate this.

 

 

Images attached to this report
Heisenberg IFO General
hudsonl - posted 15:51, Thursday 08 January 2026 (12130)
Heisenberg Marconi Cabling
Daniel and I removed some cables from Marconis used to actuate AOMs in the Heisenberg setup so we could use them temporarily to measure fiber phase noise for the OEO project. The cables for each marconi are cable-tied together and labeled according to which port they should be plugged back into to run the Heisenberg experiment. The images attached show the proper settings on the Marconis for the Heisenberg experiment. The devices should be returned to these settings when we return to collecting data on the Heisenberg experiment.
Images attached to this report
LASTI General
dennis.martynov - posted 11:09, Monday 05 January 2026 (12129)
ISI stabilisation with upgraded C-6D
After C-6D's RZ offset was removed and (as a consquence) the actuator noise has improved, C-6D started to measure the ISI tilt from 6-7 mHz (log 12128). I decided to update the C-6D filters and blend both RX and RY with CPS at 7 mHz. The updated control scheme has improved rigning of the previous 25 mHz blend (log 11906). I have also updated the RZ filter and blend C-6D with CPS at 19 mHz but we do not have a witness sensor for it.

Three attached figures show the translational degrees of freedom measured by the out-of-loop T240 sensor compared with the ground STS-2 sensor. Z is controlled with BIS / GS13, other degrees of freedom are controlled with C-6D / GS13. The suppression factors make sense as given by the blending filters. The X/Y control filters may look a bit odd because we achieve a larger supppression factor at 0.1 Hz (x20-30) than at 0.2-0.3 Hz (x10). I guess this should be close-to-optimal control scheme for BSC stage 2 because its state 1 T240 blends inject a bit of tilt to the platform at low frequencies. It might be good to suppress that and also gain another order of magnitude at microseism. We can fine tune the blends once we complete the qualitative coupling model of the platform motion to DARM.
Non-image files attached to this report
LASTI General
dennis.martynov - posted 12:29, Sunday 04 January 2026 (12128)
C-6D current angular noise
As mentioned in log 12121, Amit and George have rotated the C-6D top plate to compensate for the RZ offset (discussed in 11846) that we have introduced back in Dec 2024 likely because we overcompensated for the air currents. This time we noticed only a small change in the RZ control signal in air (when the chamber curtains were closed) and after the pump down (150 cnts with 0 coil driver impedance = 0.28 uN m of torque that is likely caused by T change or buoyancy coupling via other degrees of freedom).

We still have two items to fix / improve:

* install thermal shields

* replace the fibre to remove the RX/RY assymmetry

Still, I decided to look into the C-6D noise budget more carefully and compare its noise with BIS' (log 12123) and T240's sensitivity. The noise budget is shown in the first attached figure: the red trace shows the raw RY signal and the green trace shows the calibrated spectrum. At low frequencies, the measured motion is limited by the classical temperature noise. The actuator (DAC) noise and thermal noise are significantly lower than the classical temperature noise in this state (without the T shield). C-6D measures the platform tilt from 7 mHz. I tried to assess the C-6D own noise by subtracting the tilt spectrum measured by the X degree of freedom from RY. The result is shown by the magenta trace. Above 30 mHz, we cannot assess the C-6D motion because X is limited by the translational motion of the platform. There is likely a bit of miscalibration at 17 mHz (the RY eigen mode) and the C-6D resonance is still seen. I may try to correct this.

The second attached figure shows the C-6D RX, RY estimated noise spectra (RX - Y/g and RY - X/g) and their comparison with BIS and T240. Similar to RY, RX is likely a bit miscalibrated around its eigen mode of 9 mHz. In the 7-40 mHz band (where I think the tilt coupling is most critical), C-6D has a factor of 4-5 lower noise compared to BIS and a factor of 15-20 lower noise compared to T240's noise. The next hardware update should likely be the installation of the thermal shield.
Non-image files attached to this report
LASTI General
Rich - posted 14:32, Wednesday 31 December 2025 - last comment - 07:24, Thursday 01 January 2026(12126)
Ground STS2

Plugged in the ground STS2, not completely sure that it is working, but i think so. The alignment is going to need to be checked, and it will probably need to be recentered next time that i am in

Comments related to this report
dennis.martynov - 07:24, Thursday 01 January 2026 (12127)
Thanks, Rich. Attached are comparison figures between STS-2 and T240 with BIS / GS-13 in-loop. Also note that STS-2 has additional noise in Z below 20 mHz as shown in the attached figure STS2_Z_noise when ISI was only damped. Once stabilised, there should be no ISI stabilisation below 20 mHz in Z.
Images attached to this comment
Non-image files attached to this comment
LASTI General
richard - posted 00:04, Monday 29 December 2025 (12125)
Restart m1fb0 DAQ to add DQ channels to m1isicrs

[Arnaud] 09:00 PDT

LASTI General
dennis.martynov - posted 03:53, Saturday 27 December 2025 (12124)
ISI stabilisation with BIS
The platform is currently stabilised with BIS in loop. The blending frequencies are X: 40 mHz, Y: 80 mHz, Z: 20 mHz, RX, RZ: 67 mHz, RY: 32 mHz. I think the X/RY blends are a bit conservative with some room for improvement but I tried to avoid ringing the blends during the first iteration. The Y/RX blend is a factor of 2 higher than Y/RX because we have one of the station 2 coils shorted relative to the ISI that likely couples extra noise (as already mentioned in 12122). The RZ blend is also a bit higher than RY because the seismometers are tuned to couple 10% of the horizontal motion. The Z blend can be as low as 5-10 mHz from the noise figures but I decided to start with a more conservative value.

I also blend BIS with GS13 at 0.3 Hz. By design and from the noise figures, we should blend at 0.7 Hz. However, we had to set the seismometer force feedback UGF at 1.5 Hz vs the design value of 4 Hz because we excited the first parasitic mechanical mode of the proof mass at 35-40 Hz during the pump down. We will need to retune the low-pass of the analog servo and then shift the force feedback to 4 Hz in the future. Further analysis is needed but at a first glance it seems that BIS / GS13 blending at 0.3 Hz does not couple much GS13 noise at low-frequencies.

The sensor correction is currently off because the ground STS-2 seismometer seems to be turned off.

I have retuned the notches in the ISI isolation filters because we run up new mechanical modes around 84 Hz and around 125 Hz. It seems that the new sensors (BIS, CRS) have not added any low-frequency parasitic mechanical modes.

The attached figures show the spectra of the out-of-loop T240 sensor. Since STS-2 seems to be off, I had to take the reference trace a couple of hours in advance and the seismic spectrum might have changed a bit. However, overall, I think the spectra make sense: we should achieve a suppression of a factor of 30 at microseism for Z, a factor of 10 for X, and a factor of 2-3 for Y. There is also an enhancement in noise below 100 mHz that may or may not be related to the elevated tilt during the platform stabilisation. Will need to assess this in the future. I also think that T240 is rotated by 0.2 rad relative to the ISI X/Y axes and mixes our basis a bit. Not a big deal for now but might be good to rotate T240 in the future.

I think these slower blends compared to the GS13 ones (100 - 250 mHz) look promising and should be able to improve the suppression of the vertex motion and reduce MICH / SRCL / vertex WFS noise in DARM below 25 Hz. Similar to Hang Yu's 2017 5-Hz paper and similar to the C-6D's paper, the propagation of the platform motion via the suspension and the interferometer's global control system is required (and ongoing) to qualitatively assess the improvement.
Non-image files attached to this report
LASTI General
dennis.martynov - posted 07:07, Thursday 25 December 2025 - last comment - 05:34, Friday 26 December 2025(12122)
BIS diagonalisation
As noted in log 12121, the Birmingham / Tsinghua team has installed our 1D sensometers inside the HAM spring hutches. We call the seismometers BIS as a short of bison because the seismometers look like these horned fellows (at least we tried to design them this way). This abbreviation also stands for the "Birmingham Inertial Sensor".

Our goals were the following:

* build seismometers with a 5-10 improved noise compared to T240 in the 10 - 100 mHz band (required for optimal ISI stabilisation). It seems that all modern Trilliums, Guralps, STS-2 follow the Wielandt's approach and rely on leaf springs where antispring is created by the spring's shear force. We followed a similar approach with lots of smaller and one key difference:

* make them UHV and fit inside the spring hutches. As far as I know, T240 cannot fit in these places. UHV environment provides both opportunities and challenges. The key opportunity is that air damping in the small gap between the capacitive plates limits the thermal noise of the state-of-the-art commercial sensors. The key challenge is that it is difficult to utilise capacitive sensing in vacuum with picometer resolution because the amplifier must be near the seismometers in vacuum.

In summary and so far, the seismometers managed to achieve a factor of 3-4 improved sensitivity compared to the T240 seismometers and a factor of 2 worse than their design sensitivity. This will be investigated during the commissioning period. The first attached figure shows the noise level from Dec 24 in RY degree of freedom. The ISI tilt has been selected because this is the quitest angular degree of freedom and may reach down to the seismometer own noise:

* The red and gray trace compare BIS and T240 Z acceleration. There is some extra noise in T240 above its self-noise and above BIS. This might be due to magnetic coupling or due to imperfect diagonalisation of the seismometer.

* The green and black traces show the inferred ISI RY tilt acceleration from the two vertical BIS signals and from T240 X signal (that couples tilt with a usual g/omega^2 coefficient). The figure shows that we measure the ISI tilt above 25 mHz and sit on some other noise at lower frequency. The noise is above the design and will be further investigated.

* The cyan trace attempts to reach the BIS's self noise by subtracting the ISI tilt measured by T240 X. We reach acceleration of 0.2 nrad /s^2 / sqHz at 25-50 mHz. This is a factor of approximately 2 above the design shown by the orange trace.

* The blue and magenta curves show the T240 self noise and GS13 self noise for comparison.

The second attached figure shows the same as above but in units of rad/sqHz.

As mentioned in log 12121, we have a cabling problem with station 2: one of the coil wires is shorted relative to the ISI. This creates a ground loop in the system and Lari estimated that 1/4 of the coil current returns via the ISI ground. This likely couples noise from the ISI ground to the seismometer. Indeed, RX seems noiser than RY because RX forms by all three vertical signals.

This noise is shown in the third attached figure: RX vs RY. The figure also shows raw seismometer signals that are tuned to S_1/2 = Z +/- 0.1 H. The H-coupling coefficient is tuned as a compromise between the horizontal sensitivity and seismometer diagonalisation (we require more than 60 dB subtraction of horizontal signals from the vertical ones to reach the design sensitivity).
Non-image files attached to this report
Comments related to this report
dennis.martynov - 05:34, Friday 26 December 2025 (12123)
I've fine tuned the BIS diagonalisation matrix to include the horizontal coupling from the orthogonal direction to each seismometer station. The coupling is one the level of 0.05 - 0.1% but still adds a noticable amount of signal below 7 mHz. Will be interesting to see how stable are these coefficients over long periods of time (due to the spring relaxations) and, most importantly, over the ISI tilts on the order of sub-mrad (this we should be able to model accurately).

The gap between BIS and T240 still stays at 4-5 at 10 mHz and the gap between the measured BIS noise and its design is 1.5 - 2. T240 has managed to reach its self noise at 5-10 mHz likely because of the reduced B-field during the measurement. I briefly looked at the T240 data during the day and the noise level is around 10-30 nm/s^2/sqHz below 30 mHz. We measured the magnetic noise level of ~0.1 - 1 uT / sqHz in this frequency band during the day from the subway trains. The measurement is strongly influenced by the position of the magnetometer because the HEPI structures seem to be magnetic. Since T240 is further away from the magnetic structures than the spring hutches, I estimate the B-field near the ISI centre on the level of 0.1 uT / sqHz. This implies that T240's delta E effect is 0.1 m/s^2/T that I think corresponds to the 30% cold rolling of their (likely) Ni-Fe spring alloy. Our spring is 40% cold rolled and we measured the coefficient of 0.2 m/s^2/T in Birmingham (if installed the proper end to the proof mass) but our seismometers are inside the mu-shield with measured efficiencies of 100-300. Therefore, I would expect that BIS would be less sensitive to B-field by a factor of 50 - 100 compared to T240. Looking at the day time data, I think we are at least a factor of 5-10 lower. Will need to explore the delta E in more details during the commissioning period.
Non-image files attached to this comment
LASTI General
richard - posted 04:26, Wednesday 24 December 2025 (12120)
CRS installation on HAMX

[Tom Roocke, Arnaud] with help from Rich, Bruno (and the Birmingham team)

Summary: Last week (Dec 15-19) we assembled and installed two CRS (Cylindrical Rotation Sensors D2300100) on the LASTI HAMX-ISI. The CRSs had been shipped mostly assembled from CIT the previous week (see cit log 924). Mapping of the installation can be found under the wiring diagram and pictures under D2300465-v6 (also attached here in pdf1 and pdf2).

CRS balancing : The two CRSs were balanced on day 1 in a cleanroom in the Orange Lab. We used a +/- 1mm range IPS sensor mounted on the table with optic post, looking at the tip of the wing to measure and adjust the CRS resonant frequency (see picture1, picture2). The cleanrooms needed to be turned off to measure the resonant mode. On first hang, we used 14um flexures on one CRS and 15um on the other. After a few hours of balancing, one of the two CRS broke, and the other one next to it broke off instantly after from the shock wave through the table. We tried hanging the 3rd CRS with 13um flexures, but they tore apart at soon as we suspended the proof mass. The last two succesfull hangs were done using 18um (CRS1) and 17um (CRS2) flexures. We didn't need to adjust the vertical COM for the proof mass to be suspended with a reasonably low frequency (<25mHz) which means the COM position is very close to the design. Also, the new mass at the top of the cylinder to adjust the horizontal COM has proven very useful for balancing. While the IPS was useful for tuning the frequency, its range was too small, and the CRS wing kept touching the sensor. We should find another one with a larger dynamic range. We don't know why the first set of flexures broke. Maybe flexures <15 um are too fragile to handle? UW did their CRS locking test with a 12um flexure, so it's not clear. The ones we used were all baked at 120C for 48hours even though this should not have an effect on the material properties.

Electronics: On day 2 Tom modified the CRS preamp (D2300154) to allow the LPS readout through that same board with a 10k TIA and 1V bias voltage by respectively modifying R7 from 5kohms to 10kohms on preamps G, H, I, J, K, L and modifying R50 from 50kohms to 10kohms on all bias voltage regulators. Those gains were chosen based on the max bias voltage of 2V and max reverse current of 2mA (and a factor of safety of 2) on the LPS FCI-InGaAs-1500 pd (datasheet). The in-air breakout cables were also fabricated to connect the db25 to the various preamp inputs and dac drives. The 1550nm toptica laser is run with a laser driver D1200719, modified with a new R14 (value to check) to allow a deeper modulation depth. This chassis was borrowed from Birmingham.

CDS: A barebones m1isicrs was first created (thank you George!) and simulink blocks were added copied from CIT for the HOQI/CRS and from 6D for the LPS. The simulink models are under /opt/rtcds/userapps/release/isi/m1/models/m1isicrs.mdld which is not kept under svn revision. A HAMX-CRS screen was added to the LASTI sitemap to access to the various signals and filterbanks (see MIT_CRS_medm.jpeg)

Sensing: CRS1 is sensed with HOQIs loaned by UW (latest design D2500104-v4) and mounted on either side of the proof mass. CRS2 is sensed with LPS loaned by Birmingham, with a custom mounting plate and a 1/2" siskiyou mount as interface. As recommended by George, the retroreflector mounted on the wings of CRS2 was switched to a 1/2" 1550nm mirror epoxied with epotek-353ND cured at 150C in an oven. The displacement readout was commissioned for the HOQIs but not yet for the LPS. The tilt of the proof mass will be measured by taking the difference and dividing by 2 times the lever arm. 

Installation in chamber and sensor alignment: The two CRSs were moved from the lab to the chamber side. The sensors were mounted to the CRSs and plugged in their readout chain to check the signals were mapped correctly to their respective ADCs (see picture3). We also ran the light path to the CRS sensors chamber side to make sure it was making it all the way to the sensors. When those checks were done, we moved the CRS (while locked, very carefully) in the chamber as shown in the video linked here. The flexures did not break during this installation process. We aligned the sensors by maximizing their fringe visibility. This was a very long and tedious process for both set of sensors. We had to change the way the springs are loaded on the CRS legs to help us maintain the alignment when changing the plate position (as shown in the D2300448 redline). It is probably best to do the sensor alignment step in the lab (if a test stand with spare laser and preamps are available). The LPS were aligned by adjusting the siskiyou mirror mount pitch and yaw adjusters with the CRS locked in its middle position and modulated the laser frequency through the TEC driver by sending a 0.1Hz signal. For the HOQIs, to get fringes, we let the CRS swing by restricting the lockers to a few mm motion on the wings. After this step the CRS frequency was checked (with HOQI signals for CRS1 and IPS for CRS2) then locked for ISI balancing.

Since pump-down: On Saturday Rich and Amit balanced the ISI, closed the chamber, released the CRSs and started pumping down. The two CRS were showing signs of life, and the picomotors were remotely adjusted for rebalancing. While we have a displacement readout channel for CRS1, we rely on fringes for CRS2 to estimate motion until the demodulation scheme is comissionned. The frequencies of CRS1,2 were measured at 14.5mHz and 18.4mHz respectively (see fig1 and fig2). Damping with the capacitive plates was comissioned for CRS1 (see fig3) and a first low frequency spectra was measured, after digitally correcting for the ellipticity of the sensors see pdf3.

For CRS2, without a displacement channel and fringes visiblle for a fraction of the full CRS range it proved difficult to recenter remotely. After picoíng for about 2 hours last night, I eventually lost trace of it, even when going through a full rotation (we should see fringes since the CRS goes through a full cycle, but we don't). Something mechanical may have happened to CRS2. We should take a look in chamber via the viewports at next opportunity.

We will continue monitoring and commissioning the signals remotely. 

Here is the summary for each of the CRS: 

Images attached to this report
Non-image files attached to this report