Yulin, Denis, Rich, Bruno
I visited MIT last week to test the laser position sensors (LPSs) on the ISI table inside the HAM-X chamber. The LPS is an interferometric position sensor that will later be integrated with HoQI and COBRI on the SenIT suspension table for combined readout and noise-performance testing. Since the suspension table was not ready before my final day at MIT on July 10, three LPSs were instead installed on the HAM-X ISI table with three separate fixed mirrors, as shown in Figure 1. Each sensor and mirror is mounted on a holder secured to the platform using screws or dog clamps. The distance between the front surface of the titanium sensing head and the mirror surface is set to 48 mm, which is the nominal operating distance and gives close to maximum interferometric contrast. All three sensors can reach at least 85% contrast after alignment, and some have previously reached around 90%. For future integration with the suspension table, the simplest and lowest-risk procedure is to release the clamps, move the existing LPS with its holders to the required positions, preserve the 48 mm separation, and perform a final alignment. The correct sensor height has already been included in the holder design based on the CAD model of the suspension table.
The in-vacuum system also includes the DB-25 cable, which connects the photodiodes inside the LPSs to the vacuum electrical feedthrough and then to the transimpedance amplifier outside the chamber. The in-air electronics are shown in Figure 2 and consist of the laser temperature and current driver in the upper 19-inch chassis, the transimpedance amplifier in the lower chassis, and the laser box mounted near the 1550 nm fiber feedthrough. As shown in Figures 3 and 4, the laser box is bolted to a flange adapter immediately next to the feedthrough, giving essentially zero exposed in-air fiber between the laser box and the chamber. This arrangement is beneficial for LPS displacement-noise performance and should also be compatible with a future LIGO observatory installation.
The AA and AI filters are connected to the m1sus CDS system using channels 1–4 of ADC2 and DAC3, with DAC channel 1 connected to temperature control and channel 2 connected to current control. The LPS CDS model has been built and is running on m1sus as M1LPS and the LPS_MAIN medm screen is linked to the LASTI_OVERVIEW medm screen.
Before I left MIT, the SQS 1550 nm 1×8 splitter had been shipped from Caltech but had not yet arrived. For the initial checks of the sensors, cabling, and CDS model, I therefore used the 1550 nm 1×2 splitter previously used for Arnaud’s CRS setup and connected LPS1 (H1) and LPS3 (H3). Once the 1×8 splitter arrives, it can replace the temporary splitter so that all three LPSs can be operated simultaneously. The attached PDF shows a preliminary measurement with H1 and H3 while the chamber was still at atmospheric pressure and the ISI was locked to ground. The two sensor spectra show good agreement, and the differential signal reaches below approximately 1 pm/sqrt{Hz} from a few hertz to several tens of hertz. The remaining low-frequency noise is likely influenced by incoherent air-current fluctuations, while the higher-frequency noise is likely dominated by residual motion of the ground-locked ISI. Both contributions are expected to decrease once the chamber is pumped down and the ISI is floated.
This log is related to the ongoing discussion about in-air fibres between Yulin and Rich/Alex:
In summary, I think we should avoid any sensitive component exposed to air in a precision measurement system, if possible at all. The sensors have sub-pm precision down to 20 mHz, while the optical path inside ~meter-scale in-air optical fibres fluctuates by ~1 rad on a time scale of a few sec (depending on the air currents, insulation, fibre length, etc.). This is 6-7 orders of magnitude above the level we like to measure. Even though the coupling is small and common mode, it is still visible in noise spectra as discussed below.
We spent a significant amount of time studying this problem in Birmingham. The first attached figure shows the LPS performance achieved while sensing a suspension with 3 sensors. The noise level 0.3 pm/sqHz at 0.1 Hz, a hair above the LPS sensitivity with stationary mirrors (0.2 pm/sqHz at 0.1 Hz). The laser is installed near the feedthrough in this measurement, and no fibre is exposed to air.
We then moved the laser away from the flange and installed an in-air fibre in between. As a start, we used Thorlabs SM patch cords, inspected 1 m and 5 m long options. In both cases, we see glitches straight after the installation. The glitches are caused by the stress relaxations in the fibre jacket and decay over time, but still appear from time to time, likely due to T-induced stress relaxations in the jacket and indices of expansion mismatches. The glitches and their count are shown in the second and third attached figures.
We have then tried a different set of fibres as shown in the attached figures 4-6. We got the best results with UHV SQS fibres, likely because there is no soft jacket; the fibres are insulated in a stainless steel tube. Still, none of the curves match the 0.3 pm/sqHz performance at 0.1 Hz. We tried further insulating the fibres with pipe foam. This works fine in reducing the noise of Thorlabs patch cords, but also leads to glitches because the insulation relaxes in air, similar to the jacket.
Therefore, I think we should not compromise the performance and mount the laser on the flange as Yulin has done, if possible at all.
REMINDER: When comparing to the values at the sites, recall that this OPO has the O3 mirrors
Note: I skipped a couple weeks because I was out of town.
This week
532 Power at Fiber Launch: 46.9 mW
REFL 532 Power (Unlocked): 31.0mW
REFL 532 Power (Locked): 11.6 mW
Trans 532 Signal (Locked): 8.05 (i think this is in uW but seems low, need to check calibration)
532 Mode matching = 0.71
Seed Incident: 35 uW
Seed Trans (no gain): 0.150 V
Seed Trans (max parametric gain - no gain): 2.38V
NLG: 15.8
Optimal OPO Crystal Temp: 34.700 C
Yulin, Rich, Lari, Den
Rich mentioned to me that the BIS response has a relatively high-Q peak at around 0.95 Hz, as shown in the first attached figure, measured during the ISI RY excitation. Since the BIS / GS13 blend is at 0.3-0.5 Hz, the peak may lead to the instability of the blend. This log discusses the cause of such BIS response, the reason for the blend instability, and the implemented upgrade to the BIS analog servo to address the problem.
By design, BIS runs in force feedback mode similar to most broadband seismometers. Since we focus on low frequencies (below 1 Hz), we have not optimised the BIS mechanical structure for the high-frequency performance. As a result, the BIS body modes are at 20-40 Hz because we use M6 titanium rods to support the proof mass. The design bandwidth of BIS is 4 Hz. This worked well in Birmingham; however, the BIS feedback servo has run up the mechanical modes during the pump down at Lasti last Dec. This is likely caused by a better mechanical stability of the MIT ISI because we utilise rubber springs in Birmingham, which likely provide sufficient damping for the BIS modes.
Since the BIS mechanical modes got run up on the last day of our Dec visit, we did not have time to update the analog servo. Instead, Amit has lowered the BIS servo gains by a factor of 8. In this configuration, the mechanical modes are stable, but we approach close to the lower frequency side of the stability phase bubble of the BIS servo. This is illustrated in the attached figures 2 and 3. Figure 2 shows the design open loop and the MIT servo open loop with a decreased gain. Figure 3 shows the BIS response to external excitation; it is calibrated to the velocity readout and indeed shows a peak at 0.95 Hz.
In designing the BIS blends in Dec, I have aggressively low-passed the BIS blend at 1 Hz with an elliptical filter to make the system stable. However, since Rich uses a more accurate approach, which relies on the inertial sensor noise figures, he got another, more optimised blend. That blend takes into account the BIS response but leads to the instability of the HAM control. The reason for this is that the ISI loop works on top of the force feedback loop of any inertial sensor. Loop algebra shows that the ISI control signal adds a boost to BIS according to the equation
boost = 1 + band_pass,
where band_pass is the BIS blending filter. We high-pass the BIS signal to blend with CPS and low-pass it to blend it with GS13. The effect of this boost is rather modest, and its Bode plot is shown in the 4th attached figure. The boost claims a few extra degrees of phase at 1 Hz. However, since the BIS servo has already been marginal, Rich's optimised blend has destabilised the BIS force feedback servo and, as a result, the ISI.
After discussing with Lari, we realised that the simplest solution would be to reduce the cut-off frequency of the Sallen-Key low-pass filter inside the analog servo to reduce the open loop gain near the BIS body modes. Conveniently, Murata also started to sell 470 nF class 1 C0G/NP0 1206 capacitors that Rich acquired last week. The last attachment shows the updated schematic of the BIS Rev A servo with increased capacitance. The design has shifted the low-pass frequency from 30 Hz to 10 Hz. The new simulated open-loop and BIS responses are shown in Figures 2 and 3.
Yulin replaced the capacitors on Sunday and increased the gain by a factor of 8 back to the nominal values. So far, the mechanical modes look stable, and the BIS response is consistent with the simulated value. Figures 5-10 show the measured BIS response during the HAM excitations in X,Y,Z,RX,RY,RZ. The phase shift between BIS and GS13 is close to zero up to 1 Hz, and the BIS response has a low-pass filter. After the next pump down, we will see how the optimised BIS/GS13 blends behave and may push the BIS bandwidth a bit further if needed.
Charith, Marc, Shahzod, Bruno, Lisa
We cleaned up the left side of the pump room to prepare for the new MIT Baffle Prototype experiment for Cosmic Explorer.
More on this experiment later.
For now, the main implication for everyone is that this room, or at least the left side of this room, is not a storage area anymore!
Please refrain from putting anything there.
Picture attached -- if you are familiar with the way that room looked before, you should be impressed!
Documentation, no explanation
The Gains on the ISI loops changed, I add 1.4 to X
1.25 to Y
0.7 to rY, UUG was at 50Hz
0.8 to rX
The Z loop was running with a phase margin of 15 degrees so i redid the boost
At 100Hz there is about a factor of 16 between the noise in the vertical and horizontal directions
Thank you, Rich. This is the ADC noise at high frequency. This noise can be estimated by running ADC noise through the BIS calibration filters engaged in the BIS X,Y,Z, RX,RY, RZ filter banks. I attach one plot as an example for the relative gain of 1 (see relative gains for different degrees of freedom below). The ADC noise should not be the limiting factor at low frequencies, though it is also not very far away, because we needed to avoid BIS clipping when the ISI is not isolated. As shown in the BIS servo design, the whitening gain can be increased via jumpers.
Relative gain: ADC noise scales differently between different degrees of freedom and can be computed from the BIS diagonalisation matrices. The BIS design prioritises vertical degrees of freedom and measures = Z +/- 0.08 H because we need an accurate tilt subtraction (>60 dB) to reach the sensor noise in the presence of tilt. As a result, the horizontal sensitivity is worse than in Z but I think this is fine because H is significantly polluted by the tilt noise.
Relative gains for different degrees of freedom are listed below:
X, Y, RZ: 6.8
Z: 0.41
RX, RY: 0.86 (this is higher than in Z because the lever arm is less than 1 m and is also influenced by the sensor arrangement; I think this is quite similar to the scaling of the GS13 signals).
The Z noise is lowest with the RX,RY : Z noise ratio of 2.2 and X,Y,RZ : Z noise ratio of 16.6. I think this is quite consistent with the measured values.
I also attached the BIS noise budget model. It is for the Z degree of freedom.
Nelushi, Bruno, Evan
When the Frosti-lite nichrome windings are supplied with 40 W of power, a FLIR meter (set to emissivity 0.95) registers the temperature of the emissive edge at 95°C. This is consistent with thermistor readbacks. The ambient temperature registered by the FLIR is 23°C. The clean room fans are on. The radius of the emissive edge is 19 cm, and its thickness is 3 mm, although it sits behind the reflector opening, so not all of the emitted power from that edge makes it into the reflector—some presumably bounces around inside the reflector and is absorbed. Assuming only half the emitted power makes it through the reflector, the theoretical maximum power emitted from this emissive edge via the reflector is 1.1 W.
With the thermopile (S401C, read back with a PM100D) placed with its sensor surface parallel to the plane of Frosti and about 5–6 cm away, the maximal power reading is about 1.0 mW. The thermopile can be twisted in alignment to receive more directly the emission out of the Frosti reflector, where the maximal registered power is more like 1.8 mW. The active area of the sensor is 95 mm2, so the peak intensity received is 1.1–1.9 W/cm2.
The Frosti annular pattern at 5–6 cm distance very crudely is 9 cm in radius and perhaps 3 cm in width (FWHM). The corresponding effective area is 110 cm2. Taking 1 mW/cm2 as a typical intensity value in this annular area, the total emitted power inferred from the thermopile is 0.2 W.
Nelushi confirmed several times that the thermopile, when supplied with 1.5 µm laser light, is broadly speaking reading the expected power values. There is no filter on the thermopile. Holding a hand in front of the thermopile registers about 2 mW; just to pick rough numbers, 1 cm2 of thermal emission from a 32°C blackbody should net about 4 mW of power flux against a 23°C blackbody.
Over USB one can get screengrabs from the camera via OpenCV (example python script attached). More advanced features like radiometric data don't seem to be supported on the E8 Pro model. It is exposed in /dev/video* but not via serial ports (no /dev/tty*). flirpy could not talk to it.
I've made a set of blend filters for the BISon the HAM-ISI. Den already installed a set which work really well, so this is mostly polishing and following our procedures.
The Plot BIS GS13 Blend is GS13/BIS blend which probably could be improved, I mostly tried to minimize the high frequency noise, CYAN curve (CUR noise) is the total combined sensor noise from this blend and makeGS13_BIS_Blend.m is the relevant script
CPS BIS Blend shows the displacment sensor/inertial blend (60-70mHz)
CompFilters, show the resulting three part complentary filters (wtih a little tweaking to explain why the sum isn't 1), and the script MakeBis_inertial_Blemd
*********** THIS IS PRELIMINARY *************
- since at least the rY filter is not stable, the BIS response might not be as origionally advertised, more later
Posting this here because I seem to forget some detail of the process every time I do this and in case it's useful to anyone in the future.
You will need: Netfinder or similar application, prologix GPIB configurator application
First: set IP address in netfinder. Power the GPIB controller and connect it to a computer with whatever method you intend to use (via router, directly to usb, etc.)
For local connection use something like 169.254.1.100 (local, ie USB connection) or 192.168.1.100 (default) for eg. a router connection (or replace 192.168.1 with your subnet if they don't match)
Subnet mask should be 255.255.0.0 for local, 255.255.255.0 for router network. Gateway blank for local, or router IP for networked. Configure the device, make sure you get an success message, and close netfinder.
Plug the controller into the instrument you're using it with. Click the device in the 'select device' box and set it to controller. For a SR785, the default address is 10. The other fields seem to auto-set after connecting to the 785 no matter how you set them on my computer, but in the case that doesn't work, the 785 uses a LF EOS. Press 'Update CONNECT.INI' and then exit.
On the 785: In system> remote: set output to GPIB, addr. 10. In output, set hard copy button > ascii dump, destination > GPIB, GPIB ctrl: host
REMINDER: When comparing to the values at the sites, recall that this OPO has the O3 mirrors
This week
532 Power at Fiber Launch: 44.9 mW
REFL 532 Power (Unlocked): 27.6mW
REFL 532 Power (Locked): 11.05mW
Trans 532 Signal (Locked): 8.26 (i think this is in uW but seems low, need to check calibration)
532 Mode matching = 0.71
Seed Incident: 35 uW
Seed Trans (no gain): 0.150 V
Seed Trans (max parametric gain - no gain): 2.46V
NLG: 15.9
Optimal OPO Crystal Temp: 34.744 C
Degradation finally seems to be slowing down! To within the precision of my technique, I seem to have more or less the same result as last week for my estimate of threshold.
I need to track down the cause of my drifting demod phase, thats probably limiting my measurement now that the degradation has slowed down.
The cavity was unlocked for about a day this week because the guardian crashed when mazer.mit.edu was moved behind a new MIT firewall. Thanks Fred for fixing the issue this morning.
The fiber coupling for the pump is a bit lower this week, down to ~61%. This may be due to drift in alignment rather than degradation, but I don't have time right now to reoptimize.
REMINDER: When comparing to the values at the sites, recall that this OPO has the O3 mirrors
This week
532 Power at Fiber Launch: 41.5 mW
REFL 532 Power (Unlocked): 26.4mW
REFL 532 Power (Locked): 10.3mW
Trans 532 Signal (Locked): 8.18 (i think this is in uW but seems low, need to check calibration)
532 Mode matching = 0.71
Seed Incident: 35 uW
Seed Trans (no gain): 0.150 V
Seed Trans (max parametric gain - no gain): 2.04V
NLG: 13.6
Optimal OPO Crystal Temp: 34.767 C
Recording the wiring configuration for the OSEMs in the Mid Y HAM here. With respect to positions in the attached photo the order is: 1. upper left OSEM 2. upper right OSEM 3. lower left OSEM 4. lower right OSEM 5. side OSEM
The orientation (Left/Right) assumes that one is looking at the back side of the suspension.
It's been 150 days since the BIS diagonalisation was set (log 12122), one or two vents, and multiple thermal cycles. In summary, I think the 3 BIS stations (6 seismometers) inside the HAM platform have survived this well, and the sensor diagonalisation coefficients set back in Dec 2025 remain valid to the required precision of ~0.1%. A few more details:
The first figure shows the DC force on the inertial masses of 3 BIS stations + T240 for reference. In contrast to T240/STS-2 seismometers, BIS couples 10% of the DC integrator signal to the fast "velocity" channel to monitor the mass position via the same ADC channel. In total, there have been two "catastrophic" events during the period seen by both BIS and T240: Jan 23 and March 24/28. The latter one was a vent (log 12168), and I could not track anything obvious about the former event in the log, maybe power supply issues on Jan 29 (log 12134).
The C-6D in-vac temperature sensors (shown in orange) have been, unfortunately, disconnected on March 3rd. However, I think it is quite clear that both BIS and T240 follow a similar T profile to some degree. The integrator control signal was well within the range (~20% deviation of max): the V1-3 sensors should start to saturate at +/- ~5000 cnts in these units.
It is also feasible to spot that station 1 fluctuates on the more or less same DC level, while station 3 seems to creep down a bit during the first few months. The second attached figure shows the subtractions of the V1 signal (which contains the same T fluctuations) from the V2 and V3 signals. It seems that the V2 station, similar to V1, does not significantly creep, but the DC force shifted during the disturbances discussed above. The station 3 signal does seem to creep a bit during the observation period. In principle, we have remote recentering on BIS, but the DC offset is still rather small.
Then I looked at the BIS real-time signal to check the BIS diagonalisation. It seems that ISI has been moving quite a bit in tilt during the last few days, and T240 has been saturating. I then looked at data from May 24. The diagonalisation from Dec'25 still looks good with RX/RY signals below the BIS raw signals (that measure ~Z +/- 0.07 H) by a factor of 200 at 10-20 mHz, as shown in the third attached picture. The noise budget is shown in the fourth attached figure, and the sensors still sit close to the design sensitivity, at least, in the frequency band where we can subtract ISI's own motion, below 20 mHz.
REMINDER: When comparing to the values at the sites, recall that this OPO has the O3 mirrors
This week
532 Power at Fiber Launch: 35.2 mW
REFL 532 Power (Unlocked): 22.8 mW
REFL 532 Power (Locked): 8.32mW
Trans 532 Signal (Locked): 8.13 (i think this is in uW but seems low, need to check calibration)
532 Mode matching = 0.71
Seed Incident: 38 uW
Seed Trans (no gain): 0.155 V
Seed Trans (max parametric gain - no gain): 2.05V
NLG: 13.2
Optimal OPO Crystal Temp: 34.817 C
[Bruno, Eric]
We're making some changes along the Y-arm to support the new coupled cavity expt that will be installed between the Mid and End-Y HAM chambers in LASTI. In the past couple of weeks the following changes have been made:
Still some cleanup to do, but I think we're in good shape for going into the HAMs when Rich gets back from Vacation
REMINDER: When comparing to the values at the sites, recall that this OPO has the O3 mirrors
This week
532 Power at Fiber Launch: 33.5 mW
REFL 532 Power (Unlocked): 21.0mW
REFL 532 Power (Locked): 7.17mW
Trans 532 Signal (Locked): 8.13 (i think this is in uW but seems low, need to check calibration)
532 Mode matching = 0.78
Seed Incident: 32 uW
Seed Trans (no gain): 0.150 V
Seed Trans (max parametric gain): 2.25V
NLG: 15.0
Optimal OPO Crystal Temp: 34.967 C
Comparing this to the behavior of the crystals at the sites, they typically see a similar level of degradation after only a couple of days. https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=82196
This initial rapid degradation is followed by a slower decline in performance over the next year: https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=82080
Our initial degradation seems to be occurring less rapidly. Hopefully this remains true for the long-term degradation.