Displaying reports 1021-1025 of 1025.Go to page Start 48 49 50 51 End
Reports until 19:28, Saturday 25 March 2023
General
admin - posted 19:28, Saturday 25 March 2023 (11189)
PNG IMAGE
This is a .png image
Images attached to this report
GRAVITES General
epolini - posted 18:26, Friday 24 March 2023 - last comment - 09:58, Monday 27 March 2023(11177)
Replacement of the long fiber from the laser to the ifo
Today I replaced the 3 fibers with a new one ordered on thorlabs (SM980-5.8-125-CUSTOM) 15 meters long. I measured power losses before and after.
The power on the powermeter was fluctuating, so I considered the highest detected power.

BEFORE the replacement: 
Power on the bench before the first fiber (after the polarization tuning) : 38 mW
Power after the 1st fiber : around 38 mW
Power at the entrance of the ifo : 28 mW

AFTER the replacement I decided to connect the fiber before the polarization tuning
Power on the bench before coupling in the new fiber : 50 mW
Power at the end of the fiber, before the ifo : 45 mW

I left two x 1 meter fibers next to PDs, in order to replace the PM fibers with SM fibers. On the bench in green lab there is also a 10 meters fiber without jacket to replace the arm of the ifo.

Next steps: replace the fibers and place the ifo on two small stages at different heights to start thinking about the vertical switches and isolation stages.
Comments related to this report
yinx - 09:58, Monday 27 March 2023 (11193)
Asking to get the tag changed to GRAVITES
GRAVITES General
epolini - posted 18:25, Friday 24 March 2023 - last comment - 19:27, Saturday 25 March 2023(11176)
Recovery of fiber ifo at 1064 nm
With Dorotea and Xinghui, we restored the fiber ifo built last year by Vicky and Chris on the bench on the right side of green lab. 
We used the 1064 nm laser on the bench on the left area, since it was already coupled in fiber. We connected that fiber to two other fibers, cleaning the connectors. We measured the powers at the intersections and we were losing 50% of the power. That was not important for our purpose. 
We connected the input fiber to the ifo in the box and the two outputs to two photodiodes. We saw the signals on the oscilloscope, realizing how sensitive is the ifo to audio and seismic perturbations. We then sent the PD differential signal to an analog servo controller (we needed to remove the knobs in order to change the gain and the offset). Then we sent the signal to the piezo controller to control the ifo and lock it. By tuning the servo controller, we managed to lock the length of the ifo acting on the piezo.
Comments related to this report
epolini - 19:27, Saturday 25 March 2023 (11180)
In the figure you can see the setup (ifo is in the small box to reduce noise) and the signal of the ifo locked on the oscilloscope (in blue).