After some tinkering, I improved the OPO dither lock substantially. The dither frequency is now 22kHz, the loop's UGF is 960Hz, and the phase margin is about 55 degrees. The attached plot shows the measured transfer function around the UGF.
A few weeks ago, I switched the SHG cavity lock from dithering a cavity mirror to dithering the laser's frequency. This substantially improved the SHG's stability. Since then, I've spent a good bit of time attempting to switch the OPO over to a feedback loop that dithers the laser's frequency as well, but have run into several problems. (1) For moderate dithering amplitudes, the SNR of the error signal demodulated from OPO's green transmission PD is quite low. In general, the OPO's green transmission power is low, so the PD signal is fairly noisy. (2) If I increase the dithering amplitude enough to get an error signal with a decent SNR, this modulation level causes huge (>50%) modulations in the SHG's transmitted green power. (3) There are various PD gain and bandwidth problems. I modified one of the custom PDs in the setup (PD 14) to have a 6pF feedback capacitor and a 500k feedback resistor. This the large trsnsimpedence impedance gives close to the desired amount of amplification (5 Mega-Ohms). but has noise that rises linearly with frequency above about 5kHz. Below this corner, the noise is 1uV/sqrt(Hz). On the other hand, the PD-100A has a transimpedence gain of 5 Mega Ohms and noise of about 15uV/sqrt(Hz) at 8kHz, which is similar to the custom PD. At this gain level (70dB), the PD-100A has a low bandwidth of only 6kHz. To take advantage of dithering the laser frequency, we would like to dither at closer to 50kHz. For now, I plan to switch the OPO back to dithering one of the cavity mirrors. This may not be the optimal cavity lock, but it seems to work well enough and will allow me to begin aligning the squeezer's output beam to the homodyne.
I switched the OPO back to a dither lock. Currently, I'm dithering the piezo at 22kHz, which is a factor of about 3x my previous dither frequency. I'm not quite sure why I'm able to dither so much faster now. The dither lock catches consistently, but will break if there's too much environmental noise, as seems to be typical of these dither locks. With the OPO dither locked, the transmitted green and CLF power are stable. The next step here should be aligning the squeezer to the homodyne and looking for mean squeezing.
Mounted a retroreflector on the existing suspension in the red lab and the QPD board on a raised plate to test sensor on a free hanging suspension. Still using a passive high pass filter before the demodulation stage and no amplification in the summing/differencing stage. Saw regular motion of the suspension around 1Hz which did not settle even after leaving for ~1hr (~20 Hz range plot). Also measured noise below 1 Hz with suspension "locked" (resting on the jack) and with the LED off (~1Hz range plot). note: except for LED off measurements, could not get a quiet enough signal to leave spectrum analyzer input range at -50dBV, usually settled at -34dBV instead.
Away for two weeks but when I come back we'll look at noise below 1Hz when the retroreflector is more securely fixed with respect to the QPD. Also getting updated qpd board and looking into replacing the corner-cube retroreflector with a cats eye retroreflector which could decrease the footprint of the sensor and increase sensitivity.
The idea of the cat's eye is essentially (pardon the ASCII)
----------| / \--------------||
QPD & LED | <-- D --> | f | <-- d --> ||mirror
----------| \ /--------------||
where the lens and mirror are rigidly mounted to the suspension mass, while the QPD & LED are mounted to the suspension cage. By choosing d = f, the ray transfer matrix of the cat's eye is
[[ -1, 2 * (f - D),]
0, -1,]]
This is a retroreflector in the usual sense of the word, in that any ray arriving from a large distance D >> f will be directed back from where it came from (i.e., the ingoing ray can be described by the vector [r, -r/D] for any lateral offset r satisfying the paraxial approximation, yielding an outgoing ray [r, +r/D]).
However, for our purposes we really do want to image the LED onto the QPD, meaning any vector [r, ϕ] eminating from the LED is imaged onto the vector [-r, -ϕ] at the QPD. In order to do this, we need to null the off-diagonal element of the cat's eye ray matrix, which means enforcing D = f. In practice that might be difficult, since D probably needs to be controlled to within ~1 mm, but is probably okay to try as a first attempt. One could also rigidly mount a second lens in front of the QPD/LED (with focal length f′ equal to mounting distance d′) so that the rays going toward the cat's eye are collimated:
----------|-------------- / \ / \--------------||
QPD & LED | <-- d' --> | f'| <-- D --> | f | <-- d --> ||mirror
----------|-------------- \ / \ /--------------||
In this case, the exact distance D becomes irrelevant, though it should probably be kept small (few millimeters?) to avoid too much light lost.
[Eleonora, Rich, Regina]
Yesterday and today we continued the work started in entry #11285.
We tried to use the two T240 instruments in LASTI (one next to the squeezer and one near the BSC). We centered the masses of both of them and read the signals (X MON, Y MON, Z MON) both with an oscilloscope and on the CDS system: on the oscilloscope we detected some peaks if someone jumped next to the trillium but the spectra on CDS were flat. After several trials, including interfacing the electronics with a new instrument ( a Streckeisen STS-2 Broadband Sensor which is working), we concluded that the T240 instruments are not working. Rich will send one of the instruments to repair.
We installed a Streckeisen STS-2 seismometer under the desk in LASTI near the BSC, as shown in Figure 1 and we connected it ot the STS-2 Interface Chassis in the rack. We covered the STS-2 with a box (see Figure 2). NB. Everytime we want to move the seismometer we HAVE TO lock the internal masses with a screwdriver, otherwise we can break it.
We centered the masses by using the botton on the small green box which is connected to the instrument. The signals read by the Streckeisen STS-2 are reported in Figure 3 (M1:HPI-BSC_STSINF_B_{X,Y,Z}_OUT).
In order to check its calibration we want to compare it with the L4Cs. We connected the corner #4 of the HAMX to two L4Cs, one H and one V placed on the ground. The signals read in V and H were different as shown in entry #11285, which is weird. So we started investigating: we replaced the vertical L4C with a new one and it was showing a signal significantly lower. We measured the internal resistance of all the L4Cs and they were correct (5.5 kOhm). The L4Cs of the other corners sees a reduced motions due to the spring attuation. We decided to fix the four corners of the HEPI to ground in order to use the L4Cs already installed. The signals right after fixing the HEPI are shown in Figure 4, with H4 and V4 placed on the ground. We noticed that V3 was not connected to the electronics box and we decided to connect back the H4 and V4 channels to the original L4Cs. The signals after this actions are shown in Figure 5 and they look correct.
Now, we need to devide the signals by the response of the instrument.
Unfortunately, the L4Cs signals don't look as here anymore. Corners 3 and 4 produce signals that are one order of magnitude lower than corners 1 and 2 of HAMX. Today I took a spectrum of the L4C signals and IPS signals of HAMX since they share some electronics, shown in Figure 1. Except for IPS H1 and V1, the other IPS signals seems reasonable and consistent with each other.
If you want to access to a CDS machine connected to the controls account from your laptop, you can use the following instructions:
ssh -X -C -L localhost:2233:m1wsA.local:22 controls@athens.mit.edu
ssh -X controls@m1ws6
From a mac laptop, you should use xquartz/x2go in order to display diaggui etc.
I replaced the M3 mirror on the OPO with one coated with the design specifications. With the new M3 installed, I re-aligned the OPO and tuned up the dither lock. At the moment, the OPO dither lock is implemented by dithering a cavity mirror, but I expect to change this to a laser frequency dither lock when a pair of clocked red pitayas arrives next week. The attached files show the 00 cavity mode on a camera with a 532nm band-pass filter, the IR (purple trace) and green (yellow trace) mode matching from a cavity scan. Note: Previously, this post had a comment about IR stability in the OPO cavity. I believe this is likely resolved, but this claim wasn't supported by the plot I had included. I believe what was measured there was just the PD's dark noise and the OPO was not on dual resonance.
As expected, the OPO is now much more stable in IR than it was before. The attached image shows the OPO's green transmitted power (yellow trace) and IR transmitted CLF power (purple/pink trace) with a 500ms timescale. We no longer observe the large IR power oscillations we saw before. This time around, I was careful to check that the cavity is on dual resonance and the power level on the IR PD is well above its DC offset output voltage.
I swapped the improperly coated M3 for a new, properly coated M3, which has an IR reflectivity of 89% and a green reflectivity of 96%, compared to the opposite specs on the previous M3. With the new M3, we have more green transmission through the SHG, but less IR transmission. Since the IR transmission was already low, this makes it more challenging to identify co-resonant locations automatically. Before, the reflected IR power from a cavity scan could be immediately curve fit to find resonance locations, but now it needs to be processed first. I implemented a python function that detrends the reflected IR power data and applies a band pass filter. This makes the IR resonance locations much easier to identify. The attached plot shows an example of this.
We ordered new M3 mirrors from Intrinsic Crystal, a Chinese vendor. We had previously ordered M3 mirrors with flipped 532nm and 1064nm coating specs, and re-ordered these mirrors. The new mirrors are to-spec, within reasonable measurement bounds, I measures 12% IR transmissivity and 4.1% green transmissivity. The specs are 11% and 4% respectively. Nick and I measured the IR transmissivity at about a 10 degree angle of incidence, using his setup, so this seems consistent with a 11% transmissivity at normal incidence. To measure the green transmissivity, I used the SHG in the compact squeezer setup.
Today, I re-measured the beam profile after mirror M4 and obtained a result similar to the one reported in entry #11277 (with a confidence level of 3%). I installed the telescope described in the same entry. I took the two lenses with f = 100 mm and f = 50 mm from the 10-foot-long bench currently located in the green lab. To achieve a beam with a waist around 400 μm, I positioned the lenses at z1 = 3 cm and z2 = 20 cm from M4, respectively, as shown in Figure 1 and Figure 2. I measured the beam profile at the output of the telescope, considering z = 0 as the position of the second lens, and the data is shown in Figure 3. The beam waist is approximately 400 μm and is located 23 cm from the second lens. I placed the AOM around the waist.
The next steps will be to install the electronics as described in the entry #11269 and the BS/mirror combination shown in #11270 (but we run out of BS). The final goal is to stabilize the 200-meter-long fibers just arrived from OZ optics.
At the end, to be sure not to clip the beam in the AOM (diameter aperture = 0.5 mm), I replaced this telescope with a f = 77.3 mm lens placed 17.5 cm after the last steering mirror. The waist diameter is around 200 um and located at 12 cm from the lens. I centered the AOM's crystal around the waist position, being careful that the size of the beam was smaller than its active aperture along the propagation.
The AOM is driven with an RF generator for the moment and the optimal parameters set are: frequency = 80 MHz, amplitude = 3 dBm. The amplifier installed after the RF generator is powered with a 24V DC power. The input of the amplifier is 3dBm and its output is 29bBm, meaning that it has a gain of 24 dB (higher that the 20dB expected). Its compression point is 29.5 dBm.
I alignmed the AOM in order to maximize the power on the 1st order diffraction beam:
P(before AOM) = 63 mW
P(after AOM) = 62 mW -> P(0th order) = 22 mW and P(1th order) = 40 mW.
The diffraction efficiency is around 67%.
The shape of the beams exiting the AOM is shown in Figure 1: the 1st order is on the right and it is very elliptical and it diverges very fast. In order to measure the parameters of the beam exiting the AOM, I placed a 150mm lens distant 17.5 cm from the AOM and then I did the inverse matching on JamMt. The data on the different axis measured with the beam profiler are shown in Figure 2 (the reference position is the lens position). The average fit gives w0 (radius) = 96um and z0 = 34cm. It turns out that the waist is positioned 10 cm BEFORE the center of the AOM.
I used this beam to design a telescope on JaMmt to collimate the beam back to 2mm diameter. After several iterations I decided that the best solution (beam not propagating too long in free space, 2 lenses, 0th and 1st order beams separated enough etc) was:
where the positions are referred to the center of the AOM (z0 -10 cm). The resulting beam has w0 = 1113 um (the fixed collimator expects a diameter of 2.4mm) and z0 = 3.77m.
I also thought that it might be better to couple both path in fiber and recombine them on a fiber coupler before being detected to avoid to design and install additional telescope and other complications. The reference fiber is 10cm-long.
[Regina, Rich, Eleonora]
We started measuring seismic noise in LASTY to estimate the seismic noise that we should mitigate in GRAVITES. We conducted this work in LASTY since the sensors can be interfaced with the CDS system.
We have access to three different types of seismometers:
1. Nanometrics Trillium 240 Broadband Sensor: it has a corner frequency of 4 mHz and reads three axes. This sensor provides the best spectrum for us as we are interested in measuring low-frequency seismic noise. We properly connected the cables to the electronic boards, but the signals X MON, Y MON, and Z MON were not sensitive to seismic activity. Following the reference in DCC D0901489 (https://dcc.ligo.org/LIGO-D0901489/public) and applying 5V to pin 6, we still haven't obtained a meaningful output signal. What we need to do is to center the geophone masses using a software, and we will attempt to do this next week.
2. L4C geophone: it has a corner frequency around 1 Hz, and each sensor reads one axis, i.e., H or V. To obtain calibrated signals, we connected a vertical and a horizontal geophone to one corner (#4) of the HEPI of the vacuum chamber near the green lab (see Figure 1). We acquired the data using CDS, and the spectrum is shown in Figure 2.
3. Guralp seismometer: it has a corner frequency above 1 Hz and it is installed on Regina's bench. We haven't used this sensor yet because the seismic activity is attenuated by the bench on which it's installed, and it's not calibrated.
[Rich, Vicky, Eleonora]
We centered the masses of the T240 by sending 5V to the correct pin, as written in the manual, i.e. RX which according to this scheme corresponds to pin #18.
We currently do not have state-space filtering implemented within CDS. However, we can run a custom C-code, and the procedure to add a user C-code is described in the RCG developer's guide (see T080135).
State-space models have certain advantages over ZPK models.
Christopher Wipf previously wrote a routine to convert discrete state-space Matlab models to a CDS-compatible C function where the [A,BC,D] matrices are stored in the corresponding header files. I modified the code to work for any ZPK or StateSpace models. Users have the option to initialize the filter coefficients. I set up a 2-input 2-output filter model consisting of two separate lowpass filters, each with corner frequencies (0.5 Hz & 10 Hz), and checked the results on two inputs (random Gaussian noise in this case) and compared it with the expected one. They both match well.
Next steps:
Notes:
# While compiling with custom C-code using cdsFunctionCall block,
I've switched the SHG locking scheme to dither the laser frequency instead of dithering one of the cavity mirrors. This seems to substantially improve the system's stability; I can now tap directly on the SHG cavity and it does not loose lock. Before tapping anywhere on the compact squeezer's optics breadboard would break the cavity lock. The dither lock has a modulation/demodulation frequency of 80kHz, two integrators, a measured UGF of 650 Hz, and a phase margin of 37 degrees. The attached plot shows the inferred open-loop transfer function measured with a red pitaya. The feedback loop auto-locks successfully most of the time. Heuristically, it seems to acquire lock about as consistently as the dither lock I had previously implemented by dithering a cavity mirror.
The locking procedure fits the derivative of a lorentzian to the error function to determine the error function slope, which it uses to calculate an auto-gain parameter so the feedback loop does not need to be retuned as the power level changes. The attached plot shows an instance of this curve fit.
The SHG has now remained locked for over three hours using the scheme of dithering the laser frequency.
Dithering the laser's PZT allows us to dither at a much higher frequency than dithering a PZT on the optical cavities does. The attached plots show the error functions from dithering the laser's PZT at 80kHz, a factor of 10 greater than the previous 8kHz dither on the SHG and OPO PZTs. This should allow us to increase the loop bandwidths by about a factor of 10 for additional audio-band noise suppression. The attached plots are from the SHG. The laser's PZT can be driven by up to 100V, but is currently being driven by the Red Pitaya's output directly at 1V with no additional amplification. Considering that we want the laser's frequency range to stay well inside the cavity linewidth for the transmission dither lock, the ideal voltage level to the laser is about 6V, so the 1V drive is in about the correct range. In the future, we could dither harder with a 6x amplifier between the Red Pitaya's output and the laser if need be.
[Vicky, Eleonora, Eric]
We moved some electronics from the drawer in highbay to NW22 in order to be able to implement a feedback loop to control a long fiber length with an AOM. We took a free space AOM from the PSL table (see Fig. 1).
The electronic components are shown in Fig. 2.
[Vicky, Eleonora]
In the attached picture, the fiber length control schematic including the electronic components as they are going to be implemented in the next days.
AOM data sheet found on the DCC: https://dcc.ligo.org/DocDB/0162/T1900493/001/1205C-843.pdf.