(Hudson and Masaya) The OPO lock's stability seems to be changing quite a bit even over fairly short timescales. The transmitted power level with the OPO locked also seems to be consistently less than the peak power level when we scan the OPO, which is odd. After some experimentation, we believe this is because one of the piezos is actuated nearly perpendicular to the cavity axis, whereas the other piezo seems to be actuated at an angle to the cavity axis. The attached images show cavity scans with the piezo attached to M3 (channel 2 is scanning) and with the piezo attached to M4 (channel 4 is scanning). On both cavity scans, channel 1 shows the 532nm power on a transmission PD (a PD 100a on the 60dB gain setting). As can be seen in these cavity scans, scanning M3 results in much less higher-order mode content and the TEM00 spikes have similar power levels. Scanning M4, we see much larger higher-order mode content and highly inconsistent power levels in the TEM00 mode. We attribute the fact that the transmitted power level is below the peak power level when scanning the cavity to the fact that we were taking a cavity scan with M3, but using M4 to control the cavity length when the OPO was locked. Similarly, we attribute the fact that the OPO lock's stability was changing often to the OPO locking on different resonances which have significantly different power levels when actuating on M4. To fix these problems, we plan to use only M3 to both dither and control the OPO cavity length going forwards. The dither signal is 0.16V peak-to-peak, so there should be plenty of room to dither and actuate the same piezo.
I updated the OPO locking procedure to dither and actuate only on M3. This seemed to increase the lock's stability significantly and made it so that when the cavity is locked, it's output power is the maximum power observed from scanning hte cavity. The lock has a bandwidth of 400 Hz and a phase margin of 7 degrees, so it's very close to being unstable. I'm going to play around with the loop shape and see if I can use a lower-value of resistor in series with the PZT on M3 to increase the pahse margin.
We moved Diabolo Laser and drivers for each wavelength from NW17 to NW22 as shown in the attached photos.
Dhruva and I set the LO polarization without being careful to match it to the OPO's polarization. The LO passes through a PBS right before the homodyne's beam splitter, and the CLF beam is reflected off of a PBS, so we expected the OPO and LO polarizations to either be matched, or about 90 degrees out of phase. Masaya and I used a polarizer to make a rough polarization measurement and found that the LO and OPO polarization states are roughly matched. We did this measurement with a polarizer in a lens mount, with the polarizer loosely mounted so we could turn it, so this was not the most accurate measurement ever, but at least tells us that we should be able to see some interference between the LO and seed beam out of the OPO.
Masaya and I noticed today that the cavity OPO scan measured by the red pitaya during lock aquisition looked visibly different from the cavity scan displayed on the oscilloscope when scanning the cavity length. We determined that the difference was due to the different scanning speeds (about 0.5 Hz during lock acquisition vs about 16 Hz when scanning the OPO and viewing the scan on the scope). For very slow cavity scans, the OPO seems to have a substantial thermal self-locking effect, which can be seen in the fact that the cavity transmission increases linearly and then drops sharply as we scan through a resonance, instead of having a symmetric, lorentzian, lineshape. A similar effect seems to cause the cavity scan's higher-order-mode content to change significantly dpending on the scan frequency, which can make lock acquisition more challenging if the apparent high-order mode content is large. This through off the OPO locking script this morning, and we implemented several changes to fix this problem. (1) Scan the cavity at 16 Hz during OPO lock acquisition (2) Do not use the curvefit linewidth as tolerance metric when setting the piezo back near resonance; this number will be thrown off by thermal effects that cause the spectrum to look highly non-lorentzian.
After unclipping the homodyne's LO beam, Masaya and I re-aligned the homodyne beam splitter to achieve a 50/50 splitting ratio on the homodyne and then measured the homodyne's shot noise to verify that it's shot noise limited down to low frequencies. We had about 2.6 mW of LO power and observed broadband noise levels of 256 nV/rt(Hz) with the LO on (shot noise), 16.3 nV/rt(Hz) with the LO blocked (homodyne dark noise) and 6.2 nV/rt(Hz) with the SR785's input terminated with a 50 ohm resistor, all at 3.3 kHz. I accidentally saved the files as .eps files instead of .txt files, so I haven't nicely compiled them into a python plot, and the aLog won't let me upload them. They're saved to the compact squeezer's github repo. The homodyne detector's noise floor is about 12dB below shot noise, so we should be able to observe 10dB of squeezing with this homodyne. If by some miracle we achieve more squeezing than this, we'll need to turn up the LO power further.
We (Hudson and Masaya) noticed the other day that the LO beam for the homodyne was clipped on the PBS on the LO's path leading to the homodyne beam splitter. We re-aligned the beam to unclip it and took an image of the beam with a beam profiler to verify that it's Gaussian (see attached image).
[Rich, Eleonora]
Today we moved a spectrum analyzer from the electronic shop in NW17 to NW22 (see attached figure), since we need it for the GRAVITES protoype at 1064nm.
The OPO dither lock was very stable last week, but yesterday it seemed to drop lock very easily. This made it impossible to begin aligning the OPO's output to the homodyne since the OPO would lose lock the second I touched a steering mirror on the way to the homodyne's beam splitter. The lock still had about 1.2kHz of bandwidth, but would drop easily. After some investigation, the problem seemed to be that the dither amplitude was too small and the error signal was dominated by the red pitaya's ADC noise (see attached figure). I doubled the voltage sent to dither one of the piezos, which seemed to fix the problem. The attached figures show the measured error signal before and after increasing the dithering voltage, and the OPO cavity loop's measured transfer function after increasing the dithering strength. The loop has about 30 degrees of phase margin, 6dB of gain margin, and a UGF of about 1.2kHz.
For a while, the SHG cavity has had a tendency to ring at about 1kHz. JD did some analysis on the cavity's structure and determined that this was likely due to a mechanical resonance of the mirror mounts at this frequency. He suggested that I add some small pieces of rubber to the SHG mirror mounts to damp this motion. I did this, and the problem seems to have been fixed.
I re-aligned the CLF to the OPO using a camera and a PD with 1064nm line filters. The CLF is aligned to the TEM00 mode (see attached picture) and locks consistently. However, the lock seems more sensitive to acoustic pickup than it was last week. I'm not sure why this is -- it doesn't seem like changing the beam splitter from a 90/10 to a 50/50 on the CLF path should have any effect on the stability of the OPO's green transmission lock.
After switching out the beam splitter and re-aligning, we have about 35uW of IR power transmitted through the OPO from the CLF port. This is about the power level we'd expect since we had about 5uW of power before replacing the 90/10 beam splitter with a 50/50.
The seed light from the CLF port transmitted through the OPO is now visible on an IR card!
Today I introduced the VCO in the setup in order to be able to send a correction to the AOM and close the control loop. The previous configuration (open loop), described in entry 11303 and shown on the top of Figure 1, uses an RF generator both for the AOM and as an oscillator for the mixer. In order to send a correction to the AOM I installed a VCO, adapting the setup as shown on the lower part of Figure 1.
I removed the power splitter and I set back the RF generator to an amplitude of 3dBm. I sent a DC power of 24V to the amplifier and 12V to the VCO. As PID I used a Newport LB1005 Servo Controller. In order to adjust the VCO (ZOS 100+) RF frequency to 80MHz, I needed to send a 9.5V DC via v-tune to the VCO. I used the servo controller to do so. I adjusted the offset output (on the bottom right) in order to have an output signal of 9.5V on the oscilloscope. Then I connected the output of the mixer to the A channel of the servo and sent the correction to the VCO, which is now connected to the amplifier. The re-arranged electronics is shown in Figure 2.
I syncronized the frequencies at the RF generaror and the VCO. The VCO frequency is driven by a generator inside the PID controller by adjusting the output offset knob. Since this is cannot be tuned precisely, I moved the frequency of the RF generator. I syncronized the RF signal and the VCO signal by connecting them to the mixer and looking at the output on the oscilloscope first and the spectrum analyzer later. The RF generator's signal had an amplitude of 7dbm and I attenuated the RF power at the VCO output from 16dbm to 0dbm (to be able to see a clear sinusoidal shape on the oscilloscope). I moved the RF generator's frequency to 71.053532530 MHz which minimized the beat note frequency down to few tens of kHz. The beat note frequency changes over time, in the region between 20kHz to 60kHz. This is due to the fact that the voltage generated by the PID might not be stable (2mV fluctuation corresponds to 10kHz frequency shift). I am thinking on different ways to drive in a more stable way the VCO.
Attempt to close the loop:
To keep the power reaching the AOM < 1.3W, I needed to add some attenuators between the VCO (16dbm) and the pre-amplifier (gain 24dB): -10dB and -1dB.
By closing the loop, the error signal fluctuations' amplitude reduces and the correction is flat around 7V (same as if the loop was not). By playing with the PI corner frequency and the gain, the loop was still not properly working (Fig 1), i.e. blocking one path of the optical scheme, no changes.
Attemp to close a loop between RF generator and VCO frequencies
As a check, I then connected the RF and VCO to the mixer and I engaged the loop with the PID. The peak mentioned before at tens of kHz (Fig 1), is reduced by closing the loop but several other peaks (most of them distant around 20 kHz) stays (Fig 3). Further investigations ongoing.
I continued the work on the 1064nm prototype.
I installed the telescope described in entry #11300 and I tuned it in order to have a beam as collimated as possibile with a diameter of 2mm. I checked the size and shape of the beam with the beam profiler at different positions along the beam propagation axis. I played with the tilt of the lenses in order to reduce the astigmatism of the beam after the AOM. I placed both the lenses on the thorlabs rail after an iris.
I coupled first the short fiber into a 2-mm-diameter fixed collimator, in reflection from the BS before the AOM. The fiber port is installed on a mount where angular degrees of freedom and tilts can be controlled. I followed the procedure described in entry #11244 and I reached 83% coupling. This beam was already collimated with the initial telescope we installed.
Then I worked on the coupling of the beam into the 200-meter-long fiber after the AOM telescope just installed. I followed the same procedure as above but I also maximized the coupling by adjusting the lenses by hand: both tilt and longitudinal positions. After many iterations I reached 38% coupling, which I considered good enough to go on.
I connected the two fibers to a 50/50 fiber beam splitter.
I measured the power at the two output of the splitter with a powermeter (NB. the total power entering the setup can be tuned by rotating the HWP of the power controller placed before the two steering mirrors before the AOM) :
Then I installed a RF photodiode (Model 1811 IRDC -125 MHz low noise photoreceiver) and I checked on the oscilloscope (time/div = 10ns to see the 80 MHz) both the DC and AC outputs.
I got the following signals with a power of 200 uW read on the other BS output on the powermeter:
I decided first to check the phase noise in open loop. I used the RF generator not only to drive the AOM but also as a RF reference for the mixer. I connected the RF generated signal to a power splitter (ZSC-2-1+) before the amplifier. I increased the amplitude of the RF signal from 3 dBm to 6 dBm to keep the same the power driving the AOM. The I connected the other output of the power splitter to the mixer. I connected a low pass filter to the output of the mixer and I connected it to the oscilloscope. The signal is sensitive to setup perturbations.
In the pictures attached the complete setup and electronics are shown.
The CLF passes through a beam splitter before and after reflecting off of the OPO cavity. The beam splitter basically acts as a lossy circulator to reflect the CLF beam that has interacted with the OPO to an RFPD. In the initial design, this beam splitter was a 90/10 beam splitter so 10% of the power out of the CLF fiber went to the OPO, but then 90% of the CLF power that made it to the OPO was reflected to the CLF's RFPD. To increase the maximum seed/CLF power transmitted through the OPO for alignment, I replaced this 90/10 IR beam splitter with a 50/50 beam splitter, which chould increase the transmitted seed power by a factor of 5. Previously, we had about 5uW of transmitted seed power with the OPO locked, so it should now be about 25uW. Hopefully, this 25uW will be enough to see on a card and use to align the squeezed path to the homodyne. I used a camera and PD to re-align the CLF to the OPO. The alignment looks decent on a scope, but I'm not completely convinced I'm exciting a TEM00 mode. This needs some further investigation.
It seems one can easily convert a cheap webcam (ex: $20 Logitech) into a near IR camera. By default, these are shipped with a filter that blocks IR. So, if we remove that and also replace the focussing filter, the cam can easily be used to image the 1064nm beam. Webcams are essentially plug-and-play and work out of the box. The one I tested has a frame rate of 30fps. So, while we try to configure the Basler acA720-290gm camera for the testbedAI exp, I think we can go ahead and use this webcam for the initial neural alignment training. Images 3-4 are from the webcam for the Mach-Zehnder interferograms (misaligned/aligned), while 5-6 are from the Thor labs beam profiler.
Mounted a retroreflector on the existing suspension in the red lab and the QPD board on a raised plate to test sensor on a free hanging suspension. Still using a passive high pass filter before the demodulation stage and no amplification in the summing/differencing stage. Saw regular motion of the suspension around 1Hz which did not settle even after leaving for ~1hr (~20 Hz range plot). Also measured noise below 1 Hz with suspension "locked" (resting on the jack) and with the LED off (~1Hz range plot). note: except for LED off measurements, could not get a quiet enough signal to leave spectrum analyzer input range at -50dBV, usually settled at -34dBV instead.
Away for two weeks but when I come back we'll look at noise below 1Hz when the retroreflector is more securely fixed with respect to the QPD. Also getting updated qpd board and looking into replacing the corner-cube retroreflector with a cats eye retroreflector which could decrease the footprint of the sensor and increase sensitivity.
The idea of the cat's eye is essentially (pardon the ASCII)
----------| / \--------------||
QPD & LED | <-- D --> | f | <-- d --> ||mirror
----------| \ /--------------||
where the lens and mirror are rigidly mounted to the suspension mass, while the QPD & LED are mounted to the suspension cage. By choosing d = f, the ray transfer matrix of the cat's eye is
[[ -1, 2 * (f - D),]
0, -1,]]
This is a retroreflector in the usual sense of the word, in that any ray arriving from a large distance D >> f will be directed back from where it came from (i.e., the ingoing ray can be described by the vector [r, -r/D] for any lateral offset r satisfying the paraxial approximation, yielding an outgoing ray [r, +r/D]).
However, for our purposes we really do want to image the LED onto the QPD, meaning any vector [r, ϕ] eminating from the LED is imaged onto the vector [-r, -ϕ] at the QPD. In order to do this, we need to null the off-diagonal element of the cat's eye ray matrix, which means enforcing D = f. In practice that might be difficult, since D probably needs to be controlled to within ~1 mm, but is probably okay to try as a first attempt. One could also rigidly mount a second lens in front of the QPD/LED (with focal length f′ equal to mounting distance d′) so that the rays going toward the cat's eye are collimated:
----------|-------------- / \ / \--------------||
QPD & LED | <-- d' --> | f'| <-- D --> | f | <-- d --> ||mirror
----------|-------------- \ / \ /--------------||
In this case, the exact distance D becomes irrelevant, though it should probably be kept small (few millimeters?) to avoid too much light lost.
Today, I re-measured the beam profile after mirror M4 and obtained a result similar to the one reported in entry #11277 (with a confidence level of 3%). I installed the telescope described in the same entry. I took the two lenses with f = 100 mm and f = 50 mm from the 10-foot-long bench currently located in the green lab. To achieve a beam with a waist around 400 μm, I positioned the lenses at z1 = 3 cm and z2 = 20 cm from M4, respectively, as shown in Figure 1 and Figure 2. I measured the beam profile at the output of the telescope, considering z = 0 as the position of the second lens, and the data is shown in Figure 3. The beam waist is approximately 400 μm and is located 23 cm from the second lens. I placed the AOM around the waist.
The next steps will be to install the electronics as described in the entry #11269 and the BS/mirror combination shown in #11270 (but we run out of BS). The final goal is to stabilize the 200-meter-long fibers just arrived from OZ optics.
At the end, to be sure not to clip the beam in the AOM (diameter aperture = 0.5 mm), I replaced this telescope with a f = 77.3 mm lens placed 17.5 cm after the last steering mirror. The waist diameter is around 200 um and located at 12 cm from the lens. I centered the AOM's crystal around the waist position, being careful that the size of the beam was smaller than its active aperture along the propagation.
The AOM is driven with an RF generator for the moment and the optimal parameters set are: frequency = 80 MHz, amplitude = 3 dBm. The amplifier installed after the RF generator is powered with a 24V DC power. The input of the amplifier is 3dBm and its output is 29bBm, meaning that it has a gain of 24 dB (higher that the 20dB expected). Its compression point is 29.5 dBm.
I alignmed the AOM in order to maximize the power on the 1st order diffraction beam:
P(before AOM) = 63 mW
P(after AOM) = 62 mW -> P(0th order) = 22 mW and P(1th order) = 40 mW.
The diffraction efficiency is around 67%.
The shape of the beams exiting the AOM is shown in Figure 1: the 1st order is on the right and it is very elliptical and it diverges very fast. In order to measure the parameters of the beam exiting the AOM, I placed a 150mm lens distant 17.5 cm from the AOM and then I did the inverse matching on JamMt. The data on the different axis measured with the beam profiler are shown in Figure 2 (the reference position is the lens position). The average fit gives w0 (radius) = 96um and z0 = 34cm. It turns out that the waist is positioned 10 cm BEFORE the center of the AOM.
I used this beam to design a telescope on JaMmt to collimate the beam back to 2mm diameter. After several iterations I decided that the best solution (beam not propagating too long in free space, 2 lenses, 0th and 1st order beams separated enough etc) was:
where the positions are referred to the center of the AOM (z0 -10 cm). The resulting beam has w0 = 1113 um (the fixed collimator expects a diameter of 2.4mm) and z0 = 3.77m.
I also thought that it might be better to couple both path in fiber and recombine them on a fiber coupler before being detected to avoid to design and install additional telescope and other complications. The reference fiber is 10cm-long.