Displaying reports 781-800 of 1025.Go to page Start 36 37 38 39 40 41 42 43 44 End
Reports until 11:17, Monday 18 December 2023
DarkMatter General
swadha - posted 11:17, Monday 18 December 2023 (11445)
Search and bound threshold for ADBC data analysis
Non-image files attached to this report
cSQZ General
jacques - posted 18:29, Friday 15 December 2023 (11444)
PLL RP standalone test

I have tested the RP which will be used for the PLL. It is currently misleadingly named shg.mit.edu.

The RP outpout channel 1 is a signal bounded by 0 and 1V. 

I tested a voltage scan at output1, which was fed to a 10x amplifier, then to the Thorlabs PZT driver. The gain and offset have been adjusted so that the voltage out of the driver (which will be ultimately fed to the DBL) is always between 5V and 80V. We are well inside the PZT drive range of the laser (-100V to 100V).

Note for the future: the bandwidth of the 10x amplifier seems to be around 5-10kHz so any correction signal with a frequency higher than this will be filtered.

For amplitude/frequency reference: the signal of the figure is the output of the 10x amplifier, whose input is the RP output channel 1, with the commands  

scopeDecimation =  4096
    # Scan Cavity Mirror
    scan_pll(scanFreq =  1*1/(8E-9*(2**14)*scopeDecimation), 
         sweepAmplitude = 0.45, offset = 0.05)

Next steps: add the PID module to the RP script and close the PLL loop.

Images attached to this report
cSQZ General
hudsonl - posted 17:00, Friday 15 December 2023 (11443)
CSQZ Fiber Test Data Sheets
Attached are test data sheets from AFW for the fibers replacing the broken fiber box.
Images attached to this report
cSQZ General
jacques - posted 19:38, Thursday 14 December 2023 (11442)
PLL RF generator swap to VCO

I swapped the generation of the RF for the PLL, from the Moglabs channels 1 & 2, to the IntraAction VCO model DE-40M used here as a power RF source (see Link for the characterization of the VCO).

From the RF signal of the VCO, I split it using a minicircuits ZA2CS-251-10-WS+ RF splitter. One port goes to the AOM, the other to a -20dB attenuator then to the "LO in" port of the PFD. I turn the carrier frequency and the carrier power knobw at maximum. Note that I did not take a power RF budget of the setup.

I realigned the fiber coupler in the AOM path to be matched to the new ~52MHz sideband. There is now 37uW of power from the AOM path impinging on the PLL PD, whereas about 250uW of light gets converted to the 1st order after the AOM. I did not try to further improve the matching to the fiber coupler. The total power (AOM path + DBL laser path) is still ~220uW as before. This is done by lessening the absorption of the fiber attenuator in the DBM path. 

The figure below shows the "ERR out" port of the PLL (channel 1) and the beatnote around 52MHz (channel 2). We see that even with about 60% less power from the AOM path, the beatnote has the same voltage as before. 

Conclusion: The RF power swap to the VCO is successful.

Images attached to this report
cSQZ General
hudsonl - posted 19:43, Wednesday 13 December 2023 (11441)
SHG EOM U-Bench
The actuators on the fiber-coupler into the U-bench seem to have a lot of hysteresis, which indicates that they're old and need to be cleaned. Hopefully doing so will fix the problem that the U-bench alignment drifts considerablyover the course of a day or so.
cSQZ General
jacques - posted 16:57, Wednesday 13 December 2023 (11440)
Swap to low noise minicircuits amplifier

Following the preliminary test of the PFD (Link) I swapped the {minicircuits TVA-R5-13, 35dB amplifier and the -20 dB attenuator} with a Minicircuits ZFL-500-LN+ of design gain 20dB on the signal of the PD. The new electronic scheme is: PD-> RF Splitter (for monitoring the PD signal on an oscilloscope) -> ZFL Amplifier -> "RF In" port on the PFD. See figures 1 and 2 for pictures of the minicircuits components

Figure 3 shows the signals from the RF splitter (PD signal, channel 2 on the oscilloscope) and the ERR OUT of the PFD (channel 1). The beatnote has a period of 7.5 us so a frequency of 267kHz.

Note that the AOM and LO signals were generated by the moglabs channels 1 (+31dBm) and 2 (+10dBm) at 60MHz. 

Conclusion: there is enough power to conveniently generate the error signal at the output of the PFD.

Images attached to this report
cSQZ General
jacques - posted 14:50, Tuesday 12 December 2023 (11439)
test of IntraAction VCO model DE-40M

For the PLL RF frequency driver I will use the IntraAction VCO Model DE-40M (https://www.lambdaphoto.co.uk/intraaction-de-series-acousto-optic-deflector-driver.html). I summarize below the frequency and the output power of interest for the PLL (I don't use any voltage control or modulation on top of the naturally generated signal). 

Figure 1 shows the spectrum of the VCO output when operated at maximum carrier power and highest frequency. The sine wave is generated at 52.55MHz. The output power, as read on the spectrum analyser, is -11.516dBm, where I used attenuators of 15+20+10+4 dB at the input of the spectrum analyzer. This means that the actual VCO max output power is around +37.5 dBm.

Images attached to this report
Heisenberg IFO General
hudsonl - posted 14:39, Tuesday 12 December 2023 (11438)
Radio-Frequency Photodetectors 2-5 (RFPDs 2-5)
I built and characterized four additional radio-frequency photodetectors (RFPDs) to be used in the compact squeezer and the breadboard OPO. These devices are clones of the RFPD designed for the compact squeezer's CLF and will be used for CLF 2f detection, and PDH locking in both the compact squeezer and the breadboard OPO.

Attached files show the dark noise spectra and transfer functions for each of these four photodiodes. These measurements used a pair of ZFL-1000LN+ Minicircuits amplifiers with 23dB of amnplification to get above the spectrum/network analyzer's noise floor. The tansfer function measurements were made using a modulated diode. I did not observe any non-stationary noise, which had been a problem in the initial RFPD design, which used a higher capacitance photodiode, and caused the RF amplifier to oscillate.
Images attached to this report
cSQZ General
jacques - posted 14:17, Tuesday 12 December 2023 (11437)
Test of PLL error signal out of PFD, generated with PD signal

I used the moglabs channel 1 to drive the AOM at 60MHz sharp, 31dBm; channel 2 as the input of the "LO in" of the mixer, 60MHz sharp +10dBm. I optimized the light power arriving on the PLL PD (total power 226uW, comparable to the reference in entry Link ).  I amplified the PD signal using the +35dB amplifier (figure 1) and added a 20dB attenuator (otherwise the output of the amplifier saturates the PFD).

To get the laser's beatnote (=PD signal) as close as 60MHz as possible, I fine-tuned the temperature of the lightwave laser around 35°C (the Diabolo is kept at 35°C with a -0.59V external temperature voltage control).

Figure 2 shows the obtained signals: channel 1 (yellow) is the output "ERR out" of the PFD. It has a period of 10us so the beatnote frequency f_PD - f_LO = -2*1/10e-6 = -200kHz. To understand the sign convention and the extra factor 2, look at entry Link. Channel 2 is a pickoff from the PD signal, a zero mean sine wave at 60.2MHz.

Conclusion: following entry Link, we are able to obtain the error signal using the actual PD signal. 

Next step: test the Intra-Action VCO model DE-40M and swap it with the Moglab channels.

Images attached to this report
cSQZ General
jacques - posted 21:44, Monday 11 December 2023 (11435)
Test of the Phase Frequency discriminator

The phase frequency discriminator (dcc here https://dcc.ligo.org/LIGO-E1200114) is a kind of two-stage mixer particularly useful for generating a usable PLL error signal. It requires two inputs: "DBL PLL in" (= the PD input) whose frequency and phase are noted (f_PD, phi_PD),  and "LO in" whose frequency and phase are (f_LO, phi_LO). We assume these inputs to be sine waves at these frequencies for the sake of simplicity.

For each of these inputs, the nominal power is between 0 and 10 dBm and the mixer does not work properly if either input has less than -10 dBm. Note that as long as both inputs have more than -10 dBm of power, the output has the same shape, independent of the input powers. The relevant output is ERR OUT. 

Without loss of generality, the initial condition of the PD and LO signals corresponds to | f_PD - f_LO | != 0 and | phi_PD - phi_LO | != 0. In the following I call « beat note » the beat note between the PD and LO signals. For the figures I used RF signals generated around 60MHz from channel 1 and 2 of the Moglabs generator. The LO is always at 60MHz while the PD has different frequency offsets. Both signals' power is +10 dBm. 

 

Step 1: achieve | f_PD - f_LO | = 0 (Frequency mode). If 0< | f_PD - f_LO | <~100kHz, the output "ERR OUT" is a single-sided sawtooth at half the frequency of the PD-LO beat note (| f_PD - f_LO | /2), min/max at +-4V with a mean of +-2.1V depending on whether f_PD>f_LO or f_LO>f_PD. There is a « SIGN » input which accepts either a 50 Ohm terminator or a voltage. The sign of the "ERR OUT" mean is flipped if the voltage on « SIGN » is higher than 1V. 

For high frequency beat notes (| f_PD - f_LO | > 100kHz) the sawtooth signals becomes a sine wave at half the frequency of the beatnote (i.e. first harmonic of the sawtooth Fourier decomposition) + a DC offset of +-2.1V, to keep track of the direction in which the piezo should move. 

Figure 1 shows the sawtooth signal when f_PD - f_LO = 1kHz. Figure 2 shows the sawtooth signal when f_PD - f_LO = -1kHz. Figure 3 shows the low passed sawtooth signal when f_PD - f_LO = 1MHz. Note that the frequency of the sawtooth is always half that of the beat note. There was nothing at the « SIGN » port.

Independent of the initial beat note frequency, this sawtooth signal is the error signal which can be fed into a PID then into a piezo driver then into the Diabolo PZT control in order to obtain | f_PD - f_LO | = 0. 

 

Step 2: achieve | phi_PD - phi_LO | = 0 (Phase mode). Once the PD signal has the same frequency as the LO, the error signal now becomes a DC voltage between -4.4 and 4.4V, whose value depends on the relative phase between the two inputs. By changing the phase of either the laser or the PD, we can zero out this DC voltage. Using the moglabs, a shift of 180° between LO and PD allows to go from +4.4V to 0V (same to go from 0V to -4.4V).

 

Conclusion: the PFD works as intended.

Images attached to this report
cSQZ General
jacques - posted 16:53, Monday 11 December 2023 (11434)
Curiosity: Noise spectra of the Moglabs from channels 1 and 2

I plot the noise spectrum of the different channels of the Moglabs when if generates a -0.22dbm RF signal at 60.000000 MHz. I use a long SMA cable to connect the outputs of the moglabs to the spectrum analyzer. Figures 1 and 2 show the off/on spectra of channel 1 (thus accounting for any spurious peak in the RF domain that the cable might pick up, which we see from figure 2 that there are no such peak). Likewise, figures 3 and 4 show the off/on spectra of channel 2. 

We see that both channels have spurious peaks but channel 2 looks cleaner.

Images attached to this report
cSQZ General
jacques - posted 15:28, Monday 11 December 2023 - last comment - 21:46, Monday 11 December 2023(11433)
RF Power at the PLL PD

With 92uW of light power arriving on the PLL PD from the AOM path only (and 138uW of power arriving on the PLL PD from the Diabolo pickoff for a total of 227uW arriving on the PLL PD), the RF beatnote has a power of -17dbm, without amplification.

Note that the beatnote can lower in time due to polarization drift (on both paths) or fiber coupler mismatch drift on the AOM path. This needs to be checked each time the PLL loop is activated.

 

Images attached to this report
Comments related to this report
jacques - 21:46, Monday 11 December 2023 (11436)

Conclusion after testing the PFD is that we need to amplify the output of the PD so that it is above -10dBm, otherwise the mixer cannot produce an error signal. See entry Link

RetroSense General
msredden - posted 14:02, Monday 11 December 2023 - last comment - 14:04, Wednesday 10 January 2024(11432)
Box for RetroSense board

Includes:

Non-image files attached to this report
Comments related to this report
evanhall - 14:04, Wednesday 10 January 2024 (11468)

I reduced the depth of the #0-80 screw holes to something more like 1/8".

Marie placed with the order with Xometry. They wanted to add 3 mm fillets on the vertical interior edges and on the cutout for the dsub. We'll see if that induces mechanical interference issues.

This new zip file also contains a SolidWorks assembly of the circuit board itself extracted from Altium.

Non-image files attached to this comment
GRAVITES General
epolini - posted 15:48, Saturday 09 December 2023 (11431)
Simulink model results up to 1.6MHz : GAWBS?

I put togehter all the new elements and I run the simulink model to check if we are limited by GAWBS also at higher frequencies.

The updates wrt to the previous study are:

1. SR560 in DC with gain = 100 (before it was LP @ 3kHz and gain = 100) to reduce the noise at high frequency and avoid to be limited by its noise

2. measurements up to 1.6MHz : measurements up to 100kHz have been acquired with the SR785 and from 100kHz and 1.6MHz with the AG4395A (GPIB)

3. the noise of the marconi has been corrected as explained in the entry 11430

 

Simulink parameters:

PDH gain = 270000000

PID : zeros [], poles 2*pi*[0.5e6] (due to the SR560 behavior, see entry 11423), Gain [1]

VCO : zeros [], poles -2*pi*[0], Gain 2*pi*1.5e3

 

In Fig.1 the error signal closed loop simulink model is shown.

In Fig. 2 the correction signal simulink model is shown.

 

Conclusions:

1. the GAWBS is not the limiting factor since at high frequency we have a white noise limiting the measurement.

2. the model of the GAWBS is wrong: in order not to interfere with the correct measurement I need to divide it by at least a factor 10. Just to set it below the limiting noises, I divided it by 20 in this simulation.

3. we considered RIN and shot togheter as RIN + shot = sqrt(3) * shot (electric blue curve). 3 comes from the intensity noise due to the RIN + 2 shot noise contributions coming from the two sidebands at w0+/-80MHz. 

4. The detector noise is the dark noise of the 1811 PD and it has the same shape as RIN and shot noise. I multiplied this by 4 in order to match the model. 

Open question: why do we have this excess of white noise? Are we not considering the RIN + shot + dark noise correctly? Eric suggested that there might be an other source of white noise not considered yet.

Images attached to this report
GRAVITES General
epolini - posted 15:35, Saturday 09 December 2023 (11430)
Marconi noise up to 1.6MHz

We measured the the Marconi noise up to 1MHz using both the SR785 and AG4395A.

Since we are looking at high frequencies, we need to consider the frequency response G of the PLL loop:

Marconi noise * G / (1 - G) = ctrl

Marconi noise = (1/G + 1) * ctrl 

where ctrl is the control signal of the PLL loop composed by the two marconis with symmetric settings and two SR560 servos (one used for the TF measurement set in DC with gain = 1).

The measured PLL TF is shown in Fig. 1, the PLL control signal is shown by the blue curve in Fig. 2 and the Marconi noise is the orange curve of Fig. 2.

Images attached to this report
cSQZ General
masaya - posted 20:31, Thursday 07 December 2023 (11426)
Oscillation of Input power of 532 beam
[Masaya]

There was an oscillation of 532 laser at 27 kHz. 
The amplitude got small by decreasing the gain of SHG control of Diabolo laser.
From this, it is considered the oscillation was because the phase margin of the control of SHG was small and PZT attached to SHG mirrors oscillated.
Images attached to this report
General
hudsonl - posted 17:31, Tuesday 05 December 2023 - last comment - 11:51, Friday 08 December 2023(11420)
CSQZ Fiber Box LO Fiber
The LO fiber out of the Fiber box looks to be very lossy. There's 180mW into the fiber box, and we expect the LO fiber to have roughly 90% of this power, or 160mW. Its output is only about 35mW, and the fiber is clearly glowing near where it's spliced when viewed with an IR camera.
Comments related to this report
hudsonl - 14:48, Wednesday 06 December 2023 (11422)
The LO power is also about 25% of the expected power given the 180mW input power. The LO power is 2.5mW vs 9mW expected. This indicates that the problem is not with the new splice, but with a fiber element common to both the LO and the SHG pump paths.
hudsonl - 11:07, Friday 08 December 2023 (11427)
It looks like the 99/1 fiber splitter is damaged. Connecting the PM fiber out of the laser to the fiber splitter, we get about 80mW out of the 99% port with 220mW in, or about 36% coupling. Using a standard single mode patch cable, we get 180mW out, or 82% coupling, more in line with what I would expect. Given this, I believe we should try replacing the 99/1 splitter before we order new AOMs for the fiber box.

The fibers have a damage threshold of about 1 MW/cm^2, and a "practical safe level" of about 250 MW/cm^2 according to thorlabs (see https://www.thorlabs.com/newgrouppage9.cfm?objectgroup_id=1362&pn=P1-1064-FC-1). The fiber we're using has a mode field diameter of about 6 microns, or an effective area of about 1.1 x 10^(-6) cm^2, which gives a damage threshold of about 300 mW and a "practical safe level" of about 75 mW. Based on this, it seems likely we damaged the fiber splitter by sending too much power at it.
hudsonl - 11:17, Friday 08 December 2023 (11428)
While the fiber daage threshold is quoted at 300 mW, the spec sheet for the 99/1 fiber beam splitter says it should be able to handle up to 1 Watt of power. See https://www.thorlabs.com/drawings/830d2b7c309f8c74-E4C1FA78-A61A-FC1E-0E705545D0A363D3/TN1064R1A1B-SpecSheet.pdf
hudsonl - 11:51, Friday 08 December 2023 (11429)
Connecting the fiber-coupled output of the laser directly to the input of the fiber box, we only have about 25% transmission from the input to the pump fiber. We send in about 205 mW and get about 55 mW on the pump path. This indicates that while the 99/1 beam splitter may need to be replaced, there is a sperate problem with the fiber box as well.

Connecting the output of the 99/1 to the input of the fiber box, we have about 200mW in, about 100mW out of the "99%" path, and about 20mW out of the pump path fiber from the fiber box, so the losses all add up, and it looks like there is a problem both with the 99/1 beam splitter and with the fiber box.

The 1% pickoff port of the 99/1 has 1mW of power, so it looks like both paths have about 50% loss. This makes sense, as the damaged fiber looks to be the input to the beam splitter.
Displaying reports 781-800 of 1025.Go to page Start 36 37 38 39 40 41 42 43 44 End