I tested the "SHG" (MASTER) RP's network analyzer function, using the diagram: Excitation -> out1 -> in1 -> NA, where a short SMA cable was used for the connection from out1 to in1. The Excitation is sent directly to out1; there is no PID in the setup.
Below are figures of what I get for the frequency response, for testing out1->in1 and out2->in2 respectively.
Interpretation: looking at the magnitudes, there are weird resonances at 15MHz (out2) up to +6dB, and a double peak at 30MHz (out1) up to 20 dB. We are below the Nyquist frequency of 62.5MHz so I don't understand why these resonances exist. The phase's rolloff at >100kHz is probably due to some internal delay to the RP.
Conclusion: I don't investigate this further because this RP is used for the PZT drive only (limited at 50kHz anyways), but we should keep this in mind for future reference. Hudson showed me plots of the same measuresments on another RP where this resonance did not exist.
The loop is now stable enough so that I can start taking transfer functions using the RP. I used the same code as Hudson in his plots Link.
The figures below show some transfer functions for different values of P and I. I still need to process them to understand if the loops' UGFs and phase margins evolve accordingly.
After a quick discussion with Eric, it seems that there is some extra slope at higher frequencies, which could be explained by an extra lowpass above 10kHz. I am yet to find where this lowpass is. Furthermore the loop is very noisy at frequencies below 1kHz.
Conclusion: I can take transfer functions but there seems to be subtleties not yet understood at high and low frequencies.
Some ideas for future investigation to see if they can improve the loop:
- Take the spectrum of the error signal at low frequencies.
- Test the RP PID independently by looping its output on a 50Ohm resistor into its input and running the built-in netword analyser, to see if there are any extra lowpass (. This should also clarify the physical meaning of the parameters P and I in the code.
- Write down the analytical model of the loop and compare to the experimental data.
- Currently very low light power comes from the AOM path into the PD (30uW). Improve alignment and mode-matching in the AOM path.
- Feedback the PZT driver output on the Lightwave instead of the DBL and see if the error signal looks less noisy.
For reference, here is the shape of the PD spectrum around the beatnote carrier.
After discussion with Eric, it seems that the added slope above 10kHz is simply due to the input lowpass prior to the PID that I put at 50kHz. For reference, below is a measurement of this lowpass filter only, done by the sequence: network analyzer excitation -> out1 -> in1 -> PID (P=1, I = 0) -> out2 -> input2 -> NA's input. This lowpass was put here because the manual of the DBL laser recommanded not going above 100 kHz for the PZT drive as this could damage it (by exciting its resonances).
Conclusion: I should redo these TF without the 50 kHz, as I expect the PZT resonances to be well above the current UGF. I should plot the TF up to several MHz so that I can also see the PZT resonances. In a second step, it would then be useful to add band filters at the frequencies of the PZT resonance to selectively kill them.
I suspected that the 10x amplifier added noise to the correction to the laser PZT Link, so much that the loop could not be stable Link.
I removed the 10x amplifier before the PZT driver and tried to close the loop. I got better results than last time, with PID_response(0.5, 0.01, 5, 0) (order is offset, P, I, D). On a scope (figure 1) the error signal (channel 1) actually seemed not to saturate, while its spectrum was more reasonable than last time (figure 2). The spectrum of the PD signal is stable (figure 3) with the peak at the driving frequency,
At this stage, the drawback of this 10x-bypass scheme is that we have less range in the PZT and thus need to adjust the temperature or the offset of the PZT driver more.
Conclusion: The PLL stability was improved by removing the 10x amplifier at the cost of having less range. I believe that we can keep the setup this way provided that we also actuate on temperature.
Next step: fine tune the coefficients of the PID loop as is and actuate on the temperature.
It was noticed previously Link that when the pll is closed, there are extra sidebands in the PD signal (beatnote between the two lasers), whose frequency offsets are ~284kHz from the carrier beatnote.
One hypothesis was that these sidebands came from the VCO generator Link.
I swapped the VCO with the moglab's channels 1 and 2, driving at exactly the same frequency of 51.56875MHz, with about the same amount of power. Figure 1 is the spectrum of the PD signal using the VCO when the loop is closed, figure 2 is that using the moglab.
Conclusion: since both spectra look similar, I conclude that the sidebands don't come from the RF source.
It is not clear to me at this stage if these sidebands are detrimental to the stability of the PLL, thus if further noise hunting in that respect is useful. Furthermore it is possible that these sidebands are resonances of the PZT of the laser.
Thanks to Fred we solved the issue to connect the stage to its controller. I report here the procedure to solve the issue in case we will not be able to run the controller program again
Go to Control Panel, Network and Internet, Nerwork and Sharing Center
Change adapter settings
Go to Ethernet 2
Right click on Properties
Configure on Internet Protocol Version 4 (TCP/IPv4)
Double click
Configure ethernet 2 with a 192.168.0.1 address, netmask 255.255.255.255
The control seems to have an ip address of 192.168.0.20 port 8088
If the controller interface is still not dispayed, you should do ctrl + alt + canc and open the task manager. If Fuyu is running several times this is the issue and you should close the different sessions.
Re-measured finesse for new cavity configuration: p 253, s 6634
Fifth dataset p-s splitting = 10.3 MHz
Measured spectrum, gaγγ bounds, combined bounds from all measurements
Re-measured finesse for new cavity configuration: p 212, s 7256
Fourth dataset p-s splitting = 10.03 MHz
Measured spectrum, gaγγ bounds, combined bounds from all measurements
I found an unused small optical platform (3.2 kg) and in order to use it as a support for the spool on the stage, I asked JD to machine an adaptor plate.
JD made the plate and it fits perfectly with the stage as shown in Fig. 1 and Fig.2. It has a squared hole on one side so if we want to perform precise measurement we will want to machine an other one. It is attached to the stage via 10 screws (M6) and to the black platform (M4) with 8 screws.
I compared the performance of the PLL using a LB1005 analog servo versus the RP's digital PID. I tried to tune the gains and integrator frequency by eye to try to minimize the error signal on a scope.
Figure 1 shows the spectrum of the PD signal (should be a peak around 51.7 MHz) using the analog servo. Figure 2 shows the spectrum for the RP.
Conclusion: The PD signal from the digital actuation shows a much nicer look so at this stage, the RP seems to be performing better than the LB1005.
Yesterday with Eric we noticed a very peculiar phenomenon happening when the correction sends a DC offset (so that the PZT is at the middle of its range), then the PID is switched on, then off, while the DC offset is kept on. Figure 1 shows the error signal when only the DC is sent. This is strange because there are stable, periodic squarish oscillations at ~6 kHz, while no actuation is being done. This phenomenon appears whether the PID is being done by the RP or by a Newfocus LB1005 analog servo controller, which means that it could not have come from the PID block in itself. Furthermore it appears whether I use channel X or channel Y of the PZT driver
Today I put the correction PID back to the RP and compared, for fixed {offset = 0.5, P = 1e-4, I = 1, D = 0}, the PLL's Thorlabs PZT driver with the one used to lock the csqz. Figures 2 and 3 show the time domain error signal and the frequency domain PD signal for the csqz PZT driver, while figures 4 and 5 show them for the PZT driver used during the previous days.
Conclusion: the weird 6kHz oscillation comes from the PZT driver, though the exact phenomenon by which it appears remains unknown. We should use another PZT driver.
Hypotheses which can be tested in the future to improve the loop:
1/ On the weird sidebands which appear on the PD signal when the loop is closed with the RP: it is possible that these come from the VCO. A test could be to swap the RF generator back to the Moglabs at the same 52MHz frequency and check again for the presence of these sidebands.
2/ On the stability of the loop: one possible cause of instability is that the proportional gain is too high. I cannot reduce it further because of internal RP contraints. Furthermore the noise figure of the 10x amplifier is unknown; it is possible that this components adds too much noise to the correction. A test could be to swap out the 10x amplifier. The drawback is that we will only have ~10V of actuation range vs 75V initially but this can be managable for the short term, where we are only interested in having a stable loop for a short while.
Today, after receiving the 1km spools spliced from Gabor, we started assembling the setup at 1550nm in green lab. The first setup will be an unbalanced MZI, as the 200m prototype in NW22 at 1064nm, but using the new 1550 nm components and the electronics designed by Eric.
Eric and I measured the output power of the laser set to its minimum which is 200mW. Even if on the laser it is written output = 68mW, the measured power is around 190mW. We took care of cleaning the fiber exiting from with laser since it looked very dirty on the fiber scope. We cleaned it both with the dedicated tape and then with methanol. We installed the PM FI in fiber and before we get a power of 177mW which agrees with specs losses of 0.55dB.
After that, I installed a 90/10 PM fiber BS to dump most of the power. I connected the port 1 (177mW) to the FI and then I measured:
This gives ~6% of power losses. I connected a fiber dump to port 3.
The I connected the port 2 to the input port 1 of a 50/50 PM BS and I measured:
This gives almost 8% of losses.
I connected port 3 of the 50/50 PM BS to one of the SM 1km spools. The spool we are using now it the one were we heated the plastic tubes that Gabor put to protect the splicing connection. We should do that also with the other spool.
After the spool I measured 6.55mW, that wrt to the 7mW in input gives 6% of losses. Considering 0.2dB/km (2.3%) losses and 1-2% due to the splicing, the remaining losses should be due to the connection PM-SM.
We wanted to install also the AOM but we miss the SMB cable.
I cleaned and organized the bench and put back the stage horizontally so that we can start to think about the stage vibration performances and logistics. To do so, we need to understand which platform to place the spool on. I found a small black optical plate of roughly 30cm x 30cm, that weights 3.2kg. The stage should support 10kg, so that could be an option if we machine some holes to fit the stage support.
We will test the polarization controller soon.
Last Friday, I continued the installation of the components after we received the SMC-BNC adaptor for the AOM from Thorlabs.
I soldered and installed the RF amplifier (the high power one, up to 2W) and I measured its gain using the Marconi. I measured the gain by checking how much the 80MHz oscillation sent with the Marconi increases at amplifier output. Sending a 120mV amplidude signal, we get 5.9V output signal. This gives a gain = 49 that converted in dBm is 20log10(49) = 33dBm.
I set the amplitude on the Marconi (channel A) to 0dBm in order to have an RF power reaching the AOM of 33dBm which correspond to 2W. The AOM max power according to its datasheet is 2.2W.
I measured the optical power before the AOM : P(before) = 6.44mW and after the AOM : P(after) = 3.34W, which is roughly what I would expect. The input power sent by the laser is still 200mW (dumped at the beginning with a fiber BS).
I used the channel B of the Marconi as a LO for the demodulation and since the demod board is not ready yet, I installed a mixer instead. I set the LO amplitude to 7dBm.
I installed an 1811 PD (which I removed from the 200m prototype in NW22) and I sent the error signal (output of the mixer) to the servo board the Eric designed.
The figures atteched show the setup at the moment.
Before closing the loop I want to install the polarization controller that will convert the SM output from the 1km spool back into PM so that it can be recombined with a PM 50/50 BS to the reference arm.
Re-measured finesse for new cavity configuration: p 152, s 6043
Third dataset p-s splitting = 13.54 MHz
Measurement spectrum, gaγγ bounds, combined bounds from all measurements
I closed the loop of the PLL for the first time by coding a PID on the RP (with P = D = 0, I = 10) and sending the output to the 10x amplifier then into the PZT driver. The output voltage is between 4 and 65V to make sure that the PZT port of the DBM is not damaged.
I preemptively made sure that the beatnote of ~55MHz can be seen around the midvoltage (~34V) in order to maximize the range of the PZT drive.
Figure 0 shows the spectrum of the PD beatnote when the loop is not closed.
Figure 1 shows the spectrum of the PD beatnote signal (at ~55MHz) once the loop is closed and figure 2 shows an example of the PFD "ERR OUT" port.
To test the robustness of the loop I slightly changed the temperature of the lightwave laser by turning its temperature knob and I could see on the spectrum analyser that the plateau at 55MHz stayed in place.
For reference, figure 3 is a spectrum of "ERR OUT" after being attenuated by 6+5+9=20dB
The next step should be to:
1/ better understand these figures and improve the loop's response.
2/ send the error signal's low frequency to a PID then to the temperature control of the DBL
Re-measured finesse for new cavity configuration: p 150.5, s 5639
Second measurement: