Today I changed the optical setup of the long arm of the UMZI to introduce some pick-offs for the monitor of the polarization before and after the polarization controller.
After the 1km spool there is a SM AOM and then I added a 95/5 BS and I connected the 5% port to a SM PBS and the 95% port to the polarization controller. After the controller I added a 95/5 PM BS and I connected to the 5% port to a PM PBS and the 95% port to the other PM BSs before the 50/50 BS and the PD.
I connected two PDA20CS thorlabs PDs to the outputs of the PBSs. The gain on the PDs is 10dB. I measured the powers with the oscilloscope and I got:
The signals did not change after the polarization controller by disabling it (the change in polarization might be slower then what I observed it for).
The result of this measurement is that the controller seems to work.
I also acquired the RIN before and after the polarization controller. I measured the power spectrum with the SR785 at the 5% port of the new BSs I installed, meaning before entering the PBSs.
The power level on the SM side (BEFORE) is 336 mV and on the PM side (AFTER) is 176 mV. I used these DC powers to normalize the spectrum shown in Fig. 1. The low frequency of the spectrum is higher after the polarization controller.
Today I measured the RIN also before the spool and with the polarization controller OFF, to understand what is introducing the excess of noise at different frequencies.
Fig. 1 shows:
The structrure between 2kHz and 4kHz is amplified by the spool, since it is way smaller before and it is present independently from the PC. The noise at low frequency (< 1kHz) is excited by the PC. To address this problem we can consider to feedback a signal to control the intensity noise after the PC.
To understand if it is possible to easily reduce the 2-4kHz noise, I covered the spool with the box we built for the Michelson setup.
Fig. 2 shows:
The noise is not reduced by covering the spool with the box, so we might need to think of something better to isolate the spool.
Continuing to repeat Maya's measurements (11411) at about half the optical power (sum voltage about 6.3 V, or about 1.3 mA of photocurrent) and also now in the tank.
Measurement 1 was done in the tank with the lid on. I replaced the thick dsub cable with just a ribbon cable for less mass.
For measurement 2 I did a lot of things: resoldered the LED, cleaned flux off the PCB, reseated the whole thing, and added mechanical parts to make a stiff connection right at the dsub connector, to reduce vibrations from the cable. This last intervention helped reduce peaks above 5 Hz, but I suspect that the improvement below that may be a result of less seismic noise. The noise there is nonstationary. The dark noise is consistent with the noise of the AD630 demodulator.
In measurement 2 the sensitivity was 16 V/mm or 3.2 mA/mm, which implies a characteristic spot size of 0.4 mm. The modulation was at 10.2 kHz.
The attachment shows the dc photocurrent response versus displacement of the retroreflector along the X axis. The slope in the X photocurrent is 3.2 mA/mm; the slope in the Y photocurrent is 0.08 mA/mm (cross-coupling of about 3%).
The noise rises broadly when the X offset is further away from zero. The sensitivity in the range ±0.25 mm is equal to or better than the standard BOSEM.
By making the coherence between the X displacement noise and the sum (intensity) noise, we can see that the excess noise at nonzero X offset is indeed predominantly an intensity noise coupling, as would be expected. The solid lines are the raw spectra; the dashed lines are the projections made by multiplying the spectra by sqrt(1 − γ2). Evidently, the amount of (linear) intensity noise coupling at zero offset is not dominating the noise floor.
The RIN at 1 Hz is about 10−6 / Hz1/2, which is about 100 times above the shot noise.
I tested the susceptibility to oscillator phase noise by frequency modulating the SRS DS345 function generator. The carrier is a 10274 Hz square wave. The modulation was a 10 Hz span with a 3.1 Hz modulation frequency. A line appeared in the X spectrum at 3.1 Hz with an amplitude of 0.4 nA/Hz1/2. I changed the carrier frequency to 3274 Hz and the line amplitude decreased by a factor of 3. The noise floor did not change.
I tried making a small cylinder of tape around the retroreflector to reduce the amount of air currents. The noise floor did not change.
The voltage regulator for the LED supplies 9.4 V. The resistance is currently dialed to 420 Ω, so the LED is being driven with 22 mA.
The voltage regulator noise is about 500 nV/Hz1/2, meaning 1.2 nA/Hz1/2 of current noise being delivered to the LED, or a relative noise of 5×10−8 / Hz1/2. This is comparable to the relative noise of the photocurrent that was seen when the LED output was monitored with a PDA100A.
A selection of time-domain measurements. I did not include the time series of the final output, which is a 2-pole low-pass filter at 300 Hz.
To simulate a grossly misaligned interferometer configuration accurately using Finesse3, where misalignments are of the order of a few hundred micro-radians, we need to include a lot of higher-order modes (max_tem=20-40). The scaling between the degree of misalignment and the number of HOMs is quadratic and thus slows down the simulator tremendously. For the testbedAI case, as part of the RL model-based training, we must continuously monitor and update mirror alignments a few thousand times to learn the optimal controller.
We have two different ways to reduce the simulation time.
(1) Adaptively refine the number of higher-order modes based on the power recovered on the final photodiode. We start with max_tem=15 and update it in units of 5 till we see less than 10% improvement in the recovered power. Note for max_tem ==N, the number of higher order modes = 0.5*(N+1)*(N+2).
(2) Based on discussions with Kevin and Dan Brown (Adelaide), we figured out that Finesse3 spends a lot of time building and deleting the model per simulation call. So we can sort of hack into Finesse3 and use some undocumented low-level functions to make the simulation object persist between the calls, modify the parameter, and get the results. This trick works if we initially fix the max_tem parameter to a higher number (30 - 40) and then dynamically update the parametes. Max_tem cannot be tuned on the fly within a persisting model.
Another way would be to run many Monte Carlo simulations once, then build a reduced order model (state space) and use that within the RL environment. I will first check the results with (1) and (2).
Time comparison/scaling plot attached.
[Eleonora, Eric, Medha]
We built a Michelson interferometer as shown in Figure 1. Its schematic diagram is shown in Figure 2. Before the interferometer, we added a 99/1 beam splitter and isolator to limit the power of light entering.
Measurements:
At the input of the 99/1 beam splitter, there was 150 mW of power initially and the outputs had 130 mW and 1.3 mW. After the isolator, the 1.3 mW of power reduced to 1.15 mW. This was split between the 2 output ports of the next 50 50 beam splitter. The lower arm (shorter) had 0.49 mW and the upper (longer) arm had 0.43 mW. After the 1 km long spool on the upper arm, the 0.43 mW dissipated to 0.38 mW. The other "input" end of the 50/50 beam splitter was connected to the PD which was connected to an oscilloscope. A power meter was connected right before the PD. When the bottom arm was opened, the power meter showed 0.26 mW and when the top arm was opened, it showed 0.19 mW. Then we adjusted the VOAs on both arms until both measurements were the same (at 0.112 mW).
Once the oscillascope was connected, the visibility could be calculated through the formula: (P_max-P_min)/(P_max+P_min).
A higher visibility is better. Initially the visibility was around 50% but after changing the VOAs, adding the AOM, and removing noise, the visibility jumped up to 96%.
AOM installation:
In order to be able to see the fringes better, we added an AOM before the 50/50 beam splitter (PM AOM FibreQ 1550). We drove the AOM with a Marconi at 40 MHz and 24 dBm amplitude. The modulus of range was 10 mHz. Yesterday the visibility was around 65% but today we reached 96% after tuning the VOAs. Still, the fringes are not completely sinusoidal because of acoustic noise as shown in Figure 3. The power after the AOM was 0.5 mW.
Reducing noise:
In order to reduce the acoustic noise, we built an insulation box for the 1 km spool. We used this "sponge-isolating-material" in a cardboard box (figure 4 and figure 5) and covered the sides and top and covered the spool. We also covered the bottom with rubber but did not glue yet it to avoid damage. After the addition of this box, the fringes were definitely less susceptible to acoustic noise. However, they still do not look completely sinuisoidal.
[Eleonora, Medha]
We moved back the Marconi from NW22 to green lab since we were running out of RF generators/VCOs. Since the one we moved had the internal clock broken, we tried first to connect it to the other Marconi we are using in green lab but it was not working (also in NW22 we were using an SG384 as clock for both marconis) and then we took a DS345 synthesized function generator from the E-shop to use its internal clock 10MHz. We noticed that it is not possible to set this Marconi down to 40MHz, but it works at 80MHz. So we are using the previous Marconi to drive the SM AOM and this new one as LO for the unbalanced MZI setup.
Yesterday, I closed the loop of the unbalanced MZI with 1km spool on one arm. The scheme of the setup is shown in Fig. 1. We used as LO and VCO the two channels of a Marconi, as servo the PI PCB board (using only the first integrator with PI corner = 480Hz) and a minicircuits mixer.
Parameters used:
LO 80MHz, 7dBm
VCO settings 80MHz, -5dBm, FM modulation 10kHz (which corresponds to a calibration factor V to Hz to radians of 6.5kHz/V).
PI corner = 480 Hz, gain ~ 1000 (estimated from the OLTF)
The error signal calibration between volts to radians is givend by the amplitude of the fringes after the mixer, which was A = 25mV/rad (so I divided the spectrum by this factor).
In Fig. 2 the error and correction signals are shown and they cross around 17kHz, which is the UGF, as shown in the OLTF in Fig. 3.
Next steps:
To measure the RIN after the PM PBS with the polarization controller ON and OFF and at different times.
To complete the simulink model of this setup including all the noises to verify what is limiting us.
To measure the control and error signals while the platform is on the moving stage.
I measured the intensity noise of the RetroSense LED using a PDA100A. I don't think all the light was on the photodiode, but in any case it seems that we are far from shot noise by a factor of 5 or so in amplitude. By changing the dc photocurrent by a factor of 2, the noise floor also changes by the same amount in amplitude. The dashed lines in the attachment are the shot noise levels corresponding to the dc photocurrent values. The LED is Vishay VSMY2850G.
The next task is to measure the current noise driving the LED.
I measured the noise of the VCO board and compared it with the noise of the Marconi operated as LO (FM mod OFF).
Measurement of the noise of the Marconi as LO
The Marconi we have now in green lab has two channels. I set both channels to 80MHz but with different amplitudes (7dBm for the LO and -5dBm for the RF channel). I set them in LO mode by turning OFF the FM mod. I beated them with a minicircuits mixer and filtered the output with a 5MHz LPF. I acquired the spectrum with the SR785 and this gives us the noise of both marconis in LO mode. The two signals have a close enough phase since the output of the mixer is around zero. In order to measure the calibration V to radians, I shifted one of the two signals by few kHz and the slope of that sine wave is the calibration factor. It is given by the amplitude of the wave and in our case it is A = 200mV/rad.
Measurement of the noise of the VCO board
I closed a PLL with the servo board designed by Eric. I used the VCO as LO and I beated it with the output of the VCO (attenuated by 20dBm in order not to send ~0dBm to the RF channel of the mixer). I measured the TF and tuned the gain in order to have a loop bandwidth of ~50kHz. The loop TF is shown in Fig. 1. The UGF needs to be considered at -20dB since the servo board is designed in a way that the two signals needed have a ratio of 20dB between them. There is a weird just around the UGF. There is a weird oscillation at few MHz once I close the loop that could go away after opening/closing the loop serveral times, discharging the integrators, removing the cables attached to the servo board needed for the TF + turning off the loop etc. I tried to perform the measurement of the correction (board noise) and error signal when the oscillation was not present. To calibrate the correction, I used the frequency response of the VCO shown in the previous entry and then I divided it by the frequency.
Fig. 2 shows the comparison between the Marconi LO noise and the VCO board noise. The marconi is lower noise in this case, meaning that it can be used as a LO for the measurment.
At 10kHz, the noise is 4.3e-5 rad. If we convert this value in dB we get -87 dB at 10kHz. Comparing this number with the datasheet for 800MHz, we should have -117dB at 10kHz which would be -137 dB. This noise is 50dB lower than what we measured, so we need to check the VCO board or think about other possible problems.
[Eric, Eleonora]
Since the loop was oscillating, Eric suggested to modify the servo board, removing the part that was sending the voltage offset to the VCO board, as shown in Fig.1. One other problem identified was that the PI frequency of the second integrator on the board was at 106kHz, that was too high, so now we are turning on only the first integrator with PI corner at 480Hz. The TF is shown in Fig. 2 and the UGF needs to be considered at -20dB and it is around 70 kHz. The LO was set to 7dBm amplitude and 80.001 MHz and the RF to -5dBm and 80MHz, with FM mod at 10kHz.
We will use the Marconi as VCO for the moment.
Jacques and I had quickly set up a PDH lock for the compact squeezer, but we didn't push the modulation frequency and were not very careful about which amplifiers we used. We had used an overly bulky amplifier between the Red Pitaya and the EOM input. I've set the PDH electronics back up with the electronics rack underneath the table and using less bulky amplifiers. The current setup uses a ZHL-3A amplifier between the Red Pitaya's output and the EOM's inout. It then uses a pair of ZFL-1000-LN+ low noise amplifiers between the RFPD's output and the Red Pitaya's input. This combination seems to give a reasonably good SNR for the PDH lock. The attacked file shows the error signal as the SHG is scanned.
The compact squeezer's SHG locks consistently with a PDH lock and stays locked as I adjust other optics on the compact squeezer board. Its UGF is about 500 Hz with quite a bit of phase margin. Masaya made some measurements of the 10x amplifiers we've been using and their phase starts to die rapidly above about 80 Hz. Given that this is the case, it's not exactly clear to us how the SHG stays locked at all, since one of these 10x amplifiers is on the actuator path in the feedback loop.
Today, I measured the frequency response (Fig.1) of the VCO board designed by Eric that we assembled few weeks ago (shown in Fig.2).
The board consists in the VCO ROS-800-719+ by minicircuits that has a frequency range from 797 to 806 MHz, which corresponds to 5V tuning range. After that there is a frequency divider AD9515, set to a division ratio of 10. This has been done in order to reduce by a factor 10 also the phase noise.
To drive Peltier in the second OPO, the PCB named TEC Drive v2 is used.
The schematic diagram is shown in the attached pdf file.
Specific important numbers are selected as
R1 (= R6) = 680 [kΩ]
R2 (= R7) = 200 [kΩ]
R3 (= R8) = 330 [kΩ]
Vin = 3.225V
The opamp used in this PCB (TPS63020DSJR. see the datasheet attached) sets V_FB = 0.5 [V]
For V_CTRL as the control signal,Vout is
Vout = {(R1R2+R2R3+R3R1)V_FB - R1R2V_CTRL}/R2R3
Red Pitaya output range is from -1 to 1 V.
Vout changes from 5.29[V] (V_CTRL = -1 [V]) to 1.16 [V] (V_CTRL = 1 [V])
The input voltage to the Peltier(The used Peltier (Laird Technologies, HOT20.31.F2A.0909. Data sheet should be attached.) V_TEC is
V_TEC = Vout - Vin (-2.065 to2.065 [V])
To measure the temperature of PPKTP crystal in the second OPO, the PCB named Thermistor Readout is used. The schematic diagram is shown in the attached file. Specific values are set as below. V_DD = 9 [V] V_ref = 5 [V] R1 = 7.5 [kΩ] R2 = 1 [kΩ] R3 = 6.8 [kΩ] Resistor value of the used themistor (10k Siemens-Matsushita B57861 series. The data sheet(NTC_Mini_sensors_S861.pdf) is attached.) R_NTC = 14.962 [kΩ](T = 15 [℃]) R_NTC = 3.4033 [kΩ](T = 55 [℃]) Vin and Vout are Vin = R1/(R1+R_NTC) * V_DD Vout = (R2+R3)/R2 * Vin - R3/R2 * V_ref Vin changes from 3.02 [V] (T = 15[℃]) to 6.15 [V] (T = 55[℃]) Vout changes from -10.44 [V] (T = 15[℃]) to 13.97 [V] (T = 55[℃])
After moving the compact squeezer and the electronics box, the SHG alignment had changed, although we still saw clear cavity flashes. It was easy to re-align the beam into the SHG by adjusting the tip/tilt of the fiber collimator and the PBS. Attached images show cavity transmission scans prior to and after alignment.
I made a wooden platform for the PZT driver for the compact squeezer. Without the platform, the PZT driver was occasionally shorting to the rack it was placed on, since it was easy for the feet to fall between the grid of wires forming the rack and allow the exterior of the PZT driver to short to ground. In usual operation, the driver's exterior should be electrically isolated from ground.
Yesterday, with the help of Fred, we managed to connect the polarization controller POS-002 (general photonics) to the computer. We used a computer power supply at +/-12V since we are running out of other easy solutions. The pinage of the power supply is shown in Fig.1 and how the cables have been connected in Fig.2. The manual of the polarization controller can be found here. The black cable from the POS-002 is the pin 2 and correspond to the +12V pin, the white one to ground and the grey one to -12V. We connected the black cable to pin 10 of the power supply, the white one to pin 17 of the power supply and the grey one to pin 14 of the power supply. In order to turn the power supply on we connected pin 16 to pin 18. Fig. 3 shows how the setup looks like.
Initially, I connected the 9-pin cable from the POS-002 directly to the computer and we tried different software to communicate with it (termite, SecureCRT, YAT). We verified that the controller was powered, that the pinage was correct and that the instrument was seen by the computer. We coudn't make it work. Fred tried to connect the cable using an USB adaptor and then we were able to speak with it using YAT (a software that now is on the computer desktop). The commands are written on the POS-002 manual. I was able to enable and disable the controller via command lines. The next step would be to use the Labview scripts to change the feedback loop.
I also adjusted the setup to avoid the PD 1811 satutation.
Power budget (input power 200mW):
Since on the short arm the power is more than double the power on the long arm, and the two are recombining at the second 50/50 BS in front of the PD, I installed a 50/50 PM BS there.
The new power on the short arm is 2.78 mW. The power reaching the 1811 PD is now 2.14 mW.
The signals from the PD were saturating, so I installed a 99/1 PM BS after the second 50/50 BS. The powers at this BS are:
I connected a fiber dump to port 3 and I sent port 4 to the PD. The DC power read on the oscilloscope from the 1811 is 280 mV and the pk-pk voltage of the RF channel is 1.15 V.
The final setup is shown in Fig. 4.
We have replaced the fiber box with connectorized fiber. We send roughly 250mW of power towards the first fiber splitter in the chain, which has a coupling efficiency of roughly 50% as of the last time we measured it. On the SHG pump path, this power then goes through a faraday isolator, a 99/1 beam splitter (1% is sent to the PLL), a 90/10 splitter (10% is sent to the CLF and LO paths), a patch cable, a U-bench with a free-space EOM, polarization paddles, and finally to the fiber to free-space collimator on the compact squeezer optics board. Currently, with 250 mW of power launched at the fiber, we get about 17mW of power out of the SHG's fiber collimator. A large source of loss on this path is the U-bench. Currently, we have about 55mW sent towards the EOM out of the first fiber collimator, but only about 25mW makes it into the fiber with the polarization paddles. We anticipate that the U-bench should be able to be aligned such that it has 70-80% coupling. With the current setup, this would give about 26mW or power out of the SHG's collimator. This 26mW is still low, so there are several other ways to reduce losses. (1) Replace the thorlabs single-mode fiber through the polarization paddles with fiber from ADF, which should be higher quality. We already have this fiber, so this should be easy to try. (2) Remove the patch cable between the 90/10 beam splitter and the U-bench. This patch cable is not necessary in the current configuration. And its inclusion likely introduces about 0.3dB of loss. (3) Splice the fibers together. This should reduce losses through the fiber network, but will revert us back to the scenario where the fiber components are monolithic. We would like to avoid this if possible. (4) Remove the polarization paddles and replace them with ThorLabs U-bench mounted lambda/2 and lambda/4 waveplates after the EOM. These waveplates have less than 1% loss each, so the total coupling from the U-bench input to SHG pump light would be 70-80% efficient, compared to about 50% efficiency currently (even if the U-bench was aligned so it had 70% efficiency). This is easy to implement, but will cost about $700.
Includes:
I reduced the depth of the #0-80 screw holes to something more like 1/8".
Marie placed with the order with Xometry. They wanted to add 3 mm fillets on the vertical interior edges and on the cutout for the dsub. We'll see if that induces mechanical interference issues.
This new zip file also contains a SolidWorks assembly of the circuit board itself extracted from Altium.