Displaying reports 721-740 of 1025.Go to page Start 33 34 35 36 37 38 39 40 41 End
Reports until 16:06, Friday 09 February 2024
cSQZ General
jacques - posted 16:06, Friday 09 February 2024 - last comment - 17:02, Friday 09 February 2024(11506)
PLL better phase noise spectrum

With more tuning of the PI parameters and swap to Low voltage input on the RP, the loop can be closed and we have 11mrad of phase noise between 1kHz and 100 kHz.

P = 0.09, I = 45

However I see on the oscilloscope that the correction signal is clearly oscillating. We should take a spectrum of it at higher frequencies to see if it's due to the PZT resonance at 270kHz.

 

Non-image files attached to this report
Comments related to this report
jacques - 17:02, Friday 09 February 2024 (11507)

Spectrum on the RP. I should move to an actual network analyzer to resolve the TF at <1000Hz.

Images attached to this comment
GRAVITES Optics 1550nm
epolini - posted 20:06, Wednesday 07 February 2024 (11505)
Michelson interferometer for laser frequency noise stabilization inside the box

Today, I tried to insert all the optics after the AOM of the Michelson interferometer inside the box that we built few weeks ago. How the fibers are organized inside the box is shown in Fig. 1.

The shape of the fringes is not really improved, as shown in Fig. 2. I tuned the frequency driving the AOM in order to minimize the error signal (after the mixer) and the error signal is shown in Fig. 3. 

The visibility was around 60% and I could not it increase it further.

Images attached to this report
cSQZ General
jacques - posted 19:17, Wednesday 07 February 2024 - last comment - 19:11, Tuesday 20 February 2024(11503)
PLL phase noise spectrum

While trying to start the PLL today, I noticed that it would often lock at the wrong frequency (43.7MHz instead of the usual 51.6MHz). I realized that one of the SMA cables, connecting the output of the minicircuits amplifier (amplifying the PLL PD signal) to the PFD "DBL in" port, was broken. I swapped it out and the PLL no longer locked on the wrong frequency. I used P = 0.2 and I = 2000.

I then acquired some spectra of the error signal on the SR785 and analyzed them in python. Below are two plots of the error signal spectral density when the loop is closed, respectively between 1Hz and 800Hz and between 100Hz and 100kHz. I also computed the corresponding integrated phase noise. For this unoptimized loop, there is ~60 mrad of phase noise between 1Hz and 800Hz, while there is ~35 mrad of phase noise between 1kHz and 100kHz. I used the PFD's spec sensitivity of 36°/V = pi/5 rad/V to convert the voltage spectrum into rad.

After discussion with Eric and comparing with his previous records, it seems that, aside from the unoptimized PI corner, we have increased noise at high frequencies. He was able to get 8.8mrad of phase noise between 100Hz and 100kHz when the loop was fully optimized. It is currently not clear where this added noise comes from, the two hypothesis are either the aging of the laser, or the RP digitalization noise.

Quick win next step: improve the averaging on the SR785 to precisely determine how much noise there is at high frequencies compared to Eric's logs.

Non-image files attached to this report
Comments related to this report
jacques - 19:11, Tuesday 20 February 2024 (11527)

Here is a prettier plot with the cumulated phase noise between 100Hz and 100kHz.

Non-image files attached to this comment
Electronics General
dorot3a - posted 15:57, Wednesday 07 February 2024 (11502)
Externally clocking Red Pitaya with LTC6754 board
This is a short guide on setting up an externally clocked Red Pitaya using an LTC6754 comparator board to convert an analog clock signal to the required LVDS logic signal. 

Setting up the LTC6754:

We want a differential output on this board, which requires a few small alterations. Desolder R20 and R21 and move the resistors to the R22 and R23 pads. This removes the antennas going to J1 and J4 (out+, out- respectively) and sends the input to two of the pins on J6 (see image). I added a header to these two pins to connect a jumper cable; this is probably not the ideal connector to use permanently but it was convenient. This will be the differential clock output from the comparator board. The pin closer to the edge of the board is the negative output, although I don't think this is particularly important as long as you're consistent about the positive and negative outputs across each RP. 
We don't need to worry about J3/out-, as the board comes with this trace unconnected. Connect in+ to a BNC tee. Terminate one side with 50Ω and connect the other side to a signal generator running a 125MHz, .2Vpp signal. Finally, connect the VCC0 and VCC1 and their respective grounds each to a power supply at ~3V. These boards will short out so make sure to check the voltage levels on the power supply before connecting it and to hook up the board with the power supply off. We should now have a suitable clock source for the Red Pitayas. 

Setting up the Red Pitaya:

Follow instructions here to connect the RP to your computer: https://redpitaya.readthedocs.io/en/latest/quickStart/connect/connect.html.
I find it easiest to use a direct wired connection when first setting these up. Make sure the SD card is in the slot. The clock input pins are the second row on the RHS when looking at the RP with the RF connectors pointed up (see pinout here: https://redpitaya.readthedocs.io/en/latest/developerGuide/hardware/125-14_EXT/top.html). All we need to do is power the RP, power the LTC6754, and use two jumper cables to connect the differential output of the LTC6754 and the clock pins on the RP. 

Check your connection by typing the RP-xxxxxx.local address on the ethernet port of the RP into your browser. It should load the host page for the RP. Note that the RPs take some time to 'warm up' and connect. You can usually tell the RP is ready when the red indicator LED next to the ethernet port stabilizes to a heartbeat pattern. Once the host page loads, you have succeeded and can now use the RP to do your bidding.


I was able to load and connect to two RPs at once using the same LTC6754 board by using a splitter between the diff output and the two clock inputs. I haven't tried to clock more than 2 from the same LTC6754. 

Connecting to PyRPL:

From the host webpage, navigate to system > network manager and copy the IP address of the RP. Use this as the hostname to connect through PyRPL. The IP address will change based on how you have your network set up/how the RP is connected, so check via the host page every time you change anything about the connection. You can verify the IP by typing it into the browser; it should also load the host webpage. Do not attempt to use the 'default' IP on the RP setup website. It will not work, even though the documentation says it will.

As of today I've only been able to successfully connect to a single RP at a time through PyRPL, but I believe this is a networking issue because I'm not able to connect to either RP individually through PyRPL the way I have them networked, although I can connect to both via browser. 

Some notes: 

Sometimes the RP connections will be very slow, or not connect at all. If this is happening, some things that might help are: power cycling the RP, checking your ethernet cable/switching it out, and reflashing the SD card. If the red indicator light isn't normalizing to a consistent heartbeat pattern, it might be due to a bad connection to the clock. Check your cables.
Images attached to this report
GRAVITES Optics 1550nm
epolini - posted 15:11, Wednesday 07 February 2024 - last comment - 20:01, Wednesday 07 February 2024(11501)
Few kHz bump due to power attenuation of the fiber. Why?

In the past days, I measured a weird effect in the 1-km UMZI optical setup. I was measuring the RIN and I noticed the weird peak shown in the entries 11484 and 11488, extended between hundreds of Hz and 10kHz. The idea we had was that it was due to stimulated Brillouin scattering, caused by an excess in optical power entering the 1-km spool (https://doi.org/10.1364/OE.19.011792). The peak is evident after the spool and not before, this is why we thought about this and the input power was 7mW, same order of magnitude as in the reported paper. 
I installed a 95/5 PM BS in front of the spool to check is this was the cause of the noise. When I introduced the BS the input power was 6.5 mW and the two output ports 5.66mW and 0.307mW. I sent the 0.307mW to the 1-km spool. The peak is way bigger after the BS, so with reduced power as it can be seen in Fig. 1. 

In Fig. 2 I included also the dark noises normalized by the power measured before shutting down the laser for the measurements. The red curve is too hight and I did not understand why yet.

I ordered a high power beam dump and we already had an high power BS, meaning that I can verify if this behaviour is related to operating the laser at its minimum power and I can also verify if this might be due to the connectors between fibers.

 

Images attached to this report
Comments related to this report
epolini - 20:01, Wednesday 07 February 2024 (11504)

Today I operated the laser at 2W, adding a 90/5/5 beam splitter after the FI and dumping with the new beam dump the output power from the 90% port. The peak at kHz is still present, so this effect is not due to operating the laser at lower power.

I checked that the spectrum also right after thre FI and the peak is always there and it increases after the beam splitters. 

cSQZ General
hudsonl - posted 19:57, Tuesday 06 February 2024 (11500)
SHG to OPO Alignment via PDH Extension Board
I have initial alignment from the SHG to the OPO using the PDH extension board. This alignment was much less painful than the initial OPO alignment, mostly because the OPO itself was already aligned, and the CLF was aligned to the OPO. My alignment procedure was as follows:

- Ensure that the CLF is well aligned to the OPO
- Remove the OPO from the compact squeezer board
- Use pinholes to align the SHG beam to the CLF beam for coarse alignment
- Using the aligned SHG beam, place a 50/50 beam splitter with a lens and a PD on one path and a lens and a CCD on the other path
- Place the OPO back on the compact squeezer board
- Align the retro-reflected 532nm beam from the OPO to itself using a beam card in the near and far fields

At this point, I was able to see cavity flashes on the CCD while scanning the OPO. This allowed me to adjust the alignment of the beam to the OPO. The attached image shows the end point of this alignment procedure.

From looking at the mode scan on the CCD, the remaining discrepancy between the SHG beam and OPO appears to be due to mode-mismatch rather than from misalignment. Tomorrow, I will adjust the positions of the lenses between the SHG and OPO to improve the mode matching.
Images attached to this report
cSQZ General
jacques - posted 16:54, Tuesday 06 February 2024 (11499)
PLL swap to custom driver

The PLL has been swapped to the new driver and the loop could be closed. Everything seems to be working as intended so far. Thanks to the increased range of the PZT driver, the loop can be closed during >1h without manual intervention.

I have not computed the phase noise yet, nor have I optimized the loop.

As reference, the loop can be closed with P=0.2, I=20.

Below a screenshot of the error signal on channel 1 (scale 0.5V/div whereas the error saturates at +-2V. So the error signal is not railed). 

A more detailed analysis will follow.

Images attached to this report
cSQZ General
jacques - posted 18:43, Friday 02 February 2024 - last comment - 18:46, Friday 02 February 2024(11496)
Characteristics of the custom PZT driver for PLL

I compile here the way the custum PLL PZT driver (modified by Eric) works:

- HV IN: power supply (0-100V) connects here

- CTRL IN: RP correction signal connects here

- HV OUT: high-voltage signal which is sent to the laser's PZT. The HV OP amp has a gain of 100 at frequencies below 50Hz.

- MON OUT: a 1/100 signal which mimics the HV OUT port, for monitoring.

There is an extra potentiometer which should be set at 0V for our purposes.

The BNS port between MON OUT and CTRL IN is not connected to anything.

Comments related to this report
jacques - 18:46, Friday 02 February 2024 (11497)

After powering the driver with both a +/-18V and 0/100V power supplies, I sent a slow ramp from teh RP to the driver and looked at the monitor output. I could see that the MON OUT went from 0 to 1V during the scan. I had previously made the mistake of using a 0/18V power supply, instead of a +/-18V. 

Conclusion: we can test finally that the HV OUT is saturated at 100V, then move to the actual actuation of the DBL PZT. 

cSQZ General
jacques - posted 17:40, Thursday 01 February 2024 - last comment - 18:31, Thursday 01 February 2024(11492)
PLL recovery (1): power budget

I have realigned the PLL's optical setup, whose power budget is as follows:

700uW at the start of the AOM path (ie incident on the AOM)

AOM driven using the Moglabs at 34dBm at 51.6MHz

600uW out of AOM, incuding 200mW of frequency-shifted light.

52.4uW from AOM path, measured at PLL PD

170uW from DBL path measured at PLL PD.

I then swapped the AOM driver, from the moglabs, to the VCO. the power on the PLL PD has increased: 

75uW from AOM path, measured at PLL PD.

 

 

Comments related to this report
jacques - 18:31, Thursday 01 February 2024 (11495)

I tried to close the PLL with the RP -> Thorlabs Driver -> DBL EXT but I again have the problem of the error signal not going to zero. 

I will not investigate this further because this is not the final setup; rather I will switch the Thorlabs driver to the custom one by Eric and study the loop from there.

GRAVITES Optics 1550nm
Eleonora, Medha - posted 14:52, Tuesday 30 January 2024 (11489)
EOM

[Eleonora, Medha]

We added an EOM to the original Michelson interferomenter to check if we could see an error signal. A schematic can be seen in Figure 1. After the mixer, we added a 1.9MHz low pass filter.

Measurements:

The power before and after the EOM was 1.13 mW to 0.433 mW which is a loss of 0.7 mW. This is around a 61% loss.

The visibility is 97%.

We drove the EOM with a synthesizer DS345, with optimal modulation amplitude of 20dBm and frequency of 6MHz. After the EOM we installed a power splitter (ZC-2-1W) to provide both the signal for the EOM and the LO for the final loop configuration. The visibility became 49% after the mixer. The DC power coming and the RF power are sent to the oscilloscope which can be seen in Figure 2, yellow and green curves respectively. Ideally, the DC power should be shifted up from the y=0 axis but should still have some oscillations. However, as seen in the figure, it basically averages to 0. 

Conclusion: 

The signal is not a DC signal but it is clearly coupled to noise, so we need first to isolate our setup from the environmental noise.

Images attached to this report
Facilities General
wenxuan - posted 15:22, Monday 29 January 2024 (11487)
MIT remote control room - hardware
Myron, Fred, Peter, Wen

We decided to set up a remote control room at MIT that allows us to commission both LIGO detectors remotely. The manuals of hardware we purchased are attached here for records purposes.
Non-image files attached to this report
GRAVITES Optics 1550nm
epolini - posted 19:47, Friday 26 January 2024 - last comment - 15:32, Monday 29 January 2024(11484)
Polarization controller characterization

Today I changed the optical setup of the long arm of the UMZI to introduce some pick-offs for the monitor of the polarization before and after the polarization controller. 

After the 1km spool there is a SM AOM and then I added a 95/5 BS and I connected the 5% port to a SM PBS and the 95% port to the polarization controller. After the controller I added a 95/5 PM BS and I connected to the 5% port to a PM PBS and the 95% port to the other PM BSs before the 50/50 BS and the PD. 

I connected two PDA20CS thorlabs PDs to the outputs of the PBSs. The gain on the PDs is 10dB. I measured the powers with the oscilloscope and I got:

The signals did not change after the polarization controller by disabling it (the change in polarization might be slower then what I observed it for).

The result of this measurement is that the controller seems to work. 

I also acquired the RIN before and after the polarization controller. I measured the power spectrum with the SR785 at the 5% port of the new BSs I installed, meaning before entering the PBSs. 

The power level on the SM side (BEFORE) is 336 mV and on the PM side (AFTER) is 176 mV. I used these DC powers to normalize the spectrum shown in Fig. 1. The low frequency of the spectrum is higher after the polarization controller.

Images attached to this report
Comments related to this report
epolini - 15:32, Monday 29 January 2024 (11488)

Today I measured the RIN also before the spool and with the polarization controller OFF, to understand what is introducing the excess of noise at different frequencies.

Fig. 1 shows: 

  • RIN before the 1-km spool in blue
  • RIN after the spool but before the polarization controller (PC) in yellow
  • RIN after the PC when PC is enabled in green
  • RIN after the PC when PC is disabled in red

The structrure between 2kHz and 4kHz is amplified by the spool, since it is way smaller before and it is present independently from the PC. The noise at low frequency (< 1kHz) is excited by the PC. To address this problem we can consider to feedback a signal to control the intensity noise after the PC. 

To understand if it is possible to easily reduce the 2-4kHz noise, I covered the spool with the box we built for the Michelson setup. 

Fig. 2 shows:

  • RIN before the 1-km spool in blue
  • RIN after the PC when PC is enabled without box in yellow
  • RIN after the PC when PC is enabled with box in green
  • RIN after the PC when PC is disabled without box in red
  • RIN after the PC when PC is disabled with box in purple

The noise is not reduced by covering the spool with the box, so we might need to think of something better to isolate the spool.

 

Images attached to this comment
RetroSense General
evanhall - posted 21:22, Wednesday 24 January 2024 - last comment - 12:36, Sunday 04 February 2024(11483)
RetroSense noise measurements

Continuing to repeat Maya's measurements (11411) at about half the optical power (sum voltage about 6.3 V, or about 1.3 mA of photocurrent) and also now in the tank.

Measurement 1 was done in the tank with the lid on. I replaced the thick dsub cable with just a ribbon cable for less mass.

For measurement 2 I did a lot of things: resoldered the LED, cleaned flux off the PCB, reseated the whole thing, and added mechanical parts to make a stiff connection right at the dsub connector, to reduce vibrations from the cable. This last intervention helped reduce peaks above 5 Hz, but I suspect that the improvement below that may be a result of less seismic noise. The noise there is nonstationary. The dark noise is consistent with the noise of the AD630 demodulator.

In measurement 2 the sensitivity was 16 V/mm or 3.2 mA/mm, which implies a characteristic spot size of 0.4 mm. The modulation was at 10.2 kHz.

Images attached to this report
Non-image files attached to this report
Comments related to this report
evanhall - 08:20, Saturday 27 January 2024 (11485)

The attachment shows the dc photocurrent response versus displacement of the retroreflector along the X axis. The slope in the X photocurrent is 3.2 mA/mm; the slope in the Y photocurrent is 0.08 mA/mm (cross-coupling of about 3%).

The noise rises broadly when the X offset is further away from zero. The sensitivity in the range ±0.25 mm is equal to or better than the standard BOSEM.

Non-image files attached to this comment
evanhall - 15:32, Saturday 27 January 2024 (11486)

By making the coherence between the X displacement noise and the sum (intensity) noise, we can see that the excess noise at nonzero X offset is indeed predominantly an intensity noise coupling, as would be expected. The solid lines are the raw spectra; the dashed lines are the projections made by multiplying the spectra by sqrt(1 − γ2). Evidently, the amount of (linear) intensity noise coupling at zero offset is not dominating the noise floor.

The RIN at 1 Hz is about 10−6 / Hz1/2, which is about 100 times above the shot noise.

Non-image files attached to this comment
evanhall - 17:45, Tuesday 30 January 2024 (11490)

I tested the susceptibility to oscillator phase noise by frequency modulating the SRS DS345 function generator. The carrier is a 10274 Hz square wave. The modulation was a 10 Hz span with a 3.1 Hz modulation frequency. A line appeared in the X spectrum at 3.1 Hz with an amplitude of 0.4 nA/Hz1/2. I changed the carrier frequency to 3274 Hz and the line amplitude decreased by a factor of 3. The noise floor did not change.

I tried making a small cylinder of tape around the retroreflector to reduce the amount of air currents. The noise floor did not change.

evanhall - 16:55, Wednesday 31 January 2024 (11491)

The voltage regulator for the LED supplies 9.4 V. The resistance is currently dialed to 420 Ω, so the LED is being driven with 22 mA.

The voltage regulator noise is about 500 nV/Hz1/2, meaning 1.2 nA/Hz1/2 of current noise being delivered to the LED, or a relative noise of 5×10−8 / Hz1/2. This is comparable to the relative noise of the photocurrent that was seen when the LED output was monitored with a PDA100A.

Non-image files attached to this comment
evanhall - 12:36, Sunday 04 February 2024 (11498)

A selection of time-domain measurements. I did not include the time series of the final output, which is a 2-pole low-pass filter at 300 Hz.

Non-image files attached to this comment
cSQZ General
hudsonl - posted 10:57, Tuesday 24 October 2023 - last comment - 18:24, Thursday 01 February 2024(11382)
SHG to OPO Mode Matching via PDH Board
The attached image shows the optimized mode matching calculation from the SHG, through an EOM and Faraday, and to the OPO. The calculation includes the measured ray-transfer matrices of the EOM and Faraday (see previous posts by Masaya). The solution has a focus inside the EOM, the lens "Lens 1" is a f = 38.1mm lens to refocus the beam incident on the Faraday, the lenses "Lens 2" and "Lens 3" are then f = -100 mm and f = 88.3 mm lenses to mode match the beam into the OPO. Lenses 2 and 3 will be placed on precision z-axis translation mounts to allow for fine-tuning of the mode matching into the OPO.

To align the beam from the SHG to the OPO, we'll want to use mirrors HR2 and HR4, which are separated by 70 degrees of Gouy phase. The maximal beam size is about 1mm radius on Lens 3, which is getting large for a 6.3mm diameter optic, but should be small enough not to clip. If we have problems with clipping on this optic, we could consider using a 1 inch optic for Lens 3 instead of a 1/2 inch optic.
Images attached to this report
Comments related to this report
hudsonl - 18:24, Thursday 01 February 2024 (11494)
Measured beam parameters from the PDH extension board roughly match the theoretical mode here. I measured a waist location about 150mm from the final PDH mirror and a waist spot size of about 70um. Some of my measurements saturated the camera slightly, so the waist size estimate is likely an over-estimate. I believe the beam shape is good enough that I should see some signal when I try to couple light into the cavity, so for now I'm going to proceed and see if I can see any cavity flashes.
Images attached to this comment
Displaying reports 721-740 of 1025.Go to page Start 33 34 35 36 37 38 39 40 41 End