I put an input filter lowpass at 100kHz on the PID block and took a spectrum with the Agilent, going up to 1MHz in the spectrum. I just took some random values for the PI without trying to optimize:
P = 0.04
I = 200
See attached. Since it's not optimized I have 35mrad of phase noise from 1kHz to 1MHz, which is actually the same value as what I had on the SR785 but from 1kHz to 100kHz. See here https://emvogil-3.mit.edu/lasti/index.php?callRep=11503.
When I tried to take a transfer function, I noticed that the custom PZT driver no longer worked. It is possible that I broke it. We will check this tomorrow with Eric.
I took the PZT drive+18V+100V out on the other table and Eric tested the PZT driver. Turning the pot now seems to correctly vary the output from -20 to 100V. Everything seems to work ok (although we didn't test sending a signal at CTRL IN). It is not clear why it didn't work yesterday...
We have been investigating the laser frequency noise that presented the weird low frequency bump noise.
The frequency noise has been measured as correction signal after locking the Michelson with the AOM. The calibration factor is 6.5kHz / V given by the Marconi.
In Fig. 1 how the frequency noise changes as the time after turning on the laser passes: the bump apprears after 1-2 mins after I turn on the laser.
NB. I need to press the button outside the lab, i.e. disabling the interlock, to remove the bump once I turn the laser on again. Just stopping the emission is not enough.
In Fig. 2 the measured frequency noises, after 10 seconds in violet and several minutes in black, the phase noise converted in frequency noise given by the datadsheet both for the X15 and E15, in blue the frequency nosie measured at AEI (https://www.mdpi.com/2410-390X/6/1/15). The phase noise is expressed in dB(rad/rtHz)/m on the datasheet that is a bit surprising. Eric suggested to convert it using the formula used on the optica paper of chris et at (https://opg.optica.org/optica/fulltext.cfm?uri=optica-9-11-1238&id=518455):
S_dnu(omega) = S_phi(omega) / [(2 * pi * tau)^2 * sinc^2(omega*tau/2)]
where tau = L*n / c with L = 1m, c speed of light and n = 1.45.
We sent the plots to NKT to understand the calibration they use in the datasheet and where this extra noise at low frequency comes from.
I also noticed that the amplitude of the fringes of the error signal used to calibrate the error signal from V/rtHz to rad/rtHz changes over time after turnin ON the laser: it passes from 100mV to 500mV in a couple of minutes.
I built the second HD with a similar footprint than the first one (12x18in). See pictures.
I took the superpolished 50/50 BS from ATFilms which seems OK on eye inspection (the arrow points to the reflective side).
However, for the superpolished 1064 mirrors, I took them from the red box in the picture, but I could clearly see that the coatings have different color, which could mean that these mirrors are HR but for different wavelengths.
Next step: shine the IR beam on each of these mirrors and measure the reflectivity/ghost beam to make sure that they are HR for 1064 at 45°.
If it isn't a 532 nm optic, its likely just an AR coating on the back. Sometimes people order their 45 AOI superpolished mirrors with AR coatings on the back. Sometimes people just use a wedge only.
The PDH extension board optics have been aligned to the OPO. On the RF photodetector looking at the cavity's reflection, I get what looks like a good PDH signal. However, the beam from the SHG is not mode matched too well to the OPO, so I have a lot of power in higher-order modes. My locking script has an implicit assumption that a beam is relatively well mode matched to the cavity it's incident on, so my automated locking procedure doesn't quite work yet. So far, the setup has demonstrated that I can align the beam from the SHG, through the PDH board, and to the OPO and that the reflected beam hits the RFPD and allows me to extract a signal modulated at the EOM drive frequency. My plan is to take a careful beam profile of the beam incident on the OPO with no mode-matching lenses, calculate the optimal mode matching I expect to achieve with the current lenses, and then calculate an optimal pair of lenses for mode matching. Attached files show the OPO's IR transmission and PDH error signal as a function of cavity length. The zoomed erros signal plot shows that I do get what appears to be a high SNR error signal, but it doesn't have a clean PDH shape. I believe much of this problem is coming from the relatively poor mode matching as light from higher-order modes in interfereing with the TEM00's PDH signal.
CAUTION: MAX RF DRIVE POWER ON AERODIODE AOM IS 1.8W (=32.55dBm). DO NOT GO ABOVE AS IT CAN PERMANENTLY DAMAGE THE AOM.
I tested the Aerodiode AOM 1016 PM 100MHz for tuning freq different from 100MHz.
Test 1: RF driver at 100MHz
Input power to the AOM: 4.6mW
Output power: 3.04mW
RF driver: 100MHz, 1.8W (moglabs)
So losses at 100MHz shift with nominal power yields 1.8 dB of loss, which is within the specs.
Test 2: RF driver at 95MHz, 1.8W
I have 2.6mW at the output. So losses of 2.47dB
The attached image shows the specifications from Aerodiode. Jacques measurements seem to agree well with these curves.
Test 3: RF 92.5MHz 1.8W
I measure 1.7mW output for 4.3mW of input. Therefore losses of 4.03dB.
Test 4: RF 90MHz 1.8W
I measure 0.93mW output for 4.3mW of input. Therefore losses of 6.65dB.
Test 5: RF 105MHz 1.8W
I measure 1.6mW output for 4.3mW of input. Therefore losses of 4.30dB.
Test 6: RF 107.5MHz 1.8W
I measure 0.83mW output for 4.3mW of input. Therefore losses of 7.14dB.
Test 7: RF 110MHz 1.8W
I measure 0.35mW output for 4.3mW of input. Therefore losses of 10.9dB.
Attached is the plot of dB losses as a function of the RF frequency. The curve is clearly asymmetric, with less losses for negative detunings from 100MHz (i.e. in the 90MHz region).
Thus this is different from the advertised curves.
I moved the high-power RF amplifiers driving the EOMs used for the SHG and OPO PDH locks from the optics table to the top of the CSQZ electronics box. In their new positions, the coaxial cables between the high-powered amplifiers and the EOMs needed to be extended, so I used new cables. The first two of these significantly distorted my PDH error signal (see plots below). This problem seems to be fixed now that I have a BNC cable from the RF amplifier to the EOM instead of a coax cable with a SMA connector on one end and a BNC connector on the other end. Attached files show the error functions with two different bad coax cavles and with a BNC cable.
Today I closed the loop for the laser frequency noise stabilization analogically, then we will close it using FPGA. I used the SR560 as servo, the Marconi as VCO, SG3282 as signal generator and a minicircuit mixer. The scheme implemented is shown in Fig. 1. As a filter I am using an integrator with a gain: LP corner frequency = 1kHz and gain = 10.
To measure the OLTF I included a second SR560 set on DC and gain = 1. I sent a swept since between 100Hz-100kHz of amplitude 500mV and the OLTF is shown in Fig.2. We have UGF ~ 10kHz and gain 20dB at low frequency.
In Fig. 3, the error and correction signals closed loop are shown calibrated in rad/rtHz, in blue and orange respectively. The calibration of the error signal is done by looking at the amplitude of the fringes of the interferometer on the oscilloscope which is 50mV. The calibration of the correction signal is done considering the FM deviation which is 10kHz and then rescale with past measurement and it gives 6.5kHz/V and devide the obtained spectrum by the frequency. The gain at low frequency is not enough to correct for all the noise, but we will do a noise budget to get a worst case estimate of the frequency noise.
Moreover, I turned off and on again the laser and I checked again the error and correction spectra afterwards. Right after turning the laser on, the spectra in V/rtHz looks as shown in Fig. 4. The peak at 8kHz is the one driving fringes for the first 3 minutes after the laser is emitting and the noise bump at low frequencies disappears. It comes back after few minutes.
Below are noise spectra taken with the Agilent 4395A on the HD that was made by an undergrad and that we took from purple lab. This spectrum analyzer goes higher in frequency than the SR785 but its noise floor is quite high. I did not amplify the signal from the HD.
Notice that the y scale is the power spectrum in dBm/Hz (not the amplitude spectrum in Vrms/sqrt(Hz)) but this does not change the following conclusions of the analysis.
We see clearly that we are no longer shot noise limited from 100kHz upwards. Rather we are limited by the electronics. After discussion with Eric, we should devise and build a new board which has lower effective capacitance.
I aligned the BS of the HD to optimize the 50/50 ratio and got the following spectra for the shot noise. The lower in frequency the shot noise curve starts to plateau, the better. Here we see that we should be shot noise limited starting from 500 Hz.
To obtain the splitting ratio, I followed the procedure described in https://emvogil-3.mit.edu/lasti/index.php?callRep=11513 and perfected the splitting ratio by live comparing the spctra of the dark noise vs shot noise at low frequencies and tried to minimize the difference on the SR785.
Below are the plots of the spectra.
To officially claim shot noise scaling at 500Hz, we should see how the noise level scales as a function of light power. Currently we have ~3.16mW of incident power on the BS.
I rebuilt the HD in a more compact way on a 12in x 18in x 2cm board. I roughly aligned all the waveplates, mirrors and the BS to balance the power incident on the HD. I took the following power budget:
Power budget:
4mW at the output of the LO fiber
3.16mW after the PBS+l/4+PZT setup, going to the 50/50 BS
1.60mW transmitted from the BS
1.57mW reflected from the BS.
When I try to power up the HD and read its PD outputs on an oscilloscope, I only see a non-zero voltage on the port "PD1" (channel 2 of the picture). I don't see anything on the port "PD2" (channel 1, notice the voltage scale) even though Ì can clearly see the IR spot aligned with the PD using a beam card. I don't understand what's happening. Is PD2 = PDA broken?
The solution was to push PD2 well into its socket and then the electrical connection was ok. The goal is then to tune the angle of the BS to simultaneously:
1/ maximimze the DC power on each PD
2/ zero the DC OUT signal (differential)
Next step: The verification of how well the BS is balanced can be seen on a spectrum of DC OUT (see figure attached by Hudson), the sooner the spectrum is shot noise limited the better (here 1000 Hz).
I checked the behaviour of disappearing fringes from the Michelson interferometer by watching at the output port right after turning on the laser. I saw fringes at 7.8MHz for 3 minutes and then the usual noisy shape. Something is probably going on inside the laser box. Eric suggested that it might be some feature of the internal control loop to mitigate phase noise.
Then I tried to lower the modulation rate sent to the AOM both with the Marconi and the SG382:
The frequency of the fringes is around 4-5kHz.
I put back the EOM after the AOM. I set the DS345 to 6MHz and 10dBm because I need it both as LO and to drive the EOM. LO needs to be 7dBm since the ZDA-1-1+ is a level 7 mixer and then the RF signal is splitted by a power splitter which removes 3dBm. I added a 13dB (10+3) attenuator to the signal that goes to the EOM to get it down to (7-13) dBm = -5dBm, that was the value maximizing the fringes. I then connected LO and RF output of the PD to the mixer and I measured the error signal on the oscilloscope. It is oscillating since I am driving the AOM. The DC poutput of the PD is shown in blue and the error signal after the mixer in yellow in Fig. 1. This the minimum of the blue curve crosses the error signal at its zero point, as we want. This means that we can go on with locking this loop. i will lock the loop with the SR560 (integrator + gain) and the marconi, while waiting to do it with RP.
We noticed that the amplifier boards we had been using for the compact squeezer had very bad phase performance, which was likely limiting the bandwidth of our loops. I switched out these 10x amplifcation boards on the compact squeezer for a custom board designed by Masaya. I also moved the low noise amplifiers used to amplify the signal out of the RFPD from the optics table to the electronics box. After these changes, the SHG still locks consistently. The attached image shows its measured transfer function. It has a UGF of about 900Hz and about 40 degrees of phase margin.
I was able to connect wirelessly to two RPs at once using PyRPL on MIT-SECURE in NW22. It doesn't really involve anything more than connecting to a single RP on a network, just make sure to create separate objects for each RP in your code. See attached code. Some more observations on working with externally clocked RPs, using a LTC6754 board to produce the correct clock signal: - If the RP is not connecting, or connecting very slowly, the 2 primary suspects tend to be the OS and the clock signal. I recommend rewriting the SD card and exchanging the LTC6754 as early troubleshooting methods for most issues. The OS files are available on the RP website; the OS version currently working for me is 'STEMlab_125-xx_OS_1.04-18_stable'. Another thing that fixed some connection issues for me was creating a new config file; this can be done either by deleting the existing file or changing the name of the file in your code, it doesn't really matter. - I'm not sure why, but sometimes the LTC6754 just doesn't work right; without a high enough bandwidth scope it's hard to check the clock signal to figure out more specifically what's going on.
I moved the optics for the HD to the second optics bench and placed them in a more compact way. See picture attached. As it currently is, the size of the setup is 17in deep by 14in wide.
With more tuning of the PI parameters and swap to Low voltage input on the RP, the loop can be closed and we have 11mrad of phase noise between 1kHz and 100 kHz.
P = 0.09, I = 45
However I see on the oscilloscope that the correction signal is clearly oscillating. We should take a spectrum of it at higher frequencies to see if it's due to the PZT resonance at 270kHz.