cSQZ General
hudsonl - posted 11:42, Friday 08 March 2024 (11546)
OPO Thermal Control Electrical Insulation
The compact squeezer's ovens have problems if the 'ground' of the coax cables connecting the thermistor and thermo-electric controller (Peltier) to the controls box come into contact with electrical ground. In these circuits, a voltage is applied across the thermistor, and a current across the thermo-electric controller and in both cases the 'ground' pin is not at electrical ground. For the SHG, there was no grounding issue as the connectors between the jumper cable attached to the thermistor and thermo-electric controller and BNC cables are dangling from the table. For the OPO, this problem was fixed by wrapping the connection in electrical tape, as shown in the attached image.
Images attached to this report
cSQZ General
hudsonl - posted 15:25, Thursday 07 March 2024 (11545)
OPO PDH Locking
The OPO consistently acquires a PDH lock with the PDH extension board and an automated locking script that communicates with red pitayas via pyrpl. The scheme takes a few seconds to acquire lock and consistently identifies and locks to dual 532 and IR resonance.

The loop itself has a bandwidth of 1400 Hz and a phase margin of 85 degrees.
Images attached to this report
cSQZ General
jacques - posted 19:57, Wednesday 06 March 2024 - last comment - 16:21, Friday 08 March 2024(11544)
second HD shot noise limited

I put the BS of the second HD on a newport mount. It is mounted such that the HR side is the first side the SQZ path encounters. In other words, the LO encounters the AR side first.

 

The optimization of the BS ratio goes in 3 steps:

Step 1:

I first removed the lenses and tuned the ratio of the BS while looking at the DC channels. The objective is to minimize the DC Diff. It will be minimized at a non-zero value (typ 100mV) because the optical paths to each PD are not the same: one beam will have a larger waist than the other so we expect different photocurrents.

Step 2:

I moved the R=25mm SP lenses back by about 1 inch such that the PD are both about 2 inches from the lenses. I expect that the new waist is at the position of the PDs.

Step 3: final tuning of the BS to zero the Diff (typically target +-5mV).

 

Then I took some spectra of the shot noise. I forgot to take the electronic noise.

The input power on the BS is 3.37mW.

Non-image files attached to this report
Comments related to this report
jacques - 16:21, Friday 08 March 2024 (11547)

The shot noise level on these plots is twice as high as the one on the first homodyne (logbook entry https://emvogil-3.mit.edu/lasti/index.php?callRep=11514), which is strange since the LO power has been the same (~3mW on the BS) and the HD PCB boards should be the same. Also the SR floor levels are different. Thus we need to cross check the shot noise on both HDs and see if the difference can be simply explained by the SR having different calibration settings at the times of the measurements.

cSQZ General
jacques - posted 19:43, Tuesday 05 March 2024 (11543)
Added a pair of lenses for the HD

I added a pair of R=25mm plano-convex lenses borrowed from purple lab, seemingly superpolished for 1064nm

Assuming n=1.5, I get 1/f = (n-1)/R = 0.5/25mm so f = 50mm. Currently they are placed 1 inch from the HD so I need to move them more.

Images attached to this report
cSQZ General
hudsonl - posted 18:21, Tuesday 05 March 2024 (11542)
SHG to OPO Mode Matching
I tuned up the SHG to OPO mode matching and alignment. The cavity scan is shown below with the primary peak the TEM00 and the secondary peak a higher-order mode mismatch mode (i.e. not a TEM01 or TEM10 mode). In the scope trace, the yellow curve (channel 1) is the voltage applied to the cavity mirror's PZT, or the cavity length, the green curve (channel 4) is the power on a reflection photodiode for the OPO, and the purple curve (channel 3) is the power on a transmission photodiode for the OPO. As can be seen, the mode matching from the SHG to OPO is reasonably good.
Images attached to this report
Heisenberg IFO General
hudsonl - posted 19:36, Monday 04 March 2024 (11541)
Relative Squeeze Angle Control with CLFs
We want to control the relative phase-angle between the two squeezed beams on the first beamsplitter using a 1% pickoff after this beamsplitter. The expected power levels in the CLF beams are as follows: 

Compact Squeezer OPO:
Rin-out = 0.89
Rclf = 0.9995
Ptrans/Pinc = 0.017

So if we send in a 1 mW CLF beam, we will have 170 nW of CLF power after the 1% pickoff.

Breadboard Squeezer OPO:
Rin-out = 0.845
Rclf = 0.9996
Ptrans/Pinc = 0.01

So if we send in a 1 mW CLF beam, we will have 100 nW of CLF power after the 1% pickoff.

If we use a Newport 1811 RFPD, which has about 8 pW/sqrt(Hz), with a 100 kHz demodulation bandwidth, we expect to have a noise-equivalent-power of about 2.5 nW. As a result, we should be able to achieve a reasonable SNR by directly detecting the CLF beams with a low-noise RFPD. If we instead use a beamsplitter with a several mW LO and a pair of photodetectors, we should be able to increase the SNR further at the cost of added complexity in the setup.
cSQZ General
jacques - posted 18:46, Friday 01 March 2024 (11539)
Second HD's optics roughly aligned

I aligned the optics of the second HD. The beam is now striking each PD of the HD.

In doing so I swapped the PZT with a mirror because the PZT was too inclined.

To continue to the BS splitting ratio optimization, I need to focus the beams after the HD into each PD. It is expected that the diameter of each PD is 2r = 1000um (after discussion with Wen and looking at the logbook entries in LLO and LHO). What they did in LIGO was to focus down to a diameter of 2w=280um, so that it is more than 3 diameters from being clipped. They got 95% visibility this way: https://alog.ligo-la.caltech.edu/aLOG/index.php?callRep=60750

I took some beam profiles at the positions shown in the pictures attached. Recall that the fiber collimator is the F110APC-1064 with waist diameter 2w_0 = 1.38mm.

Images attached to this report
Electronics General
epolini - posted 20:04, Thursday 29 February 2024 - last comment - 19:30, Friday 01 March 2024(11538)
Marconi 2026 sent to repair

We sent back to repair the Marconi 2026 which already had the internal clock broken. We were using it as LO in the gravites setup. 

Comments related to this report
epolini - 19:30, Friday 01 March 2024 (11540)

With Wen and Haoucun, we moved the Marconi 2030 that I was using in the 200-m setup in NW22 to green lab.

cSQZ General
jacques - posted 17:27, Tuesday 27 February 2024 - last comment - 18:09, Tuesday 27 February 2024(11536)
fiber collimator for HD

I measured the beam width at the input of the HD I have already built and found that it has a 13.5% diameter of ~1320um. This is compatible with the model F110APC-1064 form thorlabs.

Images attached to this report
Comments related to this report
jacques - 18:09, Tuesday 27 February 2024 (11537)

the thorlabs model is confirmed as I am starting to rebuild the HD on a thiner breadboard plate.

cSQZ General
jacques - posted 17:24, Tuesday 27 February 2024 (11535)
PLL peak that spoils the phase noise, coming from the PZT Driver?

Yesterday (02 26) I realized that there is a clear resonance feature between 1kHz and 20kHz in the error signal of the PLL. It must come from the loop because I can't make it disappear by tuning the PI coefficients (P = 0.09 and I = 45 for all the experiments done here, which are the same values as the ones I used in the past and did not give any resonance of this sort). I can only shift the frequency of this resonance and the peak height/width tradeoff.

Eric proposed to swap the 100pF capacitors at the input of the PZT driver into 10pF ones, which I did today (02 27). I then put the same PI parameters as yesterday and still saw this peak, although it hass slightly less amplitude now.

Thus I don't understand where this resonance comes from all of a sudden, while I used the same values of PI in the past and did not get such a feature.

Note that this is completely spoiling the phase noise, we go from ~20mrad for freq>freq_peak, to ~150mrad once we integrate past the peak.

Non-image files attached to this report
cSQZ General
hudsonl - posted 19:41, Monday 26 February 2024 (11534)
SHG to OPO Mode Matching
The initial mode match from the SHG to the OPO was done using measured ray transfer matrices of the EOM and Faraday. This solution gave respectable mode matching, but the eventual mode matching looked to be no better than 30% or so. I measured the beam after the EOM and Faraday and used this beam to re-compute the mode matching solution. A la mode agrees that the previous mode matching solution should not have exceeded 40% or so.

The new mode matching solution is to have a f = -32mm lens placed 133mm before the PDH board's M4 and a f = 75.6mm lens placed 42mm before the PDH board's M4. With the thin lens approximation and optimized lens locations, this should give 96% mode matching, and with thick lenses placedin the positions optimal for thin lenses, this should give 78% mode matching. As a result, I expect to achieve somewhere between 80 and 95% mode matching of the beam from the SHG to the OPO. The attached file shows this mode matching solution.
Images attached to this report
cSQZ General
hudsonl - posted 10:52, Monday 26 February 2024 (11533)
CSQZ Cavity Piezos
The attached image shows the piezos used in the compact squeezer to mount and actuate the M1, M2, M3, and M4 cavity mirrors on the SHG (M1 and M2) and the OPO (M3 and M4). This post is just for documentation in case we need to order similar piezos in the future.
Images attached to this report
cSQZ General
jacques - posted 00:22, Saturday 24 February 2024 (11532)
PZT driver broken again and then fixed by Eric

While trying to take a spectrum today (Friday), I noticed that there was overall extra noise. Then I had the problem again on the PZT Driver suddently not amplyfing neither the potentiometer nor the RP signal, like in entry https://emvogil-3.mit.edu/lasti/index.php?callRep=11526

I also noticed that the CTRL IN was contaminated by a lot of the VCO's 51MHz signal (+-200mV, to be compared to the RP's 0-1V range).

Eric diagnosed the PZT driver board and found that:

1/ Resistor R9 was apparently not soldered properly

2/ The ground power input had an inductance L2 which prevented high frequencies from being grounded

3/ some other resistors and capacitors needed to be added to better isolate the amplification stage from the VCO's 51MHz crosstalk.

4/ The potentionmeter had to be removed as it was causing more trouble (drift) than what it is worth.

After these changes, he tested the VCO on standalone with a signal generator. We saw that it works properly on MON OUT (x100/100 = same output as CTRL IN).

The next step is to recover the PLL with these fixes.

When reshuffling the layout of the PLL, we should separate as much as possible the VCO from the power supplies and the PZT amplifiers, to avoid crosstalk of any kind.

GRAVITES Stage
epolini - posted 14:56, Thursday 22 February 2024 (11531)
Green lab computer connected simultaneously to the internet and the Fuyu stage

Fred managed to solve the issue preventing us from connecting the computer simultaneously to the internet and the Fuyu stage by using an USB-ethernet adaptor and by resetting the Fuyu controller using a reset hole on the controller and removing the power for some time.

GRAVITES General
epolini - posted 14:54, Thursday 22 February 2024 (11530)
Green lab wireless network for GPIB remote connection

Fred set a wireless network in green lab (name: greenlab, password: ligola$er) in order to be able to communicate with the GPIB on the network. We tested and it works. Now it is possible to be connected to the GPIB with your own laptop if connected to the greenlab wi fi.

GRAVITES Optics 1550nm
epolini - posted 14:52, Thursday 22 February 2024 (11529)
Laser 'CONTROL' software re-installed and firmware updated

[Fred, Eleonora]

Today we re-install the last version of the CONTROL software (1.3.8.1660) by NKT and updated the firmware as the company support team suggested to try to address the issue of excess of laser frequency and amplitude noise. The result of the test is negative, we have the same behavior of Michelson fringes and laser frequency noise.

cSQZ General
jacques - posted 19:03, Tuesday 20 February 2024 - last comment - 18:35, Wednesday 21 February 2024(11526)
PLL spectrum with Agilent and input filter, broken PZT driver?

I put an input filter lowpass at 100kHz on the PID block and took a spectrum with the Agilent, going up to 1MHz in the spectrum. I just took some random values for the PI without trying to optimize:

P = 0.04

I = 200

See attached. Since it's not optimized I have 35mrad of phase noise from 1kHz to 1MHz, which is actually the same value as what I had on the SR785 but from 1kHz to 100kHz. See here https://emvogil-3.mit.edu/lasti/index.php?callRep=11503.

When I tried to take a transfer function, I noticed that the custom PZT driver no longer worked. It is possible that I broke it. We will check this tomorrow with Eric.

Non-image files attached to this report
Comments related to this report
jacques - 18:35, Wednesday 21 February 2024 (11528)

I took the PZT drive+18V+100V out on the other table and Eric tested the PZT driver. Turning the pot now seems to correctly vary the output from -20 to 100V. Everything seems to work ok (although we didn't test sending a signal at CTRL IN). It is not clear why it didn't work yesterday...

cSQZ General
jacques - posted 19:17, Wednesday 07 February 2024 - last comment - 19:11, Tuesday 20 February 2024(11503)
PLL phase noise spectrum

While trying to start the PLL today, I noticed that it would often lock at the wrong frequency (43.7MHz instead of the usual 51.6MHz). I realized that one of the SMA cables, connecting the output of the minicircuits amplifier (amplifying the PLL PD signal) to the PFD "DBL in" port, was broken. I swapped it out and the PLL no longer locked on the wrong frequency. I used P = 0.2 and I = 2000.

I then acquired some spectra of the error signal on the SR785 and analyzed them in python. Below are two plots of the error signal spectral density when the loop is closed, respectively between 1Hz and 800Hz and between 100Hz and 100kHz. I also computed the corresponding integrated phase noise. For this unoptimized loop, there is ~60 mrad of phase noise between 1Hz and 800Hz, while there is ~35 mrad of phase noise between 1kHz and 100kHz. I used the PFD's spec sensitivity of 36°/V = pi/5 rad/V to convert the voltage spectrum into rad.

After discussion with Eric and comparing with his previous records, it seems that, aside from the unoptimized PI corner, we have increased noise at high frequencies. He was able to get 8.8mrad of phase noise between 100Hz and 100kHz when the loop was fully optimized. It is currently not clear where this added noise comes from, the two hypothesis are either the aging of the laser, or the RP digitalization noise.

Quick win next step: improve the averaging on the SR785 to precisely determine how much noise there is at high frequencies compared to Eric's logs.

Non-image files attached to this report
Comments related to this report
jacques - 19:11, Tuesday 20 February 2024 (11527)

Here is a prettier plot with the cumulated phase noise between 100Hz and 100kHz.

Non-image files attached to this comment