[Eric, Eleonora]
Today we tested the splicer to splice a fiber marked as broken in NW22, a P3-460P-FC-1 SM fiber with 3mm jacket. With the tools available we removed the jacket and the plastic tube around the coating + cladding. The coating is 250um and the cladding 125um. We used the Micro-Strip tool to remove the coating from the fiber. The splicer manual is attached (we received it on a DVD...).
We managed to splice the single mode fiber following the instructions reported on the manual. Before splicing we used the Fujikura CT52 cleaver to cut the two ends of the fiber.
We temporary put the splicer on the desk in green lab as shown in Fig. 1. The successful splicing is shown in Fig. 2.
Fred, Wen
The LIGO remote control room is ready for public use after months of construction.
There are total three computers in the control room. MITCR1 and MITCR2 are the two computers on the desk powered by Intel NUC mini-PC. The third computer EVA is connected to the two TVs on the wall. The third sets of wireless & mouse in the pic below are for EVA. All of the three computers have access to ndscope, diaggui, and medm for both L1 and H1 detectors. The exact specs can be found in 11487.
Here are some of the useful commands to get started. Note that medm is read-only for security purposes.
Initialize credentials:
kinit -p <albert.einstein>
Open ndscope:
ndscopel <channel> # for L1
ndscopeh <channel> # for H1
Open diaggui:
diaggui # Then select 'NDS2' in 'Data Source Selection', and pick the nds2 server of the site.
Open medm:
medm_lho.sh -u <albert.einstein>
medm_llo.sh -u <albert.einstein>
I'll add/update more commands in this alog thread.
After adapting the setup to be able to install the homodyne detector, I tested the PDA10BI adapted as explained in the entry #11561. 56uW are arriving on both PDs. The new fiber port is on the input + photodiode. I checked on the oscilloscope the channels monitor + (yellow), monitor - (blue), RF output (purple) shown in Fig. 1. Monitor + is not seeing all the power, monitor - sees 60mV. The RF signal has an offset of -128mV. Fig. 2 shows also the feature of monitor + vs monitor - and it seems more noisy. The sensive area is really small so I doubt that the problem is that the light is not reaching it. The fact that the fiber port was missing suggests that maybe it was not working in the past and people removed it for diagnostic purposes.
Yesterday I restored the original setup of the UMZI with 1km setup needed for the test of the stage. I adapted it on order to test the thorlabs homodyne that we recently interfaced with fiber ports and eventually a low noise homodyne (fempto or aLIGO-like).
I moved the 90/10 BS from the end (right before the 1811) to the beginning (right after the FI). I added a VOA in the short arm in order to balance the power inside the arms. I used the Marconi we moved back from NW22 as LO. I re-connected the polarization controller and enabled it. Fig 1 shows the power budget of the current setup.
Then I closed the loop using the 1811 PD, before testing the homodyne. The DC power arriving on the 1811 is 112mV and the RF power is 200mV pk-pk (~-10dBm). I mixed this signal with 7dBm LO and the error signal after a 1.9MHz LPF was 72mVpk-pk. This translate in a calibration factor of 36mV / rad.
The error and correction signals are shown in Fig. 2. The OLTF is shown in Fig. 3 (NB. I added 20dB to the mag as required by the design of the board). I sent a sweep sine of 300mV amplitude on the SR785. The UGF is ~20kHz.
Connecting the stage and driver: Stepper motors have 4 wires (sometimes 6 or 8, but usually 4) coupled in pairs to 2 coils (A/B). There are a few standard color codes which can be found online. In this case the colors are labeled directly on the motor as A+, A-, B+, B-. Connect the 4 wires to the correspondingly labeled screw terminals on the driver. Connecting the driver and controller: the FUYU FMC4030 is a 3 axis controller, we only need one of the 3 axes. I tend to choose the 1st set of outputs. Connect PU1, DR1 on the controller to PUL-, DIR- on the driver. Connect 5V on the controller to ENA+ on the driver, then tie ENA+ to DIR+ and PUL+ on the driver. Powering the driver & controller: Connect the 'GND' and '+Vdc' terminals a 24V power supply.
We observed that the OPO's transmission of its CLF beam is about 0.7%, whereas its expected transmission is 1.7%. At the same time, the cavity's measured IR finesse of 54 agrees well with the expected finesse of 51 and indicates that there are no large excess losses. The OPO's mirrors should be coated for 89% reflectivity on the input/output mirror, and 99.95 +- 0.02% on the HR mirror (see attached quotation from LaserOptik). Some of the disagreement in measured power likely comes from mode mismatch, but since the cavity scan looks fairly clean (see attached image), I do not believe this can fully explain the factor of 3 difference between the expected and measured OPO transmission. I believe the measured transmission is likely lower than anticipated due to the HR mirror having a slightly larger reflectivity than 99.95%. For instance, if the mirror's reflectivity is really 99.98%, the expected finesse and transmission numbers agree exactly with the measured values. LaserOptik measured the reflectivity at 99.95% (see attached pdf), but this was done on a flat witness sample instead of the highly curved cavity mirror. Given these measurements, my baseline assumption going forwards is that the OPO's HR mirror has a reflectivity of about 99.98% for IR. Note: on the mode scan, channel 1 is the cavity length and channel 2 is the transmitted CLF power. In the LaserOptik_Final document, the OPO's HR mirror is "M4".
JD cut in half a connector I bought from thorlabs to fit the old thorlabs homodyne that was missing one fiber coupler in one of the two ports. Thorlabs is not producing these pieces anymore so that was the best solution. Today, with Elio, we shaped the perimeter of the piece in order to match the aperture with some sand paper. Fig 1, on the left, shows the homemade fiber coupler. This week I will compare the Mach Zehnder loop performance using the PD and the homodyne. The two fiber ports are not totally symmetric but one is longer than the other, as it can be seen in Fig. 2. This might cause a different amount of power arriving on the photodiode itself.
[Myron, Rich, Ben, Eleonora]
Last week we moved the two empty racks from the container outside the NW17 building into lasti and then green lab. Now we placed the two racks in green lab next to the gravites table, see Fig. 1, and they will host gravites' electronics.
Lisa, with Eric
Many months ago (old log) I had to remove the SHG from the LASTI squeezer to send it to LLO, since the LLO SHG failed.
To replace the LASTI SHG, I thought we could use the "old" AEI SHG which was sent to us over a decade ago and we have been using it in several experiments since then. I took that AEI SHG (which was in the green lab, thanks to Dorotea for cleaning up the table), and moved it to the LASTI squeezer, in the center of the table, NOT ALIGNED (picture attached).
This AEI SHG looks like the LIGO ones from the outside, but it has different mirrors, and therefore it requires a different mode-matching solution.
With Eric we found yesterday the mode-matching solution that it was used before for the AEI SHG.
I also attach here the pptx of the documentation of the LASTI squeezer.
Looking at some other old log posts, it seems that the efficiency of this SHG may be in decline. Compare the most recent measurement: Pin ~300 mW, Efficiency = 26% 2017 Measurement Against this measurement from when it first returned from LHO: Pin ~300 mW, Efficiency = 48% 2014 Measurement I found our stash of SHG crystals (see attached photo) and we can refresh if needed. Comparing the quantities remaining against the list in the squiki, I suspect that the current crystal is PPKTP: H1 Sqz Crystals Wiki Page
Gabor sliced the 40-km spool and I measured the losses (Fig. 1). I connected the spool to the first BS of the Michelson setup.
Power before the spool : 105.mW
Power after the spool: 11.4mW
It correspond to a 90% of power losses. In dB we have: loss_dB = 10*log10(P_in/P_out) = 9 dB. Considering losses of ~0.2dB/km, after 40km we have 0.2dB/km x 40km = 8dB which is in agreement with what I measured.
This measurement may be limited by SBS. In the future you may wish to lower the injected power level significantly.
I measured the losses again after placing a 99/1 BS before the spool.
Input Power = 2.2mW
Power after the spool (40km) = 0.378mW.
Losses = 83%
Losses [dB] = 10*log10(2.2/0.378) = 7.65dB
In order to have the computer connected to the internet and the fuyu stage simultaneously, we need to connect the red cable to the front usb adaptor (Fig. 1) and the blue cable from the stage controller on the back of the computer (Fig. 2).
DO NOT DISCONNECT THESE CABLES.
The attached image shows an updated power budget for the Heisenberg interferometer, which should allow us to achieve about 135mW of pump power into the SHG if necessary, compared with about 44 mW in the current setup. It will also allow us to pick off all of the local oscillator beams from the Lightwave electronics laser with several mW in each beam. The main improvements will come in two places (1) swapping a fiber isolator for a free-space Faraday isolator with a measured 7% loss and (2) using a high-power fiber beam splitter for the initial 90/10 power splitter, which will allow us to improve the mode matching into the fiber. The mode-matching will need to be re-done after adding the Faraday anyways, so this is a convenient time to redo the mode matching. Moving the PLL pickoff to after the 90/10 beam splitter will also reduce losses on the SHG pump path, but by less than 10%, so this is a more minor improvement. As noted in the figure, the power budget assumes measured losses where available, and the worst-case scenario specification from AFW otherwise. AFW quotes a max excess loss of 1dB per splitter, corresponding to 7% loss per splitter. I measured the Faraday to have 93% transmission, in line with the manufacturer's spec of less than 8% loss. The assumed transmission through the fiber port is 50%, which is the current value, although this may be improved by purchasing fiber couplers with a longer Rayleigh range in the future.
Since fixing the issue with the finesse model, we have been struggling to train the agent to learn a stable policy. The main problem stemmed from requiring at least 50k agent interactions (a few days) with the environment to assess the behavior before changing the architecture/hyper-parameters, etc.
I modified the baseline RL algorithm to improve the sample efficiency, which has led to improved tracking and disturbance rejection for the RL-Agent. The trick involved multiple reuses of previously collected experience samples multiple times (M=10) to calculate the gradient and update the actor and critic networks. This led to better exploitation of the off-policy continuous control agent. This scheme has also reduced the sensitivity to the exact choice of hyper-parameters.
Now, the agent executes the alignment task for several iterations but then forgets what to do after a while and goes rogue. This issue usually goes away with longer training. The way I am dealing with this right now is via curriculum learning, where I progressinvely increase the task complexity during training and improve the agent's robustness to unexpected situations. But I will talk to a few RL experts in the coming days to see if there are smarter ways to deal with this catastrophic forgetting.
I will now test the behavior on the real system and see the actual performance.
-----------------------------
PLOTS:
-----------------------------
The first plot shows the recovered rewards during training. Unlike supervised learning, the non-monotonic behavior in learning is a characteristic of RL that makes it harder to train and decide when to stop training. Remedy is to log and look at the involved loss functions.
Second plot shows how the agent keep the two beams aligned and maximize the power at the CCD. You can see how the agent performs its task for a while and then goes unstable.
Third plot shows the actuator signals. We penalize large variations in actions, hence the don't act like a bang-bang controller.
Fourth plot shows the cumutive reward collected by the agent.
Last plot shows the individual components used to construct the reward metric (= linear-quadratic-regualtor term + penalty terms boundary constrain violations + discrete encouragements for close to optimal behavior)
-----------------------------
################################################
The new pseudo-code is given below.
##############-------Pseudo Code-------###############
Initialize Q-networks (Q1, Q2)
Intialize target Q-networks (Q1_target, Q2_target)
Initialize Policy network
Intialize temperature parameter (alpha)
Initialize Optimizer for Q-networks, policy, alpha
For each training episode:
Initialize state s
While not done:
Choose action a from policy net, a = policy(s)
Take action in the environment:
Observe: next_state s'
reward r
isDone flag
Store transition (s,a,r,s',isDone) in the experience replay buffer
Sample a batch of transitions from the replay buffer
(s_batch, a_batch, r_batch, s_next_batch, isDone_batch)
for i in range(sample_multiple_reuse_ratio):
# Update Q-networks
Q_target = r_batch + gamma * min [ Q1_target(s_next_batch, policy(s_next_batch), Q2_target(s_next_batch, policy(s_next_batch)]
- alpha*log_probability(a_batch|s_batch)
# calculate Q-network loss functions
Loss_Q = MSE(Q1(s_batch,a_batch), Q_target) + MSE(Q2(s_batch,a_batch), Q_target)
update Q1 & Q2 using Loss_Q
# update Policy network and Temperature
Q_current = min [ Q1(s_batch,policy(s_batch)) , Q2(s_batch,policy(s_batch))
# Loss-Policy
-[ alpha * log_probability(a_batch|s_batch) - Q_current]
# Loss_alpha
-alpha * [ log_prbability(a_batch|s_batch) + target_entropy ]
Update Policy and Alpha using Loss_Policy and Loss_Alpha
# Update target Q-network (Soft update) using Polyak Averaging
Q1_target = tau * Q1 + (1-tau) * Q1_target
Q2_target = tau * Q2 + (1-tau) * Q2_target
# Update States for the next iteration
s = s'
################################################
I modified the compact squeezer's homodyne in several ways. First, I removed the 50mm focal length lenses that had been used in the setup as they were not super-polished and replaced them with a pair of super-polished lenses of the same focal length. (2) I mounted all of the optics on 1 inch diameter, 3 inch tall posts, with the exception of the lenses and the beam dump for the rejected power. This will hopefully improve the detector's mechanical stability. The attached file shows the current homodyne. After making these modifications, I re-aligned the homodyne. It is currently shot noise limited above about 1kHz, and the shot noise clears the detector's dark noise by over 20 dB. The full spectra are shown in the attached plot. I also ensured that there is a power plateau as the steering mirrors after the beam splitter are scanned vertically or horizontally, so the beams should be hitting the centers of the diodes and should not be clipped. i.e. the beam spot size on the diodes should be smaller than the diodes themselves.
The LO power was 2.0 mW while measuring the shot noise spectrum.
I switched the fibers going to the SHG pump and CLF ports on the compact squeezer board so that we send 90% of the power to the CLF port and 5% to the SHG pump. This is done so that there is roughly 25mW sent to HR mirror on the OPO and the OPO's 2% on-resonance transmission corresponds to about 0.5mW of transmitted power, which is visible on a beam card. The SHG still locks consistently at these low powers, but the OPO locking script needed to be modified slightly. With these modifications, the SHFG and OPO lock consistently, allowing me to lock the OPO on dual resonance (far from the phase-matching temperature so there is no nonlinear gian). I get a stable, bright beam out of the OPO, which allowed me to roughly align the OPO to the homodyne (previously aligned by Jacques). The next steps are to take a beam profile of the OPO beam and mode match and precisely align it to the homodyne's LO. With that alignment done and the CLF and LO loops tuned up, we should be able to measure our first squeezed spectrum.
To attempt to reduce losses in the setup, I aligned a fiber bench with an EOM, and a pair of waveplates, which allows us to remove a set of fiber polarization paddles from the setup and get rid of the associated losses from this extra fiber element. Currently, there is about 88 mW out of the fiber incident on the EOM, and 43 mW out of the fiber collimator on the compact squeezer breadboard, so the overall efficiency is about 50%. Without the EOM and waveplates, there is about 47mW out of teh collimator, so the total loss from the EOM and waveplates is about 5%. In principle, fiber benches should be able to achieve about 70-80% efficiency. However, I chose fiber ports with collimated beam waists of 190 um (or 380 um diameter) to ensure there would be no clipping on the EOM's 2 mm diameter aperture. Unfortunately, I neglected to account for the fact that a 190 um waist corresponds to a Rayleigh range of 107 mm, and the fiber bench is 135 mm long to fir the EOM and waveplates. I believe my coupling is fundamentally limited by the lenses in these fiber ports. Currently, David Zhang is working to align a 135 mm long fiber port with 380 um collimated spot size couplers, which have a Rayleigh range of 426 mm. If all goes well, this will allow him to achieve the desired 70-80% coupling efficiency and we will eventually replace the fiber port I aligned with the one he is aligning. After placing the newly aligned fiber port in the setup. I also moved several fiber components around to make more space for the eventual Mach-Zehnder interferometer and to make it easier to redirect the 90% pickoff of the seed laser to the OPO's CLF temporarily for alignment, and send the 5% pickoff that would usually go to the CLF port to the SHG pump port instead. This is useful for alignment as the OPO only has about 2% transmissivity to the CLF beam on resonance. If we send in a CLF of about 2 mW, we transmit about 20 uW, which is not enough to see on a beam card with the room lights on. However, if we instead send 40 mW into the CLF port, we transmit on the order of 1mW of IR power, which can be seen on a card. This considerably facilitates initial alignment of the OPO to the homodyne.
Yesterday I tried to investigate better the issue I encountered last week of not being able to stably lock the Michelson. In mid-February I was able to lock it while recently I was not able to keep the lock.
The loop seemed not to correct entirely for the noise. Fig. 1 shows the OLTF which is losing coherence at high frequencies and Fig. 2 shows the error and correction spectra. The loop configuration is not ideal and Ben suggested to introduce the LB1005 after using the SR560 as pre-amplifier.
We set the DC gain of the SR560 two steps below the overload condition, around 50/100. Then we set to zero the input channel disconnecting the LB from the system. After playing with the servo settings we managed to keep the loop locked for most of the time, but still with some spikes as shown in Fig. 3. It shows the error signal (input, yellow), correction signal (output, purple) and the error monitor (blue) on the oscilloscope. We can see that the system is locked for most of the time if the environment is quite (no talking, no tapping, no walking etc). The spikes identify that the loop loses lock and it is therefore not stable enough. We tuned the parameters of the LB in order to get the system stable for the longer possible time.
Some reasonable parameters:
I was not able to acquire an OLTF because the system was losing lock. I acquired some spectra even if they are not perfect due to the lose of lock, which is partially filtered out by avaraging (Fig. 4).
We cannot do better with the current isolation of the Michelson interferometer, so we need the enclosure we have been working on with Medha. It is possible to lock the interferometer so I guess that RP can be currently tested.
[Haocun, Eleonora]
We connected the polarization controller to a new power supply after making a cable for it. Pictures attached. We reconnected it to the computer and it works.
I put the BS of the second HD on a newport mount. It is mounted such that the HR side is the first side the SQZ path encounters. In other words, the LO encounters the AR side first.
The optimization of the BS ratio goes in 3 steps:
Step 1:
I first removed the lenses and tuned the ratio of the BS while looking at the DC channels. The objective is to minimize the DC Diff. It will be minimized at a non-zero value (typ 100mV) because the optical paths to each PD are not the same: one beam will have a larger waist than the other so we expect different photocurrents.
Step 2:
I moved the R=25mm SP lenses back by about 1 inch such that the PD are both about 2 inches from the lenses. I expect that the new waist is at the position of the PDs.
Step 3: final tuning of the BS to zero the Diff (typically target +-5mV).
Then I took some spectra of the shot noise. I forgot to take the electronic noise.
The input power on the BS is 3.37mW.
The shot noise level on these plots is twice as high as the one on the first homodyne (logbook entry https://emvogil-3.mit.edu/lasti/index.php?callRep=11514), which is strange since the LO power has been the same (~3mW on the BS) and the HD PCB boards should be the same. Also the SR floor levels are different. Thus we need to cross check the shot noise on both HDs and see if the difference can be simply explained by the SR having different calibration settings at the times of the measurements.