Displaying reports 621-640 of 1025.Go to page Start 28 29 30 31 32 33 34 35 36 End
Reports until 11:53, Wednesday 24 April 2024
GRAVITES General
epolini - posted 11:53, Wednesday 24 April 2024 (11608)
Govee Thermometer moved to green lab

I moved the thermometer that I put in NW22 to green lab today (04/24) at 11:47am. I am acquiring data with the app, to monitor the temperature on the bench and verify if there is any correlation with the laser power fluctuation.

Images attached to this report
General
nikhil - posted 21:33, Monday 22 April 2024 (11603)
ndscope leap-second bug fix

testbedAI CyMAC Ndscope was crashing due to an error associated with leap second update in the gpstime python package. IETF is no longer serving the leap-seconds.list, so I updated the corresponding URL (as given on the IETF page). This fixed the problem.

/usr/lib/python3/dist-packages/gpstime/leaps.py

Changed: Variable: LEAPFILE_IETF_URL

From
     https://www.ietf.org/timezones/data/leap-seconds.list
To:
    https://data.iana.org/time-zones/data/leap-seconds.list

cSQZ General
hudsonl - posted 18:09, Monday 22 April 2024 (11602)
CSQZ CLF Loop
We locked the CLF loop on the compact squeezer. The loop has about a 10 kHz UGF, 40 degrees of phase margin, and 6dB of gain margin. See attached figure. The moglabs is set for frequency actuation with a FM gain of 10 kHz (I believe this means 10 kHz/V).
Images attached to this report
cSQZ General
hudsonl - posted 15:49, Monday 22 April 2024 (11601)
CSQZ SHG to OPO Polarization Control
I added a quarter waveplate immediately after the half waveplate in the compact squeezer. This seemed to fix the mode matching problems from the SHG to OPO and pump the OPO with a pure vertical polarization. The CLF 2f signal now has a power of about -28dBm, substantially above the Red Pitaya's noise floor of -43dBm.
Images attached to this report
GRAVITES Optics 1550nm
epolini - posted 19:41, Friday 19 April 2024 (11600)
PM HP FI damaged

While measuring the spool loss, I realized that the power after the high power Faraday isolator was too low. For an NKT output power of ~200mW, the output power of the HP FI is ~60mW. This is not in agreement with the expected losses. I tried to clean the connector with the fiber cleaner and the methanol but checking with the fiber scope the damage was still there. We can proceed and splice the end of the PM HP FI with the PM BS following it. 

GRAVITES Optics 1550nm
epolini - posted 19:36, Friday 19 April 2024 (11599)
Second 40-km spool spliced in green lab

[Eleonora, Eric]

We spliced the 40-km spool with our splicer in green lab following the same procedure of last time and everything went smoothly. 

Fig. 1 shows the results on the splicer for the external end of the spool - loss = 0.01dB due to axis offset

Fig. 2 shows the results on the splicer for the internal end of the spool - loss = 0.02dB due to axis offset and core bending

This time we also installed the strinking tubes bought on amazon that Gabor suggested me to get, as shown in Fig. 3.

In order to have both ends of the spliced spool exiting from the top, I copied how Gabor did with the other spool, as shown in Fig. 4.

I measured the losses of the spool (using the NKT laser):

Input power = 113 uW

Output power = 20.5 uW

Loss = 82%

Loss_dB = 10*log10(113/20.5) = 7.4dB.

This loss level corresponds to 0.185 dB/km which is in agreement with what we expect, meaning that the splicing was successful.

Images attached to this report
GRAVITES Optics 1550nm
epolini - posted 19:26, Friday 19 April 2024 - last comment - 13:46, Tuesday 30 April 2024(11598)
Thorlabs ULN15TK power fluctuates

I noticed that the new thorlabs laser power fluctuates over time. I noticed it yesterday and I checked in different locations: right after the laser, after the isolator, after couple of BSs. I left the laser on the whole night and this morning the power was 20% higher than yesterday evening. We have an appreciable 20% decrease/increase also on a scale of tens of minutes. I queried the laser status and temperature with the commands shown on the manual using pyserial and the answers are not in agreement with what stated there. I contacted the thorlabs technical support to understand how to stabilize internally the laser temperature and (hopefully) stabilize the power.

Comments related to this report
epolini - 18:19, Tuesday 23 April 2024 (11605)

I saved the time series data using the Moku:Go, sampling rate 10Sa/s. I acquired 4 sets of 5 minutes. I used the PDA20CS with gain 0dB which corresponds to a transimpedance gain (TG) of 1.51e3 V/A.

I attach the data expressed in Volts and in Watts.

Power[W] = Vout / (R(lambda) * TG)

 

Images attached to this comment
epolini - 19:12, Tuesday 23 April 2024 (11607)

The laser current is stable at 0.55 A.

The laser temperature is stable at 25 Clesius.

The fbg temperature is stable at 28 Celsius.

epolini - 20:20, Wednesday 24 April 2024 (11609)

Today we measured the power after placing a 95/5 BS and a PBS after the FI. Attached the power at the output of the two PBS's ports. Pol 2 fluctuates more than pol 1.

Images attached to this comment
epolini - 13:46, Tuesday 30 April 2024 (11612)

I acquired data last Friday during the day (~2h30) and in the evening (~12h), after noticing that the saturation of the PDA20CS was ~2V for 0dB gain setting which corresponds to a maximum input power of 1.4mW. 

Fig. 1 and Fig. show the time series of the power registered with the PDA20CS. The power is not fluctuating as before, but I am keeping the laser on in order to have the laser more stable. The power flucuates by less than 5%.

Images attached to this comment
GRAVITES General
epolini - posted 19:11, Friday 19 April 2024 (11597)
Cable trays installed inside the cleanroom structure for GRAVITES

[Myron, Eleonora]

We installed two cable trays inside the clear room structure in green lab for the gravites experiment. The cable trays allows to bring the cables from the PDs on the bench and the instruments to the new rack we recently installed. We installed 2 trays 118'' long.

NB. On mcmaster it said 10 foot and 118'', we discovered that unfortunately they are 118'' and the 10 foot is just an appoximation. 

We connected them in a T shape. I attach some pictures here to show how we connected them with the cleanroom structure and between them.

Images attached to this report
GRAVITES Optics 1550nm
epolini - posted 20:14, Thursday 18 April 2024 (11596)
Thorlabs ULN15TK laser RIN

[Eleonora, Medha]

In the past we measured the RIN of the new thorlabs laser at 1550nm. We found out that the measurement is limited by the detector noise (using both the PDA20CS and the 1811) above 100Hz. We are confident that the noise comes from the detector since it is not changing by turning off the laser. Therefore, the detector noise measured  is higher than what expected from the datasheet. 

The noise floor on the SR785 was of 2-3uV/rtHz above 100Hz that normalized by the power and converted in dBc/Hz gives ~ - 130dBc/Hz for each setting on the PD. 

The computed values from the specs are obtained with the formula

Detector Noise = NEP * R(lambda) * Gain(depends on the gain setting).

0dB setting:

det_noise_V = 5.12e-11*1.5e3 = 7.68e-08 V/rtHz 

det_noise_dBc = 20*np.log10(det_noise_V/0.72) = -139 dBc/rtHz

20dB setting:

det_noise_V = 6.54e-12*1.5e4 = 9.81e-08 V/rtHz 

det_noise_dBc = 20*np.log10(det_noise_V/7.2) = -157 dBc/rtHz

50dB setting:

det_noise_V = 2.91e-12*4.75e5 = 1.38e-06 V/rtHz 

det_noise_dBc = 20*np.log10(det_noise_0db/9.92) = -160 dBc/rtHz

For the setting 0db and 20db we had 0.45mW arriving on the detector that corresponded to 0.72V and 7.2V respectively on the oscilloscope.

For the setting 50db we had 21.1uW arriving on the detector and 9.92V on the oscilloscope. We installed a BS to decrease the power and being able to raise the gain in order to see if the detector noise was decreasing and we could get a better RIN measurement.

Fig.1 shows the intensity noise spectrum in dBc/Hz with the laser ON and OFF. The measurement for frequency < 100Hz is in agreement with the measurement provided by thorlabs where we have -100dBc at 10Hz and -130dBc at 100Hz. At 1Hz they measured -60dBc, which is higher than the -75dBc we measured but it might be due to our measurement accuracy. 

I checked the shot noise level of the measurement just in case:

Shot Noise = sqrt(2*I*e) = sqrt(2 * (0.72 V / 1.5e3V/A) * 1.6e-19A/Hz) = 1.23e-11/rtHz -> -218 dBc/Hz

for a power of 0.45mW (for lower powers it is smaller). This confirms that it is lower than the detector noise.

We are confident that the noise limiting us is detector noise since it stays also when the light is turned off. There might be a discrepancy between the data on the PD datasheet and the real parameters. 

Our conclusion is that the intensity noise is up to specs so we won't investigate this further.

Images attached to this report
cSQZ General
hudsonl - posted 15:45, Wednesday 17 April 2024 - last comment - 15:48, Wednesday 17 April 2024(11594)
Beam Polarization from SHG into OPO
Yesterday, Jacques, Eric and I noticed that we were unable to excite only the vertically polarized cavity modes by adjusting the half waveplate between the 532nm Faraday isolator and the OPO. We verified that all of the optical elements are working properly. The problem seems to be that there are several mirrors between the Faraday and the waveplate, which likely have different reflection phases for vertical and horizontal polarizations. As a result, the light is elliptically polarized by the time it reaches the half-waveplate. Given the setup's space constraints, the best solution seems to be to place a 532nm quarter waveplate next to the half waveplate on the beam path from the SHG to OPO.
Comments related to this report
hudsonl - 15:48, Wednesday 17 April 2024 (11595)
I verified that the problem is due to eliptical polarization by rotating the Faraday such that the output polarization is vertical. In this orientation, we only excite vertically polarized cavity modes. This could not be achieved for any orientation of the lambda/2 indicating that the elliptical polarization is due to different reflection phases from the mirrors.
GRAVITES Optics 1550nm
epolini - posted 17:19, Tuesday 16 April 2024 - last comment - 19:05, Tuesday 23 April 2024(11592)
Thorlabs ULN15TK laser installed in green lab

[Eric, Eleonora]

Today we installed the new 1550nm thorlabs laser in green lab. We connected it to the interlock connecting the red cable to a power strip under the optical bench where the power supplies of both lasers are currently connected. 

How to turn the laser ON/OFF

Open the Command Prompt 

>python

>>import serial

Verify to which COM port the device is connected (it might change after the PC restarts)

>>ser = serial.Serial('COM8', baudrate=9600)

>>ser.write(b'laser on\r')

Now the laser is on and it is emitting 100mW.

To turn it off

>>ser.write(b'laser off\r')

Comments related to this report
epolini - 19:05, Tuesday 23 April 2024 (11606)

How to read the quieries with pyserial

>>> import serial
>>> ser = serial.Serial('COM8', baudrate = 9600)

>>> ser.write(b'laser on\r')
9
>>> ser.write(b'read_param laser_state\r')
23

>>> out = ''
>>> while ser.inWaiting() > 0:
...     out += ser.read(1).decode()
...
>>> print(out)

000: OK

000: OK:60,0x0000

>>> ser.write(b'read_param laser.current\r')
25
>>> while ser.inWaiting() > 0:
...     out += ser.read(1).decode()  --- NB press tab before out
...
>>> print(out)

000: OK

000: OK:60,0x0000

000: OK:0.550000,0x0000

>>> ser.write(b'read_param laser_tec_ctrl.temperature\r')
38
>>> while ser.inWaiting() > 0:
...     out += ser.read(1).decode()
...
>>> print(out)

000: OK

000: OK:60,0x0000

000: OK:0.550000,0x0000

000: OK:24.999786,0x0000 

>>> ser.write(b'read_param fbg_tec_ctrl.temperature\r')
36
>>> while ser.inWaiting() > 0:
...     out += ser.read(1).decode()
...
>>> print(out)

000: OK

000: OK:60,0x0000

000: OK:0.550000,0x0000

000: OK:24.999786,0x0000
000: OK:27.999939,0x0000

GRAVITES Optics 1550nm
epolini - posted 20:17, Saturday 13 April 2024 (11591)
Measurement UMZI 1km on Fuyu linear stage pt.2

1. Low pass filters at the Moku's inputs

I acquired new data with the Moku:Go after inserting two low pass filter, with frequency adapted for the sampling frequency, at the Moku's inputs to avoid the aliasing. The maximum time serie that the oscilloscope can acquire is 100-seconds long and the sampling frequency is 100Hz (ts = 10ms). The low pass filter has been chosen to have a fc ~ 25Hz in order to filter all the unwanted folded noises. In the experimental setup, the PD should be connected to the LPF and then the signal propagates until the Moku:Go. The output resistance of the PD 1811 is 33Ohm while the Moku load impedence is 'high'. I chose a 100kOhm resistor and a 67nF capacitor. The obtained fc = 24Hz. I soldered two LPF inside small boxes, as shown in Fig. 1, and connected them to the Moku, as shown in Fig. 2. 

2. Data acquired with the Moku:Go

I used the oscilloscope on the Moku:Go. Channel 1 is connected to the error signal and channel 2 to the correction signal. The loop was locked for the entire time of the measurements. I exported the data as file.csv, settings.txt and screenshot.png. Then I imported the data in python and plotted the time series with the following script:

t = []
err_V = []
corr_V = []

# Path to your CSV file
csv_file_path = 'MokuOscilloscopeData_20240413_185302_Traces.csv'

# Read data from CSV and store in arrays
with open(csv_file_path, 'r') as file:
    csv_reader = csv.reader(file)
    for row in csv_reader:
        if row[0].startswith('%'):  # Skip lines starting with '%'
            continue
        t.append(float(row[0]))   
        err_V.append(float(row[1]))  
        corr_V.append(float(row[2]))  


t_shift = [time - t[0] for time in t]    
plt.plot(t_shift, corr_V, label = 'corr signal')
plt.plot(t_shift , err_V, label = 'err signal')
plt.legend()
plt.ylabel('Amplitude [V]')
plt.xlabel('Time [s]')
plt.grid()        

I show the time series data for three different configurations: the stage is not moving, the stage is moving towards the computer, the stage is moving towards the wall.

2.1 Stage not moving

Fig. 3 show the data without the 24Hz-LPF in front of the Moku's input. Fig.4 shows the correct time series data, after the insertion of the LPF. The effect of the filter is clear.

2.2 Stage moving towards the computer

I used the same setting mentioned in the entry 11588, i.e. position +60, acc/dec 5, speed 1, Rel. The filtered data are given shown in Fig. 5. I started moving the stage after ~10 seconds and it can be seen that the correction signal increases. The stage finished its ride after ~60 seconds and it can be seen that the correction signal desreases.

2.3 Stage moving towards the wall

The only different setting here is the position -60. The filtered time series is shown in Fig. 6. The noise of the stage moving in the two directions is a bit different and I am wondering if this effect is also shown in the time data. Also here it is clear when the stage was moving looking at the correction signal.

Next steps: to perform and tune the PSD of these data and understand the figure of merit to look at to determine the performance of the different stages (high load Fuyu and rotation actuator Zaber).

Images attached to this report
Heisenberg IFO General
jacques - posted 15:46, Friday 12 April 2024 (11590)
Test of the RP data acquisition (2)

Following the first attempt at RP acquisition, I resolved the problem by changing Ben's RP (apparently it is broken at both inputs) with another one. Then, using the streaming app (and not the SCPI server), described here, I could record a 100kHz, 0.2Vpp sine wave, on the RP's SD memory.

Conclusion: the streaming module of the RP can be used for data acquisition. Note that I was on the most recent stable version of the RP OS (2.00-30).

Also below is a code snippet from Ben which should allow to launch the acquisition on python (replace the parameters with the ones relevant to the RP):

import subprocess
subprocess.Popen(["./rpsa_client","--streaming", "--hosts=192.168.0.78", "--format=csv", "--limit=2000000", "--mode=volt", "--dir={direc}"])
Electronics General
epolini - posted 10:58, Thursday 11 April 2024 (11589)
Dark Noise of the fempto homodyne: HBPR-100M-60K-IN-FC

Attached the dark noise of the fempto homodyne measured down to 100mHz by the company.

Non-image files attached to this report
GRAVITES Optics 1550nm
epolini - posted 20:33, Wednesday 10 April 2024 - last comment - 17:51, Tuesday 16 April 2024(11588)
Measurements with the Fuyu horizontal stage

In the past days I arranged the optical setup to be able to displace the 1-km spool on the stage without encountering obstacles. I connected the Moku:Go inputs to the error and correction signals closed loop of the UMZI setup. The powers coming from the two arms and recombining in the second 50/50 BS of the UMZI, are 122uW (I tuned the VOA). The calibration of the error signal is 67 mV/rad. The loop was closed as done last time.

1. Stabilization of spool and stage

At first, since while the stage was moving the spool was not stable, I screwed a post into the small platform hosting the spool on the stage and I added some ring spacers to fit the height. The I insert the black piece shown in Fig.1 to fix the spool to the platform. I also fixed the stage to the optical bench with three clamps (see Fig. 2). NB. if the moving platform eccesses the range, the stage is over vibrating and this makes the stage itself displacing, despite the clamps. 

2. Chose the speed and acceleration of the moving platform

I tried different options for the speed and accelation. I was checking the error and correction signals on the oscilloscope, both on the analog oscilloscope and the Moku:Go oscilloscope, which I used to acquire time series data. 

How to set the controller interface:

Select the single-JOG interface. 

X Tar, X Tar Speed, X Tar Acc, X Tar Dec are respectively the target position, speed, acceleration and deceleration. Then one can chose between Abs (absolute), Rel (relative) and Stop. Abs refers to the absolute value for the position and after testing this feature, I realized that there is an hysteresis and then it is better to use the relative option. I move the stage manually to one and and then I understood that to cover almost all the linear range I should indicate +/-60 in the X Tar Pos. The sign + refers to the direction towards the computer, the sign - to the direction towards the wall. To avoid being too violent on the spool and to unlock the system, I set X Tar Speed to 1 (that should correspond to 1mm/s according to the manual) and X Tar Acc/Dec = 5 (see Fig. 3). I measured the distance and the time relative to these settings and I got: 150mm in 60 seconds that corresponds to 2.5mm/s (speed = 1). Then I changed the X Tar Speed to 2 and I got: 150mm in 30 seconds that corresponds to 5mm/s. 

3. Measurements of the error and correction signals and polarization changes during the stage motion

I installed two PDA20CS PDs at the outputs of the PBS after the polarization controller that was ON during the measurements. Since the Moku has only two input channels, I monitored the polarization time series only with a traditional oscilloscope. In Fig. 4 are shown: error signal closed loop (yellow), correction signal (blue), signals in output of the PBS after the polarization controller (purple and green) while the stage with on top the spool is moving with the settings already discussed. We can see that the purple signal is changing over time, but then it goes back to its average value. The correction signal has some spikes but overall the system seems locked for the whole measurement time.

4. Check OLTF of the UMZI before and after the measurements with moving stage

I measured the OLTF of the loop injecting a swept sine of 300mV amplitude, both before and after all the measurements with the stage and it did not change as shown in Fig. 5.

5. Power check after the measurements

The powers recombining at the second 50/50 BS are 117uW (reference arm) and 113uW (spool arm). The power decresed but it did not end up to be too different between the two arms.

Next step:

I acquired the time series with the Moku:Go and I will show the spectra while the stage is at rest and when it moves.

Images attached to this report
Comments related to this report
epolini - 17:51, Tuesday 16 April 2024 (11593)

NB. the powers in Watt need to be divided by a factor 10 here since there was an attenuation on the powermeter but no ND filter installed during these measurements.The powers recombining at the second BS are of the order of 10 uW and not 100 uW.

Facilities General
wenxuan - posted 16:13, Monday 01 April 2024 - last comment - 14:24, Tuesday 23 April 2024(11577)
Access to writable MEDM for remote commissioning

To have access to write-able MEDM: there are two ways to do it, but it requires Yubikey to authenticate. 

 

1. Launch medm from the CDS login portal, but display it on the local machine. It has everything but a bit slow

ssh -X -C albert.einstein@cdsssh.ligo-la.caltech.edu
sitemap

 

2. Remote desktop. I've installed X2Go for L1 (instructions). 

x2goclient # Launch the client

If the remote control freezes every 15 minutes due to screensaving, manually turn it off by

xset s 0 0 &
xset -dpms; xset s off

Comments related to this report
wenxuan - 14:24, Tuesday 23 April 2024 (11604)

Command for X2Go:

ssh -C -L localhost:2233:cr10:22 albert.einstein@llocds.ligo-la.caltech.edu
Displaying reports 621-640 of 1025.Go to page Start 28 29 30 31 32 33 34 35 36 End