Displaying reports 601-620 of 1025.Go to page Start 27 28 29 30 31 32 33 34 35 End
Reports until 18:31, Tuesday 14 May 2024
cSQZ General
hudsonl - posted 18:31, Tuesday 14 May 2024 (11632)
CSQZ Homodyne Re-Aligned
[Hudson, Jacques]

We re-aligned the homodyne up to 80% visibility. After cleaning the optics, the homodyne got misaligned, so we took the opportunity to rework the homodyne beam path such that none of the optics for the Mach-Zehnder interferometer are on flip mounts. When that experiment is running, other mirrors, that would direct the beam directly to the homodyne will be flipped out of the beam path.

We will now proceed to measure squeezing and anti-squeezing vs NLG and compute the total losses and phase noise to determine if these are consistent with the homodyne's imperfect visibility being the dominant loss mechanism.
cSQZ General
hudsonl - posted 12:03, Tuesday 14 May 2024 (11631)
CSQZ SHG Silicon RFPD Replaced with InGaAs RFPD
I replaced the custom Silicon RFPD with a Thorlabs InGaAs RFPD. This looks to have significantly improved the SHG lock's SNR, especially at low powers. This solved a problem that with high power sent to the OPO and low power sent to the SHG, the SHG cavity would come unlocked due to any small disturbance on the table.

The Thorlabs PD is located off of the optics table and will eventually need to be moved back to the compact squeezer board for compactness. This is a relatively low priority until the Heisenberg interferometer experiment is completed.
cSQZ General
hudsonl - posted 19:52, Wednesday 08 May 2024 (11630)
2dB Squeezing and 10dB Anti-Squeezing Measured with all Feedback Loops Closed
[Hudson, Jacques]

After replacing the custom Silicon PD measuring the CLF beam with a Thorlabs InGaAs detector, we were able to turn down the CLF power level considerably while retaining enough SNR on the CLF and LO loops to lock them.

With these loops locked, we measured about 2dB of squeezing and 10dB of anti-squeezing on the homodyne, verifying that all essential parts of the compact squeezer are working. Next, we will begin more careful diagnostics of the system and work to tune up the squeezing levels from 2dB to the 8-10dB range. The attached images show spectra with squeezing and anti-squeezing. Note that the y-axis scales are different in these images.
Images attached to this report
cSQZ General
hudsonl - posted 19:42, Wednesday 08 May 2024 (11625)
CLF RFPD Path Modifications
I added a steering mirror after the beam splitter in the CLF path acting as a circulator. The steering mirror allowed me to direct the CLF beam reflected from the OPO off of the compact squeezer board and onto the table, where I set up a Thorlabs high gain, high bandwidth InGaAs PDA10CF. See attached image.

I anticipated that this PD would have better performance than the custom PD since it uses an InGaAs diode instead of a silicon diode. However, I anticipated this improvement would be roughly a factor of 4, accounting for the quantum efficiency of InGaAs (0.7 A/W at 1064 nm) compared to silicon (0.15 A/W at 1064 nm). However, the Thorlabs detector had a noise performance almost 20 dB better than the custom Silicon PD. 

Switching out these PDs allowed us to significantly reduce the CLF power while retaining reasonably high SNR on the CLF and LO feedback loops. 

The attached file shows the beam path with the new Thorlabs PD.
Non-image files attached to this report
GRAVITES Optics 1550nm
epolini - posted 18:36, Wednesday 08 May 2024 (11629)
Aluminium spool cleaned and sent to get re-spooled

[Eleonora, Myron]

Yesterday we cleaned the spool designed by Eric following this procedure:

1. Clean it with a tissue and some isopropanol

2. Clean it with the liquinox solvent in the small ultrasonic machine in Vacuum Prep Lab - Fig. 1

We filled the tank with water and half a cap of liquinox solvent. We used the degas option for 5 minutes. Then we removed the water and washed the container. We filled again the the tank with water and we put some isopropanol on the spool. We set other 5 minutes of degas option. Then we washed again the tank. 

3. Clean it with isopropanol in the ultrasonic machine

We filled the tank with water again and we set the machine to heat mode. We filled half of the small contaner shown in Fig. 2 with isopropanol and we insert the spool in it. We set the machine to degas again and we rotated the spool during this time to be sure to clean it all. Moreover we used some tissues do further clean it. Then we put the alchol in a container saying that it has been used once. 

Finally, we sent the spool and a new 1km received today (not spliced yet) from FIS to the company Opticoncepts to get it re-spooled. Lead time 10 days.

Images attached to this report
GRAVITES Optics 1550nm
epolini - posted 18:10, Wednesday 08 May 2024 (11628)
Noise Budget of the Balanced Mach Zehnder Interterferometer with 1km arm

I report here the noise budget of the 1km-BMZIs described in the entry #11626 using both the Thorlabs and the NKT lasers.

I also used the formula (23) of Chris' optica paper to express the maximum optical power as a function of the fiber length allowed by stimulated Brillouin scattering. In the equation appears the effective length (Leff) of the fiber. 

Fig. 1 shows Leff as a function of the real length for the Corning SM28-ULTRA fibers that we are currently using. 

Leff = [ 1 - exp(-a*L)]/ a

where a are the losses expressed in a linear scale. The SM28-ULTRA fibers have 0.18dB/km losses which translates into a linear scale as : a = 1 - 10^(-0.18/10) = 0.0406.

The Brillouin scattering threshold power is given by:

P_B = k * K * Aeff / (gB * Leff)

where k = 19 (numerical factor), K = 1 (mixing efficiency), Aeff = pi * rmf^2 = pi * (46 um)^2 (effective area), gB = 1.68e-11 m/W (Brillouin gain).

Fig. 2 show the maximum power due to SBS as a function of fiber length:

Psbs( L = 1km ) = 0.075W = 75 mW

Psbs( L = 40km ) = 0.0038 W = 3.8 mW

Noise budget BMZI 1km arms

I considered the model of the frequency noise of the lasers as shown in the entry #11616. I also included the rms of the correction and error signal data.

The input power P0 to the BMZI (before the first BS) = 3.2 mW.

1. NKT laser

Fig. 3 and Fig. 4 show the correction and error signal noise budgets respectively.

2. Thorlabs laser

Fig. 5 and Fig. 6 show the correction and error signal noise budgets respectively.

There is an underestimated source of noise in the thorlabs laser case. We just received the 100m and 200m spools, so I will build an UMZI with unbalance < 329m in order to measure the frequency noise.

Images attached to this report
GRAVITES Optics 1550nm
epolini - posted 21:57, Monday 06 May 2024 - last comment - 13:25, Tuesday 07 May 2024(11626)
Balanced Mach Zehnder Interferometer Characterization: 1km (Thorlabs, NKT) and 40km (NKT) arms

Last Friday and today, I have characterized the balanced Mach Zehnder interferometer using both the Thorlabs and NKT lasers. Since the Thorlabs laser frequency noise is too high for an unbalance between the arms higher than 329 meters, we don't have at the moment a setup to measure it in this way. We ordered shorter spools but in the meantime I characterized the balanced MZI using both lasers.

1. BMZI characterization with 1km arms

It was possible to lock the loop using both the Thorlabs and the NKT lasers. I report the loop characterization for both.

1.1 Thorlas laser (BMZI 1km arms)

Sensor TF : 180 mV/rad

VCO FM dev : 6 kHz - VCO TF : 3.9 kHz/V

UGF : 32 kHz

Servo Gain : 33 dB

Fig. 1 shows the error and correction signal spectra calibrated in rad/rtHz.

Fig. 2 shows the OLTF of the loop.

Fig. 3 shows the derived TF of the servo board.

1.2 NKT laser (BMZI 1km arms)

I increased the laser power in order to have the same error signal fringes amplitude.

Sensor TF : 175 mV/rad

VCO FM dev : 6 kHz - VCO TF : 3.9 kHz/V

UGF : 31 kHz

Servo Gain : 33 dB

Fig. 4 shows the error and correction signal spectra calibrated in rad/rtHz.

Fig. 5 shows the OLTF of the loop.

Fig. 6 shows the derived TF of the servo board.

2. BMZI characterization with 40km arms

It was possible to lock the loop ONLY with the NKT lasers. The most probable reason is that the unbalance between the two 40km spools is of the order of hubndreds of meters and this prevents the loop to be locked due to laser frequency noise. I characterized the BMZI loop with the NKT in two different configurations: same settings as before that gives gain peaking and lower gain on the servo to avoid gain peaking.

2.1 NKT laser (BMZI 40km arms) - same settings as before

The error signal fringes have a smaller amplitude.

Sensor TF : 154 mV/rad

VCO FM dev : 6 kHz - VCO TF : 3.9 kHz/V

UGF : 30 kHz

Servo Gain : 34 dB

Fig. 7 shows the error and correction signal spectra calibrated in rad/rtHz.

Fig. 8 shows the OLTF of the loop.

Fig. 9 shows the derived TF of the servo board.

2.2 NKT laser (BMZI 40km arms) - lower gain

Sensor TF : 154 mV/rad

VCO FM dev : 6 kHz - VCO TF : 3.9 kHz/V

UGF : 1.56 kHz

Servo Gain : 7 dB

Fig. 10 shows the error and correction signal spectra calibrated in rad/rtHz.

Fig. 11 shows the OLTF of the loop.

Fig. 12 shows the derived TF of the servo board.

 

Next steps:

We need to measure the unbalance between the two long spools in the arm. Eric suggested to modulate the frequency of the laser and then measure the phase at the output of the interferometer. I ordered some spools (100 meters, 200 meters and 1 km) so that we can balance the arms and lock the 40km-BMZI also with the thorlabs laser. 

Images attached to this report
Comments related to this report
epolini - 13:25, Tuesday 07 May 2024 (11627)

As a comparison, I plotted on the same plot the error and correction signals calibrated in rad/rtHz of the BMZI 1km with NKT and thorlabs lasers. The loop configuration is comparable. We notice an excess in noise at frequencies > few kHz for the measurements done with the thorlabs laser. The plot is attached.

Images attached to this comment
Heisenberg IFO General
jacques - posted 20:29, Friday 03 May 2024 - last comment - 16:17, Sunday 05 May 2024(11621)
building the new HDs: origin of the 43kHz oscillation & other preliminary tests

I debugged by HD board:

- At the start of my inquiries, spontaneously, the DC power supply would show:

1/ That the +18V is drawing 0.16A while the -18V is drawing 0. The cause of this is not certain but I am suspecting the power cable to play a role in it

2/ That both +-18V are drawing 0.3A, which is much higher than the expected ~0.1A in normal regime. The origin of that is probably a bad soldering of one of the components, leading to either bad contact or shorting of a component. The way to solve this is to remove the most suspect components until the current drawn is back to a reasonable value.

- Symptom 2/ was accompanied by the U4 and U3 regulators (maintaining +-5V for the rest of the board) which became exceedingly hot. This meant that the component which was drawing all this current was one the OpAmp powered at +-5V. It turned out that this was U11, as removing it made the DC supply go down to <0.1A. 

- The 43kHz oscillation comes from the RF2 path (notch at 10MHz), as removing either the U11 OpAmp or the L11 inductor made these parasitic oscillations disappear in all of the other outputs. It is however important to note that the C49 capacitor was always removed, so the signal path is cut. This means that somehow the RF2 path picks up noise from the environment and oscillates at its frequency? Unless there is some bad soldering on the RF2 path...

- U9 can be resoldered using Slawek's heat gun (but make sure that the chip is mainted in place using a tweezer). Careful examination of the pins's sides should determine whether U9 has been soldered properly or not.

- I then did some quick tests of the outputs: I tested the PD1 and PD2 paths, simply by shining a lamp on both photodiodes at the same time and looking at the outputs on a DC scope. I do get a non-zero DC output on the scope (see attached): a periodic signal at the grid frequency of 120Hz. This tells me that the paths to the PD output are working. It is important to notice that neither signal go to zero when one I block one of the PDs from the incident light; I would have expected the corresponding PD output voltage to go to 0 but I could misunderstand these outputs.

- I also get a signal on the DC OUT path, which goes from -5V to +5V depending on the orientation of the board wrt the light source.

- The AC path also gives a non-zero output when I shine some light on the PDs

- RF1 gives a very strong and stable oscillation at 16.66MHz (see picture), even when there is no light on the PDs, which is very worrying.

Images attached to this report
Comments related to this report
jacques - 16:17, Sunday 05 May 2024 (11624)

I realize that R62 (tranimpedance) was not soldered in during these tests.

Below is a noise plot obtained at the AC OUT when all the components are stuffed except U10 and U11, there is 25dB of attenuation on the agilent input.

I don't see any noise on AC OUT of Eric's board so there is also something broken there.

Images attached to this comment
Electronics General
epolini - posted 20:16, Friday 03 May 2024 (11618)
Agilent moved from NW22 to NW17 (electronic shop)

[Eric, Eleonora]

We moved backt the Agilent 4395A that was used for the gravites 200-meters prototype in NW22 back to NW17. Now it is on a chart in the electronic shop.

Heisenberg IFO General
jacques - posted 19:07, Thursday 02 May 2024 (11617)
building the new HDs: first tests and first issues

With Eric we built a pair of HD electronic boards, one of which (mine) is fully stuffed. 

We added some PDs to it (<93% quantum efficiency so we'll need to swap them with the good ones at the end of the tests) and tried to measure the electrical noise performance on a spectrum analyzer (SR and agilent), by simply powering the HD boards and blocking the PDs.

Note that the DC OUT has a cutoff at 10kHz and that the AC OUT is internally amplified, so it should be above the noise floor of the Agilent.

The spectrum I took of my HD, on the Agilent, showed a strange comb feature (spaced by 43kHz, which may simply come from the maximum of the ambiant noise spectrum) and very large AC OUT noise (see first picture, red is the AC OUT and attenuated by 15dB to prevent Agilent saturation, yellow is Agilent noise floor). The comb is also visible in the DC OUT channel (see second image).

This comb is not seen on Eric's HD board.

Conclusion: there must be a problem in the electronics on my board, possibly from the RF resonant paths. I will need to progressively take down the components until I figure out where the problem comes from.

 

Images attached to this report
GRAVITES Optics 1550nm
epolini - posted 21:16, Wednesday 01 May 2024 (11616)
Maximum unbalance between the interferometer arms before being limited by laser frequency noise

Since one possible issue with the UMZI not locking could have been an excess in laser frequency noise at the input of the Mach Zehnder (and Michelson for the Thorlabs as well), we calculated the maximum length unbalance to keep the phase at the output below 1 radian.

I did that for both the NKT HP X15 and the Thorlabs ULN15TK. I fitted the frequency noise taken from the datasheets. I computed the rms for the input frequency noise. I used the formula to convert the input frequency noise into output phase noise used in Chris' paper:

S_phi(omega) = (2 pi tau)^2 * S_freq(omega) -> A_phi(omega) = (2 pi tau) * A_freq(omega)

where tau = n*DL / c.

We set A_rad_rms = 1 rad.

1. Thorlabs ULN15TK maximum arm unbalance

Fig. 1 shows the laser frequency noise data, fit and rms. I considered A_rms_freq = 1e5 Hz.

DL = A_rms_phi * c / (2 * pi * n * A_rms_freq) ~ 329m

This means that with the arms unbalanced by 1km, as it is in the current setups, we will have fringe wrapping at the output. 

2. NKT X15 maximum arm unbalance

Fig. 2 shows the laser frequency noise data, fit and rms. I considered A_rms_freq = 5982 Hz.

DL = A_rms_phi * c / (2 * pi * n * A_rms_freq) ~ 5505m

This confirms that having the arms unbalanced by 1km should be fine for this laser.

Images attached to this report
GRAVITES Optics 1550nm
epolini - posted 20:57, Wednesday 01 May 2024 (11615)
Lessons learned debugging the UMZI - 1km loop

The UMZI was not locking in the past days even with the NKT laser. While debugging, we measured the servo board and adjust the settings on the marconi. 

1. Eric's servo board transfer function from OLTF

The first thing we tried was to match all the parameters of the loop we had once we were able to lock it. Sensor TF ~ few tens mV/rad, LO 80MHz and 7dBm, VCO 80MHz and -5dB (sent to a HP amplifier), PI corner at 480Hz (turning integrator 1 ON on the board) and then adjust the gain accordingly (there is no indicator, so it needs to be measured). To derive the gain used when locked, I did:

TF_servo = OLTF / (TF_sensor * TF_Plant * TF_VCO) 

Using the OLTF, TF_sensor, TF_Plant, TF_VCO used in the entry 11479.

freq_27 = np.array(freq_27)
mag_27 = np.array(mag_27)
phase_28 = np.array(phase_28)
OLTF_complex = 10**(mag_27/20)*np.exp(1j*np.pi*phase_28/180)
c = 3e8
L = 1e3
n = 1.44
tau = n * L / c
P = np.exp(-1j*2*np.pi*freq_27*tau)
A = 6.5e3 / (-1j*freq_27)
S = 25e-3
F = OLTF_complex / (P*A*S)
F = np.array(F)

magnitudes = np.abs(F)
phases = [cmath.phase(element) for element in F]
phases_degrees = [phase * 180 / cmath.pi for phase in phases]

mag_db = 20*np.log10(magnitudes)

plt.figure(figsize=(9, 6))
plt.subplot(2, 1, 1)  
plt.semilogx(freq_27, mag_db, label = 'Servo Board Mag')
plt.ylabel('Magnitude [dB]')
plt.xlabel('Frequency [Hz]')
plt.xlim([100, 1e5])
plt.legend()
plt.grid()


plt.subplot(2, 1, 2) 
plt.semilogx(freq_28, phases_degrees, label='Servo Board Phase')
plt.ylabel('Phase [deg]')
plt.xlabel('Frequency [Hz]')
plt.xlim([100, 1e5])
plt.legend()
plt.grid()

The plot is shown in Fig. 1. The gain to set on the board is the one at high frequency in the TF, meaning 40dB.

2. Set 40dB gain on the servo board by looking at its TF on the SR785

To set the gain correctly, we tune it looking at the servo's TF. We send the swept sine to the input of the servo board, i.e. where we usually send the error signal. Then we take the output at the output of the board, i.e. where we usually take the correction signal. We put the signals on the SR785 and we see the gain range of the board by tuning the gain. The range is 6dB to 40dB. 

NB. be careful not to saturate the board by sending a large amplitude swept sine. 500mV amplitude was making saturating the board and I was reading a gain range of -28dB to 5dB. A good amplitude is 50mV.

We set the gain back to 40dB.

3. Coupling settings on the Marconi

I did not notice that the settings on the Marconi changed. The FM modulation was set to Ext AC mod coupled instead of Ext DC mod coupled. By checking the output of the marconi with the frequency counter it was obvious that the correction signal of the loop was saturating but the frequency was not spanning the VCO range.

 

Images attached to this report
GRAVITES Optics 1550nm
epolini - posted 20:21, Wednesday 01 May 2024 (11614)
Thorlabs laser power measurement after turning ON

Today I acquired again power data from a moment the laser was off and for 4 hours after turning it on, to understand the time we need to wait for the power to stabilize. 

Fig. 1 shows the full data points at the top and then I zoomed into the data when the power was on. There is a power drift at the beginning and then it fully stabilize after ~1h30. 

I inserted a BS 95/5 after the FI following the laser. The input of the BS is port 1 and the outputs are:

Port 2 (5%) : P = 4.5mW

Port 3 (95%) : P = 79mW.

Port 3 (95%) goes to the full setup, while the 5% pick off goes to another 99/1 BS. The 1% of this BS goes to the PDA20CS photodiode connected to the Moku:Go.

The power measured with the powermeter is 35uW while the recorded value is a bit lower. This can be due to the not perfect value of the transimpedance gain reported on the datasheet or the clipping due the beam divergence in the photodiode (since the sensitive aperture of the powermeter is substantially larger).

Images attached to this report
cSQZ General
hudsonl - posted 13:52, Wednesday 01 May 2024 (11613)
CSQZ OPO Temperature Setpoint
I had been operating the OPO around a setpoint of 37.5 C. This seemed a bit high, but I was able to see reasonable NLGs of about 3. However, it turns out this was a side-lobe of the phase-matching sinc function. After re-optimizing the crystal's temperature and position, I now see an NLG of >10 at a phase-matching temperature of about 34 C. This makes sense as it is closer to the crystal's design spec for its phase-matched temperature as well as the SHG's phase-matched temperature.
cSQZ General
hudsonl - posted 17:07, Friday 26 April 2024 (11611)
Dichroics Added to Path from SHG to OPO
I added a pair of dichroics on the path from the SHG to OPO on the compact squeezer to prevent IR light from leaking from the SHG to the OPO. Each of these dichroics is specced at <10^-4 IR transmission. 

Previously, we had about 50mW of IR into the SHG, with a finesse of 150, so we expect about 7.5 W of circulating IR power in the SHG (ignoring pump depletion for simplicity in the order-of-magnitude calculation). We then expect about 750 uW transmitted through the SHG's dichroic and 75 nW sent into the OPO's input/output coupler after the single dichroic on this beam path. This ignores losses from the commercial 532nm optics on the PDH board, which will probably attenuate the IR beam by a few orders of magnitude so we're sending closer to 1 nW into the OPO. 

After adding a pair of dichroics, we should now be sending at most 1 fW of IR power into the OPO's input/output coupler and realistically are likely sending closer to 10 aW into the OPO.

After adding the dichroics, we re-aligned the OPO but did not redo the mode matching. On resonance, we now have about 150 mV on the 532 transmission PD compared to 500 mV previously, indicating that the mode matching is substantially worse than it was previously. This is not surprising given that the beam now passes through a pair of dichroics which we expect to shift the waist position. The dichroics are 6mm thick with an index of about 1.5 and are passed through at 45 degrees, so we expect the beam's focus to shift by about 2*(1.5-1)*6*1.41 = 8.5 mm.

The thermal effects in the OPO also appeared to be worse with this configuration, indicating that improving the mode matching may also improve the OPO's thermal stability.
GRAVITES General
epolini - posted 18:06, Thursday 25 April 2024 (11610)
New Windows version for the computer in green lab

Fred updated the Windows version on the computer in green lab so that I was able to download the Moku app that allows to take data etc. 

GRAVITES Optics 1550nm
epolini - posted 19:26, Friday 19 April 2024 - last comment - 13:46, Tuesday 30 April 2024(11598)
Thorlabs ULN15TK power fluctuates

I noticed that the new thorlabs laser power fluctuates over time. I noticed it yesterday and I checked in different locations: right after the laser, after the isolator, after couple of BSs. I left the laser on the whole night and this morning the power was 20% higher than yesterday evening. We have an appreciable 20% decrease/increase also on a scale of tens of minutes. I queried the laser status and temperature with the commands shown on the manual using pyserial and the answers are not in agreement with what stated there. I contacted the thorlabs technical support to understand how to stabilize internally the laser temperature and (hopefully) stabilize the power.

Comments related to this report
epolini - 18:19, Tuesday 23 April 2024 (11605)

I saved the time series data using the Moku:Go, sampling rate 10Sa/s. I acquired 4 sets of 5 minutes. I used the PDA20CS with gain 0dB which corresponds to a transimpedance gain (TG) of 1.51e3 V/A.

I attach the data expressed in Volts and in Watts.

Power[W] = Vout / (R(lambda) * TG)

 

Images attached to this comment
epolini - 19:12, Tuesday 23 April 2024 (11607)

The laser current is stable at 0.55 A.

The laser temperature is stable at 25 Clesius.

The fbg temperature is stable at 28 Celsius.

epolini - 20:20, Wednesday 24 April 2024 (11609)

Today we measured the power after placing a 95/5 BS and a PBS after the FI. Attached the power at the output of the two PBS's ports. Pol 2 fluctuates more than pol 1.

Images attached to this comment
epolini - 13:46, Tuesday 30 April 2024 (11612)

I acquired data last Friday during the day (~2h30) and in the evening (~12h), after noticing that the saturation of the PDA20CS was ~2V for 0dB gain setting which corresponds to a maximum input power of 1.4mW. 

Fig. 1 and Fig. show the time series of the power registered with the PDA20CS. The power is not fluctuating as before, but I am keeping the laser on in order to have the laser more stable. The power flucuates by less than 5%.

Images attached to this comment
Displaying reports 601-620 of 1025.Go to page Start 27 28 29 30 31 32 33 34 35 End