Heisenberg IFO General
hudsonl - posted 16:05, Wednesday 29 May 2024 (11654)
Diabolo Laser Interlocked
[Hudson]

I interlocked the diabolo laser. Both lasers on the squeezing table, the Lightwave electronics for the compact squeezer and the Diabolo for the breadboard OPO are now interlocked.
cSQZ General
hudsonl - posted 20:16, Tuesday 28 May 2024 (11651)
CSQZ Squeezing Levels
After cleaning the optics the other week, I took a more careful squeezing and anti-squeezing spectrum today. The pump beam's power is still quite unstable, so it seems like it will be challenging to make a careful squeezing and anti-squeezing vs NLG plot until that is fixed. However, even with the unstable pump, we can start to investigate which parts of the setup are limiting our measured squeezing levels.

Currently, we measure 2.6 dB of squeezing and 6.73 dB of anti-squeezing measured between 55kHz and 70kHz, where we have a clean spectrum. I expect the squeezing measurement is more reliable here as it is less sensitive to pump power fluctuations. The last time we measured the homodyne, it had a visibility of 80%, and the OPA's escape efficiency is 0.95 giving about 40% total loss from these sources. As a result, I expect to see 4.0 dB of squeezing in the absence of phase noise, any additional sources of losses, and with the OPA approaching threshold. In the absence of phase noise and with the OPA near threshold, a maximum squeezing level of 2.6 dB is indicative of about 55% total loss, about 15% higher than that from sources identified so far. As this number assumes the OPA is very near threshold, which it is not, the unexplained losses are somewhat below 15%.

In conclusion, measurements of the compact squeezer so far are roughly compatible with squeezing levels limited by homodyne visibility.

We note that for given effective optical losses, L = (Power Out / Power In), the maximal observable squeezing level is -4.343*ln(L) [decibels]. This is plotted in the attached figure.
Images attached to this report
GRAVITES Optics 1550nm
epolini - posted 13:10, Tuesday 28 May 2024 (11650)
UMZI 1km NKT with new isolation spool

I made some measurements with the UMZI 1km NKT including the new vibration isolation spool (see entry #11649), using the servo board designed by Eric (PI corner at 482Hz). I made the set up as simple as possible by removing the polarization controller and including a SM 50/50 BS at the output.

Power budget

The power before the first PM 50/50 is 0.79mW.

On one arm we have 302uW and then the 1km spool. The power after the spool is 290mW. This gives 0.176dB/km of losses that is in agreement with what expected and confirms that the splicing worked. The spool is followed by the AOM and the power recombining at the second SM 50/50 BS is 59uW. 

On the other arm, I installed a 50/50 PM BS followed by a VOA to balance the power recoupling in the second BS. 

Measurement UMZI 1km with old spool

Fig. 1 shows the OLTF (source 300mV) and the UGF is ~33kHz. The fringes here are not nice looking as already noticed in the past (see Fig. 3).

Measurement UMZI 1km with new spool and isolation box

Fig. 2 shows the error and correction signals for the two cases and it can be seen that the new spool and the isolation box I assembled, are reducing  the acoustic noise below 4kHz. In particular the structure around 300Hz is reduced by 1-2 orders of magnitude. Fig. 4 shows the shape of the fringes in this configuration that is way clearer than with the old spool.

Images attached to this report
GRAVITES Optics 1550nm
epolini - posted 20:27, Wednesday 22 May 2024 (11649)
Isolation box for the new 1km vibration insensitive spool

Today we received the 1km fiber re-winded around the vibration insensitive spool designed by Eric.

The company re-winding the spool is Opticoncepts and the tension or re-spooling is (0.65 +/- 0.05)N.

I spliced it with an SM patch cord and then I built an isolation box to host it.

I found a robust metal box sitting in my office that Rich helped me to cut in order to let the fiber pass through it on both sides. Then I cut and glued with some double sided paste a rubber layer on the bottom of the box as shown in Fig. 1.

Then I cut and pasted some acoustic foam on the four lateral walls and on the box lid as shown in Fig. 2 and Fig. 3. Then I tightly screwed the four screws on the lid to seal the box (Fig. 4). I will test the UMZI with this new box (with the NKT laser) tomorrow.

Images attached to this report
Heisenberg IFO General
jacques - posted 20:11, Wednesday 22 May 2024 (11648)
Building new homodyne detectors: fix to the old problem and new problems

[Eric, Jacques, Malo]

We had previously concluded that the AC OUT paths on both Eric's and my HD boards were not working properly. Eric debugged it by noticing that the PD sensors were shotcircuiting the auxiliary pins when they were fully shoved into the board. Thus the fix consisted in adding some electrical tape to insulate the aforementionned zone of the board and resoldering the PDs. 

But even with this fix, Eric's board's AC OUT was showing some increased electronic noise above a few hundreds of kHz, see pictures of the agilent. This was not noticed on the SR because of its limited bandwidth to 100kHz.

The DC path looked flat in terms of noise up to 10s of MHz, both on the Agilent and the SR, so the problem seems to come from the U8 op amp on the AC OUT (doing the inverting x10 gain). 

Thus we can still go on by desoldering the AC path's input capacitance and only using the DC path (which then also inherits the AC components), at the expense of losing the internal x10.

Next step:

1/ Do a TF measurement my shining modulated light on it.

2/ Debug my board by removing the AC path.

Images attached to this report
LASTI General
wenxuan - posted 15:37, Wednesday 22 May 2024 (11647)
VOPO platform out of HAMX chamber

+ Dhruva, Eric, Myron, Rich, Lisa

 

We moved the VOPO platform and miscellaneous optics out of HAMX chamber in LASTI. We tried to keep all the hardwares on VOPO such that the alignment is not too far off.

 

Eric and Lisa organized all the removed optics that are not on the VOPO platform. The indexing can be found below.

 

Activity log:

1153 Disconnect OSEM cables from the in-chamber small cable rack.

1154 Disconnect TEC, etc. cables from the OPO.

1158 Disconnect SHG fiber and CLF fiber. They are left in chamber.

1205 Removing dog clamps that fix VOPO on HAMX

1213 Disconnect the cable of the on-VOPO lens stage. It's at the place where SQZ vacuum leaves VOPO.

1225 All dog clamps off

1228 Remove periscope

1247 Remove optical mounts 1-12 on fc side of table. They are put in the box and labeled. See this picture for the original locations of the optics. 

1250 The lens stage mode match to FC is now on fixed on VOPO because it shares cable with OPO

1307 Optical mounts 14-17 removed from table. The 2-inch optics on mount #17 is accidentally broken by a senior scientist due to the stress release of the setscrew. 

1311 Two PZT optical mounts are removed from table. 

1326 VOPO out

1330 lunch break

Images attached to this report
GRAVITES Optics 1550nm
epolini - posted 20:14, Tuesday 21 May 2024 (11646)
UMZI 23m characterization with laser current = 0.540 A

Today I tried to check if with the new laser current set (0.54 A), the frequency noise of the laser would have gone down to specs.

I characterized the UMZI 100m and the BMZI 40km.

Thorlabs UMZI 100m 

The fringes amplitude was fluctuating a lot and this is also evident in the error and correction signals shown in Fig. 1.

I tried then to isolate the spool a bit better by fully closing the metallic box and placing a rubber inside the box and below the spool. Myron helped me drilling some holes to allow the fiber to exit the box on both sides and cutting the rubber. We did that for both the boxes (100m and 200m spools). Fig. 2 show the interior of one of the boxes.

I made again the measurement and the amplitude of the fringes was still fluctuating... 

Thorlabs BMZI 40km

I adapted the setup in order to have a BMZI 40km. I removed the polarization controller and the splitter on the 'shorter arm'. I replaced the second PM BS with a SM BS. Input power of the BS1 = 0.4 mW. 

Fig. 3 shows the error and correction signal for this configuration. At high frequency we still notice the step behaviour due to the amplitude fluctuations. At lower frequency we see the peaks introduced by the coupling of acoustic noise.

The good news wrt the previous measurement campain is that we can lock the BMZI 40km and the UMZI 100m. The terrible news is that these amplitude fluctuations will prevent us from using this laser for the experiment. 

Images attached to this report
GRAVITES Optics 1550nm
epolini - posted 19:46, Tuesday 21 May 2024 (11645)
Faraday Isolator after NKT laser replaced

Today I replaced the HP fiber isolator (up to 2W) since it was damaged with a lower power one. This new one can take up to 300mW so we need to be careful not to increase the laser power above this threshold. 

I checked if the peak in the intensity noise (1kHz-tens kHz) was somehow caused or increased by such issue but it is not. The intensity peak is still there, as shown in the attached picture.

Images attached to this report
GRAVITES Optics 1550nm
epolini - posted 19:42, Tuesday 21 May 2024 (11644)
Thorlabs laser not operated in the single frequency region?

Yesterday, I took several measurements with the UMZI 23m with the Thorlabs laser and different laser currents. I summarize in the following table the important parameters for each configuration:

Laser current [A] V_DC output [mV] Vpkpk output [mV] VCO FM dev
0.548 880 270 5kHz
0.546 867 (875 after) 396 (300 after) 2kHz
0.544 877 (870 after) 180 (130 after) 2kHz
0.542 860  250 (388 after) 2kHz
0.540 850 (855 after) 270 (208 after) 2kHz
0.552 879 (868 after) 220 (200 after) 2kHz
0.550 (previous) 570 170 (130 after) 5kHz

The values *after* mean after the measurement to show that the amplitude of the fringes oscillates over time. Going lower in laser current, i.e. 0.539A, gives smaller fringes amplitude (difficult to lock) and going higher was introducing evident frequency noise visible in the fringes.

I also characterized the same loop with the NKT laser - UMZI 23m:

These values are constant after the measurement, the fringes DON'T FLUCTUATE with the NKT LASER (as it is supposed to be).

Fig. 1 shows the error and correction signals for differen laser currents for the Thorlabs laser. The orange curve is the NKT measurement. The dark green curve is the VCO noise calibrated in rad/rtHz for a frequency deviation = 2kHz. The curve with lower power noise seems the one corresponding to a laser current = 0.540A. 

Fig. 2 shows the referred laser input power for the cases discussed above, the VCO noise and the Thorlabs lasers frequency noise specs.

It seems that the laser frequency noise is still higher than what it is supposed to be. There are some structrures at lower frequencies that could be a good sign. 
I set the laser current = 0.540 A and I did some measurements shown in the following entries.

Images attached to this report
cSQZ General
jacques - posted 11:41, Tuesday 21 May 2024 (11643)
Solution to the SHG lock problems: demod phase. Other optimizations and polarization controller DC supply issue

[Malo, Jacques] Date: yesteraday.

After a discussion with Eric, the most probable cause of the bad SHG PDH error signal is the demod phase which hasn't been adjusted since I added the polarization controller on the input fiber path. 

Sure enough, I adjusted the "phase=60" parameter in the iq_shg.setup(...) function in the setIq() function of the lockShg file (see code snipet below). Then the error signal scan made sense (see first screenshot) and the lock could be aquired with the correct amount of IR.

# Set IQ modulation/demodulation
def setIq():
    # Names to simplify later function calls
    asg_dither = r_shg.asg1
    iq_shg = r_shg.iq0
    # Add a constant voltage to the output so we never drive the PZT with a
    # negative voltage
    #asg_dither.waveform = 'dc'
    #asg_dither.offset = 0.21 #0.21
    #asg_dither.output_direct = 'out1'
    #asg_dither.trigger_source = 'immediately'
    # Start the IQ modulation/demodulation
    iq_shg.setup(frequency=50e6, bandwidth=[1e6], gain=0.0, 
             phase=60, acbandwidth=1E6, amplitude= 0.4, # 0.2 , #min((1-asg_dither.offset),asg_dither.offset), 
             input ='in2', output_direct="out1", 
             output_signal='quadrature', quadrature_factor=1)

However, even with an optimized demod angle and the SHG loop closed, I noticed that the measured GR on the OPO reflection's PDH PD DC (the PD set vertically on the FI to the OPO), was lower by 20-30% than the peak of GR on this same PD when the SHG was scanned. Tuning the SHG temperature while the loop was closed did not seem to improve this. Therefore we only produced 23mW of GR out of the OPO instead of the 30mW measured last week. It is not clear to me where this discrepancy comes from. The input IR to the SHG was 60mW, same as measured last week.

We then tried to lock the OPO by tuning its demod phase. The PDH error signal scan seems skewed (see second picture) but after discussion with Eric, he suggests that this is due to the scan being too slow, such that the OPO suffers some thermal transient while scanning. Thus the error signal should look more symmetric if we scanned faster. This is yet to be tested. Nevertheless, the OPO could be locked with the current code and demod phase of -45. See code snipet below, which is the lock(...) function on the lockOPO python file. Notice that the argument which corresponds to the demod phase is now called "demod_phase" and is now an internal variable of the lock(...) function, which means that changing it live is tedious without copy-pasting the script in a Jupyter notebook/Jupyterlab.

#%%
def lock(tolerance = 10, stepsPerRes = 20, dual = True):
    start_time = time.time()
    ################################
    #           INPUTS             #
    ################################
    # PDH modulation amplitude [Volts]
    modulation_amplitude = 0.2
    # PDH modulation frequency [Hz]
    modulation_freq = 50e6
    # PDH Demodulation Phase [Degrees]
    demod_phase = -45
    # Scope Decimation (sets scanning speed)
    scopeDecimation = 16384 # max is 16384
    ################################
    #         END INPUTS           #
    ################################
    stop()
    #shortcuts
    iq_opo = r_opo.iq0

    print("Take a scope measurement (Green)")
    pztG, pdhG = measureErrorSignal(demod_phase = demod_phase, 
                                    modulation_freq = modulation_freq, 
                                    modulation_amplitude = modulation_amplitude,
                                    make_plot = False)

    print("Take a scope measurement (IR)")
    pztIr, pdIr = measureIrResonance(make_plot = False)

[...]

At this point, the intensity fluctuations of the GR made any CLF amplification fluctuate by about 50% of its max amplitude (measured on the PD to the HD path). So instead of continuing with the loops, we decided to recover the good temperature/OPO crystal setpoint with maximized NLG. Scanning the OPO, and with the CLF still at 3MHz, we could achieve an amplification factor of ~14 (1.06V on the DC power of the "PD to HD", when CLF was amplified, vs 76mV when the temperature was way off its optimal setpoint). This should convert to a good NLG (I forgot to put the CLF at 0Hz detuning to make the actual measurement). This was obtained with ~20mW of GR and tec.shgSetpoint(33.15), which is quite lower than the ~34°C that was previously found.

Finally we tried to set up the polarization controller for intensity stabilization but we suffered from a DC supply issue, where the controller was drawing too much current and too little voltage than its intended setpoint (max 1.2A and +-12V), as soon as the DC supply was turned on. It is not clear how one should fix this.

 

Next steps:

  1. scan the OPO faster in the PDH locking sequence to make the automatic slope fit more robust,
  2. redo a measurement of the NLG with the CLF offset set at 0,
  3. solve the issue of the polarization controller,
  4. Do a TF and spectrum measurement of the SHG and OPO loops to see where the 150Hz noise comes from
  5. Debug the HD board
Images attached to this report
GRAVITES Optics 1550nm
epolini - posted 16:41, Monday 20 May 2024 (11641)
Change Thorlabs laser current

[Eleonora, Eric]

Commands to change the laser current with pyserial:

>>> ser.write(b'write_param laser.current 0.555\r')
32
>>> out5=''
>>> while ser.inWaiting() > 0:
...     out5 += ser.read(1).decode()
...
>>> print(out5)

000: OK

>>> ser.write(b'read_param laser.current\r')
25
>>> out5=''
>>> while ser.inWaiting() > 0:
...     out5 += ser.read(1).decode()
...
>>> print(out5)

000: OK:0.555000,0x0000

cSQZ General
jacques - posted 21:00, Sunday 19 May 2024 - last comment - 11:04, Tuesday 21 May 2024(11640)
shg lock issue: bad pdh setpoint?

I tried locking the SHG using the usual python script. But the PDH error signal plotted by the red pitaya looks very strange and the subsequent lock is not good: I have only ~5mW of green coming out of the SHG (measured at the entrance of the OPO) and the PD in reflection of the OPO confirms this problem (virtually zero 532nm light reaching it when SHG is locked). See below for pictures of the scope when the SHG is scanned, after I execute the shg.lock() command and the plot of the PDH error signal.

On the scope, channel 1 is the SHG PZT driving voltage, channel 2 is the incident IR power on the SHG, channel 3 is the transmission PD of the SHG cavity (transmitted IR), channel 4 is the PD in reflection of the OPO (i.e. measures the generated GR out of the SHG).

The scan looks normal: there is a clear dual resonance between the IR and the GR at the expected crystal temperature, the alignment of the IR into the cavity is good.

The locked scope trace looks bad: the transmitted IR power (channel 3) is twice as low as the peak value when scanned, there is no GR detected (channel 4). Changing the temperature of the SHG does not make the GR peak appear on channel 4.

The PDH error signal looks bad: instead of a linear region around the expected resonance setpoint, we have a strange flat region.

It is not clear to me what's going on on the PDH lock that would explain this kind of error signal.

Images attached to this report
Comments related to this report
jacques - 11:04, Tuesday 21 May 2024 (11642)

Comparison of two SHG scans between sunday evening and monday afternoon. There is a slight nonlinearity as can be seen on the slope at the zero crossing.

Images attached to this comment
GRAVITES Optics 1550nm
epolini - posted 11:32, Friday 17 May 2024 (11638)
UMZI 23m Thorlabs Laser and Frequency Noise Investigation

In the past days I have been trying to measure the frequency noise of the new ULN15TK thorlabs laser, using the UMZI. I tried with 1km unbalance and it was too much for fringe wrapping issues as I mentioned in a past entry. I spliced and tried the 100m and 200m spools but that was also too much: the fringes were mostly hidden by frequency noise on the oscilloscope and that made it impossible to lock the UMZI. Moreover, with such unbalance, the fringes were oscillating in amplitude as well. I looked around for shorter fibers and I found some thorlabs P3-1064Y-FC and P3-980A-FC fibers of 2m and 5m used for 1064nm experiments, that are also good at 1550nm.

The final configuration that allowed me to lock the UMZI was with a total unbalance between the two arms of 23 meters: 

2 x 2m (AFW 1550nm patchcord - SM1-22-T) + 5m (P3-1064Y-FC) + 2m (P3-1064Y-FC) + 2 x 5m (P3-980A-FC) = 21m 

Then the long path counts also the AOM and the polarization controller and a BS that gives ~6m. The short path has 2 BS that gives ~4m. This gives 2m extra unbalance between the arms.

tau = 1.45 * 23m / 3e8 m/s ~ 1e-7 s

UMZI 23m Thorlabs Characterization:

VCO frequency deviation = 10kHz (VCO TF = 6.5 kHz/rad)

Error signal calibration = 75 mV/rad

DC output power 1811 = 570 mV

UGF ~ 22kHz (swept sine 200mV)

Fig. 1 shows the error and correction signals calibrated in rad/rtHz.

Fig 2 shows the OLTF of the loop.

The curves are not spoiled by acoustic peaks like the measurements taken with the NKT and this suggests that we could be limited by laser frequency noise.

UMZI 23m Thorlabs Noise Budget:

I also computed the noise budget. Fig. 3 shows the error signal and Fig. 4 the correction signal. I included the ULN15TK laser frequency noise fitted by the measurement provided with the laser. It seems like if the VCO noise limits the low frequency of the error signal but the correction signal is flat and seems like it could be limited by laser frequency noise. 

Thorlabs Laser Frequency Noise Comparison:

I decided to calibrate the correction signal into referred laser input noise and compare it with the ULN15TK provided date (fitted) and the DFB15TK datasheet (fitted).

To convert the correction into input noise, since the AOM in this case is on one arm and not at the entrance of the interferometer, I did:

asd_freq(omega) = 1/ (2 * pi * tau) * (1+G) * VCO / freq * 1 / G * asd_phi(omega)

where tau = n * L / c  with n = 1.45, L = 23m, c = 3e8 m/s; G = OLTF and asd_phi(omega) is the correction signal calibrated in rad/rtHz shown in Fig. 1.

Fig. 5 shows the comparison of the ULN15TK and DFB15TK laser frequency noise and the correction propagated to the input of the system. I did not correct for the OLTF in this case since the UGF is around 22kHz.

The laser frequency noise computed from the data is 10-100 time higher than the specs and lower than the model DFB15TK above 50Hz. 

I added the rms of the reconstructed input noise and I belive it to be around 1e6Hz - 2e6Hz. If I convert it into maximum differential arm length I get: DL = 16m - 32m, as I expected from the measurement I made. 

Why do we see this excess in noise? Are we sure that it is laser frequency noise? The fact that we don't see the peaks as in UMZI NKT in the audio band suggest that there is a noise that covers it and it can be luckly to be frequency nosie. Are there other sources we are not considering? I am sending the plot to the Thorlabs Technical support.

Images attached to this report
GRAVITES Optics 1550nm
epolini - posted 18:08, Thursday 16 May 2024 - last comment - 17:14, Friday 17 May 2024(11637)
UMZI NKT 1km to Measure NKT Frequency Noise

The other day I locked the UMZI 1km with the NKT, I took the Corr, Err, OLTF and I made a noise budget. Since I was supposed to be limited by frequency noise, I thought it was a good idea to re-measure it with a loop that is more reliable than the unbalanced Michelson. 

UMZI 1km NKT Loop Characterization:

VCO frequency deviation = 5kHz (VCO TF = 3.9 kHz/rad)

Error signal calibration = 104 mV/rad

DC output power 1811 = 200 mV

P0 (before BS1) = 139 uW

Pout = 9uW

UGF ~ 30kHz

Fig. 1 shows the error and correction signals calibrated in rad/rtHz.

Fig 2 shows the OLTF of the loop.

UMZI 1km NKT Noise Budget:

I also computed the noise budget. Fig. 3 shows the error signal and Fig. 4 the correction signal. I included the E15 NKT laser frequency noise fitted by the AEI data. The noise budget suggests that the laser frequency noise limits the high frequency.

NKT Laser Frequency Noise Comparison:

I decided to calibrate the correction signal into referred laser input noise and compare it with the X15 datasheet and E15 data.

To convert the correction into input noise, since the AOM in this case is on one arm and not at the entrance of the interferometer, I did:

asd_freq(omega) = 1/ (2 * pi * tau) * (1+G) * VCO / freq * 1 / G * asd_phi(omega)

where tau = n * L / c  with n = 1.45, L = 1km, c = 3e8 m/s; G = OLTF and asd_phi(omega) is the correction signal calibrated in rad/rtHz shown in Fig. 1.

Fig. 5 shows the comparison of the NKT X15 and E15 laser frequency noise and the correction propagated to the input of the system. I did not correct for the OLTF in this case since the UGF is around 30kHz.

The noise at high frequency is in agreement with both the X15 and E15 specs, while below 1kHz the system couples with other sources of noise difficult to measure. The bump between 1kHz and 10kHz could be due both to a cross-coupling of the excess amplitude noise already shown in the past or also to a bump that is present in the data that can be found on the NKT note (https://www.nktphotonics.com/wp-content/uploads/2022/01/frequency-stability-application-note-version-1-1-.pdf) at around 5kHz. In order to reduce the excess noise below 1kHz, we need to use the vibration insensitive spool that has been sent to re-wind with a 1km new sm28-ultra fiber and a decent acoustic isolation. 

Images attached to this report
Comments related to this report
epolini - 17:14, Friday 17 May 2024 (11639)

I also compared it with data acquired in March with the Michelson 1km unbalances interferometer used as pre-stabilization. Attached the comparison with the most recent data (none corrected by the OLTF). They are in agreement.

Images attached to this comment
GRAVITES Electronics
eoelker - posted 19:12, Wednesday 15 May 2024 (11636)
GRAVITES low power homodyne
I put together a fiber coupled homodyne for stabilizing the GRAVITES interferometer for low power operation.  Due to issues with Raman scattering, we will likely have to operate with ~100 nW of light at the output and good noise performance in the 10-100 mHz band.  With a transimpedance >= 1 MOhm, this should be no problem.  I modified one of Tomoki's aLIGO diagnostic homodyne boards (D1300671-v6) to work with low power/high transimpedance.  

Changes: I unstuffed the resonant paths, the DC lowpass was unstuffed and inductors replaced with 0 ohm resistors, U9 -> OPA828, R29 -> 1 MOhm (feedback cap also removed), D11/D14 -> C30617BFCH ingaas diode, FC fiber recepticle.  This diode has the same pin configuration as the ETX-500 , so it actually fits in one of the existing PD footprints on the board.  However it's base is larger than the ETX-500 and I had to cover the thru-holes for the high QE diodes on D11/D14 with tape to avoid a short to the diode case.  I have not yet attempted to increase the transimpedance gain of the PD1 and PD2 outputs.  They are currently useless for 100 nW input power. 

The noise performance looks sufficient. From about 0.5 Hz-10 kHz it is Johnson noise limited. Should be just shot noise limited for 100 nW at these frequencies.  At high frequencies, we see a roll-up in noise.  This does not seem to be due to the photodiode capacitance.  Their total capacitance is way too low (The fiber coupled diodes only have ~0.5 pF each).  Tomoki's board likely has several pF of stray capacitance, but this is also insufficient to explain what we're seeing based on my circuit noise budget.  Interestingly, we see the same noise when I remove the last 0 ohm resistor on the input path before U9.  This should get rid of almost all input capacitance, so the frequency dependence of the noise seems to be from the op amp itself.  I'm not certain, but this could be a roll-up in current noise at high frequencies, a common issue in low current noise JFET op-amps.  There is no plot of current noise vs frequency in the OPA828 datasheet, just a value at 1 kHz (my circuit noise budget currently assumes that the current noise is frequency independent).  We don't really care about the high frequency noise performance for GRAVITES, so this isn't an issue.  At low frequencies, the performance looks good: ~400 nV/rtHz at 100 mHz and ~2 uV/rtHz at 10 mHz.  

There's a bit of room to increase the transimpedance more if we'd like.  Haven't plugged this pd spectra into our interferometer noise budget, but I suspect it won't be our limiting factor at our desired measurement frequency (likely 10-100 mHz).  
Images attached to this report
cSQZ General
hudsonl - posted 17:11, Wednesday 15 May 2024 (11635)
CSQZ Powers
[Hudson, Jacques]

After switching back to the CSQZ's normal configuration, we have 60 mW IR out of the SHG pump fiber, 120 mW before the EOM, and 30 mW of green pumping the OPO. This is a 50% increase in IR power from before, likely due to cleaning the fibers.
cSQZ General
hudsonl - posted 16:59, Wednesday 15 May 2024 (11634)
Swapping CSQZ Between Homodyne Alignment and Squeezing Configurations
This post will detail the changes needed to switch the compact squeezer from a state useful for aligning to the homodyne to one useful for generating high levels of squeezing.

Note: The OPO is dual resonant with large NLG at about 34 celsius and dual resonant with low NLG at about 36.5 celsius.

[1] Unscrew the PM fiber connected to the fiber port's input with the EOM for the SHG and connect it to the input to the AOMs
[2] Unscrew the 90% pickoff going to the fiber polarization control paddles on the CLF path and connect this to the fiber bench
[3] Connect the output of the AOMs to the variable optical attenuator and fiber polarization controller on the CLF path
[4] Disconnect the 23dB signal amplifier after the RFPD for the OPO lock.
[5] Add an ND filter in front of the RFPD for the SHG lock. This should have a total attenuation of ND = 1.0
[6] Flip up ND filters in front of the CLF RFPD so they are no longer in the beam path
[7] Re-optimize waveplate orientation on SHG pump path and polarization paddles on CLF path. Before doing this, turn down the gain on the PD monitoring the rejected polarization out of the CLF fiber.
[8] Switch the csqzControl.py script to read in "lockOpo" as opo instead of "lockOpoLowPower"
GRAVITES Optics 1550nm
epolini - posted 22:36, Tuesday 14 May 2024 (11633)
New 100m and 200m fiber spliced and characterized

I received and spliced a 100m and a 200m SM28-ultra fibers from corning. The losses displaced on the splicer are 0.00dB for both of them. NB: when you splice be careful to make the fiber holder 250 touch the blue holders.

I characterized the losses using both the NKT and the Thorlabs laser (to double check due to laser amplitude noise/fluctuations).

NKT - 100 meters

P(before) = 54uW

P(after) = 52.8uW

Loss = 2.2% - Loss_dB = 0.098 dB

Thorlabs - 100 meters 

P(before) = 180uW

P(after) = 174uW

Loss = 3.3% - Loss_dB = 0.15 dB

NKT - 200 meters

P(before) = 54uW

P(after) = 48.5uW

Loss = 10% - Loss_dB = 0.47 dB

Thorlabs - 200 meters 

P(before) = 180uW

P(after) = 154uW

Loss = 14% - Loss_dB = 0.68 dB

These losses are way higher than expected. 

I inserted the spools into small boxes I found in Rai's office as shown in the attached photos.

Images attached to this report