Anthony Amorosi - posted 10:57, Thursday 19 September 2024 (11761)
HAMX-ISI control update

This takes over the updating of the HAMX-ISI control system done by Morgane Zeoli from Aug. 5 to Aug 23 after the CDS upgrade. 


•    Adding DQs channels:
T240 and ground STS2 sensors signals have been added in the model as DQ channels. The procedure is as follow:

1/ Kill the model by ssh-ing to m1hamx and using the following commands, in this order, 
rtcds stop m1isihamx    →   rtcds stop m1hpihamx   →  rtcds stop m1iophamx                
(killing the iop may not be necessary)

2/ From a machine with matlab 2019, enter the name of the channel to add as DQs in the comment window in the simulink .mdl file, followed with its sampling rate. Update the .mdl file.
Example: SENSCOR_GND_STS_X_MATCH_IN1_DQ 2048

3/ Restart the model with the following commands, in this order, 
rtcds start m1iophamx    →   rtcds start m1hpihamx   →  rtcds start m1isihamx                

4/ Reboot the front end by ssh-ing to m1fb0 and using the following command,
sudo reboot 

The STS2 signals can be accessed at the following name
M1:ISI-HAMX_SENSCOR_GND_STS_{X,Y,Z}_MATCH_IN1_DQ              , it include calibration
The T240 signals can be accessed at the following name
M1:HPI-HAMX_SENSCOR_{X,Y,Z}_IN1_DQ                                                 , it include calibration


•    Gains
Gains on some sensors have been noticed wrong. Some indications were:
- The unity gain frequency being 10-ish Hz instead of 30 Hz, filters have not being changed.
- Signals amplitude between GS13 and CPS after blending were not matching.

Transfer function from ISO_X_Exc in ISO_X_IN1 was showing a DC offset by a factor of 1/2.27. Assuming the CPS and actuators are calibrated, it has been decided to apply a gain of 2.27 on the signal of the GS13. This gain is added as “Gain_corr” in the M1:ISI_HAMX_GS13INF_{H1-3,V1-3} filterbank.

Accrodingly, the STS2 sensor on the ground for sensor correction also has a 17.93 gain correction switch in the  M1:ISI-HAMZ_SENS_COR_GND_STS_{X,Y,Z}_IIRHP filterbank under the name “Gain”.

Accordingly, the T240 semsor on the ISI table also has a 0.432 gain correction switch in the M1:HPI-HAMX_STSINF_B_{X,Y,Z} filterbank under the name “Gain_corr”.

With all these gains ON, the control are stable and with the expected unity gain frequency. The origin of these gains is still unknown and I would recommend carefully identifying the acquisition gain and ensuring that these are correct.

 
•    Tilt-coupling compensation
Morgane previously identified the coefficients that minimized the  X to RY coupling. These were implemented in the CPS_ALIGN matrix as X to {RX,RY} elements. These has been switched for {RX,RY} to X, as it makes a little more sense: i.e. X motion sensed by the CPS is sent to the X actuotor, bu also in RX and RY, instead of sending the the X actuator motion sensed in RX and RY. 

The values have been fined tuned, currently sitting at X to RX = -0.004 and X to RY = 0.000. 
Values updated in the /home/controls/Data/Populate_Hamx_ISI_MEDM_Screen.m script.

Y to RX and Y to RY have not yet been investigated.

•    Sensor correction:
 Sensor correction has been implemented and optimized to a certain level. 
The gain leading to the best performance was found to be 1.125. It has been implemented as the “Boost” control switch in the m1:ISI-HAMX_SENSCOR_GND_STS_{X,Y,Z}_MATCH filterbank.

Details of the measurement cam be found in the diaggui 24_08_30_Senscor.xml file in /home/controls/Data/2024_09_18_Anthony/, where the sensors signals has been recorded stored for different sensor correction gains.

References are as follow:
  0 to 11:  control off
12 to 23: control on, no sensor correction
24 to 35: sensor correction gain =  1.0
36 to 47:                                     =  1.5
48 to 59:                                     =  0.5
60 to 71:                                     = -1.0
71 to 83:                                     =  1.25
84 to 95:                                     =  1.125
currrent:                                     =  1.075

/!\ This was done with the 250 mHz blend. The ISI is running now with the 100 mHz blend, I am unsure of if the blend frequency could impact the optimization of the sensor correction gain. The ISO controllers in X,Y,Z have also been upgraded after this test.

•    ISO controllers:
It has been seen that the ISO filters (Controllers and boost) could use some improvement in X,Y,Z. New versions have been produced and are stored in the “Cont_4” and “Boost_4” switches in M1:ISI-HAMX_ISO_{X,Y,Z}.

Related to the gain issue mentioned above, the controller gains had to be tuned so the unity gain frequency measured on diaggui was matching the designed value of 30 Hz. The correction is directly implemented in the “Cont_4” switches, and the value was different for the three designs.

The measurements that served the design of the controllers can be found running the Load_OL_TF.m matlab script in /home/controls/Data/2024_09_18_Anthony.

An attempt of a v4 version of the controller for RY has also been produced, but the closed-loop system rings at a frequency of 40-ish Hz probably due to a wrong notch in the controller desgin. This version is aborted for now.

•    Feedforward
Transfer function from feedforward sensors to (ST 0 L4Cs) to the platform have been identified. 

The set of measurement can be found in /home/controls/Data/2024_09_18_Anthony/24_09_09_FF_TFID.xml
It contains the set of transfer function from the M1:ISI-HAMX_FF_{X,Y,Z}_EXC drive to M1:ISI-HAMX_BLND_GS13{X,Y,Z}_IN1 and from M1:ISI-HAMX_FF_{X,Y,Z}_IN1 to M1:ISI-HAMX_BLND_GS13{X,Y,Z}_IN1 with no drive.

The feedforward transfer function is obtained by taking the ration of these two quantities. The /home/controls/Data/2024_09_18_Anthony/FeedForward_FilterDesign.m matlab script loads the data and attempt a fit on the FF transfer function.

The Y and Z filters have been implemented in CEDM in the M1:ISI-HAMX_FF_{X,Y,Z} filterbank under the “Cont_1” control switch, performance are not satisfying as the phase of the fitted filter only matches the experimetnal phase in a limited frequency range from 5 to 10 Hz.

The main assumption that may explain the poor performance I that the HPI stage is not operating (not damped), this causes a strong resonance peak at 10 Hz making the fitting of the FF transfer function difficult. We might consider locking the HPI stage for now and retake the measurement.


•    Set of ISI ASDs and TF
A set of reference ASDs have been taken with the current design of the control filters. The measurements can be found in /home/controls/Data/2024_09_18_Anthony/24_09_09_RefASD.xml

The references are as follow:
00 to 02: Control OFF
03 to 05: controller version 3, sensor correction, 250 mHz blend.
15 to 17: controller version 4, no sensor correction, 100 mHz blend.
18 to 20: controller version 4, +sensor correction, 100 mHz blend.

The /home/controls/Data/2024_09_18_Anthony/ISI_Perf.m matlab script loads data of the ISI with different isolation levels and plots spectrum and transmissible transfer functions with the nost recent designs of the controllers/blends/gains.

The /home/controls/Data/2024_09_18_Anthony/ISI_Spectrums.m matlab script allows to fetch data at a specied GPS time and plot ASDs.


•    Vacuum
The chilling system for the vacuum pumps has been repaired and the pumps are now operating normally. The rough and turbo pumped has been turned on few times for a couple of hours in the week Sept 09 – Sept 15 and 1e-2 Torr vacuum has been reached. No strong evidence of leak was observed.

During the first pump down of the system however, it was observed that the ISI control loop went unstable at a frequency of 68 Hz. It was suspected that the removal of air made a rocking mode very high Q, which is driving the ISI into instability. The door has been opened and the T240 vacuum pod, which was initially clamped on the ISI, has been placed on 3 thin shims to make it kinematic. This assumption turned out right as the 68 Hz mode disappeared after closing the tank and pulling vacuum again, and the control loop were stable.

•    Next steps
For the next steps, I would suggest:
- closing investigating the cause of the gains mismatch, and ensuring that the corrections that were applied were correct;
- Have a second look into the tilt-compensation matrix, identify the couplings in Y direction;
- Redesign the version 4 of the RX, RY and RZ controllers;
- Lock down the HPI stage and redesign the feedforward filters;
- Have the Turn_on. M matlab script working.
 

Images attached to this report
Heisenberg IFO General
jacques - posted 00:04, Thursday 19 September 2024 (11760)
Breadboard SQZ's LO locked using the pair of AOM instead of the PZT

Jacques, Hudson, Melissa

We previously had issues on the CLF-LO loop because we were running out of range with the PZT (3um of actuation range), due to the natural drifts of the optical paths between SQZ and LO. Furthermore the PZT's actuation motion was affecting the alignment of the LO. To mitigate the range of the drifts, Hudson removed the green fiber but it did not seem to significantly improve the range issue.

We then used a pair of fiber AOMs (from the old cSQZ "fiber box") in place of the PZT to close this loop. We swapped the 90/10 BS for CLF/LO, with a 50/50 BS. We have to reoptimize the CLF's incident polarization, as well as the LO's as the fiber AOMs have an efficiency which is polarization dependent. We used the moglabs's channels 1 and 2 to send two signals at 150MHz to upshift and downshift the LO, with actuation (freq mod) on ch 1. This means that the cSQZ no longer has a RF source to drive its AOMs.

Today, I sucessfully locked the CLF-LO loop of the BSQZ using the setup consisting of two AOMs on the LO path and actuation on the frequency modulation. For the non-CLF related locks, I used the jupyter notebook "BSQZ general script". I used the jupyter notebook "debug AOM lock" to close the CLF loops, this is a copy of "clf2.py" with minor modifications. After optimization it will be copy pasted back into clf2.py. See picture attached for the TF and noise spectrum of this new loop. The main change was to increase the gain to ~500.

Using this I was able to get 4.5-5dB of SQZ vs 13 dB of aSQZ. I verified that the NLG temp was optimal and realigned the HD2 visibility to be 96.8%. So I don't really understand why we can't get more squeezing right now.

The next goal is to fine-tune this new loop and do a CLF calibration sweep on the BSQZ and compare it to the model. It would also be good to calibrate the noise spectrum in phase noise units.

Images attached to this report
cSQZ General
jacques - posted 23:47, Wednesday 18 September 2024 (11759)
Shot noise clearance of both HDs using RP acquisition

I have now connected both HDs to the acquisition RPs and have taken shot noise vs electronic noise spectra. They are taken with the usual 2.6mW of LO on HD1 and 1.1mW of LO on HD2. I have properly isolated the HD board using baluns and 50 Ohm adapters. 

They are here.

Conclusion: on both HDs, there is about 20dB (21dB for HD1, 19 dB for HD2) of clearance below 1MHz. The electronic noise has some periodic pulses. It's not clear whether they come from the environment (we have already put all the electric plugs onto the same set of wall plugs, but it is not isolated from the rest of the grid), or from the setup itself.

I used the jupyter notebook "CLF calibration acquisition" to acquire the data and "CLF calibration data/CLF calibration data analysis" to analyze and plot the data.

Non-image files attached to this report
Heisenberg IFO General
jacques - posted 21:37, Monday 16 September 2024 (11758)
Shot noise clearance on HD1 using RP acquisition

After discussion, the extra noise that was observed on the RP's acquisitions for the HD1 probably came from ground loops between the HD1 and the RPs. Furthermore i noticed that the inputs of the RP were not terminated by 50 Ohm. I then proceeded to

1/ Add baluns on RF2 (f_low = 200kHz) and DC_OUT (f_low = 10kHz) to isolate the HD1 from the rest of the setup.

2/ Terminate the two inputs of the RP using SMA tees and SMA terminators.

I then put a 1.9MHz lowpass on the DC_OUT and an SR560 preamp. It's a pure gain of 50x. (max 3dB bandwidth 1MHz). I then acquired data on the RP, sampling at a decimation of 32 (i.e. 3.9MHz). I recorded the electronic noise (without LO) vs shot noise (with 2.6mW of LO). See plot here.

Conclusion: with this setup, we have >20 dB of shot noise clearance below 1MHz.

Non-image files attached to this report
cSQZ General
jacques - posted 19:24, Wednesday 11 September 2024 - last comment - 22:06, Wednesday 11 September 2024(11755)
Recovery of cSQZ to 7dB SQZ

Because of the execrable temperature changes which have plagued our lab for the past 10 days (It was 21.2°C on the cSQZ table at the beginning of last week, then it became 23.4°C at the end of last week, then it became 24.8°C today), the cSQZ got misaligned. I had to waste two days to realign it, just to get back to where we were two weeks ago.

I first started with the alignement of the input laser beam into the SHG. 

- Symptom: When scanning the SHG and comparing the peak heights between FSR, we see that they are not constant (envelope effect). Furthermore, we are not able to minimize the relative peak height between of 02 (first mismatch mode) divided by the fundamental 00, below 40%. See here.

- Solution: Tune the dichroic in the SHG (tip tilt mount). This reduces the 02 height. It does not get rid of the envelope effect, which is likely due to some clipping on the input SHG mirror (we can see that the beam arrives close to the edge of the mirror and that there are two spots inside of the cavity, between the input mirror and the dichroic). See here for the decrease in mode mismatch on the highest mode.

Conclusion: With this tuned, and sending 60mW of IR, I measured 26mW of GR at the output of the SHG.

- Symptom: sometimes there is less IR power at the input of the cSQZ (drops from 60mW to 50mW).

- Solution: Tune the free-space to fiber coupler on the lightwave's board and/or the polarization paddles.

Future work: should try to reduce the 1.7kHz noise on the SHG's IR and GR when it's locked. This can cause extra phase noise on the SQZ. I tried adding an iir notch on the SHG RP at 1.7kHz but it did not help reduce the oscillations. See here for the error signal spectrum.

Since the dichroic moved, the path from the SHG to the OPO got misaligned. I had to realign it, first by making sure that the reflected GR is overlapped with the incident GR (i.e. obtain a signal on the PD GR REFL DC, then looking at both the GR Trans PD and camera. Doing so I was able to recover the same scan shape of the GR into the OPO (i.e. misalignement completely cancelled, 30% relative peak height for 02).

Future work: the lock of the OPO is not so stable currently (possibly because Hudson is working on the table in parallel). Optimize its stability.

I then closed the CLF loops and got 7dB of SQZ for >13dB of aSQZ on the SR. See photo.

Images attached to this report
Comments related to this report
jacques - 22:06, Wednesday 11 September 2024 (11756)

It now looks like the GR and IR TRANS powers are extremely stable over ~1h (no noticable fluctuations on the scope), so much so that the power stabilization loop can be engaged on the GR TR (OPO) and this does not stabilize the CLF amplification. This means that the SQZ stabilization issue really comes from the temperature controller. It is possible that there is some thermal hysteresis/bistability which prevents the OPO oven loop from locking at the correct setpoint.

Heisenberg IFO General
hudsonl - posted 16:49, Wednesday 11 September 2024 (11754)
Tune Up Breadboard OPO CLF Path
We noticed that several of the optics on the path from the IR fiber coupler to the breadboard OPO were loose. We noticed this after one of these optics was bumped and the path was misaligned. I went and tightened all of the optics on this path, re-aligned the beam through the AOMs, and re-aligned the beam into the OPO. 

To align the beam to the OPO, I sent green light through the OPO's input/output coupler. I then aligned the CLF beam reflected off of the CLF's input mirror to the green beam. This gave me clear cavity flashes on a camera and clear transmission peaks on a transmission PD. From there, it was straightforward to fine-tune the alignment to the cavity. Currently, the mode matching is nearly 100% for the 532 resonances corresponding to the lowest piezo drive voltages (the mode matching degrades for larger PZT voltages as the mirror is not actuated exactly parallel to the cavity axis).

For the AOMs, we have 12 mW of incident IR power before the first AOM and about 6.3 mW in the first diffracted mode after the second AOM, so the efficiency is about 72% per AOM.
Heisenberg IFO General
jacques - posted 16:02, Sunday 08 September 2024 (11753)
AM modulator installed and intensity stabilization loop implemented

(Saturday)

Following the previous logbook entry, I installed a New Focus 4104 amplitude modulator between a half-wave plate and a PBS on the laser path, before it being coupled to the fiber.

The AM's manual is here. Without any other optics, the AM is actually a voltage-dependent waveplate. It is converted to an intensity modulator by placing it between two polarizers (the first one to align the polarization on the AM's prefered axis, the second one to do the intensity filtering). I use a half-wave plate as the first polarizer and a PBS as the second polarizer, which amounts to the same. 

I then realigned the path to the fiber coupler and obtained the same coupling ratio as before (between 170mW and 200mW arriving on the Ubench). 

I then tested that the setup worked properly to control the laser power, by plugging REDPITAYA27.MIT.EDU's "out1" port to a Thorlabs PZT driver (the +-1V range of out1 gets converted to +-15V, which is enough for our stabilization range) then to the AM, then I took the PD on the cSQZ board after the first PBS to monitor the power and plugged it to the RP's "in1". I plot the normalized power vs RP's output voltage here. I see that the AM's response (W/V) is actually linear in the range of the RP, and that it has enough range (6% of the total power) to actuate the expected power fluctuations.

The control loop is then closed on the fluctuations of the PD after the PBS, and the results are plotted and compared using a scope acquisition over several minutes, as shown here. After computation of the normalized (unitless) standard deviations when the loop is open vs when it is closed, I obtain 5.2e-3 when the loop is open vs 1.65e-3 when the loop is closed, thus showing that the AM loop indeed manages to reduce the power drifts.

The PID parameters for this PD are P = 20 and I = 100. The RP did not saturate over the ~10 minutes this loop was closed.

After these tests, I unfortunately realized that the PD after the PBS was not well positionned in the path of the IR beam and thus most likely received a stray beam. Subsequent tests can be directly done by choosing another PD to close the loop on such as the trans GR after the OPO, in order to directly actuate on the power that matters for squeezing production.

This scheme can also be implemented on the polarization controller instead of the AM, if we observe that the SQZ generation drifts are coming from the cSQZ setup itself and not the main laser, thus not disturbing the power of the LO fields.

Images attached to this report
Non-image files attached to this report
cSQZ General
jacques - posted 01:45, Sunday 08 September 2024 (11752)
Investigating SQZ drifts

(Friday)

It is currently impossible to acquire quality data on the compact squeezer because all the monitor PDs show that the powers (incident IR, GR, CLF) are drifting too much and too quickly (timescale 1 to 10s). This means that we cannot have a squeezing measurement which is stable for more than a few minutes. We have to constantly readjust the OPO temperature and recalibrate the CLF's resonance.

To quantify this I took scope traces of different monitor PDs along the path of the cSQZ.

- Ch 1: I plugged a PD on the fiber which is going to the VOA for the MZI LO and monitored the power. This should be an indication of the power at the level of the lightwave's fiber coupler, as well as the power that is arriving on the cSQZ before the PBS. 

- Ch 2: I plugged a PD on the fiber which is going to the PD on the cSQZ board after the first PBS. This is an indication of the power going into the SHG.

- Ch 3: PD of the GR in transmission of the OPO

- Ch 4: PD on the SQZ path after the OPO (monitors CLF amplification).

I took two batches of data over >400s each time, montioring either all four channels, or the first two channels (with a higher internal gain on Ch 2 to increase the SNR on the scope). I subtracted the DC biases of all photodiodes. I then apply a lowpass (butterworth) of cutoff frequency 0.5Hz to further filter the noise.

I obtain the following curves. I see (and I can compute) that all these channels are correlated, which leads me to believe that there are effective power drifts of the NPRO of the order of several percents over minutes. This in turn changes the power of produced GR, which changes the OPO's temperature control setpoint (recall that this needs to be controlled down to 0.02°C), and thus leads to CLF amplification fluctuations and SQZ drifts.

Another cause of SQZ drift could be the ambiant temperature which has increased by 2°C over two days, going from 21 to 23°C on the optical table. This has plagued us since May and still hasn't been reliably fixed, which is unacceptable.

Non-image files attached to this report
Heisenberg IFO General
hudsonl - posted 17:17, Thursday 05 September 2024 - last comment - 09:57, Friday 06 September 2024(11750)
Tuning Up Breadboard OPO PDH Lock
[Hudson, Melissa]

We tuned up the breadboard OPO's PDH lock. It now has a UGF of about 1 kHz and a phase margin of 55 degrees. We removed some analog electronics and the loop is now entirely controlled by a red pitaya and a -10x inverting amplifier. A digital filter notches out the first piezo resonance. The loop now has an integrator for improved low frequency gain.
Images attached to this report
Comments related to this report
hudsonl - 09:57, Friday 06 September 2024 (11751)
In order to notch out the piezo resonance at 8 kHz using the red pitaya's IIR module, it was important to test the transfer function of this module using another red pitaya connected to a SR785 and to adjust the "loops" parameter to set the decimation properly for an audio-band notch filter.
RetroSense General
evanhall - posted 12:10, Sunday 01 September 2024 (11749)
RetroSense v2 with SFH 4851 LED

I outfitted a RetroSense v2 (MIT:11483) with an SFH 4851 (drop-in replacement for OP232/OP132, but supposedly with greater total power output and central intensity). The LED is given about 30 mA of forward current, but is directed through a vented 1/4–20 screw (vent diameter 1.6 mm, vent length probably about 10 mm). This results in about 0.24 mA total photocurrent, and the displacement spectrum see in the attachment. The transimpedance is 25.5 kΩ and the clock frequency is 30.0 kHz. The sensitivity is about 0.6 mA/mm, implying a characteristic spot radius of about 1.2 mm (not sure how that gels with the vent diameter, but the light seems to enter and exit the retroreflector cleanly). This was optimized by varying the retroreflector distance from the LED/QPD with the third axis of the translation stage.

The coupling of LED intensity noise into the displacement readout at a static displacement offset is totally consistent with the theoretical expectation. The SFH 4851 has a RIN of about 5×10−7/Hz1/2 at 1 Hz (MIT:11727).

The measured displacement noise below ~3 Hz is not totally stationary and has some contributions from vibration and from air currents (enclosing the setup in a plastic bin reduced currents to the level seen here). The setup is not vibrationally isolated but is roughly on a three-point contact with the table.

The retroreflector is still an f = 15 mm lens mounted about 15 mm from a flat mirror, both 1/2″ diameter.

Non-image files attached to this report
Heisenberg IFO General
masaya - posted 13:08, Saturday 31 August 2024 (11748)
Mach-Zehnder Interferometer Control
Mach-Zehnder Interferometer for Heisenberg limit scaling by two squeezed states is locked with 1 kHz UGF using 1% pick off(1 % Reflection) Beamsplitters and the high gain DC differential receivers.
Non-image files attached to this report
Heisenberg IFO General
masaya - posted 12:55, Saturday 31 August 2024 (11747)
Two phases of squeezed fields are successfully controlled through Local Oscillator
Two squeezed fields are controlled successfully controlled. Transfer functions are shown in the attached phohos. UGFs are 1 kHz for both.
Input power is about 1 uW from each.

Regarding optical setups are available in the post: https://emvogil-3.mit.edu/lasti/index.php?callRep=11725
Images attached to this report
Non-image files attached to this report
Heisenberg IFO General
Malo - posted 06:08, Thursday 29 August 2024 (11746)
Latest HD plots and data uploaded in dropbox
I've uploaded the plots and data of the HD boards M1, M2, M3 taken with the Agilent in the dropbox. Data is missing on M1 RF2 (I had changed a component) yet I had checked directly on the Agilent that this data was similar to the other boards. As for board J, a lot needs to be resoldiered and data should be retaken since I haven’t followed every change and I guess some components differences can be the cause of a sharper drop in transfer function. As for M4, it consumes 0.15A instead of 0.09A; I haven't tried to debug it even roughly but this might be a quicker way to have another board than building another one from scratch.

As for the data files and their titles: 
-they refer to RF output ports when "RF" is written; to DC OUT (main) port when not.
-some values of incident light power is written for the transfer functions yet the spot was larger than the photodiodes so we should approximately divide this power by a factor 10^3
-the transfer functions were done with one of the PD inputs only, cf a previous logbook post on the symmetry in the PD inputs of these boards
Heisenberg IFO General
hudsonl - posted 17:33, Wednesday 28 August 2024 (11745)
Homodynes Oscillate at about 19 MHz with More than 3 mW of Incident Power
[Hudson, Jacques]

The homodyne detectors develop a strong oscillation at 19 MHz if there's more than 19 MHz of incident LO power. We balanced the homodyne detector such that there was no more than 5 mV of DC signal on the DC differential path and still saw this strong oscillation when using a strong LO beam. This problem is solved by turning down the LO power.
Heisenberg IFO General
hudsonl - posted 21:46, Tuesday 27 August 2024 (11744)
Spectrum Acquisition with Red Pitaya Validation
I tested the planned data acquisition setup for the homodyne spectra in the Heisenberg interferometer setup. I connected the output of the compact squeezer's homodyne's DC differential port to the input of a 9 MHz bandwidth -25x amplifier circuit and then the output of this circuit to the input of a red pitaya. The amplifier is needed since the homodyne's output voltage noise is about 23 nV/rt(Hz), whereas the red pitaya's input-referred ADC noise is a few hundred nV/rt(Hz). 

With this electronics chain, I measured the dark noise (orange curve) and shot noise (green curve) of the homodyne. The blue curve shows the red pitaya's ADC noise with its input terminated by a 50 ohm SMA terminator. The mean values of these spectra between 100 and 700 kHz are 
ADC noise: -139.8 dB
Dark noise: -126.7 dB
Shot noise (0.8 mW): -109.8
This gives a dark noise clearance of 16.9 dB. The compact squeezer's local oscillator can be adjusted up to 5 mW, which would give an extra 8 dB of dark noise clearance, such that the total clearance between shot noise and dark noise is about 25 dB in this frequency range. This should be sufficient for the measurements we want to make.

Note: The average values computed above were computed by averaging the PSDs in this frequency range in units of V^2/Hz and then converting these averaged values to dB units. This gives a different average value than converting to dB units first and then averaging. I believe the choice made here is the correct one: for averaging, variances should be summed, not ASDs or the logarithms of variances.
Images attached to this report
Heisenberg IFO General
hudsonl - posted 17:28, Tuesday 27 August 2024 (11743)
High Bandwidth Analog Preamp Circuit
I soldered up a -24x gain (nominal) analog preamplifier circuit using an AD829 op amp. I configured this as suggested in fig. 36 of the data sheet: https://www.analog.com/media/en/technical-documentation/data-sheets/AD829.pdf with a -24x small signal gain. With the agilent, I measured a gain of 24 dB at low frequencies, but the Agilent has a 50 ohm input whereas this amp will eventually be connected to a high impedance red pitaya input, so I expect the true gain is 27 dB, (-23x amplification) which is consistent with the design and resistor tolerances.

The circuit's 3 dB bandwidth is 9 MHz, this is lower than the 40 MHz predicted by the op amp's data sheet, but I attribute this difference to the fact that I have wired the circuit up on a test board with through-hole components instead of with surface mount components on a PCB. At 4 MHz, the gain is 23.45 dB, and overall this circuit is sufficiently fast to serve as a preamplifier between the homodyne and red pitaya.
Images attached to this report
Heisenberg IFO General
hudsonl - posted 13:35, Sunday 25 August 2024 (11742)
Homodyne RF Resonant Frequencies
The broadband homodyne detectors have resonant frequencies of about 8.9 MHz on the RF2 port and 13.7 MHz on the RF1 port (see attached plot). We should use these frequencies for the two CLF tones.

This data was measured on the PCB labeled M3 using high QE diodes and while it was in the plastic enclosure. Measured transfer functions on the M1 PCB were similar (the resonance locations were in about the same place). These transfer functions were measured using the Agilent network analyzer, the Rio fiber laser, a fiber amplidude modulator, and a few uW of optical power on one of the two homodyne diodes. I used a switchable attenuator (one of the ones on a rotatable mount with various attenuation levels) to control the incident optical power. To ensure the circuit was not saturated, I looked at the DC differential path on an oscilloscope and ensured its DC value was not more than several hundred mV.
Images attached to this report
cSQZ General
hudsonl - posted 13:34, Sunday 25 August 2024 (11741)
CSQZ Switch Between Alignment and Squeezing Modes
Here are the steps to switch the compact squueezer between alignment and squeezing modes. The whole process should take about 30 minutes and only requires swapping two fibers, which should both now be easily accessible.

To go from "squeezing" to "alignment":
(1) Swap the 90% output from the fiber bench with the EOM for the SHG to the input of the polarization paddles on the CLF path. Turn off the AOMs on the CLF path and cover the output of the second AOM with a fiber cap.
(2) Check the DC power on the OPO reflection RFPD for the CLF 2f signal. The PD saturates at 10 V and ND filter should be added as needed to prevent saturation.
(3) Connect OUT 2 of the RP 28 red pitaya to the J8 pin on the analog 10x amplifier
(4) Send the output of the CLF RFPD through a DC block and then to the input of RP 28 instead of to the bias tee
(5) Replace the 60k resistor in series with the OPO mirror's PZT with a barrel

To swap from "alignment" to "squeezing":
(1) Replace the barrel in series with the OPO mirror's PZT with a 60k resistor
(2) Send the output of the CLF RFPD to the bias tee instead of through the DC block and the input of RP 28
(3) Connect the output of the OPO control red pitaya to the J8 pin on the analog 10x amplifier
(4) Swap the 90% output from the polarization paddles on the CLF fiber path to the fiber bench with the EOM for the SHG. Clean the fiber at the output of the AOMs and connect this to the polarization paddles.
(5) Remove any unneeded attenuators from the CLF 2f RFPD. Use the polarization paddles to maximize CLF power, then adjust the RF level out of the Moglabs RF driver to attenuate the CLF power to about the right level and then use the polarization paddles to fine-tune the incident CLF power. Note: this could instead be done with a VOA if we want to add one.