Displaying reports 401-420 of 1025.Go to page Start 17 18 19 20 21 22 23 24 25 End
Reports until 13:29, Friday 20 December 2024
TIR General
evanhall - posted 13:29, Friday 20 December 2024 - last comment - 17:26, Monday 13 January 2025(11844)
Notes for assembling picomotor rig

New Focus 8853 needs ball bearings for kinematic contact, in addition to custom mount.

New Focus 8742 needs new fuse (Littelfuse 229) and a new USB micro-B surface mount connector, which has been inexplicably ripped off the board (candidate replacement on Digikey: UJ2-MIBH-G-SMT-TR). Momentarily bypassing the fuse, the controller lights up but I did not attempt to drive it. The software installs and at least opens on Windows 10. We need a 3-to-1 picomotor concentrator cable (NF8725) unless we can find one around.

Comments related to this report
evanhall - 17:26, Monday 13 January 2025 (11863)

The kinematic mount fits fine into the picomotor portion of the mirror mount with 0.25″ steel balls.

The USB connection on the picomotor driver is not repairable and we'll have to use the ethernet port instead with a switch.

Images attached to this comment
Non-image files attached to this comment
General
evanhall - posted 16:38, Tuesday 17 December 2024 - last comment - 18:05, Monday 13 January 2025(11842)
Interference of Koheras and Teraxion lasers

Sasha, Evan

We are trying to measure the frequency noise of a new Teraxion laser by beating it against the 1550 nm Koheras used by Gravites.

We started with 1.9 mW from the Koheras and 1.8 mW from the Teraxion. We interfered them with a PM 50/50 splitter and found 1.3 mW at each of the two outputs.

We then started scanning the Koheras wavelength while watching for the beat note on an rf spectrum analyzer. At the start, the Koheras was at an offset of 129.6 pm relative to 1550.12 nm. We tuned to 1550.2924 nm (offset of 172.6 pm) and found the beat note at 58 MHz. The nominal frequency of the Teraxion is 1550.27 nm.

The dc out of the NF1811 was 1.5 V, which was enough to totally saturate the rf output with the beat note present. We steered the beam almost completely off the photodiode in order to recover a linear beat note waveform.

We set up an rf frequency synthesizer (IFR 2026) and attempted to make a PLL feeding back to the synthesizer frequency, but it would not lock even at an FM deviation of 100 kHz. We may try to instead lock the lasers to each other.

Comments related to this report
evanhall - 17:01, Wednesday 18 December 2024 (11843)

We are now trying to lock the PLL by feeding back to the Teraxion.

Enabling the modulation input, even when 50 Ω terminated, kicked the Teraxion wavelength by about 7 pm (we tuned the Koheras down from 173 pm of offset relative to 1550.12 nm down to more like 165).

We mixed the beat note (around 80 MHz) with the Aeroflex output using a ZRPD-1+ and a 5 MHz low-pass filter. The LO and RF were both about 7 dBm. This went into an LB1005 with a gain of -20 dB and a PI corner of 1 kHz, then into a series of rf attenuators amounting to -34 dB, then into the Teraxion modulation port. The Teraxion test report gives 5.0 GHz of modulation for a 5 Vpp excitation.

The PLL locked with a bandwidth of 85 kHz and about 60 degrees of phase margin. We measured the spectra of the control and error signals and data analysis will be forthcoming.

evanhall - 18:05, Monday 13 January 2025 (11864)

Fiber collimator: TC06APC-1064 (wrong wavelength)

Focusing lens onto NF1811: LB1901-C 75 mm

PM fiber beamsplitter: AFW PFC-15-2-50-FB

General
evanhall - posted 16:40, Friday 13 December 2024 - last comment - 13:31, Tuesday 17 December 2024(11839)
Teraxion laser initialization

Installed Teraxion LXM-S laser on breadboard in green lab. (LXM-S1550.00-040-03-I1-W0, serial number LXM533725, LIGO-S2401070)

Running the laser requires the installation of some software libraries and a National Instruments based control software with GUI. (This requred first installing the Microsoft Visual C++ Redistributable codebase in both x64 and x86.) The software runs on Windows 10 (maybe also 11, have not checked).

The measured output power is 42 mW.

The manufacturer confirmed there is internal isolation of at least 40 dB (this is probably what the test report means by return loss, which is measured at 60 dB). The interlock connector is MMCX series.

Comments related to this report
evanhall - 20:45, Monday 16 December 2024 (11840)

I measured the RIN by collimating the light with a 15 dB attenuator onto a PDA20CS.

It is probably better to remeasure with a custom photodiode, but that will require some retooling to get the correct transimpedance gain.

Non-image files attached to this comment
evanhall - 13:31, Tuesday 17 December 2024 (11841)

I measured this with a custom low-noise photodiode with a transimpedance of 2.4 kΩ. (https://dcc.ligo.org/LIGO-D2400261)

I am not sure why the RIN is 10× higher below 10 Hz compared to the PDA20CS measurement.

Non-image files attached to this comment
TIR General
evanhall - posted 13:22, Friday 13 December 2024 (11838)
1550 nm phase modulator test report

From Optilab

Non-image files attached to this report
SQZ Site VOPO
lisabar - posted 13:01, Friday 13 December 2024 (11837)
LIGO SQZ VOPO components in NW22 clean area

Slawek, Lisa

The components needed to glue the mirorrs of the LIGO VOPO cavities are now in the NW22 lab, in the clean space area that Slawek has prepared close to the CTN experiment.

The gluing procedures that we are going to review and update are: LIGO-E1800010

Historical notes:

  • Glue on the back of the mirror motivated the switch back to EP30-2 but Slawek has a vision for how to fix that problem so we have more consistent gluing results.
Images attached to this report
Heisenberg IFO General
hudsonl - posted 18:41, Thursday 12 December 2024 (11836)
Automatic Compact Squeezer Squeezed State Tomography
After getting the data acquisition set up and running relatively quickly on a red pitaya, we can systematically measure the squeezing level as a function of LO demod phase. This will be essential to locking the homodynes on the correct phases for the full experiment. The following scan measured spectra at 100 data points evenly spaced between 0 and 180 degrees and took 542 seconds = 9 minutes. If possible, we would like to speed up this process so that we are confident that the state in the interferometer is stable over the duration of our measurement. This stability should itself be measured carefully.

This measurement used the damaged compact squeezer homodyne after verifying that its response is still linear.
Images attached to this report
Facilities General
evanhall - posted 14:19, Thursday 12 December 2024 - last comment - 19:28, Friday 10 January 2025(11834)
Particle counts in NW22

Particle counts in three areas in NW22 are as follows, in units of 1/m3. Each is the average of three measurements using a Lighthouse Handheld 3013. This is with one person (me) standing about 2 m from the counter.

  Room CTN ADBC ADBC "night mode"
>0.3 µm 15 600 4300 6300 33 000
>0.5 µm 2900 890 970 6800
>5 µm 0 0 0 0

Can't really explain why ADBC night mode is so horrible. I think from the TUI documentation this is supposed to be a 25% reduction in the amount of blowing, by some measure.

We do not have readings at the >1.0 µm threshold, but the data we do have suggest the two clean rooms (on their "normal" setting, when the LEDs read 02 and ––) are better than class 100, worse than class 10.

Images attached to this report
Comments related to this report
evanhall - 18:06, Wednesday 08 January 2025 (11856)

Sławek, Evan

We measured the flow bench and some other areas with a different particle counter (Airy brand). Counts in 1/m3. The CTN clean room measurement was directly under one of the HEPA filters.

  Room Flow bench Room ADBC (off) CTN
>0.3 µm 9200 0 6400 13400 0
>1.0 µm 6700 0 1060 5300 0
>5 µm 3500 0 710 2500 0

 

evanhall - 19:28, Friday 10 January 2025 (11859)

I vacuumed in and around the flow bench's blower cabinet. The blowers are surrounded by porous black foam, which may be casting off black soot into the blowers. These should be replacted with proper pleated HVAC filters. We need two 20×24×1 filters and four 16×20×1 filters. Some effort is probably also needed to wipe out as much of the soot as possible.

LASTI General
lisabar - posted 10:34, Thursday 12 December 2024 (11833)
Filter cavity tip-tilt removed from HAMX

Rich, Lisa

To make space for Den's 6D sensor in HAMX, we removed the unused tip-tilt with the filter cavity input coupler, as well as 4 additional 2" optics (2 lenses and 2 steering mirrors), as well as 1" optic.

These compononets are visible in the top-right corner of the picture below.

They are temporarely stored in the VOPO clean area.

 

Images attached to this report
Heisenberg IFO General
hudsonl - posted 21:09, Wednesday 11 December 2024 - last comment - 15:16, Thursday 12 December 2024(11832)
Heisenberg Interferometer Homodynes Broken?
As of yesterday (Dec 10th), the breadboard OPO's homodyne seemed not to work and had considerable excess dark noise. The board was using the M1 PCB initially. I swapped in the M2 and M4 boards, and neither of these behaved as expected either. However, the compact squeezer's homodyne board seemed to be working well and had about 30 dB of clearance between shot noise (with about 1 mW LO power) and dark noise. 

I left the electronics connected and turned on overnight and as of this evening when I came down to the lab, the compact squeezer's homodyne does not appear to be working either. There is only about 15 dB of clearance between shot noise and dark noise with the same optical power as before, and the spectral shape of the shot noise has changed (it is flatter out to higher frequencies). This leads me to believe that keeping the electronics turned on overnight caused the homodyne board to overheat, damaging the board. If this is the correct diagnosis, we should find a way to heat-sink the homodyne boards' voltage regulators without electrically connecting them to each other or to other grounds. 
Images attached to this report
Comments related to this report
hudsonl - 15:16, Thursday 12 December 2024 (11835)
[Hudson, Melissa]

Although the CSQZ homodyne appears to be at least partially broken, we measured its linearity to determine if it is at least usable for the time being. The attached plots show these measurements, which suggest its response is still reasonably linear.
Images attached to this comment
SQZ General
wenxuan - posted 22:07, Sunday 08 December 2024 (11830)
LASTI VOPO dichroic replacement

Jasper, Nikhil, Wen

 

We spent some time on Thursday to fix the dichroic filtering of the pump in the squeezed vacuum path. The pump power was noticably high on the pump reflection path because I put a dual superpolished HR mirror temporarily for rough alignment. We replaced it with a superpolished dichroic optics, but there's still fair amount of green (25 uW) coming out of chamber. 

To resolve this problem, we did some testing on the superpolished dichroics available to us. This was done on the sqz table right after SHG. We tried three different AOIs (roughly) and measured the reflected power with 60 mW input power. 

  22.5 deg 45 deg 67.5 deg
Dichroic E1600367 17 mW 1.2 mW 17 mW
Dichroic E0900491 2.2 mW 1.1 mW 12 mW

At the optimal AOI of 45 deg, the dichroic has a power transmission of >95%. However, the transmission reduced to ~70% at angles of 22.5 deg or 67.5 deg. 

We also tested the optics to make sure which side has the HR coating for 1064. The power reflectivity for 1064 was measured to be >99% for all dichroics (<1% loss). 

 

For the pump in VOPO chamber, the power was measured at the VOPO REFL (4.8 mW), after 1st dichroic (50 uW), and after 2nd dichroic (25 uW). The low transmission of pump on the 2nd dichroic is very likely to be the AOI, although its pretty close to 45 deg. We will need to move the optics a bit to find the optimal output path. 

DarkMatter General
evanhall - posted 18:17, Friday 06 December 2024 (11831)
Removed ADBC vacuum pipe

Swadha, Evan

We removed the pipe from the ADBC (formerly 2 m filter cavity) vacuum chamber. Knife edges on either end of the pipe are still intact. There is a ~1 mm, more or less round pit on the angled side of the knife edge on the rotatable flange, but it doesn't look like this will impact the integrity of the seal. The pipe has been foiled off, as have the flanges on the vacuum chambers.

SQZ General
wenxuan - posted 16:05, Thursday 05 December 2024 (11825)
LASTI VOPO locked on 532

The LASTI VOPO is locked quite stably on green. 

 

The OPO servo board was measured first. The filters on the common path has the following pole/zero.

COMPENSATION: 300 Hz/ 3kHz

BOOST: 1 kHz / 15 kHz

FILTER: does nothing

Daughterboard slot is disconnected

 

The filters on the slow path:

BOOST: 30 Hz / 360 Hz

COMPENSATION: 30 Hz

FILTER: does nothing

 

The servo board has the correct flat gain of unity with all filters disengaged. I've only enabled the BOOST filter on the slow path to give us the desired OLTF. The error signal still has too low of SNR, so I've increased the power a bit more (now 20 dB into the EOM, maybe too much but the EOM is not heated up). The final OLTF is shown here. The phase margin is plenty (83 deg). We could be more aggressive on the filter, but the lock is very stable; it remained locked when I opened the tank lid and tapped the breadboard. 

The green TRANS scan is shown here. Yellow is PZT input; blue is PDH signal; green is transmission. It looks like the small peak is first-order misalignment, but I couldn't fix it with alignment. We will investigate why there is an extra tiny beam spot on the OPO cavity. 

 

Images attached to this report
GRAVITES General
eoelker - posted 22:13, Wednesday 27 November 2024 (11829)
Prestabilization setup

[Eric, Dorotea]

Our new fiber feedthrough arrived this week, so we're finally able to recomission the laser prestabilization setup.  Below, some detail is included on the current status.  Note: so far, we've done no careful noise budgeting or characterization since this is only the 2nd time I've locked this loop.  However, I want to document everything in detail since we probably won't get back to this for a couple months.

The design for our prestabilization setup is based on this paper from ANU.  Since the NKT BASIK E15 has slightly worse free running low frequency noise than the laser used at ANU (Orbits Ethernal Slowlight), I've added feedback to the laser piezo to avoid range issues at low fourier frequencies.

The requirement for GRAVITES is to stabilize our laser frequency noise to below 2e4 Hz/rtHz at 10 mHz.  The performance attained at ANU is about 2 orders-of-magnitude better than what we require.  In principle, our design should be an improvement.  Our thermal time constants should be about a factor of 3 larger than reported by ANU, though we havent measured them directly yet.  We are also using a vibration-insensitive fiber spool which reduces phase noise from seismic vibrations by roughly 4 orders of magnitude according to our shaker tests.

Jaime's MSRP poster gives a nice overview of the setup.

 

Instructions for locking the servo

To lock the laser, the laser modulation must be enabled and set up with the proper parameters.  I attach a few screenshots showing the current settings: 1, 2, and 3.  Afer the laser is turned on, one needs to allow it to thermalize for about 15 minutes before trying to lock it.

The EOM modulation frequency has been set to 15.1 MHz, a frequency that gives decent sideband transmission for a 1km arm length imbalance when the carrier is locked to a dark fringe.  The demod phase was set to ensure that the servo has the overall correct sign.  We temperature stabilize the fiber EOM to limit RFAM issues.

The Marconi VCO is currently set to a frequency deviation of 20 kHz, DC coupled, the servo board has been set for maximum gain, and the laser piezo is currently being driven by a separate SR560, set as shown.  The Michelson is currently outputting ~50 uW of power onto an 1811.

Free running, we see the following on an oscilloscope.  Here, channel 1 (yellow) shows the DC power at the dark port, channel 2 (blue) shows the error signal after demodulation, and channel 3 (magenta) shows the correction signal being fed back to the PZT.  When the servo is unlocked, all 4 switches on the servo board are in the down position.  When you're ready to lock the servo, turn on the enable switch, then the two integrators.  When locked, the oscilloscope traces look as shown.  Ideally I'd like to see more suppression in the error signal to get a tighter lock, but im limiting the bandwidth to ~15 kHz because of the delay introduced by the 1 km arm length.  If I were doing things over again, I would probably lower the length imbalance by a factor of a few to increase the capture range and improve the bandwidth. 

 

Characterization of the servo (very limited so far)

Currently, I'm locking with a roughly 14 kHz bandwidth and 50 degrees of phase margin.  I'm being a bit conservative since the phase rolls off rapidly and I want to have enough gain margin.

The error signal spectrum may be bottoming out at a high white noise floor, though we need a noise budget to start making sense of things.  A back of the envelope callibration says this corresponds to about 2 Hz/rtHz.  The only loop component we've characterized is the servo input referred noise, which is a few orders of magnitude below this floor (see Jaime's Poster for a spectrum).  However we havent characterized any of the other components in the loop yet.  Everything else is second-hand and there could be some surprises waiting for us.  

Images attached to this report
Non-image files attached to this report
SQZ General
wenxuan - posted 15:43, Monday 25 November 2024 (11827)
VOPO TEC fixed with Newport controller

We plan to measure the nonlinear gain of the LASTI VOPO given that both pump and seed are aligned to the cavity. The cabling of the LASTI VOPO oven is here:

(pin #1, pin #6): Peltier

(2,7): Thermistor 1

(3,8): Thermistor 2

(4,9): Thermistor 3

pin #5 is a floating pin. Note that this D9 pin diagram is NOT necessarily the same with other VOPO!

 

After some digging of the old ilogs, I found the OPO oven pin diagram of the Beckhoff from Kentaro's post:

(pin #1, pin #9): Thermistor 1

(2, 10): Thermistor 2

(7, 15): Thermistor 3

(pin #{3,4,5,6}, pin #{11,12,13,14}): Peltier

pin #8: GND

Note that the EtherCAT TEC driver from LASTI is D1201359, which is different from D1201482 used at LLO. See llo63007. After making a cable to adapt from D9 to D15, I connectorized the VOPO oven to the Beckhoff interface. However, the Beckhoff failed to read any thermistor value after resetting. The TEC board is well connected to the Beckhoff chassis, like the functional SHG TEC board.

 

Anyway, I've just used a Newport 3040 TEC controller to replace Beckhoff. It's D15 pin layout is:

(pin #{1,2}, pin #{3,4}): Peltier

(7,8): Thermistor

After making another adapting cable, the controller reads back thermistor and controls temperature properly now.

Heisenberg IFO Relative Squeeze Angle Control
hudsonl - posted 17:31, Wednesday 20 November 2024 (11828)
Relative Squeeze Angle Control with Fast and Slow Actuation
[Hudson, Melissa]

We tuned up the relative squeeze angle control loop using the PZT mirror on the compact squeezer's path for high-bandwidth, low dynamic range actuation and pair of 532 AOMs for low-bandwidth high-dynamic range actuation. The PZT mirror can lock the loop stably with a UGF of about 1 kHz, but runs out of range after about 10 minutes. On the other hand, actuating with the AOM pair with high bandwidth interferes with other locks in the system, so this combination of actuators seems optimal.

We set the Marconi actuating the AOMs to 100 Hz FM frequency deviation, and it is set to external modulation with DC coupling. It is important to Null the FM deviation offset by using ""select coupling" --> "DCFM Nulling" (Visible when "Ext DC Coupling" is selected. The loop is setup to feedback to the PZT with a PI corner of 200 Hz and a proportional gain of 16. This error signal is then fed through a -10x analog amplifier and then the PZT driver. It is also fed to the input of a second PID block with a gain of -0.2, the output of which is sent to the AOM pair.

Using the Red pitaya's ASG module, the PZT is set to a nominal voltage of 75 V. The combination of the two loops has been observed to keep the PZT voltage within +- 10 V of this value over the course of at least 20 minutes, a substantial improvement over the situation without the AOM feedback in which case the lock would drop entirely after about 10 minutes.
SQZ General
wenxuan - posted 18:38, Friday 15 November 2024 - last comment - 14:12, Monday 18 November 2024(11824)
LASTI VOPO red and green setup

Eric, Wen

We've received all the hardware and mounted the 4-inch breadboard to level the beam height to the viewports (thanks to Kellen's help!) This setup avoids any periscope. We installed pump TRANS/REFL camera & PD, CLF REFL camera & PD, and seed TRANS camera & PD on the breadboard. With these sensors, we get the pump and seed pretty-well aligned into the VOPO cavity. 

We then tried to lock the OPO loop. The demod PDH signal was too weak so we removed the 6dB attenuator of the 80-MHz injection to EOM. After that, we can see the signal pretty clearly on the scope. Locking with the current OPO servo board was not successful. We have to re-do some loopology as we did with SHG lock acquisition. 

Images attached to this report
Comments related to this report
Lisa - 14:12, Monday 18 November 2024 (11826)

I am attaching here the ICS record with the successful mirror gluing process, with screeshot of the procedure:

https://ics.ligo-la.caltech.edu/JIRA/browse/Bake-10201

Images attached to this comment
Displaying reports 401-420 of 1025.Go to page Start 17 18 19 20 21 22 23 24 25 End