M. Richardson, M. Todd, K. Kuns, E. Oelker, R. Mittleman
We put the Y-HAM far end door on. Erik torqued the feedthrough flange and Kevin and I torqued the viewport flange. We used 180 in.lbs. After closing the vent-valve by the Y-HAM to 170 in.lbs. (last vac notebook entry had it around 172 in.lbs), we began to pump down with the two roughing pumps at 16:45.
Approx 17:10, the roughing pump out flow temps are around 77degF and the YARM pirani gauges are reading around 80torr.
Approx 17:50, the roughing pump out flow temps are around 82degF and the YARM pirani gauges are reading around 2.5torr.
At 18:00, we've closed all the valves and shutoff the roughing pumps, and left the water running for ~15min before shutting off. We'll come back in an hour and check the pirani gauges to see if we've got any obviously bad leaks.
At 21:00, 3 hours after shutting off the roughing pumps, the pirani gauges are reading 3.02 and 2.86 torr.
I ran a quick seismic measurement with the L4Cs in the Y-HAM, and they do not show much difference to the pre-door-on/pre-pumped state; so the motion between 2-10Hz may really be the stack resonances moving things around.
Following up from 12279
Today I added the two LPS installed on the CRS to the m1lps model, as H4 and H5 (chan 28, 29 from ADC2) - see fig1 and fig2. This model was rebuilt, installed and restarted without issues. The LPS_MAIN medm screen was also modified (fig3) to add the two additional LPS. The new additions are highlighted in green boxes. All antiWhitening filters and demodulation LP filters were copied from the Sensit LPS and loaded in the appropriate filterbanks.
trying to update m1lps model from lightwave :
matlab license needed update: I used my caltech license to get it working.
all CSD library parts were missing a link from the user model
found the location of the CDS_PARTS and added it to the matlab startup script - hopefully this fix will be applied to anyone login in as controls :
I'm in the process of determining which electronics chassis we need to have populated for the new experiment. I'm taking another look though our existing stock and am brainstorming ways to reduce our needs to limit how much we need to procure. Some progress detailed below:
ASC Electronics:
CDS Electronics:
I am also in touch with the Caltech Electronics team to see if we can coordinate with them on joint orders for anything else. The spreadsheet below has a running list of our other needs:
https://docs.google.com/spreadsheets/d/1Tj7x6LSWZRILykRMQ9FO5DNx8-9EQ4WdR53vpVZqvb0/edit?usp=sharing
Wanted to test whether I could use the VM high-bandwidth photoreceiver for phase noise measurements of Titanium:sapphire microlasers (NASA collaboration with Brightlight Photonics)
The current setup is a cleaved SMF-28 fiber mounted on a nanomax in front of the PD [photo attached] -- InGaAs PIN photodiode, 100 um diameter, TO package with a lens cap
Previously I was using a 2.6 um tapered lensed PM fiber (Corning PM15-U25D, FC/APC) , but was measuring QE ~ 45% at 1550nm. The lensed fiber has a divergence of about ~0.4 rad, which apparently exceeds the acceptance of the lens cap. The optimal position was several mm from the cap... the cleaved fiber has divergence about ~0.1 rad at 1550 nm
With cleaved fiber (taking transimpedance gain 103.27 kΩ):
1550 nm | 170 uW | R ~ 1.07–1.19 A/W from shot noise (photocurrent 182–203 uA) | QE ~ 86-95%
780 nm | 390 uW | R ~ 0.27–0.28 A/W from shot noise (photocurrent 104–111 uA) | QE ~ 42-45% [NOTE: the 780 is frequency-doubled 1560 and there may be some small leakage of 1550 light]
Attaching dark noise and shot noise of the 1550 measurement. With 170 uW, we have ~7dB clearance 40-70 MHz, 6 dB 100 MHz, 5 dB 200 MHz... Good enough clearance at low power to measure soliton quantum noise of our sapphire microrings (recently metallized by Daniil! so now they have microheaters, we need to test those).
Conclusion: this photodiode will work for 780 nm measurements, no need to swap in silicon PD
To-do: cleave 780 nm SM fiber, possibly switch to APC fiber to avoid back reflections
Consolidating the specs of our VM high-bandwidth photoreceiver for future reference
Optimized band: 10-200 MHz
Photodiode and optical input
AC output [attachment 1]
RF output [attachment 3]
DC output [attachment 3]
M. Richardson, M. Todd
We've taken some seismic measurements that we have more confidence in, after Mitchell and I confirmed the cabling for each L4C in the Y HAM. We also spotted an error in how I wired up the c2sei model, which previously had YAW and ROLL flipped. We saved this data to /home/controls/Templates/sei/l4c2dof.xml . All three translation DOFs have an integrated RMS of around 1um down to 200mHz. PIT and ROLL are around 0.4urad integrated RMS at 200mHz, while YAW is down aroud 1urad. Noticeably, there is quite a bit more motion between 1--10Hz, especially in the rotational DOFs then expected.; this could be due to the door being off, and air currents causing a bunch of motion.
After confirming each L4C output, we plugged them into the CyMAC in order (CH1 - CH6), without using CH0. There is some unexplained issues with CH0 ADC channel, so we avoided using it. We calibrated the ADC channel from counts to meters using four FM. They are: the reciprocal of the L4C TF, an AC filter and undoing the preamplification of the L4C signal and undoing the ADC (16bit, +/-15V). We did not assume any deviation from the spring parameters quoted in the design, though the springs have surely sagged over time (20 years).
These outputs were then fed through a sensing matrix to recover the DOFs. The channels correspond numerically to the L4Cs as configured in the layout sketch. The middle left L4C (L4C-4) was moved previously to route the transmission path to the periscope, and so it's separation from the central axis is assumed to be 8 inches shorter than L4C-2.
The response of each sensor can be expressed as t + theta x r, where t is the relevent translational DOF and theta x r gives the two relevant rotational DOFs from the cross product with the lever arm. The inversion of this matrix yields the sensing matrix.
Data from this morning ~9am. Also attaching the l4c calibration.
Following up from 12275
Today we completed the alignment of the two CRS LPS, and rebalanced CRS2 to ~18mHz. It would be good if we have a chance to measure a ISI tilt to CRS transfer function before chamber closeout to verify the d-values. Otherwise, the work on this side of the table is now complete, and the two CRS were left unlocked.
The fringe visibility after alignment were the following :
Max fringe visibility (fig1): 90% LPS1 / 82% LPS2
Min fringe visibility (fig2): 52% LPS1 / 38% LPS2
The variability of the fringe visibility is because of large amplitude of the CRS in air, swinging by ~mm relative to the LPS. In normal operation, the CRS wing is only moving sub micrometers, so we should be able to park it in a good LPS spot.
To get the resonant frequency, we measured the time between two slow fringes (when the velocity is zero) which corresponds to 1/2 a CRS cycle. 28s*2 = 56 s = 18mHz (fig3)
Next the LPS signal chain needs to be commissioned (model change, medm screens, filter settings etc...) but all of this can be done remotely.
Otther things to note:
* The LPS 1 (left) and 2 (right) were respectively connected to chan 28 and 29 of m1sus ADC2.
* The remote balancing picomotors were tested (for both CRS) from the PC sitting in the cable tray, they each rotate as expected in both directions.
* The capacitive drive chain for the damping was verified for CRS2 (voltage sent from dac and measured with the multimeter on the plates themselves).
* A new cable was added between the output of the Birmingham TIA chassis (db9, Output 1) to the last channel of the m1sus ADC2 AA chassis, where all the SensIt signals are routed.
* Since the CIT Toptica is not in used anymore (replaced with the Birmingham laser used for SensIt) we removed it from chamber side to reduce cable clutter. It was stored in a box, alongside with all the CRS tools in the pink lab NW17-009
* Pictures related to this work can be found in this album
Arnaud, Marie, I think the aligned LPS-CRS contrast of 82-90% is consistent with the Birmingham measurements.
I would like to add a note on the added vertical offset of the retroreflector and, related to that, on the lowest observed contrast when the CRS moves in angle. The LPS contrast can be approximated as
C ~= C0 - (alpha / theta)^2, where C0 ~ 0.85-0.9, alpha is the CRS tilt angle, and theta determines the angular range of the sensor and is given by the equation
theta w / 2 / (x_beam + h), where w = 0.5 mm is the beam radius, x_beam = 2.5 mm is the offset between the input and returning beam, and h is the vertical offset between the CRS pivot axis and the retroreflector.
By design, we like h = 0 and achieve theta = 100 mrad. In your case, it seems that you have lowered the retroreflector by ~20 mm to achieve the LPS-retroreflector separation of 50 mm. However, this also leads to a reduced angular range of LPS: theta = 11 mrad. I think this is consistent with what you currently measure.
The reason why a non-zero vertical offset of the retroreflector position, h, reduces the contrast is because the CRS motion shifts the retroreflector horizontally according to the equation x_retro = h * alpha, and this shift significantly reduces the contrast. I think in the ideal case, you could remove the vertical offset, h, and move LPS closer to the retroreflector. This case would preserve the angular range of the LPS-CRS sensors at the design value.
S. Park, M. Todd
After replacing some voltage regulators and op amps I was able to get the 6channel pre-amp working. We moved the CyMAC over closer to the PSL racks and started connecting cables from the L4Cs to the preamplifier to the CyMAC. CH1 of the pre-amp seemed to not behave as nicely as I originally thought, and seemed to trip, or affect in some way, the ADC out -- so we elected to sub in the SR560 for it but the 560 we chose seems to be limited to a gain of 10 for the signal coming from the seismometer (even though it should be like mV).
Using the sei model and repopulating the sensing matrix after rebooting the CyMAC, we took some spectra of the various DOFs. The DOFs are in m/rtHz and rad/rtHz, with the ADC Noise calibrated in the brown trace. It seems we are still limited by ADC Noise below 1Hz, though until I fix the pre amp we are stuck at a gain of 10.
This week, we were able to get the NPRO lasing (current setup). In this setup, the NPRO is end pumped along the same path as the previous seed beam. The two beams are combined using a 808/1064nm dichroic mirror before sharing a common lens right before the crystal.
Only preliminary data exists at the moment, but the output power was roughly ~4mW without very careful alignment of the pump beam. Here is a photograph of the output on a beam card.
A very rough spectrum was taken of the laser's output intensity noise. This is not calibrated, but shows the expected resonant-relaxation oscillations. We should be able to use this data (once calibrated) to determine more of the solid-state parameters of the laser.
In the next sequence of logs, we will be measuring
1.) Pump RIN and wavelength
2.) Lasing mode profile vs. seed mode profile
3.) NPRO RIN measurement
4.) Mapping to noise model
JD, Richard, Arnaud, Vermeer, Alexandra
HAMx front side door was opened yesterday, see timelapse.
The x2 CRS LPS from Birmingham were installed on CRS2 today. Those were built to give more range for the CRS with retroreflectors. LPS1 was installed on the left wing, LPS2 on the right wing. The 1550nm 1/2" mirrors from last install (log 12120) were unglued (snapped off), and retroreflectors were reinstalled offseted from the wing with long M2 screws (air baked overnight) so the distance is closer to the nominal 50mm (the closest we could do previously was 59mm given the old requirement of 70mm). For the light path and readout chain, we will use the spare cables from the Sensit experiment.
The 1550nm 1x8 splitters was installed in chamber, for SensIt and will be connected to the two LPS tomorrow. There is still some CDS issues and a long db9 cable to run from the LPS TIA to chaa 29-31 in the Sensit AA chassis in the m1sus rack.
More pictures can be found here: https://photos.app.goo.gl/zKqvGNVRHcv3qyeCA
Vermeer, Arnaud
Summary: CRS2 flexures is replaced, LPS coarse alignment complete.
While I replaced the CRS2 flexures, Vermeer connected the input of the 1x8 splitter (received from Birmingham) to the peek fiber coming out of the 1550nm SM feedthrough, 3 of its output to the Sensit LPS and 2 to the CRS LPS. We verified there was power getting out of each Sensit LPS visually with an IR card and on the DC power channel in CDS (channels 0,1,2 on ADC2 of the m1sus frontend or M1:LPS-SUS_H{1,2,3}_DC_IN1). We then borrowed 2 of those 3 readouts (H2,3) to align the CRS LPS. The process of aligning the LPSs was easy, we first found the return beam with an IR card and adjusted the LPS plate sideways to get fringes on CDS. We freed up the CRS to allow fringing and started the fine tuning process using the adjustable optic mount. Fringe visibility on LPS 1 : 85%, and LPS 2 : 61% (see ndscope screenshot attached, H2 channel = LPS1 and H3 channel = LPS2 ). We will continue fine tuning LPS2 tomorrow morning.
Left to do:
*cable up the CRS LPS to its AA chassis and label connectors
*balance and fine tune CRS2 frequency
*modify the m1lps simulink model to add the new LPS channels that will be connected to chans 29-32 of that same ADC card.
*pack and store the previous LPS + 1550nm mirrors
*verify CRS1 HOQIs are still working, now that HOQI SensIt is using a split of the light chain
*pack and store CIT 1550nm laser now that CRS is sharing Birmingham's setup
*update the CRS MIT wiring diagram
Spent some time try to build a model that can take some of the ADC channels and pop them through a matrix to spit out the different DOFs of interest. Eric helped me with the simulink model and I built and installed the model. After some fiddling around and some medm screen tuning, it's possible to pull up the DOFs on dtt, and easily accessing the necessary filters and matrix from an medm screen.
The dof channels are C2:SEI-L4C_DOF_<dof>_OUTPUT
And the medm screen is in 'userapps/cds/c2/medm'
I spent some time yesterday to debug the newer CyMAC which lives in NW17. This unit is currently in the highbay where we plan to use it for measurements until our new LASTI CDS node is operational. Previously it lived in the Yellow Lab, but Evan told me that they never got it working properly. I had to debug a few things, but it seems to be working OK now.
I found that AI is now pretty knowledgable about our RTCDS software and can give fairly coherent debugging advice if you give it output from the various journalctl logs for the RTCDS daemons. This was not true 6 months ago when I tried something similar. It frequently gets a bit confused and makes suggestions based on older CDS features which are deprecated or obsolete, but to be fair the same is true of the dcc and CDS gitlab wiki (it seems to be searching our documentation, so I imagine that the two issues are related). Still, it was a huge improvement on spending days digging through the documentation myself.
Basic summary:
Yesterday I took some time to try and diagnose why the slow control did not seem to be feeding back correctly, I wanted to look at the DC set point of the temperature. I saw that for some reason it seemed to be set to 59C, which was way too high and more than likely at the very top of its range. I had some trouble actually fixing the setpoint, but by turning the tuning mode from "Extended" to "Min" with the dip switch in the back, I was able to put the set point to 47.5C. I had to turn the laser diode ON to do this, which I was very nervous about, but it should hopefully save this state for the next power cycle. With this setting, I fed the slow error signal through a 100mHz LPF to the temperature control and saw some of the behavior I was looking for.
Watching the PZT driver monitor channel on a scope, you can see it swing around its set point at low frequencies quite a bit, and then when the slow control is engaged it is much more stable around its setpoint. We probably want some gain adjust on the slow control error signal, which will benefit from an integrator with a variable input resistor.
I'm working on getting the new(ish) cymac working so we can take some seismic data with the L4Cs in HAM-Y. I spoke with Evan about it, and evidently they never got it working properly in the Yellow Lab.
The workstation computer wasn't set up properly. After an hour or so of troubleshooting, I decided to just wipe the computer and start over with a fresh debian 12 + cds-conda install. I also added the required NDS variable settings to the .bashrc and configured a network for the cymac + workstation using the usb-to-ethernet adapter and a switch:
cymac1550: 192.168.1.140
cymac-workstation: 192.168.1.136
The workstation seems to be working fine now. All the cds software is functioning and we can see the cymac channels. There still seems to be an issue with streaming fast real-time data. ndscope will connect and grab data, then drop intermittently after about 10-15 s, then start again after a few s. Got some advice from the site CDS folks, hopefully can track that down tomorrow.
Checked the pirani gauges this morning: TCA=3.22, TCB=3.07torr
Noticed that the valve between the YARM chamber and the foreline was left open, so these pressure readings include a small section of the foreline and a few KF blanks.