I've been looking into all things temperature since arriving in Vienna. A few points of interest:
I'm on the fence about the best route forward. I've achieved much better stability with good thermal filters under vacuum, but that will be somewhat cost and labor intensitve to implement. We're already quite close, so we might be able to get away with staying in-air. I think we need a bit more experience with thermally length matching the two spools to make sure that process is robust and reproducable when done in air. This absolutely essential for doing the measurement. I think this is likely to be the deciding factor.
This week, I've been designing new copper pods for our vibration-insensitive fiber spools that should be much easier to stabilize than the current (poorly designed) aluminum pods we're using at present. All components will be vacuum compatible so we can go that route if we decide to. One actually needs a fairly decent vacuum to achieve good thermal isolation. We plan to do a pump-down test with the chamber Chris bought to see what sort of vacuum we can achieve here. Hopefully the blank flanges will arrive soon and we can complete the test before I leave.
Swadha, Evan
We continued coupling experiments between two prisms with copper leaf and achieved transmission as high as 3%.
We are still not happy with our procedure, but it is as follows (with critique):
We improved this procedure through the following:
Our procedure to prepare the faces between tests is:
After this procedure, we were able to get coupling in excess of 30% in the p polarization. We mounted the two prism stages on a rotating optical breadboard and took transmission readings as a function of angle of incidence. An MCMC analysis of the data suggests the gap size is 1.5 µm, which is still larger than the 600 nm of the leaf itself (measured with a micrometer). The MCMC determines the index of refraction to be 1.43, which is lower than the 1.445 expected for Corning 7980, but the constraint is not too tight and is also susceptible to any difference in power efficiency of the reflection and transmission photodiodes, which we did not set up too carefully. In contrast, the gap size is determined by the falloff in transmission with angle of incidence, which is well-calibrated.
There is a question of why, if the copper leaf is supposedly 600 nm as measured by micrometer, the gap size here appears to be about 1.5 µm.
I hypothesized that it might be due to insufficient pressure applied from way the prisms were clamped (each on a platform with a clamping arm, as you might do for a PBS). The clamping arm provides a restoring shear via its nylon-tipped setscrew.
I replaced this clamping geometry with three angle brackets for each prism, with 8-32 tapped through-holes that can accommodate setscrews. I repeated the contacting procedure described above and was not able to significantly raise the transmission higher than the values previously reported (0.3 to 0.4 in power at best). I did not take detailed data as a function of angle, but it seems this altered clamping geometry did not attain an appreciably smaller gap size.
It is also worth remarking that both clamping geometries don't really provide proper kinematic contact; the point of this two-platform geometry was to bring the prisms nearly into contact in some controlled (perhaps interferometric) fashion, rather than relying on actual contact via a spacer material (leaf in this case).
Swadha, Adam, Evan
We turned the entire coupling setup vertical so that we could let gravity maintain an approximately kinematic contact.
Simplify placing the upper prism on the spots of leaf results in a coupling of a few ppm or less, so presumably many microns of gap size. We suspect the leaf is wrinkled in some way when cut and/or placed, although we have taken some care to avoid obvious folds or scrunches when we place it.
Applying pressure at the interface seems to bring the gap size down to what we saw previously, probably 1–2 µm as inferred from the 30% or so transmission in p. (We have some transmission vs. angle data that we have not analyzed yet.)
This setup has the nice feature that when the flow bench lights are on, they produce obvious fringing at the interface, so it is possible to see the parallelism of the faces down to the level of a few fringes.
Slawek, Lisa
Slawek has designed a new process to improve the gluing process of the OPO mirrors so as to control the thickness and uniformity of the glue layer -- he suspects that not controlling these aspects was is the main reason why some of the past glued optics failed.
In revising the process, we are switching (again) from epoxy EP30-2 to EPO-TEK 353ND, with 4h curing at 85C in situ.
After two successful tests on low quality optics, today Slawek glued the mirorr that failed a few months ago during lab testing in the High Bay -- one of the curved HR optics of the spare OPO (M3).
The mirror was extremely dirty, with a visible scratch on the back side -- we cleaned it with methanol and the HR side was ok after cleaning.
The scratch is on the unusued back side, and not in the center of the mirror, so we went ahead with using this optic.
I am going to write down the new process in DCC E2500027, with a few pictures here for posterity.
By gradually tuning the temperature of one of the 50 km fiber spools, I was able to get the zero arm length difference within the photon coherence length (around 3 mm, corresponding to a 3–5 mK temperature range). This allowed for initial benchmark measurements of photon phase noise at low frequency with interferometer locked.
I injected HH photon pairs into one input of the MZI before the first BS. With a 50% probability, the photons separated, each traveling along an arm of the interferometer before rebunching at the second BS. This resulted in single-channel counts [prop to 2sin4(phi/2), 2cos4(phi/2)] and coincidence counts [prop to sin2(phi)] that varied with phase fluctuations. (This setup is different from the final measurement scheme planned; it was chosen for convenience to take an initial measurement using the current MZI setup, without a PBS or polarization control.)
Continued to interfere the Teraxion and 1550 nm Koheras to repeat December 18th measurements with a greater goal of measuring the Teraxion frequency noise. Interfered the two lasers with a PM 50/50 beam splitter (AFW PFC-15-2-50-FB) and focused on PDA05CF2. The Teraxion had a 20 dB attenuator. Started with a 164.8 pm offset for the Koheras (relative to 1550.12 nm) and then tuned to 163.3 pm and 161.8 pm. Tried to feed back to the Teraxion and lock the loop by mixing the beat note with Marconi 2030 output using a ZRPD-1+ phase detector, 5 MHz low-pass filter, LB1005 (gain -20 dB, PI corner of 1 kHz), and a series of attenuators (-34 dB total). Did not achieve lock and will need to try again to make the measurement.
Swadha, Sasha, Matt, Evan
We have begun studies of FTIR coupling between right-angle prisms using kinematic mounts. We are using Edmund Optics prism #47-785, Corning 7980, 20 mm height, uncoated.
We started with copper leaf from Barnabas gold. We folded a square of this leaf on itself to create 16 layers on top of each other and measured with a highly accurate micrometer to have thickness 10(2) µm. This implies a thickness of the single layer 630(130) nm.
Swadha cut the copper leaf into strips with a razor blade and placed them on the face of one of the prisms (they stick electrostatically). We then use a 5-axis translation stage to press one prism against the other, sandwiching the strips. Using the translation stage screws by themselves does not give sufficient compression. We had to press the stage by hand, and then we achieved the following coupling: on a PDA10CS in reflection, 2.8 V of power; on a PDA20CS in transmission with 60 dB of gain, 220 mV of power, implying a power transmission of 80 ppm in the p polarization. Repeating with the s polarization: 2.9 V in reflection, 125 mV in transmission, implying 45 ppm of transmission in the s polarization.
For glass we would expect that a gap of <1 µm would give transmissions exceeding 0.1, so this would imply we are not reaching a gap consistent with the thickness of the leaf. However, inferring the gap size is difficult because it is exceedingly sensitive to the value of the index of refraction, and to a lesser extent the angle of incidence.
I had the two Marconis under the LASTI SQZ table sent out for repairs a couple of months ago. Both have now been re-installed and hopefully reconnected correctly.
One note: On the Marconi 2026, I connected a cable labeled "ALF VCO Mod" to the modulation input on channel B which is labeled as "RLF VCO". I'm guessing this was just a typo on the label (we have an RLF and an ADF, but no ALF on the squeezer). If something isn't working, its worth a double-check.
I did a series of measurements to study how temperature fluctuations of the 50km fiber spools influence the interferometer arm length difference, i.e. photon travel time difference. The goal is to measure the actual coversion coefficient of temperature fluctuation noise to length/phase noise in the noise budget, and to prepare for using temperature to control arm length difference in photon measurements.
1550nm HH entangled photons are sent into the MZI through one of the inputs (the same one used for senting the 1550nm CW laser), and two outputs are directed to photon detectors. When there is a length difference (greater than the photon coherence length) betweent the two MZI arms, three coincidence peaks appear in the photon histogram. The time difference of the two side peaks from the center peak tells the photon travel time difference in the two arms.
Fig. 1: Spool1 temperature stepped to different temperatures. [inner spool, 24ohm heating resistor]
Fig. 2: Spool2 temperature increased by 0.5°C. [outer spool, 16ohm heating resistor]
Fig. 3 & 4: Readings from inside and outside out-of-loop temperature sensor of the two spools with Spool1 temperature reduced by 0.2°C. Note that it took ~17 hours for the temperature inside the pot to fully thermally stabilize to 1mK (we need better design for future spools). The out-of-loop sensors of Spool2 appear to be not working/poorly attached.
The resulting coeffifcient is approximately 0.8~1e-5 (dL/L)/K, which is roughly consistent with dn/dT for optical fibers, which is the dominant factor in these plastic fiber spools. This translates to a change of 40-50cm/K, and we should be able to find the photon interference (0.5cm coherence length) by tuning the temperature to that 10mK range. A side note: These coincidence peaks arequitewide (~800 ps), which is 40-50 times of detector jitter (timing jitter < 18 ps per photon detector spec), not sure whether that is from interferometer noise or some kind of dispersion, need some further investigation.
Paul, Evan
Paul has assembled a new cymac that we received from Caltech. This cymac has been set up so that the front end (cymac1550) and the workstation (cymac-workstation) are separate machines connected by an ethernet switch. Both computers run Debian 11 (Bullseye).
On cymac-workstation I added the LIGO CDS repository http://apt.ligo-wa.caltech.edu/debian and proceeded to install the cds-workstation packages (GDS tools and others... see attached apt log).
The front-end has at least some of the advligorts packages already and runs the rtcds service. It is unclear whether this was installed according to the CDS wiki instructions.
It is possible to run diaggui and ndscope from the cymac-workstation if these programs are pointed to cymac1550:8088 as the NDS host. The c2tst model does not have any channels saved to frames yet.
For the layout of the analog filters chassis: LIGO-D2300052
[Adam, Nikhil]
After the RL training with 2 degrees of freedom (DOF), we are now working towards controlling more DOFs (pitch & yaw for two optics).
So we recently updated the layout (attached) with the following modifications:
Beam ellipicity is around 90%. 50 mW in each arm. S-pol.
Manual alignment now gives us a visibility close to 96%. The beam walk procedure involves first minimizing lateral mis-alignment by ensuring the overlap of the two beams on the near-field camera using the M1 mirror. This step is followed by fringe-based optimization using the far-field camera. We then adjust the fringe on the near-field camera using BS2 beam-spliiter (and repeating the three steps if necessary based on visibility metric).
Mach-Zehnder is now dither locked to dark fringe using the output PD & piezo via CyMAC with a simple Foton-based proportional integrator feedback loop (needs some fine-tuning). We use the spare Thorlabs three-axis piezo controller (MDT 693A) from Hudson. We will follow up with some open loop measurements to design a more optimal filter.
The last several days, I've been putting together the second laser prestabilization loop for the GRAVITES moving platform test setup. I stuffed another servo board and put together a chassis. The noise performance is comparable to the previous unit. I now am able to lock both the 1550 nm and 1545 nm lasers to the prestabilization interferometer simultaneously, each with about 10 kHz bandwidth, which should be more than sufficient to suppres noise at the platform modulation frequency. See (11829) for a more detailed write-up on the prestab setup.
Still need to set up the prestabilization temp controller again, just waiting for a new cable from thorlabs. We're more or less ready to close up the acoustic enclosure. Tomorrow I'll start assembling the rotating stage now that all of the components Elio ordered have arrived.
I measured the seismic noise of the Vienna interferometer setup in the 4th-floor lab under various conditions: with the suspension table on and off; with the box open (unstable temperature) and closed (after the temperature was stabilized); seismic sensor placed directly on the optical table (without floating). All data were taken with the cryostat pump on.
The seismic data have been calibrated to g/√Hz to be inserted into the noise budget.
I measured the free-running frequency noise of the NKT laser by calibrating the control signal for the GRAVITES prestabilization interferometer. The goal is to get a stand-in trace for the vienna noise budget which uses an almost identical laser at 1550 nm (NKT E15 Adjustik) since they don't have a setup for characterizing laser noise at 1550 nm. Looks roughly consistant with the measurements published by the AEI group above 1 Hz (their measurement did not look at lower frequencies).
I set up a free-space Michelson with a one-way asymmetry of 10 cm and then aligned it with light collimated from the Teraxion. The fringe contrast was 0.53 mW maximum and 0.14 mW minimum (58% visibility). This used a PDA05CF2. With the interferometer near the half fringe, I drove the Teraxion's modulation with 0.1 V pp, which produced 0.104 mW pp. Using the relation Pmod / Pfringe = visibility * tau * 2 * pi * mod_depth (tau being the round-trip asymmetry 2L/c), this implies a modulation depth of 0.127 GHz pp, or a modulation coefficient 1.27 GHz/V. In comparison, the manufacturer's test report gives 0.99 GHz / V. Changing L to 7.5 cm, the resulting power modulation was 0.09 mW pp. The fringes went from from 0.10 to 0.56 mW (70% visibility). Changing L to 5 cm, the resulting power modulation was 0.06 mW pp. The fringes went from 0.09 to 0.57 mW (73% visibility). Changing L to 2.5 cm, the resulting power modulation was 0.04 mW pp. The fringes went from 0.08 to 0.57 mW (75% visibility). Changing L to 0 cm, the resulting power modulation was 0.01 mW pp. The fringes went from 0.00 to 0.63 mW (100% visibility). Changing L to 12.5 cm, the resulting power modulation was 0.12 mW pp. The fringes went from from 0.15 to 0.49 mW (53% visibility). A linear fit to these, correcting for the nonunity visibility and allowing for a free intercept to account for any systematic error in asymmetry, gives a coefficient 5.1 / m = 4 * pi * mod_depth / c, and hence mod_depth = 0.122 GHz pp, still with 0.1 V pp modulation depth, consistent with the above. The coefficient is therefore 1.2 GHz / V. If either of the beams in the arms is blocked, no power modulation is visible on the scope.
Compare to noise during O3: LHO:65572
I've been working to install the feed forward filters on the ISI (Stage 0 L4Cs -> Actuators). At the sites Jim has these working around 10Hz (maybe 3-30Hz? need to double check).
This is done by measuring the L4C to GS13 TF and the Actuator to GS13 TF, and then fitting the ratio (GS13/L4C)/(GS13/ACT_ = ACT/L4C
The ZFeedForward plot shows three different filters and their predicted performance (Suppression), the blue line is the data (fitting Goal), so you can see that the amplitudes are pretty good from ~1 to 20Hz, phases are mediocre.
The first two (green and olive) which map to Blue and green in performance, sorry about that, I did myself. The third (red and olive) was done using a fitting tool that Nikhil has written.
On the second plot you can see the results of the first filter which pretty much does just what is predicted
[Chris, Eleonora]
Yesterday we installed a microphone (Razer Seiren v3 mini) inside the acoustic enclosure in the lab in Vienna (where the 50km interferometer is), between the two spools as shown in Fig. 1.
We acquired the noise floor spectrum turning off the microphone (using the Razer Synapse software) and the acoustic noise spectra with the AC on / off.
Fig. 2 shows the data from the microphone with AC on and off in Volts.
Fig. 3 shows the data from the microphone with AC on and off in Volts/rtHz.
Fig. 4 compares the control signal from the loop (fast out) with the microphone data, with AC on and off. The microphone data are NOT calibrated, meaning that these data are useful only for a qualitative comparison of the peaks.
[Chris, Eleonora]
Today we also acquired the spectrum of the acoustic noise inside the box after turning on the compressor of the nanowire detectors.
The attached figure shows the comparison of the past measurements with the one of today with AC on, compressor on and pump on.
I took also the measurement with the compressor on and the pump off. The structure at 1200Hz is gone as expected.
This is the acoustic noise spectrum when we want to measure single photons with the detectors in the current lab.
I also investigated the shift discepacy between the curve taken at different times. I connected an other microphone and changed the volume level and it makes the level of the curve shift, as shown in the second plot attached.