SQZ General
dmdesant - posted 20:25, Wednesday 30 April 2025 (11946)
LASTI VOPO CLF beam path re-alignment and profiling
[Wen, Daniel]

On Monday, we performed some preliminary alignment and profiling of the beam through the two CLF AOMs. Last week, both AOMs were aligned into the beam path and had their phase-matching angles set to produce the maximum diffraction efficiency. The optical path is depicted in the first image attached here.

First, the RF drive power to the first CLF AOM was characterized as a function of signal generator drive power at 200 MHz to calibrate out the nonlinear effects of the Mini-Circuits benchtop RF power amplifier used (TVA-R5-13). We then characterized the diffraction efficiency for the first CLF AOM (IntraAction ATM-200-2DA6) as a function of drive power as it's datasheet lacked drive specifications for the 1064nm wavelength used in this setup. The resulting diffraction efficiency is shown to exceed 70% with RF drive above approximately 2 watts at 200 MHz. Also plotted is a normalized figure of merit for the AOM conversion efficiency, which is the ratio of diffracted power to input optical power (diffraction efficiency) normalized by input RF drive power (P_optical_diff/P_optical_input) * (1/P_rf_drive). This ratio in dB has the units of dBW^-1, and is useful for modeling small-signal conversion losses in RF-optical systems like this. From this figure of merit, we can see that the AOM starts to compress with about 30 dBm (1 watt) of drive. So for any future linear single-sideband conversion applications of this modulator, the total RF drive should be kept well below 1 watt at an operating RF frequency of 200 MHz and operating wavelength of 1064nm.

With the upconverted CLF beam back in action, we aligned it through the second CLF AOM configured to shift the frequency back down by approximately 200 MHz +/- our CLF frequency and approximately profiled the beam transmitted through the PBS in the beam path before the second CLF AOM. Using the BeamScanner model 1280, we measured the beam diameter as a function of distance from the PBS center (plot attached). This revealed that the beam has some amount of astigmatism, but should allow us to start mode-matching into the fiber soon. 
Images attached to this report
SQZ General
dmdesant - posted 19:39, Tuesday 29 April 2025 (11947)
LASTI VOPO Phase Matching Spot Exploration
[Wen, Daniel]

This is my first log post as I just got my aLOG account - so this information is a bit old!

A couple weeks ago, we swept the parameter space of the LASTI VOPO PP-KTP crystal temperature and lateral wedge displacement to find the possible 523nm/1064nm co-resonant conditions where nonlinear gain can be observed. In this test, we scanned the cavity while pumping the OPO cavity with 532nm light along with a weak 1064nm seed signal. The cavity was scanned over at least a full FSR while monitoring the transmission of both the seed and pump transmission. 

Next, we moved the crystal using the piezo stage to the furthest negative displacement that still allowed the intracavity beam to pass through it. This was found to be around -2mm using the zero origin that was persistent in the stage controller at the time. At this position, we swept the crystal's temperature control setpoint until co-resonance was observed. At co-resonance, we observed the sporadic gain peaking of the seed transmission and noted the relative amount of nonlinear gain qualitatively. Next, another resonance condition could be found by incrementing the crystal position and increasing the crystal temperature to recover co-resonance. This process was repeated until the temperature exceeded around 35.0C, at which point we would move to the next co-resonant cavity mode by incrementing the crystal position and reducing the temperature back to approximately 31C. These loops in temperature are repeated as the crystal is translated from left to right until it falls outside the intracavity beam. This way, each loop represents a "group" of possible co-resonant conditions within the ~30-35C temperature range scanned in this experiment. The exact range of the crystal's span over stage translation are depicted as dashed gray lines in the plot below. 

These groups are plotted and fit in the graph attached. Along with these trends, we note some of the locations where exceptionally high nonlinear gain was qualitatively observed. We plan to revisit these locations in the parameter space when optimizing the nonlinear gain quantitatively at a later date. Since we expect there to exist an optimal phase-matching temperature for the PP-KTP crystal, these curves of possible co-resonant conditions should tell us where a discrete set of 13 phase-matched and co-resonant locations (or spots) exist on this particular crystal. In this initial experiment, there was only one spot where we noticed that the crystal was damaged and no resonance could be observed at all. This was noted in the plot as a red star.

Unfortunately, the power surge that occurred a few weeks ago happened right after this data was taken. The surge caused the piezo stage to move past the stage's limit switches and collide with the VOPO assembly's base. Consequently, the zero origin of the stage was lost, so we will need to re-zero the controller before we can revisit this data.
Images attached to this report
TIR General
evanhall - posted 16:32, Saturday 26 April 2025 - last comment - 17:52, Thursday 01 May 2025(11944)
TIR mode found in 1064 nm

Swadha, Evan

We had struggled for a while to align into the cavity by coupling in reflection with a single prism only, and with the entire cavity mounted vertically (letting gravity set the kinematic contact between the prism and three small squares of copper leaf). The key changes this time were to mount the cavity horizontally (with kinematic contact described below) and to couple with two prisms (one input one output) so that transmission information is available for alignment.

The key points of the optical setup were:

  1. The cavity was mounted on three nylon-tipped 8-32 setscrews. The flat optical faces of the cavity were prepared with three small squares of copper leaf each as described earlier. (MIT:11889)
  2. Each prism is let to rest on two nylon-tipped setscrews, then pressed by hand against the cavity, flattening the copper leaf. This contact is then maintained with a third nylon-tipped setscrew aimed horizontally. This should give kinematic contact (three points of contact from the copper leaf, two points from the setscrews supporting the bottom of the prism, and another point from the horizontal setscrew).
  3. The laser beam emerges from a fiber collimator F220APC-1064 and is steered with two steering mirrors. The first steering mirror is placed between two 150 mm FL lenses so that its action is approximately conjugate to the second steering mirror.
  4. We did not attempt any mode matching into the cavity's mode, and the laser mode right now is grossly oversized compared to the cavity mode.

The optical alignment procedure went like this:

  1. We started with a few milliwatts of p polarized light going into the cavity, with a CCD camera on the transmitted output.
  2. When the beam is sent into the cavity even roughly at the correct angles and locations, a ring of points emerges in transmission that is visible on the card (and too big for the CCD — probably about 1 cm by 1 cm in extent). We were initially confused about what these points were, compounded by the fact that at 1064 nm there are non-negligible AR beams bouncing around everywhere, leading to stray beams emerging from the optical faces of the cavity at strange angles.
  3. We started driving the laser current with a 4 V pp triangle wave at 5 Hz. Some small, high-order fringing was visible more or less everywhere on the CCD, but nothing clearly resembling a Hermite–Gauss mode.
  4. In any case, we adjusted the steering mirrors in yaw until the ring of points in transmission could be made into a more or less vertical line of points.
  5. Adjusting the steering mirrors in pitch was more difficult as the line of points would expand out into a ring again. Nonetheless, by iterating (including with yaw) we were able to condense the line of points down into a wad of points as viewed on the card. On the CCD, the wad of points appeared as a more or less as a mess of light and fringes. However, it was evidence that the interference was stronger now than before, hinting that we were closer to something modal.
  6. After some more fine iteration with the steering mirrors, a TEM00 mode was evident, although it was surrounded by a strong background of messy fringes. Since the finesse in p is presumably small and the mode matching is so poor, this is not surprising in hindsight but it probably slowed us down a bit, and higher-order Hermite–Gauss modes would not have been evident.
  7. We switched input polarization to s, where the finesse is higher. The TEM00 mode was more evident now, and the background fringes smaller. We declared success.

Our estimate of the finesse in s is based on the following:

  1. The FWHM of the 00 mode corresponds to a slew of 0.78 V into the laser current driver modulation.
  2. The current driver is LIGO-D1200719-v1. My reading of this schematic is that the relevant transconductance is set by the op-amp current source U7, which appears to be 1/(3 kΩ) ≈ 0.33 mA/V.
  3. David Yeaton-Massey gives the current-to-frequency coefficients of these lasers as roughly 30 MHz / mA, but this is going to depend on the dc current drive to the laser, which I have not worked out yet.
  4. Putting these numbers together implies a finesse in s of a few hundred.

The key points for future iterations of this procedure are as follows:

  1. The horizontal cavity setup made by pressing the prisms against the cavity with fingers (and then maintaining with setscrews) seems to have worked well. We may want to formalize this with a jig because the prisms are not totally aligned with the optical face of the cavity.
  2. Importantly, the mode seems to have formed at least a few millimeters off vertically from the midline of the cavity. This is not unexpected given the toleracing we specified for the curved faces (they are 5 m ROC nominal), but it justifies our decision to design large prisms with wide clear aperture.
  3. In hindsight, the optical alignment is probably achievable without a CCD in transmission, which is important for silicon at 1550 nm. This also probably hinges on having better initial mode matching.

Attachments show the optical layout, and CCD videos of the p and s polarized transmissions as the laser frequency is swept.

Images attached to this report
Non-image files attached to this report
Comments related to this report
evanhall - 07:02, Wednesday 30 April 2025 (11948)

Matt, Swadha, Evan

We were able to PDH lock the cavity with a bandwidth of about 20 kHz. In the kilohertz range the driver and the laser current actuation have a transfer function that is neither flat nor 1/f (see attachment for David's measurements of this from Caltech — driver and laser should be identical to ours). We probably want to think carefully about pole placement here. Using 𝜈FSR = 2.9 GHz and finesse of 370 implies an optical pole at 𝜈HWHM = 3.9 MHz, which is currently irrelevant but could shift down by an order of magnitude if we redo the coupling with a larger gap.

We applied 49 MHz PDH sidebands with a fiber phase EOM that Melissa lent us (LNY10D2A 200 MHz X-cut). We are detecting in reflection with a New Focus 1811. The PDH error signal is not clean because (to give one reason at least) we are desperately in need of better mode matching. In transmission, the second-order HOMs are about 50% in height compared to TEM00.

Images attached to this comment
evanhall - 17:52, Thursday 01 May 2025 (11951)

Swadha, Evan

We continued to study the transverse modes of the cavity. We attempted better mode matching with a different lens arrangement but were only able to achieve a transmitted power of TEM02 = 96 mV and TEM20 = 112 mV versus TEM00 = 270 mV (dark noise: 18 mV), which is modestly better than before but not spectacular.

We were driving the laser current with a triangle sweep at a rate of 1 V / 4.2 ms. We applied phase sidebands at 49.2 MHz and found they appeared around the carrier at ±3.05 ms, enabling a calibration of the sweep time into frequency. The TEM00 FWHM was 3.24 ms, implying a FWHM of 52 MHz, and hence a finesse of 57 (s polarization). The laser current knobs were set at 6′00″ coarse and 7′30″ fine; temperature readback 7.27 kΩ.

Similarly, the TEM02 mode appears offset from TEM00 by 10.2 ms ( = 164 MHz) and the TEM20 mode appears offset by 13.4 ms ( = 216 MHz). The mode identifications were confirmed with the camera in transmission.

We loosened the setscrew pressing the transmission prism against the cavity, the finesse rose. The transmitted TEM00 power was now 205 mV with a FWHM time of 2.52 ms ( = 40 MHz) or finesse 72. Re-tightening the screw lowered the finesse again.

GRAVITES General
Haocun - posted 19:56, Thursday 24 April 2025 (11941)
GRAVITES CW and Single Photon Measurements [Vienna]

[Dorotea, Eric, Haocun]

We did a series of interferometer sensitivity measurements with 1550nm CW light and single photons from the Type2 source.

We explored the possibility of tuning the lock point at 1542 nm by adding an offset, to ensure that the locked 1550 nm light was centered on the fringe. Spectra with different offsets were taken but we found that introducing the offset added excess noise in the 10-100 mHz range (green and red traces in this Fig). Then we decided to adjust the lock point of 1550nm by gentally tapping the enclousure box.

The measurement taken on April 21 showed improved sensitivity compared to earlier data, probably because the temperature finally get more stabilzed. It nearly hits the dark noise limit of the Femto homodyne, so we had the low power customized homodyne set up. The measurement taken with on April 22 [Fig] gave the best sensitivity we've ever observed so far. We then injected a 1mV pk-pk dither at 125mHz, attenuated by 10dB (with an additional factor of 10 attenuation on servo board for excitation). The dither line appeared clearly in the spectrum (red) without any degradation in sensitivity.

Next, We switched to single photon interferometry setup using the Type2 source. We took an overnight measurement ~8.5Hr, using Dorotea's new script to record data lively. A similar dither at 100mHz was injected (amplitude corresponds to ~0.6mrad rms), and the dither line clearly appear in the photon spectrum [ASD and Spectra]. We also tried with the excitaiton amplitude reduced by a factor of 10 and took an 15-Hr measurement, unfortuntely we couldn't see that excitation showing up.

Measurements with higher photon rates using the Type-0 source will be very helpful for the next step. We also plan to use optical switches to alternate between CW and single-photon inputs, as well as between homodyne and photon-counting detectors.

Images attached to this report
General
sasha - posted 15:42, Wednesday 23 April 2025 - last comment - 11:51, Thursday 01 May 2025(11939)
Teraxion Laser Linearity Measurement Continued
I fit four oscilloscope time series measurement of the Teraxion 1550 nm laser in an asymmetric Michelson (arm length difference 17.5 cm) modulated with a sine wave (f = 1 kHz, A = 1V). The fits and corresponding residuals are shown in the first four figures. Fit parameters up to the fourth harmonic are shown in the table in the last file. 

m_i are effective modulation depths defined as m_i = 2*pi*(\Delta f)*\tau, where \Delta f is frequency modulation depth, and \tau is the optical pathlength delay, and m1 corresponds to the carrier frequency. The number of significant digits for the fit parameters corresponds to the fit errors from the covariance matrix (curve_fit). All measurements were taken with the same modulation, so parameter averages of the four measurements are given in the table and below. 

m1 = 2.81 +- 0.09
m2 = 0
m3 = 0.004 +- 0.008
m4 = 0.006 +- 0.007

If needed, a more precise linearity measurement can be made with cymac. 

Images attached to this report
Comments related to this report
sasha - 11:40, Thursday 24 April 2025 (11940)
Raw data from RIGOL oscilloscope and Python script that generates the plots are attached below.
Non-image files attached to this comment
sasha - 15:01, Friday 25 April 2025 (11943)
We noticed that the oscilloscope's ADC quantization steps are large for these measurements, so smoothing the data gives better fit results and more information. Attached are updated plots, fit parameters up to m6, and Python script. 

New modulation depths: 
m5 = 0.005 +- 0.004
m6 = 0
Images attached to this comment
Non-image files attached to this comment
evanhall - 11:51, Thursday 01 May 2025 (11950)

The attached script fits the nonlinearity using MCMC instead of least squares. It is slower and requires some babysitting of the sampler burn-in, but should explore the whole parameter space in the case that least squares might fail.

Neither of our scripts yet account for the modulation-induced AM, which should be added before making firm conclusions about the nonlinearity of the FM drive.

Non-image files attached to this comment
GRAVITES General
begum - posted 18:11, Tuesday 22 April 2025 (11938)
TEC recovery for prestabilization in-chamber temperature

After the power outage, Meerstetter TEC-1163 HR-2 (high resolution) input stopped reading resistance, so I was unable to recover the temperature control loop with the previous configuration. It seems this is an issue about the HR-2 input specifically, which coincides with the power outage for some reason. Low Res input 3 shows -90 °C all the time, although there is nothing connected to it. This threw me off for some time.
I changed the input to Connector 14 [attached photo top left connector], Pins 1&2 (instead of 3&4) and the sensor output (or HR input) is online on the software again. In case we need to recover this again, screenshots of the main screen, the HR1 input settings, TEC channel settings and PID controller screen are attached. This device is supplied with 32 V bias.
Note that the cables coming from the sensor are not directly soldered onto the pins but rather soldered in crimp terminals, so a firm pull will dislodge them, beware. 

Thanks to Eric for remote support from Vienna.

Images attached to this report
SQZ General
wenxuan - posted 13:27, Thursday 17 April 2025 (11934)
OPO loop recovered after broken PZT chassis

Daniel, Wen

We noticed the PZT driver chassis was no longer working when we tried to sweep OPO M4 PZT. All other driver ports of the same chassis were also dead. We switched to the chassis above (2nd from top) to recover the OPO loop. 

The loop was able to lock, but the loop stability was noticably worse than before. The OLTF looks fine but the gain setting has to be reduced 10 dB. The phase margin of the new loop was similar with before. We also checked the PDH error signals and phasing, which also looked fine. After some searching, we found the new PZT chassis is HV version and clips at high voltage. We then switched to the 1st chassis from top and solved the problem. The OPO CMB gain is now 12 dB with UGF 1.3 kHz and 60 deg phase margin (as seen in the plot).

The green in-vacuum fiber collimator was shaky as mentioned in previous logs. The problem is the loose screws of the collimator holder (we didn't have the set screw so used a flat head screw). It's now fixed. 

We noticed another resonance peak near fundamental resonances. It's likely to be a p-pol. We will examine the polarizer and waveplates in the green input path. 

The home position of the PI stage controller was also changed after power outage. We'll re-reference the position after fixing the polarization. 

Images attached to this report
RetroSense General
sasha - posted 17:10, Wednesday 16 April 2025 (11936)
Fixed RetroSense Grounding Issue + New Noise Plots
Sasha, Evan 

We fixed the grounding issue in RetroSense v3 and there is now no current leakage. The solution is using plastic screws on the optical head for isolation, but for future designs it would be preferable to use non-conducting components in the parts surrounding the LED. 

I centered the sensor and chose the distance between the LED and the retroreflector such that the diagonal signal in the centered position is equal to zero. The most recent noise measurement with this configuration is shown in the first figure. It includes the measurement of the SR560 preamp noise, since our dark noise is now at picometer level at frequencies above 1 Hz. 

The second figure shows the RetroSense sensitivity, range, and how voltage from different channels depends on displacement along the x-axis. Linear range is about [-0.2 mm, 0.2 mm].

This RetroSense version could already be placed on a suspension prototype for testing, so I will be looking into connecting it to the required readout. 

Measurement details:
dV/dx = 51.6 V/mm
V_sum = 14.0 V
I_sum = 549 uA
Shot noise = 8.36 pm/rtHz
Images attached to this report
LASTI General
wenxuan - posted 16:59, Wednesday 16 April 2025 (11933)
Recovery from power outage on 4/9

There was a power outage happened last Wednesday. It shut down all of the electronics in LASTI.

The model is restarted on m1sus, m1boot, and m1bsc, after rebooting them again after outage. All models are running now except for m1susfcquad. 

The Beckhoff was also turned off abruptly. I restarted twinCAT model and tcIOC batch file, but it was not working. The system manager complained about broken modules on SQZ chassis I, right rail. Those modules were indeed not blinking correctly as others. The Ethernet cable and power supplies are running OK. I've restart Beckhoff according to log11802. It took some time for the medm to be refreshed. The broken modules eventually went back up online after me disabling and re-enabling them and restarting. The common mode servo went back online and can lock the cavities. 

The power outage also broke the third LV PZT driver chassis. We'll report more details in the next log. 

LASTI General
lisabar - posted 13:37, Tuesday 15 April 2025 (11935)
Shipping two tip-tilts from MIT to LHO

Rich, Lisa

We are shipping two tip-tilt assemblies to Betsy at LHO.

These are:

https://ics.ligo-la.caltech.edu/JIRA/browse/Shipment-13851

and

https://ics.ligo-la.caltech.edu/JIRA/browse/Shipment-13852, which is the former input coupler of the LASTI 16m filter cavity.

We still have a tip-tilt in LASTI, inside the BSC, and that is 

https://ics.ligo-la.caltech.edu/JIRA/browse/ASSY-D1001396-12

SQZ General
wenxuan - posted 17:21, Wednesday 09 April 2025 (11931)
VOPO crystal position characterization

Daniel, Wen

We did a characterization of the crystal position while optimizing temperature given the translation stage is working. The stage has a limit of [-6 mm, 4.5 mm] range. The edge of the crystal was measured to be [-2 mm, 2.7 mm] by sweeping the cavity and finding the crystal position where it lost resonance. The stage position readback is very stable and repeatable. 

We first attempted to perform the optimization while keeping the VOPO locked, in hope to see more sensitive NLG changes at a slightly different crystal position. The OPO loop could stay locked with the stage moving less than 0.01 mm. It turned out the NLG was not sensitive at the level of 0.01 mm, or roughly 0.1 C change in temperature (while keeping the other variable optimized). So we went back to rough characterization by sweeping the VOPO cavity. 

We started sweeping by sending in 5 Vpp with 2.5 V offset at 5 Hz. We were able to see the coresonance fairly quickly, and recorded the best temperature for all the positions of the crystal. We were able to see a very linear and repeatable function of the optimized temperature versus the crystal translation. The plot will be uploaded once Daniel's alog account is active. 

Today we went back to examine some of the outlier/missing data points from the last experiment. This time we increased the sweep range of the OPO PZT by decreasing the scanning frequency (getting away from the pole). We were able to pinpoint some of the missing data from last time. We do notice that there's a damaged spot on the crystal (labeled plot to be uploaded), which misaligns both green and red beams. We plan to do a finer NLG comparison at those good positions we found during the measurement. 

GRAVITES Electronics
begum - posted 00:12, Monday 07 April 2025 (11930)
1550 HD new board, 1550 nm noise measurements through MZI

[Begum, Haocun]

We tried to debug HD board #4 for a while this week, the bias paths looks fine, problem unclear as of now, although diodes have been replaced with new ones, to be tested further later.
Haocun and I then built two new homodyne boards, #5 and #6. #5 works, so we assigned it to be the homodyne readout board for the 1550 nm path of the MZI. #6 has an unknown error, presenting as lower gain and small V_avg offset from 0, to be debugged later.

We measured the 1550 nm HD spectra, with the 1545 nm feedback loops (VCO + PZT stretcher) off and on, attached here. The noise falls below ~10^-2 rad/sqrt(Hz) above 0.2 Hz. The acoustic peak around 5 kHz shows up around the same strength for the 1550 nm noise and the 1545 nm with the feedback on. It is worth talking about the calibration for the 1550 nm noise units from V/sqrt(Hz) to rad/sqrt(Hz). The attached plot shows the 1545 nm HD readout signal in yellow and 1550 nm HD readout signal in purple. There is significant drift on the 1550 nm HD readout signal with the feedback on, so the calibration varies between Vpp/2 and 0. In the plots presented here, all 1550 nm HD signals are calibrated with Vpp/2 factor, corresponding to the lower limit for the noise curve. To see the effect of this calibration drift, we also took faster measurements (102 kHz bandwidth, and 10 kHz bandwidth on SR785) when the 1550 nm HD signal is ~0 V. The attached plot shows these measurements with purple and green traces. The high frequency portion of the composite measurement for 1550 nm signal matches well with these. I will add acoustic shielding boxes around the fiber spools, which will also act as a temperature enclosures. This should help with the stability. We are waiting on the new fiber strecher with broader range and lower pzt capacitance (11823). This will affect the low frequency noise suppression.

I also measured the 1550 nm HD spectra, while rotating the MZ arms to 340 degrees and back. The attached plot shows the composite trace measured during rotating in yellow. I slowed down the stage to a maximum of 1 degrees/s, which seems to have helped with the acoustic coupling I saw before with the 1545 nm signal, see here for past measurement. Noise injection due to rotation is more apparent >10kHz for the 1550 nm trace.

Images attached to this report
GRAVITES Electronics
haocun - posted 05:10, Sunday 06 April 2025 (11929)
Abnormal 5 kHz Oscillations with SR785 at EE Shop

[Begum, Haocun]

We assembled two additional homodyne boards for GRAVITES, labeled #5 and #6.  #4 has been fitted with new diodes and requires further testing to confirm proper functionality. #5 is performing well and is currently in use. #6 appears to have lower gain, but I haven’t finished debugging yet.

The 1550 nm homodyne is now online for the rotating interferometer. We took dual-wavelength measurements with interfeormeter locked; Begum will follow up with detailed results.

[General] We saw abnormal 5 kHz peaks and their harmonics in the dark noise spectrum of the homodynes when measured with the SR785 in the EE shop, see here. I repeated the measurement using a different power supply, but the peaks remained. These oscillations do not appear when using the SR785 in the green lab, see here. The SR785 in the EE shop probably needs fixing.

Images attached to this report
SQZ Site VOPO
eoelker - posted 16:18, Friday 04 April 2025 (11928)
VOPO Spare Oven D9 wire swap

[Slawek, Eric]

We've been having a lot of issues with the D9 for the oven.  The Peltier leads have broken off and the thermistor leads have also come disconnected several times.  One issue is the fact that the oven wires (uhv compatible Kapton coated) are way too stiff and can easily apply too much tension to the solder connections.  We decided to swap the wires for CZ-1105 wire from Crooner.  This is the wire used for the OSEM assemblies.  It is teflon coated and is much more flexible.  The gague is only AWG-28, but we contacted the manufacturer and they claim it can support up to 3 Amps of current and 300V.  The peltier uses < 1A of current, so this should be sufficient.

I rewired all of the thermistors this afternoon and Slawek will repair the peltier connections next week.

SQZ General
eoelker - posted 16:10, Friday 04 April 2025 (11927)
Pinout for VOPO build station D25 breakout cable

I am using an old in-vacuum cable from HAM X to connect to the VOPO spare (the two D9s on the VOPO base and the D15 for the PI stage).  This cable has 2x D9s and a D15 on one side, and a D25 M connector on the other side.  I couldn't find a pinout in the LASTI SQZ documentation, so I measured it myself:

 

Left Side:  D25 Connector Pin number,    Right Side:  VOPO side connections

1 - > PZT D9, Pin 1:  M2 PZT -

2 -> PZT D9, Pin 6:  M2 PZT +

3 -> PZT D9, Pin 5:  M4 PZT -

4 -> PZT D9, Pin 9:  M4 PZT +

5 ->  NC

6 -> D15, Pin 1

7-> D15, Pin 2

8-> D15, Pin 3

9-> D15, Pin 4

10-> D15, Pin 5

11-> D15, Pin 6

12-> D15, Pin 7

13-> D15, Pin 8

14 -> Oven D9, Pin 1: in-loop thermistor

15 -> Oven D9, Pin 6: in-loop thermistor

16 -> Oven D9, Pin 3 (not used)

17 -> Oven D9, Pin 9:  Oven TEC

18 -> Oven D9, Pin 5:  Oven TEC

19-> D15, Pin 9

20-> D15, Pin 10

21-> D15, Pin 11

22-> D15, Pin 12

23-> D15, Pin 13

24-> D15, Pin 14

25-> D15, Pin 15

GRAVITES Electronics
Haocun - posted 07:57, Thursday 09 January 2025 - last comment - 07:49, Friday 11 April 2025(11857)
GRAVITES Servo Board Updates [Vienna]

[Eric, Haocun]

We were able to lock the whole interferometer (50km fiber on each arm) with the servo board duplicated and shipped by Eric 11773.

Finding that we were limited by VCO noise by using 50kHz span [Fig. 1 VCO noises vs. different modulation deviation span], the following modifications were made to increase the servo gain for using 500Hz VCO span:

Homodyne:

Common Servo:

Split Servo:

       VCO Path:

       PZT Path:

With these cnfigurations, we have OLTF transfer function and crossover measured [Fig 2 & 3]. OLTF UGF is around 15kHz, and the crossover was set to around 200Hz.

TODO:

 

Images attached to this report
Comments related to this report
eoelker - 07:49, Friday 11 April 2025 (11932)

[Dorotea, Eric]

After turning down the input power at 1542 nm to reduce coincidence counts from raman scattering, we found that the overall gain of the servo board is now too low.  This causes the servo to become unstable when the 2nd common integrator is turned on.

To fix this, we are increasing the fixed gain on the common stage of the board by another +20 dB to compensate for the drop in optical power.  This should give us a UGF in the 5-10 kHz range, which should roughly match what we were operating with in January/Febuary.  I think we could push it up further, but we'll start with this to restore the old behavior.  

We increased the gain of the 3rd stage (U6 on the board) by 20 dB while keeping its integrator corner roughly the same.

R18:  10k -> 976 ohm

R31:  0 -> 8.87k