[Eric, Daniel]
Last week we achieved a state in the LASTI squeezer where we observed about 7dB of squeezing. The conditions for this were as follows
| Pump Power | 18 mW |
| VOPO Crystal Temp | 32.32 C |
| CLF Power | 60 uW |
| LO Power | 1.5 mW |
| Stage Position | -0.690 mm (-0.780 mm Neg Lim Sw) |
| Homodyne Visibility | 98.6 % |
Here are the captured homodyne spectra (raw measurement, squeezing plot)
We saw around 13dB of antisqueezing in this state, but unfortunately no trace was captured as the VOPO TEC leads disconnected again mid measurement. They have been temporarily repaired, but it seems like the TEC leads may be long enough that they are rubbing on the lid of the cavity so we are leaving that off for the time being until we replace them.
We also captured the I/Q data from the LO and CLF loop demodulators to get an idea of how the phase noise looked at the same time. Here is the post-processed phase noise breakdown. So far, it seems that over 100kHz of bandwidth our integrated RMS phase noise for both loops is as follows
| Loop | Locked In-Loop Phase Noise (RMS) |
| CLF | 9 mrad (mostly from servo bump) |
| LO | 0.9 mrad |
Inbetween this and yesterday, Eric noticed that the SHG cavity was misaligned and in fixing that, he was able to get much more pump power over to the squeezer table. Characterizing the NLG our current (seemingly good) spot as a function of pump power incident is shown here
| Pump Power (mW) | NLG | Temp (C) | Stage (mm) |
| 6.8 | 3.0 | 32.32 | -0.781 |
| 18.0 | 10.57 | 32.12 | -0.781 |
| 26.4 | 36.4 | 31.96 | -0.781 |
| 26.4 | 38.6 | 32.08 | -0.7693 |
We are now no longer limited by pump power and can achieve very high NLG with more modest pump powers. With this data, we plan to characterize squeezing as a function of NLG to determine whether we are phase noise or loss limited and by how much.
My first theory: the beam isn't centered vertically on the crystal and I'm getting some diffraction loss at 1064 nm. Based on my video evidence, this doesn't seem to be the case at 532 nm. I assume that the two modes are well overlapped given how great the NLG is.
I popped off the lid of the VOPO and took some videos of the front and back surface of the crystal when the cavity is flashing for 532 nm. In the front video, the spot is clearly centered vertically. In the 2nd video (which is a bit off axis because I have less room in the cleanroom on that side), I initially thought that it wasn't centered vertically, but then I zoomed in on the playback video and realized that I was getting confused by the indium foil sticking out underneath the crystal.
My second theory: During the build for this VOPO, three of the mirrors fell off due to aging in the glue joints (see log ). Slawek believes that the original gluing procedure used to assemble the mirrors in 2020 contributed to this failure. We cleaned the mirrors and slawek re-glued them. We think we did a decent job, but this is a likely cause of excess loss. The status of all 4 mirrors:
M1: This was the only mirror with a good glue joint. This should be as pristine as it was when glued in 2020.
M2: This is the 2nd M2 assembly that did not fail. The one that fell off will be used for the second spare cavity.
M3: This mirror was cleaned and re-used after falling off (see log). This one could have excess loss. I believe we do have one more M3 mirror that was glued together using a new optic. Perhaps we could try swapping.
M4: This one was glued together using a brand new mirror from the box
This issue is likely to be a bigger problem for the 2nd spare cavity. Both M2 and M4 for that cavity will come from salvaged mirrors that fell off.
Third theory: 532 nm loss and NLG are properties that depend more on the crystal bulk rather than the surface. However, excess loss at this level for 1064 could easily be coming from a bad crystal surface. If it were surface damage, I'd expect some variability in escape efficiency across the crystal surface. If it is due to a larger than anticipated reflection from the AR surface, however, that would be more uniform. This is probably the scariest theory, since it would mean that all the crystals will yield a similar escape efficiency since they were all coated in one run (we do have some new ones at ANU which were coated in November).
Based on the AR coating scans here https://dcc.ligo.org/LIGO-E1800242, we should have something like 0.14% loss from the two AR coated surfaces, so I'm only looking at like 0.1% of excess loss. Probably too large to be absorption in the crystal (spec for KTP is 1<100 ppm/cm at 1064 nm)
I'll note that this is the first time we're using this batch of crystals to build VOPOs (they've been our spares since 2018, though i think Hudson's mini OPO is using this batch as well). All previous builds uesd the 2015 crystals. Perhaps then it isn't too surprising that the 532 and 1064 losses are a bit different than before?
The O4 mirror reflectivity design is really unforgiving in terms of escape efficiency (we nearly doubled the 1064 nm finesse relative to O3 to lower threshold). To summarize, our known losses are:
M1: 92.7 - 93.3% reflectivity based on MIT measurement (unfortunately I don't know which mirror I'm using) https://dcc.ligo.org/LIGO-T2000534
M2: 99.9 +/- 0.025% reflectivity based on spec (MIT measurement agrees but has large uncertainty) https://dcc.ligo.org/LIGO-T2000534
M3 & M4: 99.993% reflectivity based on MIT measurement https://dcc.ligo.org/LIGO-E1800011
Crystal AR coating: 0.07% reflectivity per surface in the vendor curve https://dcc.ligo.org/LIGO-E1800242
With these values, our range for the maximum possible escape efficiency is 96.3-97.1%. Without the crystal AR losses, the maximum escape efficiency is 98.1-98.8%.
I'll note that the current H1 VOPO has an escape efficiency consistant with effectively no loss from the crystal AR coatings using the 2015 crystals: alog. This is despite the fact that the vendor curve looks similar for that batch (0.06% per surface).
I may be able to constrain this better by calibrating my reflection and transmission signals in terms of power. then I can solve for both loss and the input mirror reflectivity.
Figured out that my issue yesterday was clipping on the refl PD for 1064
Now I'm getting something that makes more sense for the escape efficiency
mode mismatch was constant at 100 mV, so the change is in TEM01
Unfortunately our escape efficiency sucks right now. Values here assume an M1 reflectivity of 93% and an M2 reflectivity of 99.9%
| Position | Reflection off of Resonance (V) | Reflection Dip (V) | trans TEM00 (V) | trans HOM Sum (V) | Excess Loss | Escape efficiency |
| -1100 | 5.28 | 0.880 | 2.24 | 0.220 | 0.27% | 95.4% |
| -2000 | 5.32 | 0.880 | 2.20 | 0.220 | 0.27% | 95.4% |
| -2900 | 5.32 | 0.880 | 2.20 | 0.220 | 0.27% | 95.4% |
| -3800 | 5.32 | 0.880 | 2.20 | 0.220 | 0.27% | 95.4% |
| -4700 | 5.32 | 0.840 | 2.12 | 0.310 | 0.26% | 95.5% |
| -5800 | 5.32 | 0.840 | 2.12 | 0.310 | 0.26% | 95.5% |
| -6900 | 5.32 | 0.840 | 2.12 | 0.310 | 0.26% | 95.5% |
| -8000 | 5.32 | 0.840 | 2.15 | 0.280 | 0.26% | 95.5% |
| -9000 | 5.32 | 0.840 | 2.14 | 0.290 | 0.26% | 95.5% |
I'll note that we see the same HOM spacing (0.27 FSR) and change in alignment across the wedge (~3%) as in the 532 nm scan. I estimate that the finesse is roughly 86
Did a reflection dip measurement to determine the 532 nm losses. Solving for the excess loss actually gives a double valued result, either an excess loss of 5% or an excess loss of 0.8% assuming an input mirror reflectivity of 98%. However, the latter value is more consistant with the measured finesse for 532 nm (measured 190, calculate 220 with 0.8% and 90 with 5% excess loss). I could constrain by using the transmitted power value as well, but haven't calibrated these voltages in terms of power yet.
This is quite good compared to some values observed towards the end of O3 (was 2-3% with an older crystal): https://dcc.ligo.org/LIGO-T2000150
Presumably this is just because this is a brand new crystal?
Again we see about a 3% change in misalignment across the entire wedge.
I tried to measure escape efficiency at 1064 nm as well, but things aren't making sense yet. Need to debug tomorrow.
| Position | Off Res | On Res |
TEM00 trans |
TEM01 Trans |
02+20 trans |
Excess Loss |
| -200 | ||||||
| -1100 | .222 | .094 | 1.86 | .140 | .460 | 0.8% |
| -2000 | .224 | .092 | 1.92 | .120 | .440 | 0.8% |
| -2900 | .226 | .090 | 1.86 | .120 | .460 | 0.8% |
| -3800 | .226 | .090 | 1.94 | .60 | .460 | 0.7% |
| -4700 | .226 | .090 | 1.98 | .60 | .440 | 0.7% |
| -5700 | .226 | .090 | 1.98 | .60 | .440 | 0.7% |
| -6800 | .226 | .090 | 1.96 | .100 | .420 | 0.8% |
| -7900 | .226 | .090 | 1.96 | .160 | .420 | 0.8% |
| -9000 | .228 | .090 | 1.98 | .160 | .420 | 0.8% |
PZT scan reveals the following HOM spacing for 532 nm
| HOM | Delay from TEM00 | Fraction of FSR |
| TEM00 | 0 | 0 |
| TEM40 | 17 ms | 1.08 |
| TEM10 | 58 ms | 0.27 |
| TEM20+02 | 116 ms | 0.54 |
| TEM00 | 215 ms | 1.0 |
This should be OK. The first 10 modes are several FWHM away from the TEM00 resonance.
The PZT resonances are a bit lower than some that Maggie measured for O3: 4.8 kHz for M4 and 5.6 kHz for M2 ( maggie was getting more like 6-7 kHz). They seem to be OK, however. I can still lock with a UGF > 1kHz for the OPO with an appropriate notch filter.
This morning I did a scan across all the co-resonance positions and characterized the NLG at each for the phase matching temp I measured yesterday 35.25C. I believe that I was at point -1800 yesterday
Green power incident 2.2mW and was fairly constant across the measurement.
Peak power off co-resonance for 1064 nm = 320 mV
| Stage Position | Delta V | NLG |
| -200 | 1064 losses high | |
| -1100 | 2.58V | 8.1 |
| -1800 | 2.56V | 8.0 |
| -2700 | 2.44V | 7.6 |
| -3600 | 2.40V | 7.6 |
| -4650 | 2.44V | 7.6 |
| -5800 | 2.64V | 8.3 |
| -6900 | 2.80V | 8.8 |
| -8000 | 2.54V | 8.0 |
| -9000 | 2.58V | 8.1 |
Crystal edges (where cavity flashes go away) are at 0 and -9700
NLG looks fairly uniform. I did not optimize the phase matching temp at each position, i just used 35.25C for all.
I did a scan on Friday to find all the co-resonance positions on the crystal when the temp was set to 35.25C. The edges of the crystal were at 0 and -9700 counts on the translation stage software (it does not keep track, so I did this manually). I found 10 coresonances in all, though the one at -200 had very poor transmission for 1064 nm and is likely not useable. I was curious to see how much the 532 nm cavity misalignment changes across the entire crystal. I started w a slight misalignment (a couple percent) so i could keep track of the mode. I saw values between 2.3% - 4.7% across the entire face of the crystal. The data was a bit noisy so i suspect some of this is just drift from the temp fluctuations/air currents in the cleanroom or from me bumping the mirrors when I was touching up the fiber polarization. I haven't found a super precise record of this anywhere, but this doesn't seem like a huge level of change over the entire wedge. Its not obvious to me that there is even a crystal alignment that would get rid of this effect entirely. I'll have to look back at John's design document for the VOPO to see if he estimated the magnitude of this effect.
All values here are for 532 nm. I'll do this for 1064 when I'm characterizing the escape efficiency tomorrow.
| Stage pos | TEM00 | TEM01 | TEM02+20 | % misalignment |
|---|---|---|---|---|
| -200 | 1.00 | 0.040 | 0.240 | 3.13% |
| -1100 | 1.03 | 0.050 | 0.250 | 3.75% |
| -2000 | 1.10 | 0.040 | 0.260 | 2.86% |
| -2900 | 1.10 | 0.032 | 0.260 | 2.30% |
| -3800 | 1.10 | 0.056 | 0.260 | 3.95% |
| -4750 | 1.14 | 0.056 | 0.270 | 3.80% |
| -5700 | 1.15 | 0.064 | 0.270 | 4.30% |
| -6800 | 1.18 | 0.072 | 0.280 | 4.70% |
| -7900 | 1.18 | 0.064 | 0.280 | 4.2% |
| -9000 | 1.18 | 0.064 | 0.280 | 4.2% |
I did a temp scan with the H1 Spare VOPO today to further optimise the NLG. Turns out that my starting value was already pretty much right on (35.25C)
Input green power 1.7mW. Seed w/o amplification 320 mV. At each temp i moved the stage a bit to recover co-resonance. I'm at one of the middle co-resonances at the moment. FWHM for phase matching appears to be about 4C, which is roughly consistant with my recollection for this length of PPKTP crystal.
| Temp | Delta V | NLG |
|---|---|---|
| 32.25 | 0.624 | 1.95 |
| 33.25 | 0.912 | 2.85 |
| 34.25 | 1.36 | 4.3 |
| 35.25 | 1.84 | 5.75 |
| 36.25 | 1.38 | 4.3 |
| 37.35 | 0.872 | 2.72 |
| 38.25 | .628 | 1.97 |
At the peak phase matching, this corresponds to an incident power of about 5.1 mW for 532 nm at threshold. Since I only have about 80% mode matching, this means that my threshold power is 4 mW for TEM00. This agrees with my estimation at a much higher nonlinear gain on friday (3.4 mW incident, nlg =29.5).
Now that I have the tranlation stage working, I am able to set the cavity to co-resonance. I realigned the rear seed beam, set up a camera/PD in transmission, and tuned the stage position so the TEM00 modes for the Pump and Seed fields would be co-resonant at ~35C. Immediately I saw the 1064 peak being amplified by a factor of about 3.5 with 1.5 mW of pump power when i scan through resonance.
Last week, I set up an improved locking setup using one of my old LabJack servo boards. I will lock tomorrow and finish optimising the NLG. So far, while scanning the crystal position, I do not see any obvious misalignment of the cavity in the cavity scans. This suggests that the crystal may already be fairly well aligned, but I'll characterize more thoroughly tomorrow.
[Daniel, Eric]
We've now observed up to 6dB of squeezing with the LASTI VOPO setup. We're in the process of characterizing our losses to track down where our degredation is coming from at present. Today we decided to measure the escape efficiency of the cavity. We set up a front seed path. I had a mode matching solution based on the rear seed path for the O4 cavities, but it didn't work so well with the O3 collimator and fiber we're currently using. We actually found that our mode mismatch wasn't too bad with no lens in the path: 82.8% of the power in the TEM00 mode compared to a sum of all modes.
By measuring the reflection dip, we were able to estimate the intracavity loss and the escape efficiency. We saw 7.24V off resonance and 6.82V when on resonance for TEM00. We had to turn the seed power down quite a bit to avoid generating 532nm. Assuming an M2 transmissivity of 0.15%, we estimate that we have an excess loss of roughly 0.09%, likely just limited by the AR coatings on the PPKTP crystal. This corresponds to an escape efficiency of 98.2%.
Along with our relatively high homodyne visibility (98%), high PD quantum efficiency (~98%), and low propagation losses (need to measure more precisely but not more than a percent or so), we feel like we should be seeing more squeezing.
A couple of things to check: We haven't done much optimization of the loops yet. Should check RMS levels to make sure we don't have a massive phase noise issue. Also, we suspect we're using a bit too much CLF power at the moment.
I'm ready to characterize the nonlinear gain and escape efficiency of the H1 Spare VOPO, but ran into a roadblock when I couldn't move the translation stage! Turns out that the stage without the sensor requires a different controller. Fortunately, I was able to find one buried in the Orange Lab. After contacting PI for the pinout, I was able to make a cable to connect to the site stages and move them. Everything is set up to run on the Beckhoff machine now.
There was no manual for the controller available online. I had to get that from the manufacturer as well. Its now uploaded to the DCC with the rest of the stage information. Tomorrow we'll start scanning the crystal position.
Eric, Daniel
Over the past week, we have been optimizing the nonlinear gain of the OPO in the LASTI squeezer. Last Friday, we pretty quickly realized that the pump polarization had been rotated completely into p-polarization. This was preveting us from seeing nonlinear gains much greater than ~1.8 (as the pump and seed were not co-polarized) and hence limited the amount of squeezing we saw. After correcting this polarization, we could immediately attain nonlinear gains around 10, so we moved on to sweeping over all the co-resonance conditions of the crystal to find the best spots on the crystal to more carefully optimize. Following the same procedure as our previous alog, we found a collection of coresonant conditions over multiple FSRs and crystal positions and plotted these here. Unlike the previous measurement, this time we tried to roughly quantify the nonlinear gain in a given state by noting the peak amplitude of the seed transmission w.r.t the seed transmission amplitude with no pump applied. Here, this NLG estimate is shown as a color code on the scatter data. The best spot we happened across was as a stage position of 2.0mm and at a temperature of 32.51C. In this case, the rough NLG estimate was around 27, but when we actually lock the OPO at this point, we find that it is much lower (around 11). We think this may be due to thermal lensing in the crystal due to the high pump power (14mW) and will be trying a few things to confirm this soon. Regardless, locking the OPO in this state yielded us ~4dB of squeezing and around 10dB of antisqueezing across a band from 400Hz on up through the bandwidth of the homodyne detector.
Moving forward, we will be checking the losses in our homodyne setup as well as the escape efficiency of the OPO to see if we are currently limitted by losses. Once we have optimized this setup to its limits, we then plan on switching out the old PPKTP crystal for the new ones we purchased.
Jaime, Begum
This is for the ADTR project that aims to generate higher order modes for active PI damping (https://dcc.ligo.org/LIGO-L2400062). The goal of this script is to generate two phase masks to convert a gaussian mode to a higher order mode. I used the code from this paper (https://www.nature.com/articles/s41467-019-09840-4) and this repo (https://github.com/nicolassmith/alm) as a starting point.
The code starts with defining the grid which depends on the chosen pixel count and size as these will vary for different SLM's. Next the gaussian beam and target modes are chosen and the fields are placed in an array. I then defined a padded grid and padded transfer function of free space to avoid wrap-around on the edge of the simulation during the FFT propagation. Finally, I implemented the Gerchberg-Saxton algorithm for generating the phase masks needed to convert the gaussian mode to the desired mode.
Algorithmic workflow:
Forward propagation:
1) Update the first plane's phase mask by computing the local overlap ( dMask = [FwdField*conj(BwdField)]/sqrt(PwrFwdField*PwrBwdField) ), then finding the phase difference between dMask and the current mask ( dPhi = sum(dMask*conj(MASK)) ), and finally completing the update by adding the local overlap into the mask accumulator with a phase correction to ensure constructive interference with current mask ( MSK = MSK + dMask*exp(-i*np.angle(dPhi)) )
2) Apply the mask at the first plane
3) Pad the field (with lossy edges to get rid of numerical reflections from the edges of FoV) and propagate over the plane distance using the padded transfer function of free space
4) Crop the forward field back to original size and store it's values at the second plane
Backward propagation:
1) Update the second plane's phase mask
2) Apply the mask at the second plane
3) Pad the field and propagate backward over the plane distance using the conjugated transfer function of free space
4) Crop the backward field back to original size and store it's values at the first plane
Finally, I plot the masks and beams and calculate the overlap efficiency. Typically, after about 20-30 iterations, we get above 95% overlap between the propagated and target beams. I am attaching an example of calculated phase masks for converting to the LG(7,3) mode and an animation of the beam propagation including a line cut of the amplitude and phase along the x-axis. I also included a beam propagation animation with and without padding for converting to the LG(1,0) mode to illustrate the reduction of wrap-around in the simulation.
Calibration of frequency modulation:
Test after SHG:
Test with waveguides:
Other remarks:
This late alog is on noise measurements done on April 24th and April 26th using the 1550 nm HD and 1545 nm HD error signals, with the old pzt in the path vs the new pzt in the path.
Before installing the new PZT strecher, I did a quick measurement of the noise on the old one. Note the old one had shown some low pass behavior, apparent in the homodyne pzt and vco loops crossover frequency measurement (see plot from 11914). The previous measurement with the old stretcher had showed 0.9 mm for 150 V pzt calibration(11823).
The spectra measured on April 24th is shown in this plot, with blue (green) and red (purple) traces showing the 1545 nm HD and 1550 nm HD signals respectively, for April 24th (April 4th). While the 1550 nm HD signals look similar for the April 24th and the April 4th reference, the 1545 nm HD signals (SR785 noise floor limited at low frequency) on April 24th show lower noise floor <100 Hz. The Vpp value for April 24th was 1.7 V while for April 4th it was about 1.3 V.
Another thing to note here is that the the relative lock-point drift between the 1545 nm HD and the 1550 nm HD on the April 4th measurement (see plot) is gone! 1550 nm HD now lives near 0 pretty consistently, see purple trace on the oscilloscope screen.
On April 26th, I swapped out the pzt driver with the new one (same as Vienna now), and coiled 16 m of fiber around it, using double sticky tape on every layer. I haven't measured the pzt efficiency and the new crossover frequnecy for the 1545 nm HD. The spectra of the locked MZI are attached here. Comparison plot with Apth 24th measurement with the old pzt shows pretty consistent behavior (see here). The relatively higher noise for the 04-26 measurement seems to be consistent with the Vpp (1.5 V), see SR785 dark noise in plot.
Note, all April 24th and 26th measurements are done with the rotation stage powered on but not rotating.
SHG temperature:
Conversion efficiency measurement:
| input power (mW) | 135 | 200 | 265 | 330 | 390 | 460 | 520 | 585 | 650 | 720 | 785 | 840 | 905 | 980 | 1000 | (1135) | (1305) |
| output power (mW) | 6.7 | 14.2 | 24 | 35.1 | 48 | 63.1 | 79.3 | 96.4 | 115.2 | 134.8 | 152.1 | 171.2 | 194.5 | 214.7 | 221.5 | 262 | 311.4 |
Eric, Wen, Daniel
We have been neglecting alogs for LASTI over the past couple weeks, so this one is pretty long. The TLDR is good news: we are fairly confident that we are observing squeezing levels of ~0.6dB and anti-squeezing levels of ~1.2dB in the LASTI lab with all control loops and subsystems functional at some capacity. The chamber is not pumped down and alignments are still very rough, so these values are for high frequencies well above the bulk of acoustic noise (see screenshots below for squeezing and here for anti squeezing homodyne spectra).
Control Loops (CLF, LO, Red PDH)
Our last update mentioned some brief notes about the state of the control loops in the LASTI setup. Since then, we were able to reinstall all the VCOs necessary for the CLF loop, including an LO signal generator in the controls rack which we phase-locked to the VCOs that drive the AOMs in the CLF path using it's internal 10MHz reference. This IFR generator in the controls rack generates both the 9 MHz and 18 MHz local oscillator signals for LO and CLF demodulation respectively. It should be noted that our LO generation path no longer includes a PZT for LO phase actuation. Instead, it uses a pair of series AOMs much like the CLF path with the only difference being that the two AOMs are driven at the same frequency (80MHz) with one driven by a VCO that we actuate with the LO demod error signal. This allows the LO phase to track and lock over an infinite phase drift, whereas the PZT has a limited phase tracking range. With all of this in place, we were able to successfully generate error signals for both the LO and CLF loops. These loops are now locking without having been tuned or optimized much. The CLF loop OLTF seems to be okay without much adjustment. Eric pointed out that the transfer function for the LO CMB is likely no longer stable due to the additional phase integration added resulting from frequency actuation as opposed to the old phase actuation approach. This is likely the case, however we have been able to get by with reduced loop gain to compensate.
In addition to these critical loops, we also found the red PDH loop we temporarily installed for locking the OPO for homodyne alignment to come in handy quite often, so we semi-permenantly installed it to share the OPO CMB and reused the filter-cavity PDH demodulator to generate an error signal. The dither frequency is selected to be 18 MHz to make it possible to hijack the CLF PD signal for feedback. In our current configuration, the 18 MHz LO signal generated by the IFR in the rack is sent to an unused RF distribution amplifier and the 26 dBm output of this drives the seed EOM on the table. The lower-level 13dBm outputs drive both the red PDH demodulator as well as the CLF loop demodulator. To use this loop, the CLF PD must be disconnected from the CLF demodulator and re-routed to the red PDH demodulator. The red PDH error signal is connected to the OPO CMB channel-2 input. So locking either can be achieved quickly by choosing which error signal to use (i.e. green or red). In the coming days, we will update this post with a detailed block diagram of the setup to make future use easier.
Pump Power Increase Attempts
Throughout a lot of this setup, we were operating with reduced pump power due to excessive loss in the pump fiber feedthrough on our vacuum chamber. In order to maximize squeezing, we needed to reduce this loss to access more pump power at the OPO. During some debugging of this feedthrough it failed completely, so we needed to run a temporary in-vacuum patch cord through the chamber wall with the feedthrough removed. To keep everything clean, we foiled over the patch cord and the feedthrough flange. Eric was able to find us a spare green feedthrough and we were able to measure an 80% transmission through it. However, it appears that it's ferrule is damaged with the fiber core unsupported at the tip. The mode output from the fiber seems okay, but it remains unclear if this will be usable for us or if it will need to be reconnectorized.
70 KHz Coherent Noise in OPO REFL BBPD and CLF BBPD
After turning up the pump power into the OPO, we discovered some mysterious noise in almost every RF error signal in the setup. While debugging the PDH and CLF demod error signals, we noticed a large amount of high-frequency noise on both demodulations. This manifested as a 70 KHz signal superimposed on the error signal and was most pronounced near OPO resonances. Upon further investigation, we observed strong sidebands 70 KHz from the PDH dither carrier on the OPO REFL BBPD signal here. These sidebands were so strong that at the time we observed them they were preventing us from locking the OPO. To at least get the OPO PDH loop stable, we made a 70 KHz notch filter and put it between the OPO PDH demodulator and the OPO CMB. The schematic for this filter is shown here and has this transfer function. Placing this in the loop allowed us to lock the OPO, but we still see these 70 KHz sidebands on the OPO REFL BBPD regardless. Similar sidebands are present on the CLF PD RF signal as well. We aren't yet sure of the origin of this oscillation, but we will be investigating these PD's power supplies for ripple as well as the possibility that back reflections from the OPO are causing this.
Homodyne Repeat Diff Channel Failure
Once this problem was at least temporarily resolved, we discovered that our homodyne detector's differential channel had failed yet again. Like the previous time, the 100k feedback resistor on the difference channel TIA had blown open. My only guess for why this is happening is that something is physically wrong with this batch of 100k 0805 resistors in the lab that causes them to be particularly unreliable. It shouldn't be possible for these to be overdissipated by the voltages present in the HD circuit. To prevent this failure from occuring again, we instead replaced the 100k feedback resistor with a 1/4 watt through-hole one (see bodge here).
At this time, the gain configuration for this homodyne is as follows:
DIfference TIA Gain: 100k
PD1/PD2 TIA Gain: 5k
We only quickly realigned the homodyne detector after repairs, but we think the visiblity is still probably > 90%. We are using 1mW of LO on each PD for now.
Squeezing Observation
With everything roughly set up, we visually/approximately recovered the crystal position reference from our previous sweep of the OPO's coresonant conditions in this alog. We found a fairly good spot on the crystal at a temperature of 35.02 C and a stage displacement of 1.0mm. We measured our pump input power to be 8.6 mW (12.6 uW transmitted) and our CLF input power to be 290 uW. After locking all loops, we were able to observe ~1.2 dB of anti-squeezing with the LO-locking demod phase angle set to -25 degrees relative to the CLF 2F phase (using IFR's carrier phase adjustement). Here is the homodyne difference noise spectrum from this state. Rotating the LO demod angle by -90 degrees to -115 degrees resulted in this noise spectrum showing ~0.6dB squeezing. In both screenshots, the dim trace is the shot-noise floor of the LO only (SQZ blocked) on the homodyne detector. There isn't much measurable squeezing or anti squeezing at frequencies lower than this most likely due to acoustic noise.
Next Steps
Moving forward, we have several improvements planned. Eric purchased some PM fibers for the pump and CLF signals to run between tables. We expect that this will make it much easier to maintain pump and CLF power as the fiber polarization will no longer be changing widely as the lab temperature fluctuates. We also likely have a lot of alignment tweaking to do on the homodyne detector and OPO phase matching to maximize nonlinear gain, which seems to be low at the moment. Once we get as much as we can out of the current configuration, we also plan on throwing in a newer OPO crystal to compare it's performance to the old one.
Swadha, Evan
We measured the load impedance of a Brimrose AMF-27.5-1064 AOM.
This AOM is impedance matched for a 27.5 MHz drive, but we might like to match it to 36.7 MHz instead. In that case it may be possible with a simple CL network: 220 pF capacitor from source to ground, followed by a series inductor tuned to around 70 nH.
The NW17 Agilent 4395A high bandwidth network analyzer is not working again. I'm very annoyed. This guy just came back from being repaired like 6 weeks ago.