Displaying reports 221-240 of 1025.Go to page Start 8 9 10 11 12 13 14 15 16 End
Reports until 16:43, Friday 15 August 2025
GRAVITES General
dorotea - posted 16:43, Friday 15 August 2025 (12042)
Input segment for 2fl measurement spliced and tested

This week I spliced all the components for the input to the 2fl measurement, put them in an enclosure, and tested the loss. I didn't prioritize keeping the fibers as short in order to leave room for redoing parts of the setup. A diagram of components is attached; note that the 1st switch (which switches between CW/photon source, and would be located before the '1550 in' line on the diagram) is not shown in since that component is being tested in Vienna-- we eventually will want to add that switch to this assembly, and I lean towards using the CrystaLatch switches that we have here, since we can run multiple off of one function generator to keep everything well synced. Currently I just have all the components laid out between pieces of acoustic foam and shoved into the box-- if we need better isolation than this we can add more insulation to the outside. I haven't set up any kind of temperature control yet but plan to just stick an omega pad on top of the box, put a thermistor inside, and wrap the whole thing in thermal insulation. I also redesigned my optical switch drivers in a split configuration so that the majority of the circuit (especially the voltage converter) can stay outside of the box and the signal just goes in to the box and connects to the switch through an sma cable. 

Begum and I tested the loss for both wavelengths on each output (results are shown in an attached image). Note that the loss looks higher for the 1542 because of the 50/25/25 beamsplitter being used. In actuality the loss is the worst on the channels with the switch. The switches seem to have around half a dB of input loss. Each channel has at most 1.1dB of total loss between the input and output (with the 'outputs' here being where the light would enter the arms). 

Next steps are to assemble the output arm and new switch drivers and finish getting our servo working-- once that's done Begum and I will test the whole system to see how long we can stay locked with these sections of unsecured fiber. 

Images attached to this report
SQZ General
dmdesant - posted 11:15, Thursday 14 August 2025 (12039)
LASTI Squeezer Recharacterization After HD Redesign - 9dB SQZ Observed

This week after finishing the realignment of our HD, we went over the squeezing characterization as a function of pump power/NLG again to see how much of an improvement we made to our readout efficiency. From our previous alog about this, we had an estimated detection efficiency of 85% with the previous HD configuration. At the time, our HD visibilities were around 96% and we reasoned that improving this to around 98% would potentially get us pretty close to being able to see 10dB of squeezing with our readout. Before taking this data again, we realigned the HD to arolund 98% visibility, but to be conservative the actual visibility is likely just under 98%. The results of this are listed in the table below. The squeezing level is derived from an average of the squeezing level over all frequencies above 1 kHz omitting the spurious noise peaks marked in red that exceed 2-3x the standard deviation of the sample sets (plots attached). Towards the end of the measurement, the crystal temperature controller was becoming a bit drifty, so I don't trust the exact accuracy of the peak squeezing level as much as I do the other data points. For this reason, I fit the squeezing level data to the theoretical lossy detection curve to estimate the detection efficiency with just the first three data points. 

Pump REFL (mW) Temp (C) Stage (mm) NLG Squeezing (dB)
21.3 31.79 -0.8090 21.6 9.0
14.0 31.93 -0.8090 8.20 8.0
8.0 32.06 -0.8090 3.80 6.5
3.0 32.15 -0.8090 2.15 4.1**

The fit suggests that our detection efficiency is somewhere between 87.4% and 89.2% (95% Confidence Bounds) with minimal phase noise degredation at these NLG levels. This is unfortunately not quite enough to observe 10 dB of squeezing. However, since we plan on swapping out our crystal and realigning the cavity slightly anyway, we will wait until after that is finished to try and push the readout efficiency up a few more percent. This may include adding a mode cleaner cavity to the LO path as well to improve the beam quality as to better match the VOPO's cavity mode M^2. 

Images attached to this report
GRAVITES General
Dorotea - posted 10:39, Tuesday 12 August 2025 (12037)
Optical switches and drivers tested

I designed a driver to allow multiple crystalatch optical switches to run off of a single function generator and power supply-- tested it today and it seems to work well, the only issue I've noticed so far is that the footprint I made for the switches is about a mm off-- works ok for now but I'll fix it in a future version. The pulse width is adjusted to within the switch specs using the 2 10k pots on the pcb. There does seem to be a small spike in optical power right when the switch changes states, but I'm not sure if this is a feature of the driver, the switch itself, or the PDs, and I don't think it's an issue either way as we'll just gate that out as part of the dead time. The dead time is <400us.

Images attached to this report
Non-image files attached to this report
GRAVITES Electronics
Begum - posted 10:57, Monday 11 August 2025 - last comment - 10:38, Tuesday 12 August 2025(12036)
OLTF and Cross-over frequency measurement for the Split Servo

I had swapped the piezo stretcher on the MZI (to model H01.20, same as Vienna) and coiled 16 m of fiber around it (see alog 11942 for noise floor comparison with other pzt). Initial measurements for the crossover showed 20 dB more gain (there is 10 dB attenuation difference on Out 1 and Out 2, so 10 dB is the real 0 dB here) at 35 Hz compared to the old pzt setup (see measurement). Increased gain is consistent with the length of fiber wound on the pzt stretcher. However, the coherence was lost >100 Hz, so measuring the cross-over was an issue. I used the maximum excitation SR785 can give, 5 V.

So I opened up the box and realized the board configuration was that of FNCServoSplitV2 without the changes in alog 11857 being incorporated. I made the following changes to the board:

The issue seems to have been due to the excitation signal on the pzt path being insufficient. Now the crossover is visible at 220 Hz (PM 70º), see attached plot. The open loop transfer function for the common path is also attached, with UGF at 14.4 kHz (PM 50º). Gain pots on both common and pzt paths were adjusted in situ before these measurements, input to the MZI servo board is 120 mVpp. Note a difference of 20 dB was reported on the OUT1 and OUT2 paths of the common board before, combined with the 10 dB attenuation of the path I mentioned above, -10 dB is the real 0 for this measurement.

I also redesigned of the notch filter on the slow path. The old filter is for 37.5 kHz, with Q ~ 8. I pulled it to 6 kHz with Q~12, attached plot shows the new design. It will eat up a few degrees of phase margin at 200 Hz, which is inconsequential. 

The changes in 11857 are now added to the schematic in FNCServoSplit_v3. I also made a new PCB file with the Chack and Rhack drawn with 805 footprint, in the same v3 version. 

Images attached to this report
Comments related to this report
dorotea - 10:38, Tuesday 12 August 2025 (12038)

Adding: the dead time is <400us. 

Images attached to this comment
GRAVITES Electronics
Begum - posted 08:54, Monday 11 August 2025 (12031)
TEC recovery

I reallized on July 29th that once again, the Meerstetter TEC-1163 stopped measuring a resonable resistance through its HR-1 channel. (This device is the temperature controller for the prestabilization setup in the vacuum tank.) I tried using a different pin pair in X-14 (this worked the last time this device had issues 11938), as well as the usual resetting things in different order. The resistance from the in-chamber NTC sensor was around 12 kΩ, which is very resonable, although the value read thru the 23-bit ADC was MΩ. Magically, the device now works. The only thing I can say that helped it start working is replacing the NTC sensor pins with the two legs of a standard 13 kΩ resistor, at which point the device started reading reasoable resistance values again. After this, I plugged the NTC pins back in and suddenly it started working. I had reset the 23-bit ADC through the software before, with the and without the resisitor attached, without any success. 
Some notes:

Images attached to this report
SQZ General
dmdesant - posted 17:46, Friday 08 August 2025 (12034)
Reprofiling LASTI HD Beams

Over the past few days, we have been reprofiling the beams and redoing the mode-matching setup in LASTI. We used the Phasics camera to perform all measurements

With phasics camera base at 19" from holes with screws holding the LO collimator, the LO beam parameters from the collimator are

  Major Minor Radial
M^2 (mm) 1.15 1.13 1.14
2W_0 (mm) 2.062 1.841 1.961
Z_0 (mm) -590.8 -983.5 -806.8

With phasics camera base at 7th hole back from far edge of table on SQZ path after first mirror, the SQZ beam parameters from the VOPO are

  Major Minor Radial
M^2 (mm) 1.12 1.13 1.13
2W_0 (mm) 0.164 0.170 0.170
Z_0 (mm) 203.6 196.8 200.3

Relative to HD BS, LO collimator screws are 29" away (-z direction)

Relative to HD BS, back of table holes are 28" away (-z direction)

Detector plane of the camera is 38.5mm forward of the mounting hole position (-z direction)

Relative to the HD BS, the LO waist positions are -(29inches-19inches)*25.4mm/inch - 38.5mm offset from measurement (-292.5mm)

  Major Minor Radial
M^2 (mm) 1.15 1.13 1.14
2W_0 (mm) 2.062 1.841 1.961
Z_0 (mm) -883.3 -1276 -1099.3

Relative to the HD BS, the SQZ waist positions are -(28inches-6inches)*25.4mm/inch - 38.5mm offset from measurement (-597.3mm)

  Major Minor Radial
M^2 (mm) 1.12 1.13 1.13
2W_0 (mm) 0.164 0.170 0.170
Z_0 (mm) -393.7 -400.5 -397

Using these beam parameters, a new mode-matching solution was found with one less lens in the SQZ path and one more lens in the LO path.

LO Lens Z Position (Relative to HD BS in meters) ROC (m)
1 -0.6582 0.100
2 -0.3908 0.050
3 -0.1202 0.150

 

SQZ Lens Z Position (Relative to HD BS in meters) ROC (m)
1 -0.6273 0.100
2 -0.1201 0.100

The theoretical beam propagation model is as follows.

After correcting mode matching, we measured the following visibility

  PD1 PD2
DC Offset (mV) 5.6 -5.6
Bright Fringe (mV) 2080 2080
Dark Fringe (mV) 25.6 14.0
Visibility 98.1% 98.1%

In theory, we may now be able to observe 10dB of squeezing with this readout. Next steps include recovering our last-known good crystal configuration and measuring squeezing again as a function of NLG

Images attached to this report
SQZ General
dmdesant - posted 13:26, Tuesday 05 August 2025 (12032)
LASTI HD Visibility Improvement Efforts

Over the past few days, we have been trying to improve the visbility of the HD for the LASTI squeezer. One potential source of mode mismatch was thought to be LO polarization purity, so we went ahead and removed the HWP after our p-pol PBS and just rotated the PBS to transmit s-pol light. Checking the output polarization purity with a reference polarizer confirmed that the polarization is correct and very linear with little p-pol field. My best alignment attempts have yielded us visibilities between 94-96%. To spot check the mode matching, I aligned the HD as best as possible to a visibility of ~94.6%. Then, I used the Phasics camera to profile the beams from the POV of the photodiodes on the homodyne detector. We observed the following beam parameters

SQZ beam with VOPO locked to seed using red PDH loop (radial measurement on Phasics camera)

2W_0 96.4 um
Z_0 89.3 mm
M^2 1.15

LO (radial measurement on Phasics camera)

2W_0 129 um
Z_0 91.8 mm
M^2 1.21

In this measurement, the SQZ beam matched fairly closely to what I remembered measuring before. However, the LO beam has changed substantially since our last measurement. I am growing suspicious of the mating repeatability of the OZ-optics PM fiber we are using for the LO in the particular collimator package we used at the time. Furthermore, calculating the theoretical visibility limit for these two beams tells us that we should expect a visibility no higher than 95%, so this change in LO beam profile seems to be the majority of our limitation on detection efficiency currently. At this point we are looking into replacing the LO fiber and collimator combination with something more stable, as well as reoptimizing the LO path to correct the mode matching.

 

 

Images attached to this report
CTN General
eoelker - posted 13:21, Friday 01 August 2025 (12029)
NPRO Borrowed

Slawek is loaning me a refurbished laser so I can build a small setup on the PSL to characterize optics for the squeezer.  I should be done with it in a couple of weeks.  

Its a refurbished Lightwave 126, MIT Property code 0300431 sticker on the laser head.

Facilities General
evanhall - posted 12:14, Friday 01 August 2025 (12028)
Table float

We floated the ADBC/TIR table again using a fresh bottle of nitrogen at 2200 psi.

The north leg gauge reads 25 psi, the two south leg gauges read 15 psi. There is more weight on the north side than the south.

CDS General
evanhall - posted 11:13, Thursday 31 July 2025 (12024)
SR785 floppy

SR785 s/n 46238 seems to be tempermental with floppies. We have seen it fail to write data, complain about bad sectors on fresh disks, and also someimes throw an error when asked to format a floppy. When the formatting fails, the floppy may become unreadable.

Nonetheless, it seems possible to fix such a floppy as follows. On Debian, to format a 1.44 MB floppy and write a filesystem:

ufiformat -f 1440 -v <block-dev> # <block-dev> being the block device in /dev corresponding to the floppy drive
mkfs.fat <block-dev> # make a FAT file system

The man page for ufiformat admonishes the user that floppies are usually sold pre-formatted, but nonetheless, this seems to be a necessary step to prepare a floppy that has been screwed up by this particular SR785.

The disk drive passes the system diagnostics. (For the last two tests, "disk changed sensor out" and "disk changed sensor in", it wants you to eject and reinsert the disk).

This SR785 was last serviced in 2025 by BRL Test. The entire CRT display is tilted upwards and the plastic screen protector is askew (...did someone punch this?).

We may want to replace the floppy drive (with a USB drive). The manual describes a replacement procedure.

SQZ Site VOPO
eoelker - posted 18:47, Wednesday 30 July 2025 - last comment - 14:02, Saturday 02 August 2025(12023)
M1 Replaced for H1 Spare VOPO (SN3). Escape Efficiency up to 97%

Today I tried to do some loss hunting with the spare VOPO.  As I diagnosed yesterday, M1 appeared to be scattering some excess light.  I tried drag wiping the optic in-situ, but wasn't able to make much progress on the escape efficiency.

I then attempted to measure the escape efficiency without the crystal by moving the crystal out of the beampath using the translation stage.  I wanted to see whether the excess loss was coming from the crystal or the mirrors.  This misaligns the cavity, of course, so I tried to recover alignment by moving M4 a bit.  Unfortunately I wasn't able to align to a reasonable mode and I suspect that the cavity is unstable.  According to the design document, it should be possible to form a stable cavity w/o the crystal and Masayuki was able to do it at LLO it in 2020.  However, the design document says that the cavity is quite sensitive to the M3-M4 distance.  I tried varying this distance a bit, but gave up after an hour or so. 

Finally, I decided to swap out M1 with the second M1 assembly.  I was then able to inspect the old M1 mirror more carefully using a green flashlight.  Sure enough, there does appear to be a very faint cloudy spot in the center of the optic that doesn't come off with drag wiping.  Tomorrow I will inspect with Slawek's microscope and may try First Contact.

Finally I realigned the cavity with the new M1 and measured the escape efficiency.  I was able to get the coupling for the front seed path up to 97%.  Now I'm measuring a reflection dip of 88.5%, which corresponds to an escape efficiency is up to 96.9%.  Together with M2 (1000 ppm transmission) and M3/M4 (70 ppm transmission each), this corresponds to an excess loss of around 1250 ppm.  This would be well within spec for the AR coating on this batch of crystals, though the previous batch of crystals were further below the spec value (Masayuki's measurements suggests that they had < 360 ppm per surface).  

I then locked the cavity and took some photos.  The green scattering looks much improved.  However, I noticed that the beam is no longer centered vertically in the crystal after realigning the cavity.  I will attempt to raise the oven tomorrow to center the beam.  Hopefully, the AR coating is under spec and I'll be able to recover more of the missing efficiency by centering the beam.  

Also, while I was aligning the cavity, one of the peltier leads broke off.  We will repair this tomorrow.  

Images attached to this report
Comments related to this report
eoelker - 15:12, Thursday 31 July 2025 (12026)

This afternoon, i loosened the screws attaching the oven mounting brackets to the oven baseplate and adjusted three point support to raise the crystal slightly.  I centered the 532 nm beam on the crystal and am still measuring an 88.5% reflection on resonance with the same mode matching as yesterday, so the escape efficieny is unchanged. 

 

Next things to try:

  • Replace mirror M3:  This is the one that fell off and was reglued by Slawek and Lisa back in February.  Probably the most likely optic to have issues.  Slawek plans on gluing the other M3 mirror this afternoon.
  • Measure the AR coating on the other crystals from the same batch.  I think I can manage this with a beam chopper and a 50/50 beamsplitter placed before the crystal to pick off the back reflection and measure on a photodiode.
  • Examine the other spare crystals under a microscope.

If any of the other crystals seem promising, I might try swapping the crystal at some point to see if I can boost the escape efficiency further. 

eoelker - 14:02, Saturday 02 August 2025 (12030)

Slawek managed to clean the M1 mirror we removed with some more forceful "drag" wiping.  We'll try this mirror again at some pont with the second spare.  

SQZ General
dmdesant - posted 18:14, Wednesday 30 July 2025 (12022)
Checking LASTI HD Alignment

This week I have been trying to optimize the detection efficiency of our readout in LASTI. Initially, I discovered that the path loss on the SQZ beam from the HD optics alone was too difficult to measure accurately. The loss I was able to measure varied between 2 and 4%. This seemed quite high despite the measurement accuracy problems and I later discovered that the SQZ beam was slightly clipped on the flipper mirror motor that we use to block the SQZ path. I have not tried to measure this loss again, but that was certainly not ideal.

Since that point in time, I have been trying to optimize the visiblity of the HD above our previous measurement of 96.5% with no luck yet. Next week when I return, I will be taking another look at the beam profiles for the LO and SQZ path to see if there was an error in the lens solution for either of the two paths.

Images attached to this report
SQZ Site VOPO
eoelker - posted 17:37, Tuesday 29 July 2025 (12021)
H1 Spare VOPO (SN3) M1 dirty?

Before I left town last week, I measured the escape efficiency of the new VOPO for Hanford and found that it was quite low (95.4%).  This afternoon I turned up the green power to a bit below threshold and took some photos of the cavity with the lid off and locked at 532 nm.

It seems like my M1 mirror front surface might be a bit dirty.  I can't see anything obvious by eye, but I do seem to be getting a fair amount of scatter at 532 nm.  The scattering is less obvious in the photo than by eye, but you can clearly see the speckle pattern from the M1 scatter in the photo of M2 and in the photo of the front side of the oven.  The problem is quite obvious.  I hadn't done this test until now because the OPO is tricky to lock when the cover is off and the cleanroom blowers are running. 

Fortunately, this is one of the easier optics to clean in-situ.  Tomorrow morning I will attempt to drag wipe M1 and repeat this green scattering test.  If it looks improved, I will remeasure the escape efficiency.  

Images attached to this report
SQZ General
dmdesant - posted 18:36, Friday 25 July 2025 - last comment - 16:09, Thursday 14 August 2025(12019)
LASTI VOPO Escape Efficiency Remeasurement

Today, I remeasured the escpape efficiency. To do so, I assembled a more permanenant back seed path which is similar to the rest of the fiber paths into the cavity. After alignment, I checked the mode matching with a cavity scan and noted the peak transmission values for the TEM00 mode and all higher order modes visible on the CLF PD DC monitor voltage

Mode Overlap Measurement (Transmission Scan)

Mode Detector Volts
TEM00 0.3900
TEM02/20 .07425
HOM1 .02775
HOM2 .02445
Detector Baseline .02100

This corresponds to a mode matching (overlap) of about 85.3%. Then looking at the reflection dip for the TEM00 mode on the reflection scan and following the same approach we did before in this alog, we collect the following data

Reflection Scan

Measurement Detector Volts
Peak Reflection (Off Res) 1.369
TEM00 (On Res) 1.264
Detector Baseline .00093

This measurement of the reflection dip corresponds to a calculated escape efficiency of 97.6%, which is a bit lower than we had before and could be due to the fact that the reflection PD was clipped in that measurement. Moreover, we did not note where on the crystal we had taken the previous escape efficiency measurement. This measurement was taken with a crystal temperature of 32.1C, and a stage position of -0.7695mm. With this back seed path in permenantly now, it should be relatively easy to repeat this measurement a few times in different spots on the crystal as we continue to optimize.

Comments related to this report
dmdesant - 16:09, Thursday 14 August 2025 (12041)

Note: There is an error here... back/rear actually should be front...

SQZ General
dmdesant - posted 17:19, Thursday 24 July 2025 - last comment - 15:19, Friday 25 July 2025(12017)
LASTI Squeezing Level Characterization (and 8dB observed)

Today I realigned our homodyne to 96.5% visibility before characterizing our squeezing as a function of NLG. It seems like something has shifted from previous alignment attempts or I was not getting as much visibility as I had thought initially.

I also roughly calibrated the measured power on the pump reflection beam coming out of the chamber relative to the pump input power off resonance. With 26mW pump incident on the OPO, we have ~24.6 mW measured on refl beam outside the chamber.

Then, by adjusting this power to a few points over a wide range, I quickly measured the roughly optimal NLG at a given pump power and then locked everything and measured the squeezing and anti-squeezing levels on the homodyne. This data is shown below.

Pump REFL (mW) Temp (C) Stage (mm) NLG Squeezing (dB) Anti-Squeezing (dB)
7.0 32.45 -0.7695 3.36 5.5 8
12.2 32.36 -0.7695 6.26 7 11.5
18.0 32.24 -0.7695 12.8 7.8 15.3
25.0 32.10 -0.7695 39.3 8.0 20.3

I also noticed that having the SR785 on DC coupling greatly reduced its dynamic range and its own noise floor was beginning to impact our measurements. This was fixed pretty easily by just AC coupling the input, which yielded much more predictable and consistent data over a wide set of input attenuator settings and shouldn't matter at the frequencies I am looking at here anyway. These squeezing levels are an "eyeball" average over the spectrum the analyzer was displaying, but I would say that they are a low conservative estimate of the real average above 400 Hz. Once this data is fit to the theoretical curves for this, we will begin hunting for losses and additional noise that could be impacting our ability to reach 10dB of squeezing.
 

 

Comments related to this report
dmdesant - 15:19, Friday 25 July 2025 (12018)

Today I post-processed this data and got an estimate for the total detection efficiency and phase noise, along with a sanity check for the threshold power of our OPO at the moment. Attached are plots with fits for the nonlinear gain of this cavity as well as the squeezing and anti-squeezing levels we saw yesterday.

The results of these fits predict our threshold power to be around 36 mW with 95% confidence bounds (35.17 mW, 36.18 mW). The squeezing fit predicts the total detection efficiency to be around 85% with 95% confidence bounds (81.6%, 88.2%). The phase noise is so small that this fit is insufficient for estimating phase noise (as there is no inflection point in the experimental data). If we trust this characterization, the current 85% detection efficiency of this setup limits us to a maximum theoretical measured squeezing of 8.2 dB, so now we may be up against the loss limits of our detection system. 85% detection efficiency is also fairly believable given some ballpark estimates and loss measurements we have already made.

At the moment, our estimated losses are as follows

Homodyne Visibility 96.5% Meas. 20250724
SQZ Beam Path Loss 98% Ballpark guess
PD Quantum Efficiency 98% Based on past detectors
Escape Efficiency 98.1% Meas. 20250717
TOTAL (n_tot) 87.7%  

Given that we need our detection efficiency to be clear over 90% in order to see 10dB of squeezing, we stand to gain the most efficiency in reoptimzing our homodyne detector, which currently contains more lenses in the sqz path than are strictly necessary. Moreover, if the seemingly low homodyne visibility we were seeing yesterday were improved to 98% or better, it would push this current estimate up to around 91.4%, at which point seeing 10dB could be possible.

One final note is that our escape efficiency measurement from last week may be slightly off as Eric noticed that his escpae efficiency measurement on the spare site VOPO was affected by the reflection of the rear seed path being clipped by the PD he was using at the time. We has a similar problem on the LASTI cavity, so will be remeasuring the escape efficiency again to pin this down as accurately as possible.

Images attached to this comment
Displaying reports 221-240 of 1025.Go to page Start 8 9 10 11 12 13 14 15 16 End