Bruno Becher - posted 12:57, Wednesday 05 November 2025 (12098)
Vacuum Bake Oven report
The Vacuum Bake Oven in the orange lab is an Ultra High Vacuum system to clean and bake components to be used in Lasti and potentially sent to the sites.

9/3/25 - Rich, Slawec, and I disassembled the vacuum system and replaced the 13.25” copper gasket on the rear ConFlat flange of the main chamber where the primary leak was detected. The knife edge on both faces of cf flange was inspected. Tightened flange back on with a torque wrench set at 18ft*lbs tightening every 13th bolt, going up 4ft*lbs each time the bolts refused to twist. Ending at a torque of 34 ft*lbs. We also replaced the copper gaskets on Valve 4 to the main turbo, and the RGA flange where leaks were also detected. All internal surfaces were wiped with isopropanol. We also took apart the Forelines and cleaned the Viton seals in the hand crank valves. 

9/7/25 - Air leak in system at back flange still detected with helium though less intense than before gasket replacements.  

10/19/25 - RGA scans showed a usable Hydrocarbon signature at 1.5e-11 Torr*L/s despite the leak. 

10/29/25 - After a 24 hour 150°C empty bake the leak "opened up" as shown in the attached RGA scans. Helium detector showed leak is coming from same rear chamber ConFlat flange.

11/4/25 - Disassembled the Vacuum system to thoroughly inspect knife edges on rear flange and perhaps replace 13.25" 275CF CU gasket.
Images attached to this report
SQZ General
eoelker - posted 12:23, Monday 03 November 2025 (12097)
LO Mode Cleaner Servo

I modified one of my old Labjack servo boards to use as a servo for locking the mode cleaner we're adding for the LO path. I will post a schematic when uploads are working again on our aLog. It is loosely based on the old iLIGO PMC servo. I also repaired one of our PZT Driver chassis to drive our old PMC.  

I changed a few things on parametric and /opt/rtcds/ to accomodate the new hardware:

Schematic for the servo board is attached

Non-image files attached to this report
SQZ Site VOPO
eoelker - posted 13:49, Tuesday 28 October 2025 - last comment - 12:15, Monday 03 November 2025(12063)
Installation instructions for SmarAct Software

Instructions below copied from SmarAct email:

 

Pardon the missing USB drive, I will include all the relevant information here:

MCS2 Windows Software Installer 2.2.12: https://depot.smaract.com/s/7wgqLx5xACjqNLZ

MCS2 Firmware (should already be installed): https://depot.smaract.com/s/kAy4dTSAkEcGYFn

 

Once installed you will be able to navigate to your C: drive and find the SmarAct folder, in which can find all the documentation for the MCS2 along with our “Getting Started” document and programmer’s guide (attached for convenience) as well as programming examples.

I’ve also included our “Stick-Slip Positioner Assembly Instructions” which gives you handling and operational information regarding our piezos. One key note is that if you mount something on the SLC-1720, ensure that you take note of the recommended screw lengths as if you use screws that are too long it will impact and damage the sensor scale below, I have attached the technical drawing and CAD model for reference.

Lastly, you will have to set the positioner type corresponding to your stage on your MCS2 controller, this will be positioner type 300 (it is most likely the default).

Comments related to this report
eoelker - 12:15, Monday 03 November 2025 (12095)

Updated link for installer (old link was broken)

 

MCS2_Installer_2.2.15: https://depot.smaract.com/s/pQzrFewiZZmrLaa

Firmware MCS2_Release_2025_08_15: https://depot.smaract.com/s/6MC4mpPonmBd8aX

Facilities General
eoelker - posted 16:55, Monday 22 September 2025 (12059)
Vacuum prep lab DI filters replaced

[Bruno, Eric]

After spending some time in recent weeks cleaning out the vacuum prep lab, the last remaining chore was replacing the filters for the type III DI water system in the fume hood.  This afternoon, Bruno and I replaced both filters.  Based on the expiration dates on the spare cylinders we had on the shelf, I'm guessing that they were last replaced about 5 years ago.  I doubt that they were used heavily used during and after the pandemic, so this was probably fine.  I hadn't noticed any spotting yet on components, but figured we might as well do it now before we move on to other tasks.

 

We did not replace the pre-filter outside of the cleanroom.  I wasn't able to find replacement filters online.  FWIW, the indicator on the prefilter says that it is still OK.

 

 

SQZ General
eoelker - posted 16:49, Monday 22 September 2025 (12058)
LASTI VOPO Escape efficiency recovered

[Daniel, Eric]

Today we were able to recover a decent escape efficiency on the LASTI VOPO with the SPPKTP crystal #3.  We did a quick measurement to confirm Daniels numbers from friday and were still seeing a low value.  After scanning the crystal a bit and finding that the escape efficiency was fairly uniform across the crystal (except for a few spots where we believe that the crystal is chipped), we suspected that we were hitting the bad spot on M3 and decided to realign the cavity.

I was able to recover an alignment that looked comparable to what I had when I first aligned the cavity a month ago by repeating my standard alignment procedure.  The cavity beam is still tilted because, as I noted then, M3 and M4 do not have the vertical adjustability found in the site version of the VOPO tombstones, so the crystal is tilted a bit to compensate.  Adjusting the vertical position of M4 is usually the final step of my alignment procedure.

We measured (1064 nm):

6.4 V on resonance

5.8 V off resonance

97% mode matching

 

This corresponds to a 90.6% reflection on resonance, or about 1800 ppm of excess loss, or an escape efficiency of around 97.7%.  This roughly agrees with the loss we expect based on Daniel's measurement of the AR coatings for Crystal #3 (~1800 ppm according to this alog)

 

Since we cant do any better than this loss-wise, we're going to go ahead and reinstall the cavity in the chamber.

SQZ General
dmdesant - posted 17:38, Friday 19 September 2025 (12057)
Preparing LASTI VOPO For Reinstallation - SPPKTP Crystal Installation and Alternate Escape Efficiency Measurement Technique

This week we continued reinstalling the "better" SPPKTP crystal (#3) into our cavity in LASTI. We hit a snag earlier in the week with the keepers for the crystal, but that has been resolved and we are seeing flashing in the cavity once more. After realignment of the red/green input beams, we achieve the following mode matching as witnessed by a cavity scan.

Mode Green Trans Voltage Red Trans Voltage
TEM00 278 mV 4.64 V
TEM01 72 mV 92 mV
TEM10 72 mV 84 mV
TEM02/20 110 mV 232 mV
Higher Order 72 mV --
Background (Offset) 66 mV 36mV
Mode Matching (%) 77.4% 93.9%

In our typical fashion, I also measured the reflection scan of the cavity to work out or excess loss from the crystal and the escape efficiency at both red and green. No particularly important crystal position was chosen for this sweep.

Measurement Green REFL Volts Red REFL Volts
Off Resonance 820 mV 6.00 V
On Resonance 368 mV 5.00 V
Background 4 mV ~0
Escape Efficiency 77.1% 95.7% 
Excess Crystal Loss 0.46% 0.5%

This excess crystal loss for green is not corroborated by the green single-pass loss measurement, which suggests a green loss ~2x larger than what I measured with the green single-pass setup. However, as we noted before, this measurement technique relies on both the mirror reflection coefficients being known prior, and becomes particularly sensitive to variation in these mirrors as their added loss to the cavity becomes comparable to that of the crystal (i.e. near critical coupling). In this case, our predicted loss yields a predicted finesse that is much higher than what we actually see (in this case it predicts F = 238, we see closer to 140). So, I wanted to try an experiment where I fit the normalized reflection coefficient scan data (captured by the scope over time with "linear" PZT sweep) in such a way that the finesse is also taken into account. Taking this approach, we need no prior information about the mirrors to fit the reflection data. We only need prior information about the high-reflecting output coupler (R2 - to distinguish its effective loss from that of the rest of the cavity), the mode-matching coefficient, and to assume that the sweep is mostly linear (which is a stretch, but let's see how we do).

Essentially, scope data of both the reflection and transmitted light through the cavity is brought in to MATLAB, normalized and scaled correctly by the mode-matching coefficient, and fit against the theoretical Lorentzian line shape of the cavity. After normalization, the normalized power reflection coefficient can be expressed as PREFL = |(r1 - r2*beta*exp(i*phi))/(1 - r1*r2*beta*exp(i*phi))|2, where phi = 2*pi*t/TFSR, beta is the intrinsic field transmission coefficient through the rest of the cavity, and r1 and r2 are the field reflection coefficient magnitudes of M1 (the input coupler) and M2 (the output coupler) respectively. After scaling and fitting, the resulting fit against the reflection and transmission data from the scope for green is here. The fits on each resonance in the scan (raw here) give us estimated parameters for the reflectivity of R1, the intrinsic loss, the finesse, and the escape efficiency of the cavity. These are tabulated here for green assuming the output coupler (M2) has a 99.85% reflectivity and a mode matching of 77.4%.

  Intrinsic Loss (%) R1 (%) Finesse n_esc
Resonance 1 0.95% 96.9% 141 75.5%
Resonance 2 0.93% 96.8% 144 75.5%
Resonance 3 0.87% 96.9% 152 75.5%

There is some variation in the fits from resonance to resonance. This is probably due to the limited memory depth of the scope (not enough data in the resonance to fit super accurately), small drifts over the sweep time, and the fact that the PZT actuation is not perfectly linear with applied voltage. Moreover, it seems like there is a strange thermal dragging/time delay between the transmission and reflection resonances. Nevertheless, the other two data points predict a loss much closer to what we measured in our single pass setup (~1%). As we stated before, it seems like this can be accounted for by M1 having a green reflectivity of ~97%, not 98%. To see if this is plausible, we will try to measure this mirror in situ the next time we take data on the cavity. For good measure, we also fit the reflection scans of the cavity at the seed wavelength (red) to compare with our typical escape efficiency calculation approach. The raw data and fits are attached here and here. The extrapolated parameters are listed below for red assuming the output coupler (M2) has a 99.85% reflectivity and a mode matching of 93.9%.

  Intrinsic Loss (%) R1 (%) Finesse n_esc
Resonance 1 0.56% 86.7% 42 95.6%
Resonance 2 0.48% 88.3% 48 95.5%

Despite the slower scans taken here potentially worsening the drift-susceptibility of the data, the escape efficiency extrapolated from this data is fairly close to that of the simple dip-fraction calculation. In addition, the predicted reflectivity for M1 is on average about what we expect (~87.5%). However, the crystal loss is surprisingly high here. In the data that Maryrose measured last week, we were able to see much higher escape efficiencies for red than we are seeing now after re-installing the crystal. It is entirely possible that we are just currently measuring at a location in the crystal where there is some excess red loss (there are several cracks in this crystal due to the coating debacle), so we will try redoing the reflection scan over the crystal next time.

In any case, once we confirm that we have similar escape efficiencies as before, we should be getting pretty close to reinstalling the VOPO in the LASTI chamber and resuming squeezing experiments.

Images attached to this report
Electronics General
evanhall - posted 14:44, Tuesday 16 September 2025 (12033)
Measurements on Caltech servo box #1

Nelushi, Evan

We have a Caltech servo box / PDH box (refer to D1800273 for block diagram, D1700219 for assembly drawing) with a variant loop-shaping transfer function board (D1900437-v1). It works okay but needs at least one more revision.

Salient points:

The loop-shaping transfer function board comprises four amplifiers:

We measured the transfer function from servo in to servo out at several VGA setpoints. Setpoints above 800 counts or so seemed glitchy and possibly slew-rate limited, so are not shown. We also measured the output noise in the 768 ct gain setting and input-referred it. There seems to be some kind of... touchiness with the output offset adjustment, and the noise can be in a good or bad state depending on the setting (there doesn't seem to be a threshold digital pot number that delineates good or bad).

Appraisal and next steps for a revision of the transfer function board:

Images attached to this report
Non-image files attached to this report
SQZ Site VOPO
- posted 21:56, Friday 12 September 2025 - last comment - 18:59, Tuesday 16 September 2025(12054)
PPKTP Crystal Characterization for LLO

[Daniel, Eric, Maryrose]

Initial Escape Efficiency Measurments for sPPKTP (#3) and PPKTP (circa 2018, #1 and #2):

We have been looking to test several crystals as possible candidates for the LLO site. We started (09/05) with testing out one of the sPPKTP crystals to confirm Daniel's measurements by measuring the escape efficiency and estimating the losses (table summary below). Later Eric provided some of the PPKTP crystals from the 2018 batch. They had previously been unopened, and one of them was exceptionally good (and was used in the H1 OPO we are planning to use as a replacement for the sites). We wanted to try out the rest of this batch in the LASTI setup to see if we could find another exceptionally good one. After testing two of them out in the LASTI cavity (numbered "1" and "2"), we found them to have nearly identical escape efficiencies and estimated losses. We proceeded to leave the "1" PPKTP in the LASTI and do further tests. A summary table of values is shown below for the measured mode matching percentage, escape efficiency, and estimated losses for red and green. Finesse measurements were also made for two of the crystal-cavity pairings.

Initial Loss Estimates and Escape Efficiency Measurements
Crystal Dip Fraction Mode Match % Mloss Mode Match with Loss est. Escape Efficiency % Finesse
sPPKTP, #3 0.9 91.5% 0.9975 0.901 97.4%  
  0.62 77.7% 0.99815 0.621   150
PPKTP, #2 0.93 96% 0.9989 0.931 98.3%  
  0.47 69.6% 0.9945 0.467    
PPKTP #1 0.96 93% 0.9988 0.931 98.2%  
  0.44 73.9% 0.9946 0.438   130

 

NLG Measurements with 2018 PPKTP Crystal #1:

We moved onto making NLG measurments across the cavity. These measurements showed that the PPKTP (#1) was pretty well aligned and had pretty uniform NLG across the crystal. Several scans were taken using estimated NLG measurements and NLG measurements with the cavity locked and optimizing the temperature and stage position. Finally, a rough scan near the center of the crystal was performed to obtain a range of acceptable temperatures for maximum NLG. Plots and tables are shown below for these measurements.

Initial NLG Rough Measurement Across Crystal
(T=32.23 C for all points)
Position (mm) Red w/ green Red no green NLG
-2 X X Edge of Crystal
-1.95 X X No useable co-res.
-1.45 -- -- Data deleted (~2.5)
-0.95 0.297 0.117 2.54
-0.453 0.299 0.113 2.65
0.04 0.253 0.111 2.28
0.537 0.253 0.111 2.279279279
1.028 0.259 0.111 2.333333333
1.521 0.243 0.111 2.189189189
2.019 0.257 0.111 2.315315315
2.505 0.245 0.107 2.289719626
2.786 X X Edge of Crystal

Average NLG from these measurements was around 2.36. Plot attached of this data.

 

Optimized NLG values for Position and Temp

Position (mm)

Temp (C) red w/ green red no green NLG    
-1.51 31.7 0.305 0.113 2.699115044
-0.5 31.88 0.309 0.113 2.734513274
0.483 31.83 0.257 0.113 2.274336283
1.483 31.89 0.281 0.115 2.443478261
2.464 31.8 0.269 0.113 2.380530973

Average NLG from these measurements was around 2.50. Plot attached of this data.

Finally, temperature vs NLG was measured. Scanning around -0.5 mm was done to find co-resonances at different temperatures. Plot of this data is also attached. We found the range of temperatures from 31-33 C to be a good setting for maximum NLG. Temperature controller was calibrated to the thermistor coefficients before this scan was done.

Discrepancies with Estimated Losses Between Escape Efficiencies, Finesse Measurements, and Daniel's Single Pass Crystal Tests:

We encountered a large discrepancy between the losses given by the escape efficiency measurements and those found by Daniel's single pass crystal measurements and the finesses we measured. We tried looking into this to see if we could nail down any inaccuracies in our measurments. We redid the escape efficiency measurment and found it to be in great agreement with the measurements taken earlier in the week.

Crystal Dip Fraction Mode Match % Mloss Mode Match with Loss est. Escape Efficiency % Finesse
PPKTP #1 0.96 94.7% 0.9989 0.932 98.3% 44
  0.42 73.9% 0.9942 0.423   126

To pin down what we think the green losses are in the previous batch of aLIGO crystals, we measured them in our single-pass setup from this alog. The full loss transmitting through the crystals (only #2 and #3 measured due to the fact that #1 is currently in the LASTI cavity) are listed here:

Crystal Trans PD Voltage (w/ Crystal) Trans PD Voltage (w/o Crystal) Loss
#2 0.459 V 0.468 V 1.92 %
#3 0.463 V 0.469 V 1.28 %

Using this as a baseline, we can predict what our green finesse should be given our mirror design (R1 = 0.98%, R2 = 0.9985%). F = -2π/ln(R1R2(1-Lxtal)). In the case where we believe we know all of our dominant loss terms (excluding possible excess loss from M3 and M4), we can compute a range of finesses in the LASTI VOPO to expect for all the aLIGO crystals in this batch (assuming they are similar). These are as follows:

Crystal Excess Loss (Lxtal) Finesse
#2 1.92 % 153
#3 1.28 % 182

We then measured the green finesse of the LASTI VOPO with crystal #1 installed. This turned out to be approximately ~120. Which is quite low and a far cry from the expected calculated values for the cavity finesse given our assumptions on the mirrors and measurements of the crystal loss out of the cavity (Maryrose mentioned that the finesse of crystal #2 and #1 looked very similar).

To further constrain the possible errors in this measurement, we made another green escape efficiency measurement of the VOPO with crystal #1 installed. In the process of this calculation, we estimate the total excess loss in one round trip through the cavity. Typically, we attribute this loss to the crystal. From this excercise we find that with our assumed mirror reflecitivites, we would expect the crystal's added loss to be on the order of 5000ppm, which does not agree with our single-pass measurement in the slightest.

Given that we know both our coupling condition and finesse to some degree from the reflection dip measurement associated with our escape efficiency and a simple cavity scan respectively, we hypothesized that the estimations of both the finesse based on measured single-pass crystal loss, and the excess loss from the crystal based on the cavity's green escape efficiencies may be flawed due to a flawed mirror. Next, we tested some different possible values for R1 to see if we could find a plausible problem with that mirror that might lead to this discrepency. If we suppose that M1's reflectivity was actually, 96.3%, we can recalculate the excess loss of the cavity (from the reflection dip) and the finesse estimate (with known crystal loss) again and find that the two estimates are closer to what we measured in the lab. In this case, our reflection dip measurement predicts the loss of the crystal to be closer to those measured directly in the single-pass setup. Moreover the additional coupling from the drop in R1 reduces the finesse to something closer to what we estimated with a cavity scan.

R1 R2 Calc XTAL Loss (Lxtal) Calc Finesse (Based on calc loss)
98% 99.85% 0.58% 228
96.5% 99.85% 1.2% 122
    Measured XTAL Loss Measured Finesse
    1.28-1.92% ~120

We also measured the reflection dip of the cavity @ 1064nm using the front seed path. Using this, we estimate the red escape efficiency, crystal loss, and finesse to be:

R1 R2 Escape Efficiency Calc  Red XTAL Loss (Lxtal) Calc Red Finesse (Based on calc loss)
87% 99.85% 98.1% ~800ppm 44.3

Our measured finesse at 1064nm is also around 44, so things make more sense with red than with green. When measuring the cavity with green light, it seems that we are in a regime where the reflection dip measurement is particularly sensitive to the reflectivity of the input coupler since it's coupling strength is smaller.

Images attached to this report
Comments related to this report
dmdesant - 16:01, Tuesday 16 September 2025 (12055)

Yesterday we pulled aLIGO crystal #1 back out of the LASTI VOPO to reinstall the SPPKTP crystal we (sort-of) like for more testing. In the process, we measured the loss of the aLIGO crystal on the single pass setup and found it to have a loss of around 1.8%. As far as loss goes, this crystal seems to be pretty good.

Unfortunately, in the process of reinstalling SPPKTP #3, I dropped the PEAK keeper for the crystal on the ground so Eric and I ran a batch of them (along with some other tools) through the air-bake oven. These are finished and we are now ready to continue with the reinstallation of the SPPKTP crystal.

dmdesant - 18:59, Tuesday 16 September 2025 (12056)

Note: Maryrose Barrios took a lot of this data and typed this nice summary up for us to have for our records. She and I just posted it with my alog account after we brainstormed a bit about our hypothesis that M1 may have a strange reflection coefficient.

SQZ Site VOPO
eoelker - posted 16:48, Thursday 11 September 2025 (12052)
O5 Oven assmblies soldered and glued

Today I soldered together the new oven assemblies for the SmarAct translation stages and glued the thermistors in place.  

For the soldering process, I did something losely based on E1700369, but modified John and Fabrice's procedure in a few ways.  The procedure:

  1. Class B clean all the required components (Peltier, thermistors, and copper oven componets.  Checking their ICS record, John and Fabrice air baked all components after ultrasonic cleaning.  DO NOT air bake the copper components.  This will cause the copper to become heavily oxidized and make it very difficult to adhere the InSn solder to the copper.  This may be why they needed to use HCl to get the indium solder to bond properly.  I air baked some of my copper wire a couple of months ago and could never get indium solder to stick to it after that.
  2. I used a hot plate covered in UHV foil like they did to heat up the copper components just enough to melt the InSn solder (melting point 118 C) but not high enough to melt the peltier solder (melting point ~160 C), placed the peltier on top, and then ramped down the temperature so the solder could set.
  3. I reversed the order of the soldering.  I attached the peltier to the top piece first (D1600067-v3) and then attached the combo to the base piece (D1600066-v4).  This allowed me to avoid using the soldering iron to attach the top piece.  Was a bit less cumbersome w/o having an iron involved.

I tugged on the components a bit and the assembly seems to be pretty strong.  I also glued in the thermistors to the top and reservoir; that process was straightforward.

Once the epoxy for the thermistors cures overnight, I will connectorize and then air bake the assembly for a few hours as a test to make sure it can withstand a thermal cycle to 100C.  We plan to ship the oven assemblies to Caltech and have the entire oven assembly vacuum baked at 100 C  (For the O3 oven build, we vacuum qualified the parts individually and the oven assembly was not re-baked after soldering and gluing the thermistors).  

 

Images attached to this report
SQZ General
dmdesant - posted 14:16, Tuesday 02 September 2025 (12050)
SPPKTP Crystal Red AR Coating Measurement

Eric, Daniel

After observing that crystal #3 was a little lower loss than the other crystals last week, we decided to measure both crystal's AR coatings at red (1064nm) to see if they will have any appreciable effect on our escape efficiency. We measured this with Eric's setup on the PSL table in LASTI. In this experiment, the incident power monitor (PDA100A) is set to a gain of 10dB and samples the incident power through a 50/50 beam splitter and a 99/1 beam splitter (i.e 1% of the power incident on the crystal). The reflection power monitor (PDA100A) is set to a gain of 30dB and is placed on the reflection-coupled output of the 50/50 beam splitter to sample the reflected power from the crystal's AR coatings (50% of the power reflected).

New SPPKTP Crystal #3

Incident Power Measurement

Mesaurement PD Voltage (V) Power (mW)
Inc Tap Background 0.015 0.32
Incident Tap 3.38 71.2

Reflected Power Measurement

Mesaurement PD Voltage (V) Power (mW)
Refl PD Background 0.036 0.0015
Refl PD (Left Spot) 1.47 0.0619
Refl PD (Right Spot) 1.65 0.0695

Reflection Coefficients

Reflection Reflection Coeff. (PPM)
Left Spot 959
Right Spot 852

New SPPKTP Crystal #4

Incident Power Measurement

Mesaurement PD Voltage (V) Power (mW)
Inc Tap Background 0.015 0.32
Incident Tap 3.36 70.7

Reflected Power Measurement

Mesaurement PD Voltage (V) Power (mW)
Refl PD Background 0.036 0.0015
Refl PD (Left Spot) 1.41 0.0594
Refl PD (Right Spot) 1.14 0.0480

Reflection Coefficients

Reflection Reflection Coeff. (PPM)
Left Spot 823
Right Spot 661

Not too unsurprisingly the red reflection coefficients for these coatings are not too great either. According to Eric, the crystals for the site-spare VOPOs have been measured to have around ~100 ppm reflection per interface. Nevertheless, we will still give crystal #3 another try to see if better caivty alignment can alow us to achieve enough gain for the 10dB goal while having a lower bulk loss as measured in our previous alog.

SQZ Site VOPO
eoelker - posted 09:42, Tuesday 02 September 2025 (12049)
H1 Spare VOPO crystal scan

[Maryrose, Eric]

On Friday, we scanned across all the co-resonance positions of the crystal to check the escape efficiency and cavity alignment at each position.  We saw no evidence of misalignment from translating the crystal and the escape efficiency was roughly equal at all 9 usable co-resonance positions.  The tenth position (-700 in the table) was too close to the edge and was not usable.  This suggests that the crystal is very well aligned and 1064 nm losses in the crystal are fairly low everywhere.

We did, however, see some slight decrease in the nonlinear gain across the crystal when keeping a constant temperature of 35.25C.  At the final position (-8700), we locked the cavity and measured an NLG of 7 with 1.33 mW in the TEM00 mode for 532 nm.  This corresponds to a threshold power of about 3.2 mW.  At the first good position (-1560) the same measurement yielded an NLG of 4.5 and threshold power of 4.5 mW.  This may simply mean that the phase matching temperature across the crystal isn't quite uniform, since we optimized our phase matching temperature near the final position.  From past conversations with vendors, my understanding is that some index variation across the crystal is common in PPLN.  Perhaps PPKTP is similar.  A very rough NLG scan where we measured the amplification in the seed power while sweeping through resonance was done on all 10 positions.  This showed a roughly linear decrease in NLG as we stepped through the intermediate points.  

This isn't particularly concerning, as these threshold values are much lower than what we've seen with previous crystals.  This cavity is operating quite efficently since we believe that the 532 nm losses are quite low (a more precise measurement of this is on our checklist for this week).

Some numbers for the escape efficiency measurement:  The crystal edges (no more cavity flashes) were at 0 and -9200 counts on the tranlation stage (the offset here is arbitrary since this stage has no position sensor).  All voltage numbers here correspond to the trans and refl PDs for the front seed path.  

Position TEM00 trans (V) TEM01 Trans (V) TEM02 Trans (V) Refl On Resonance Refl Off Resonance Esc Efficiency
-700           Not usable
-1560 3.96 0 0.110 1.67 1.82 97.9%
-2460 3.92 0 0.110 1.67 1.82 97.9%
-3400 3.96 0 0.110 1.67 1.82 97.9%
-4300 3.96 0 0.110 1.67 1.82 97.9%
-5100 3.96 0 0.110 1.67 1.82 97.9%
-6000 3.92 0 0.112 1.67 1.82 97.9%
-6900 3.80 0 0.102 1.66 1.80 98.0%
-7800 3.86 0 0.108 1.66 1.80 98.0%
-8700 3.80 0 0.100 1.66 1.80 98.0%

 

Relevant formulas (position -1560 shown as an example):

Mode matching:  3.96/(3.96 + 0.110) = .973

Reflection Dip:  1.67/1.82 = .918

Solving for intracavity loss from reflection dip and mode mismatch:   0.918 = 0.973*(sqrt(R1) - sqrt(Rloss*R2))^2/(1-sqrt(R1*R2*Rloss))^2  + (1-0.973)

assuming R1 = 0.93,  R2 = 0.999, we get that Rloss = .9994.  This corresponds to 600 ppm of excess intracavity loss.  

Escape Efficiency:  0.979 = (1-.93)/(1-.93*.999*.9994)

 

SQZ Site VOPO
eoelker - posted 12:16, Thursday 28 August 2025 (12046)
H1 Spare VOPO mirror swap: 97.9% escape efficiency

[Maryrose, Eric]

Slawek finished gluing together the 2nd batch of spare VOPO mirrors on Tuesday.  Yesterday, we managed to form a stable cavity with just the mirrors and measure the escape efficiency from the mirrors alone.  More detailed post is coming, but for now i'll summarize by saying that we saw about 500 ppm of loss in excess of the design values for the mirror reflectivities(M1: 93% and M2: 99.9%)  This corresponds to an escape efficiency of roughly 98% with just the mirrors alone.  Unfortunately, our spare mirrors are all a bit worse for wear at this point.  Some are nearly a decade old and were only salvaged by very careful cleaning from Slawek.

 

Today we decided to take our oven + crystal from the other spare cavity and move it into this cavity.  After realigning, we were able to measure an escape efficiency of 97.9%, suggesting that our crystal is only adding about 150 ppm of loss on top of what we were seeing with the mirrors alone.  This tracks with the fact that the green losses for this crystal are also quite low (~1% whereas 2% is more typical based on our measurements from O3).  We probably arent hitting exactly the same points on the mirror given that we had to change the length between the two curved mirrors quite a bit to form a stable cavity w/o the crystal.

Some values:

Reflection off resonance = 3.22V

Reflection on resonance = 2.94V

Dip = 280 mV

Dip (fraction):  91.5%

Mode matching for 1064 front seed = 97.5%

Escape efficiency  = 97.9%

 

We will try to optimize the crystal position a bit more to minimize loss, misalignment from tranlation, and NLG.

SQZ General
dmdesant - posted 12:07, Thursday 28 August 2025 - last comment - 16:52, Friday 29 August 2025(12045)
LASTI SPPKTP Crystal Green Loss/Coating Measurement

Eric, Daniel

We have been constructing a setup to measure the 532nm single-pass transmission and relection coefficients of our new SPPKTP crystals outside of the OPO cavity to confirm the loss estimate from Eric's finesse measurements of the cavities at 532nm. The setup uses a chopper wheel and a lock-in amplifier to remove background light from our measurement. A 150mm focal-length lens is used ahead of the crystal to focus the beam down to a ~100um beam diamater at the center of the crystal.

In this test, we compare the total transmission loss of the least-damaged crystal from the batch (#4) to that of the old LASTI PPKTP crystal which we removed from the cavity a couple weeks ago. In each measurement, the incident beam on the crsytal was aligned as best as possible to the center of the crystal.

New SPPKTP Crystal (#4)

Transmission Measurement (PDA100A Gain = 0dB)

Measurement Voltage (mV) Power (mW)
Lock-In Readout (No XTAL) 323 0.718
Lock-In Readout (w/ XTAL) 314 0.698
  Corresponding Loss 2.79%

Reflection Measurement (PDA36A Gain = 40dB)

Reflection coefficient is calculated relative to estimated incident power from 8% tap PD at the time of the measurement (781uW)

Measurement Voltage (mV) Power Reflected (uW) Reflection Coefficient (ppm)
Lock-In Readout (surface 2 thru 1) 21.3 0.94 1200
Lock-In Readout (surface 1) 21.8 0.96 1230

OLD LASTI PPKTP Crystal

Transmission Measurement (PDA100A Gain = 0dB)

Measurement Voltage (mV) Power (mW)
Lock-In Readout (No XTAL) 313 0.696
Lock-In Readout (w/ XTAL) 307 0.682
  Corresponding Loss 1.92%

Reflection Measurement (PDA36A Gain = 40dB)

Reflection coefficient is calculated relative to estimated incident power from 8% tap PD at the time of the measurement (2.7mW)

Measurement Voltage (mV) Power Reflected (uW) Reflection Coefficient (ppm)
Lock-In Readout (surface 2 thru 1) 1.96 0.087 32?
Lock-In Readout (surface 1) 0.83 0.037 13?

One thing I think we can be certain about is that the old LASTI crsytal has lower overal loss than this new SPPKTP crsytal we measured, which is rather disappointing. The old LASTI crystal's AR coatings are shockingly better than the new SPPKTP's. The reflected beams from the old crystal are super difficult to measure amongst the scatter from the rest of the measurement setup, even with an iris on the reflection detector. As a result, I am not super confident in the accuracy of our reflection coefficients for the old crystal. However, even accouting for the ~0.3% loss from the AR coatings on the new crystal, the SPPKTP crystal specs from Raicon (linked here) would suggest that the total overal loss would be something like (2*0.12%)+1cm*(0.42%/cm) = 0.66%. Our measured loss seems to be more than 4x this amount. On the other hand, the old LASTI crystal's loss (which is seeimngly entirely dominated by bulk losses) is in fact near the specification for the standard Raicon PPKTP crystals (2.2%). Today, I will go back and remeasure this a few times and leave a comment if I can spot a problem with this data.

Images attached to this report
Non-image files attached to this report
Comments related to this report
dmdesant - 18:20, Thursday 28 August 2025 (12047)

I re-measured these again today and found a spot on the new crystal that looks a bit better in terms of transmission losses. I also remeasured the old LASTI crystal for good measure. I also think I had a unit correction missing in the reflection power calculation for the new crystal. Those numbers are in nanowatts (updated to correct), and the same facet was accidentally measured twice (now it seems clear that one facet is almost 2x lower loss than the other). That being said, the reflection coefficients seem to still be correct. In this particular measurement, the old LASTI crystal and the new SPPKTP crsytal are more comparable (around 2% loss).

New SPPKTP Crystal (#4)

Transmission Measurement (PDA100A Gain = 0dB)

Measurement Voltage (mV) Power (mW)
Lock-In Readout (No XTAL) 499 1.11
Lock-In Readout (w/ XTAL) 489 1.09
  Corresponding Loss 2.00%

Reflection Measurement (PDA36A Gain = 40dB)

Reflection coefficient is calculated relative to estimated incident power from 8% tap PD at the time of the measurement (5.06mW)

Measurement Voltage (mV) Power Reflected (uW) Reflection Coefficient (ppm)
Lock-In Readout (left spot) 152 6.71 1330
Lock-In Readout (right spot) 236 10.4 2010

OLD LASTI PPKTP Crystal

Transmission Measurement (PDA100A Gain = 0dB)

Measurement Voltage (mV) Power (mW)
Lock-In Readout (No XTAL) 501 1.09
Lock-In Readout (w/ XTAL) 490 1.11
  Corresponding Loss 2.20%

Reflection Measurement (PDA36A Gain = 40dB)

Reflection coefficient is calculated relative to estimated incident power from 8% tap PD at the time of the measurement (5.10 mW when left spot was measured, 5.16mW when right spot measured)

Measurement Voltage (mV) Power Reflected (uW) Reflection Coefficient (ppm)
Lock-In Readout (left spot) 2.2 0.00971 19?
Lock-In Readout (right spot) 5.4 0.0238 46?
dmdesant - 16:52, Friday 29 August 2025 (12048)

Today, I measured the final crystal that was sitting in the LASTI VOPO. This is XTAL #3 from the SPPKTP batch

New SPPKTP Crystal (#3)

Transmission Measurement (PDA100A Gain = 0dB)

Measurement Voltage (mV) Power (mW)
Lock-In Readout (No XTAL) 505 1.12
Lock-In Readout (w/ XTAL) 499 1.11
  Corresponding Loss 1.19%

Reflection Measurement (PDA36A Gain = 40dB)

Reflection coefficient is calculated relative to estimated incident power from 8% tap PD at the time of the measurement (5.33mW)

Measurement Voltage (mV) Power Reflected (uW) Reflection Coefficient (ppm)
Lock-In Readout (left spot) 245 10.8 2030
Lock-In Readout (right spot) 296 13.1 2450

This crystal actually has about half the loss of XTAL #4, but unfortunately the worst AR coatings yet. That being said, the Raicol-specified loss with these AR coatings should be 0.42% + 0.45% = 0.87%. It seems that the combination of the damage this crystal endured during the coating process and the poor quality coating that resulted likely account for almost half the loss in this sample. We may revisit this crystal in the LASTI VOPO again to see if our initial alignment was at fault for the poor performance we initially saw last week.

SQZ General
eoelker - posted 16:41, Wednesday 20 August 2025 (12044)
2nd SPPKTP crystal is less of a dud

I put in the 2nd SPPKTP crystal today.  The green finesse looks similar to the first, so still quite disappointing.

However, the NLG is a bit better.  Threshold seems closer to 20 mW in TEM00.  This one is at least behaving like a newer crystal.  The escape efficiency is a bit lower ~97.5%, so perhaps there's still some optimization on alignment to be done. 

 

Still, I'm a bit perplexed by the high losses at 532 nm.  Both crystals light up even with the beam fairly well centered on the AR faces in my photographs.  Seems like there's some excess scatter compared to the crystal I'm using for the H1 spare cavity.  However, I noticed that M3 is a bit dingy, we had to walk the cavity alignment to avoid an obvious defect on one part of the mirror.  Hard to tell if the excess scatter is actually from the crystal or if its from the M3 scatter spreading out as it propagates. 

SQZ General
eoelker - posted 14:40, Tuesday 19 August 2025 (12043)
SPPKTP crystal is a dud

[Daniel, Haocun, Eric]

On friday, we decided to swap out the old crystal in the LASTI VOPO with one of the new SPPKTP crystals.  Based on Wen's investigations, we decided to start with crystal #4 from the recent Raicol order, as crystal #3 appears to be damaged along a portion of the ar coated surface.  Before installing the crystal, we decided to change some of the wiring for the oven to address the issues with wire tension causing the solder joints to break.  Seems to be much better now.

Unfortunately the crystal swap doesn't seem to have gone as hoped.  The SPPKTP crystals are supposed to have significantly lower green absorption (<0.5% as opposed to 2.2% with regular PPKTP).  However, the green finnesse doesn't appear to have improved.  I was expecting a value around 250 with the new crystal, but we're still sitting around 140.  Also, the NLG of this crystal seems to be quite low.  Based on our initial measurements, we seem to have a NLG of around 4 with about 11 mW incident and 65% mode coupling.  This corresponds to a threshold power of around 30 mW in TEM00.  Comparing with measured values from O3 (the LASTI VOPO uses the O3 edition of the input/output couplers),  this is much worse than we'd typically see for a new crystal.  Usually a threshold power at this level isn't observed until the crystal is quite old.  A threshold in the 15-20 mW range is more typical for a new spot on a normal PPKTP crystal.  We tried sweeping the temp from about 22C to 40C just to make sure we weren't sitting on a local maximum, but we didn't see much NLG outside of the peak centered at 32.6 C. 

We believe that we are well centered on the crystal and it is well aligned.  We are able to measure a 98% escape efficiency, are seeing 11 coresonance positions across the face of the crystal, and the 532 nm beam appears to be centered on the crystal face.  This is usually a good indication that things are set up properly.

This is quite disappointing.  The lower 532 nm absorption is the reason why Raicol expects them to be more damage resistant.  Given the high losses, I'm not sure if it's worth doing any long-term testing of this crystal.  We will try the 2nd crystal next.  Only a portion of the AR surfaces are damaged, so we should be able to test the remaining regions.  Perhaps we will try to measure the 532 nm AR coating for crystal #4 just to make sure that everything is up to spec.