mrtodd - posted 17:57, Thursday 12 November 2026 - last comment - 18:14, Thursday 12 November 2026(12307)
Mid-Y chamber work

M. Todd, E. Oelker

I re-ran a PDH scan and analyzed it to see if our calibration factors were still consistent for our noise measurements. It seems like we've lost a good chunk of our PDH gain, but it definitely be from a different input power / alignment as I have not been rigorous maximizing the aligned TRANS peaks. Regardless, I used the calibration factors (PDH gain, PZT gain, cavity length [did not change]) to get the suppressed frequency noise and the control signal. With some basic loop algebra, the suppressed frequency noise is just the loop suppression times the displacement noise and the control signal should be the close loop gain times the displacement noise. Looking at these, this seems to be sensible, as the control signal seems to mostly follow the table motion as we would expect for our high gain loop.

This afternoon, Eric and I moved three of the L4Cs into the Mid-Y chamber and plugged the cabling into the feedthrough panel. We ID'd each L4C signal so we know which sensor is connected to each ADC channel. We need to put the three cables left for the other L4Cs as far to the other side of the table as we can so that we can put the back Mid-Y door on and take the front one off. 

Non-image files attached to this report
Comments related to this report
mrtodd - 18:14, Thursday 12 November 2026 (12308)

I've got it setup to the cymac right now that CH1 is L4C0 (in X), CH2 is L4C1 (Z), CH3 is L4C2 (Y). The next ones should be Z, Y, Z, as was setup in the End-Y chamber.

Displaying report 1-1 of 1.