Reports until 11:57, Wednesday 11 November 2026
CoupledCavity General
mrtodd - posted 11:57, Wednesday 11 November 2026 (12306)
Frequency Noise Measurement of Test Cavity

M. Todd

I realigned the cavity now that we let the air back in and took some frequency noise measurements along with some seismic data. I used the SR785 to take the error signal spectra which I calibrated into Hz with our PDH gain disriminant. I calibrated the longitudinal (dof Y, in this case) table motion into Hz using the cavity length. This PDH scan is a little old but hopefully still mostly fits since we're using all the same settings. For reference I plotted the free running NPRO heuristic noise curve as well. From the OLG, our UGF is around 20/30kHz.

Some interesting features are these 10.5 Hz harmonics, which I reckon are most likely fans spinning or something. Additionally, around 7-8 Hz we see a broad peak which could be from the vertical resonances of the stack. I'm interested in getting the table motion data for the Mid-Y, as it may lend a clue to the broad peak at 2Hz.

 

Non-image files attached to this report