dennis.martynov - posted 09:31, Thursday 29 October 2026 (12297)
C-6D discharge time vs humidity

During the C-6D installation at LASTI, we encountered a fused silica charging problem in the cleanroom (log 11846), which Peter & Rich helped us resolve with an ion gun.

As we have not observed this issue in Birmingham, we measured the fused silica discharging time at various humidity levels, keeping the temperature at 20.5 C. Under typical laboratory conditions (45–50% RH), the discharge time is 1-6 minutes. We then placed the fused silica ring in a box containing silica gel beads to reduce the humidity to 39–40% RH. Under these conditions, the discharge time increased to 1.7 hours.

This lower humidity range may be representative of the conditions in the Lasti cleanroom. Does anyone have a measurement or estimate of the relative humidity and temperature in the Lasti cleanroom? With these values, we could normalise the water vapour pressure to our measurement temperature for a more direct comparison.

It might also be worth considering installing an ioniser bar in the cleanroom, similar to the example shown in the attached picture. In our cleanroom, the discharge time decreases to around 17 seconds when the ioniser is switched on.

 