Displaying report 1-1 of 1.
Reports until 20:18, Monday 26 October 2026
IntegratedPhotonics Chips
melissa , chaejin - posted 20:18, Monday 26 October 2026 (12293)
InGaP squeezer microring chips
(melissa, chaejin)

PhM 1 device 18: 780 resonance and 1560 nonlinear gain vs pump power

Chip ID: PhM 1 device 18 (InGaP microring from our collaborator Lilli Thiel from UCSB. Designed to pump at 780 and generate squeezing at 1560, has a separate bus waveguide for each pump and squeezed light)

Setup: one 1560 laser -> SHG for the 780 pump, and -> AOM for a weak 1560 seed. We sweep the on-chip heater across both resonances and record the 780 transmission and the 1560 transmission at the same time, for pump powers 1–240 µW. The AOM makes the seed phase sweep, so the 1560 trace oscillates between the amplified and deamplified quadrature; Melissa and I call its envelope the "NLG envelope" (because it is non linear gain envelop). setup sketch attached

Model: Chi(2) parametric gain g = g_0 * sqrt(n_780) where n_780 is number of 780 pump photons, plus thermal red-shift linear in n_780 and power-dependent 780 intrinsic loss \kappa_i(n). For each power we read T_min (minimum transmission of 780) and its detuning from the 780 dip (as a function of \kappa_i, n_780) and the two envelope values at the 1560 resonance (where we model g).

Results: cold cavity 780 and 1560 fits. 780 shift is linear in power where we fit thermo optic pulling coefficient K. 780 T_min is non-monotonic in power; this is not loss but a facet Fabry–Perot whose phase drifts with heating — with it included \kappa_i is constant and no TPA/FCA up to 240 µW. 1560 amplification reaches +6.6 dB at 240 µW, proximity to OPO threshold here is \rho ^2 ≈ 0.2, g₀/2π ≈ 2–4 MHz/√photon depending on the loss model.

Next: model and remove the facet FP (Lilli is measuring the 780 bus FP for 4 waveguide lengths). We can consider that only absorbed pump photons heat the ring, so the thermal shift should be written vs absorbed power rather than n_780

Images: setup sketch, cold cavity fits (780, 1560), 780 T_min and detuning vs power, facet FP fit, max NLG vs power, NLG envelopes, Proximity to threshold \rho vs n_780.
Images attached to this report
Displaying report 1-1 of 1.