Reports until 19:10, Thursday 22 October 2026
IntegratedPhotonics General
melissa - posted 19:10, Thursday 22 October 2026 (12285)
QE measurement @ 1550nm / 780nm of the VM High-bandwidth photoreceiver

Wanted to test whether I could use the VM high-bandwidth photoreceiver for phase noise measurements of Titanium:sapphire microlasers (NASA collaboration with Brightlight Photonics) 

The current setup is a cleaved SMF-28 fiber mounted on a nanomax in front of the PD [photo attached] -- InGaAs PIN photodiode, 100 um diameter, TO package with a lens cap

Previously I was using a 2.6 um tapered lensed PM fiber (Corning PM15-U25D, FC/APC) , but was measuring QE ~ 45% at 1550nm. The lensed fiber has a divergence of about ~0.4 rad, which apparently exceeds the acceptance of the lens cap. The optimal position was several mm from the cap... the cleaved fiber has divergence about ~0.1 rad at 1550 nm

With cleaved fiber (taking transimpedance gain 103.27 kΩ):
1550 nm | 170 uW | R ~ 1.07–1.19 A/W from shot noise (photocurrent 182–203 uA) | QE ~ 86-95%
780 nm | 390 uW | R ~ 0.27–0.28 A/W from shot noise (photocurrent 104–111 uA) | QE ~  42-45%   [NOTE: the 780 is frequency-doubled 1560 and there may be some small leakage of 1550 light]

Attaching dark noise and shot noise of the 1550 measurement. With 170 uW, we have ~7dB clearance 40-70 MHz, 6 dB 100 MHz, 5 dB 200 MHz... Good enough clearance at low power to measure soliton quantum noise of our sapphire microrings (recently metallized by Daniil! so now they have microheaters, we need to test those).

Conclusion: this photodiode will work for 780 nm measurements, no need to swap in silicon PD

To-do: cleave 780 nm SM fiber, possibly switch to APC fiber to avoid back reflections


 

 

Images attached to this report