Displaying report 1-1 of 1.
Reports until 17:10, Tuesday 20 October 2026
CoupledCavity General
mrtodd - posted 17:10, Tuesday 20 October 2026 - last comment - 09:41, Thursday 22 October 2026(12281)
CyMAC seismic measurement

M. Richardson, M. Todd


We've taken some seismic measurements that we have more confidence in, after Mitchell and I confirmed the cabling for each L4C in the Y HAM. We also spotted an error in how I wired up the c2sei model, which previously had YAW and ROLL flipped. We saved this data to /home/controls/Templates/sei/l4c2dof.xml . All three translation DOFs have an integrated RMS of around 1um down to 200mHz. PIT and ROLL are around 0.4urad integrated RMS at 200mHz, while YAW is down aroud 1urad. Noticeably, there is quite a bit more motion between 1--10Hz, especially in the rotational DOFs then expected.; this could be due to the door being off, and air currents causing a bunch of motion.


After confirming each L4C output, we plugged them into the CyMAC in order (CH1 - CH6), without using CH0. There is some unexplained issues with CH0 ADC channel, so we avoided using it. We calibrated the ADC channel from counts to meters using four FM. They are: the reciprocal of the L4C TF, an AC filter and undoing the preamplification of the L4C signal and undoing the ADC (16bit, +/-15V). We did not assume any deviation from the spring parameters quoted in the design, though the springs have surely sagged over time (20 years).

FM1: L4C2m = zpk([0.28+i*0.96;0.28-i*0.96],[0.01;0.01;0.01], 580,"n")
FM2: AC         = butter("HighPass", 2, 0.02)
FM3: 1/pre    = gain(0.01)
FM4: 1/ADC  = gain(0.0004578)

These outputs were then fed through a sensing matrix to recover the DOFs. The channels correspond numerically to the L4Cs as configured in the layout sketch. The middle left L4C (L4C-4) was moved previously to route the transmission path to the periscope, and so it's separation from the central axis is assumed to be 8 inches shorter than L4C-2.

horizontal x at (x, y, h):   s = X + h·RY − y·RZ
horizontal y at (x, y, h):   s = Y − h·RX + x·RZ
vertical         at (x, y):        s = Z + y·RX − x·RY

The response of each sensor can be expressed as t + theta x r, where t is the relevent translational DOF and theta x r gives the two relevant rotational DOFs from the cross product with the lever arm. The inversion of this matrix yields the sensing matrix.

Images attached to this report
Non-image files attached to this report
Comments related to this report
mrtodd - 09:41, Thursday 22 October 2026 (12282)

Data from this morning ~9am. Also attaching the l4c calibration.

Non-image files attached to this comment
Displaying report 1-1 of 1.