Following up from 12275
Today we completed the alignment of the two CRS LPS, and rebalanced CRS2 to ~18mHz. It would be good if we have a chance to measure a ISI tilt to CRS transfer function before chamber closeout to verify the d-values. Otherwise, the work on this side of the table is now complete, and the two CRS were left unlocked.
The fringe visibility after alignment were the following :
Max fringe visibility (fig1): 90% LPS1 / 82% LPS2
Min fringe visibility (fig2): 52% LPS1 / 38% LPS2
The variability of the fringe visibility is because of large amplitude of the CRS in air, swinging by ~mm relative to the LPS. In normal operation, the CRS wing is only moving sub micrometers, so we should be able to park it in a good LPS spot.
To get the resonant frequency, we measured the time between two slow fringes (when the velocity is zero) which corresponds to 1/2 a CRS cycle. 28s*2 = 56 s = 18mHz (fig3)
Next the LPS signal chain needs to be commissioned (model change, medm screens, filter settings etc...) but all of this can be done remotely.
Otther things to note:
* The LPS 1 (left) and 2 (right) were respectively connected to chan 28 and 29 of m1sus ADC2.
* The remote balancing picomotors were tested (for both CRS) from the PC sitting in the cable tray, they each rotate as expected in both directions.
* The capacitive drive chain for the damping was verified for CRS2 (voltage sent from dac and measured with the multimeter on the plates themselves).
* A new cable was added between the output of the Birmingham TIA chassis (db9, Output 1) to the last channel of the m1sus ADC2 AA chassis, where all the SensIt signals are routed.
* Since the CIT Toptica is not in used anymore (replaced with the Birmingham laser used for SensIt) we removed it from chamber side to reduce cable clutter. It was stored in a box, alongside with all the CRS tools in the pink lab NW17-009
* Pictures related to this work can be found in this album
Arnaud, Marie, I think the aligned LPS-CRS contrast of 82-90% is consistent with the Birmingham measurements.
I would like to add a note on the added vertical offset of the retroreflector and, related to that, on the lowest observed contrast when the CRS moves in angle. The LPS contrast can be approximated as
C ~= C0 - (alpha / theta)^2, where C0 ~ 0.85-0.9, alpha is the CRS tilt angle, and theta determines the angular range of the sensor and is given by the equation
theta w / 2 / (x_beam + h), where w = 0.5 mm is the beam radius, x_beam = 2.5 mm is the offset between the input and returning beam, and h is the vertical offset between the CRS pivot axis and the retroreflector.
By design, we like h = 0 and achieve theta = 100 mrad. In your case, it seems that you have lowered the retroreflector by ~20 mm to achieve the LPS-retroreflector separation of 50 mm. However, this also leads to a reduced angular range of LPS: theta = 11 mrad. I think this is consistent with what you currently measure.
The reason why a non-zero vertical offset of the retroreflector position, h, reduces the contrast is because the CRS motion shifts the retroreflector horizontally according to the equation x_retro = h * alpha, and this shift significantly reduces the contrast. I think in the ideal case, you could remove the vertical offset, h, and move LPS closer to the retroreflector. This case would preserve the angular range of the LPS-CRS sensors at the design value.