mrtodd - posted 17:30, Friday 16 October 2026 (12278)
CyMAC seismic measurement

S. Park, M. Todd


After replacing some voltage regulators and op amps I was able to get the 6channel pre-amp working. We moved the CyMAC over closer to the PSL racks and started connecting cables from the L4Cs to the preamplifier to the CyMAC. CH1 of the pre-amp seemed to not behave as nicely as I originally thought, and seemed to trip, or affect in some way, the ADC out -- so we elected to sub in the SR560 for it but the 560 we chose seems to be limited to a gain of 10 for the signal coming from the seismometer (even though it should be like mV).

Using the sei model and repopulating the sensing matrix after rebooting the CyMAC, we took some spectra of the various DOFs. The DOFs are in m/rtHz and rad/rtHz, with the ADC Noise calibrated in the brown trace. It seems we are still limited by ADC Noise below 1Hz, though until I fix the pre amp we are stuck at a gain of 10.