CoupledCavity General
mrtodd - posted 09:52, Thursday 08 October 2026 (12272)
Test Cavity Slow Ctrl Setup

Yesterday I took some time to try and diagnose why the slow control did not seem to be feeding back correctly, I wanted to look at the DC set point of the temperature. I saw that for some reason it seemed to be set to 59C, which was way too high and more than likely at the very top of its range. I had some trouble actually fixing the setpoint, but by turning the tuning mode from "Extended" to "Min" with the dip switch in the back, I was able to put the set point to 47.5C. I had to turn the laser diode ON to do this, which I was very nervous about, but it should hopefully save this state for the next power cycle. With this setting, I fed the slow error signal through a 100mHz LPF to the temperature control and saw some of the behavior I was looking for.

Watching the PZT driver monitor channel on a scope, you can see it swing around its set point at low frequencies quite a bit, and then when the slow control is engaged it is much more stable around its setpoint. We probably want some gain adjust on the slow control error signal, which will benefit from an integrator with a variable input resistor.