LASTI General
bruno - posted 09:54, Tuesday 22 September 2026 (12263)
Cobri Splitter passes RGA, SENSIT
The 1-4 1550nm splitter cleaning process was more complicated than we thought. 

P/N: PM-PLC-B-1X4
TYPE: PM1550 250µm bare
Connector FC/APC
S/N: 10260717060001

Inside the splitter housing are a few different epoxies which seem difficult to remove. The white epoxy is likely a low outgassing epoxy NTT OS5958.
The black rubber alongside the ends of this splitter remained ambiguous so it was carefully removed with a scalpel and tweezers. 

The RGA results show a HC outgassing rate of 5e-11 Torr-liter/s which comfortably passes under the 4e-10 Torr-liter/s threshold. 

Attached are the RGA results and some photos of the splitter. 