Reports until 18:45, Thursday 17 September 2026
CoupledCavity General
mitchell - posted 18:45, Thursday 17 September 2026 (12261)
Test Cavity Locked
Mitchell R, Matthew T

After connecting the PZT driver to the laser PZT, we were able to lock the cavity. All in all we could get it to stay locked for about 5 minutes, which meant we could measure the open loop gain on the SR785. We took a measurement at 20dB of gain on the controller and using the Boost 1 filter. The UGF is at about 46kHz with a phase margin of 150 degrees.