CoupledCavity General
mrtodd - posted 09:47, Wednesday 16 September 2026 - last comment - 13:13, Thursday 17 September 2026(12257)
Test Cavity Work

M. Richardson, M. Todd

Mitchell spent some time yesterday aligning the newly input 21.5MHz EOM, and did some cavity alignment to recover the TRANS peaks. Today we wanted to look at the error signal. We are using +7dBm for the EOM and mixer LO drive, and already seeing a massive increase in modulation depth relative to the 532nm EOM (as expected). We did some beam walking to minimize the misalignment peaks, and took a trace of the error signal while scanning. We had to change the LO phase a bit but we were able to get 3FSRs in this trace. Roughly with +7dBm, we are now seeing 15mVpp through the carrier resonance.

 

Analysis of the data will follow in the comments.

---------------  Measurements --------------

RFPD Power: 1.1mW

RFPD DC voltage: 970mV

RFPD DC "Gain" [V/W]: 0.880 V/mW

Comments related to this report
mitchell - 09:53, Wednesday 16 September 2026 (12258)
Here are the saved traces from the oscilloscope.
Non-image files attached to this comment
mrtodd - 13:13, Thursday 17 September 2026 (12260)

I temporarily swapped the 7dBm mixer for the 17dBm mixer and turned up the EOM drive correspondingly. I walked the beam alignment a bit in both pitch and yaw and was able to lower the misalignment peaks quite significantly. Retaking the data from the scope (using the script on the iMac) I was able to get a pretty good fit to our parameters. Plugging the printer output (back) of the oscilloscope into the usb of iMac allows users to grab the data that's on the screen. Just run

cd git
python capture_all_traces.py

As of right now our settings are such:

  • ~1.2 mW of power on the RFPD
  • 17dBm of RF power to EOM and LO port of mixer
  • 16Vpp for PZT scan @ 100Hz
  • 13.3MHz f_mod, even though the RFPD is tuned for 18MHz.

Knowing the modulation frequency well (13.3MHz) you are able to get a very good calibration of the true FSR and corresponding PZT gain (MHz/V). From there you can fit the other parameters, with the uncertainty arising from the multiple FSRs which will not exactly line up with each other.

  1. Cavity Length: 7.4704m +/- 0.9mm
  2. FSR: 20.065MHz +/- 3kHz
  3. TMS: 4.3 +/- 0.1 MHz
  4. Finesse: 15.03 +/- 0.07
  5. Pole: 667kHz +/- 3kHz
  6. Omega (fmod): 13.3 MHz
  7. Mod. Depth. 0.259rad +/- 1mrad
  8. Demod Phase: 9.2 +/- 0.8 deg

 

This morning I also found a cable to act as an adapter of the PZT driver output (SHV) to BNC of the laser PZT input. We have it all hooked up now, so we should be able to feedback to the PZT and try locking.

Non-image files attached to this comment