Power Meter Issue and Replaced EOM
Mitchell R.

Swapped out the EOM with one for the correct wavelength which required fine tuning input alignment to the cavity again. In the process of trying to measure the error signal with the new EOM, I discovered a problem with the power meter we have been using. The "500mw" and "5mW" settings on the sensor were giving different readings. Comparing measurements from a second power meter and sensor showed that the "5mw" setting on ours is wrong for some reason, but the "500mw" agreed. The comparison also involved using our sensor connected to the second meter and vice versa to distinguish between an issue in the sensor vs an issue in the power meter. It appears the sensor has the issue, not the power meter itself.