Reports until 17:52, Monday 14 September 2026
OEO General
dmdesant - posted 17:52, Monday 14 September 2026 (12255)
LWE-120 NPRO Cavity Response Fit

We were able to fit the cavity response data we took this past week and estimated the output coupler reflectivity and intracavity loss. The rough fit is shown here.

Out-Coupler Reflectivity 97.9%
Intracavity Loss 0.30%
Finesse 42
FWHM 357 MHz
Storage Time 2.8 ns
Injection Efficiency 87.5%

 

Images attached to this report